Sensoray 7409TB Owner's manual

Type
Owner's manual
Sensoray Model 7409TB/7409TF
Termination Boards
01/10/2000
INSTRUCTION MANUAL
Sensoray Co., Inc.
7313 SW Tech Center Dr., Tigard, Oregon 97223
voice: 503.684.8005, fax: 503.684.8164, e-mail: [email protected]
www.sensoray.com
M A N U A L 7 4 0 9 T B / 7 4 0 9 T F
Sensoray models 7409TB and 7409TF (both referenced simply
as “TB” unless otherwise specified) are robust, small footprint
termination boards that break out flat cable conductors to
screw terminals. A 40-pin header on the back side of the board
facilitates panel mounting and supports direct connection to
any of the following Sensoray products:
q All Smart A/D™ boards.
q Multi-function interface cards that utilize a 40-pin analog
I/O connector, including models 421, 425 and 7421.
q Analog output interface cards that have a 40-pin analog
I/O connector, such as model 7405.
Both TB models include an integral temperature transducer.
When used with Sensoray Smart A/D™ boards, this sensor is
connected to a dedicated measurement channel that is used for
thermocouple reference junction compensation. No wiring or
configuration of this sensor is necessary; it is hard-wired to the
Smart A/D™ dedicated measurement channel.
Model 7409TF also includes programming shunts for enabling
Open Sensor Detection (OSD) when used with Sensoray Smart
A/D™ boards.
Fig. 1 — Block Diagram
7409TF Configuration Shunts
On model 7409TF only, programming shunts are provided to
independently enable the OSD option for each Smart A/D™
channel. The option shunts are located as shown in the
following diagram:
Fig. 2 — Open Sensor Detection (OSD) programming shunts
Some Smart A/D™ products have built-in OSD option shunts.
Do not install any OSD option shunts on the TB if your Smart
A/D™ has built-in OSD option shunts; instead, install OSD
shunts only on the Smart A/D™. As for which channels should
have OSD enabled, refer to your Smart A/D™ instruction
manual for the recommended practice.
Field Wiring
Field wires need only be stripped of insulation, inserted into
the terminal receptacles, and the screw tightened for a secure
connection. The following sections discuss field wiring
specifics for the various types of supported interface boards.
Smart A/D™ Connections
8- and 16-channel cards—Five screw terminals are provided
for each channel: two for voltage sense (V+, V-), two for
sensor excitation (I+, I-) and one for shield (S).
32-channel thermocouple cards—Two screw terminals are
provided for voltage sense on each channel: either (V+, V-) or
(I+, I-). In addition, one shared shield (S) terminal is
provided for every two channels.
Two TB boards are required for each 16- or 32-channel Smart
A/D™ interface card. Connect all sensors as described in your
Smart A/D™ documentation.
Multi-Function I/O Card Connections
Two screw terminals are provided for voltage sense on each
analog input channel (V+, V-) and two for each analog output
channel (I+, I-). One shared shield (S) terminal is provided
for each pair of analog input/output channels.
Analog Output Card Connections
Five screw terminals are provided for each channel: Two for
voltage output (V+, V-), two for current output (I+, I-) and
one for shield (S).
Calibration
TB boards may be freely interchanged because their onboard
temperature sensors are calibrated. Although the sensor is
factory calibrated, it may be necessary to recalibrate it after
extended time in service or due to exposure to extraordinary
environmental stress.
Calibration Procedure
1. Connect a calibrated Smart A/D™ to the TB. Connect a
calibrated thermocouple to the TB and immerse the “hot”
end into a precise, thermally controlled environment.
Allow at least ten minutes after power-up for the system to
warm up and stabilize.
2. On the TB, adjust potiometer R4, reset and re-initialize the
Smart A/D™ and read the thermocouple data.
3. Repeat step 2 until the Smart A/D™ indicates the correct
thermocouple temperature.
Temperature
Transducer
40-Pin Header
to I/O Interface Board
Screw Terminal
OSD Programming
Shunts (7409TF only)
Blocks
Sensor
Field
Wiring
Important note for thermocouple users:
Exposing the TB to thermal transients can severely
degrade thermocouple measurement accuracy. If you
are measuring thermocouples, insulate the TB from air
flows such as ambient drafts and cooling fans. For
best results, encase the TB in a protective enclosure.
Models 7409TB/7409TF Instructions
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
7409TB.DSN A
TERMINATION BOARD, MODEL 7409TB
SENSORAY COMPANY, INCORPORATED
B
1 1
21:23:04
Title
Size Document Number Rev
Date: Sheet of
PL0PL0
PH0
SH0
SL0
PL1PL1
PH1
SH1
SL1
PL2PL2
PH2
SH2
SL2
PL3PL3
PH3
SH3
SL3
PL4PL4
PH4
SH4
SL4
PL5PL5
PH5
SH5
SL5
PL6PL6
PH6
SH6
SL6
PL7PL7
PH7
SH7
SL7
SL4
SL1
SL0
SH3
PH1
SL6
PL2
PH6
PH2
SL2
PL3
PH3
SH4
SH1
PL1
PH7
PH4
PL4
SL7
SL5
PL6
PL0
SH6
SH5
SH0
PL7
SL3
PH0
SH7
PH5
PL5
SH2
R3
10K
R1
3.9K
R6
3.3K
R5
12K
R2
1K
I+
I-
V+
V-
S
V+
I+
V-
I-
S
V+
I+
V-
I-
S
V+
I+
V-
I-
S
TB1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
U1
LM335
I+
I-
V+
V-
S
V+
I+
V-
I-
S
V+
I+
V-
I-
S
V+
I+
V-
I-
S
TB2
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
P501
2
6
10
14
18
22
26
30
4
8
12
16
20
24
28
32
1
5
9
13
17
21
25
29
3
7
11
15
19
23
27
31
36
37
38
33
34
39
40
SH0
SH1
SH2
SH3
SH4
SH5
SH6
SH7
SL0
SL1
SL2
SL3
SL4
SL5
SL6
SL7
PH0
PH1
PH2
PH3
PH4
PH5
PH6
PH7
PL0
PL1
PL2
PL3
PL4
PL5
PL6
PL7
+12V
TREF
GND
GND
GND
SHLD
SHLD
R4
1K-POT
1 3
2
PWB
PWB7409TB
RevA
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
0500095 rev A A
TERMINATION BOARD, MODEL 7409TF
SENSORAY COMPANY, INCORPORATED
B
1 1
20:52:52
Title
Size Document Number Rev
Date: Sheet of
PL0PL0
PH0
SH0
SL0
PL1PL1
PH1
SH1
SL1
PL2PL2
PH2
SH2
SL2
PL3PL3
PH3
SH3
SL3
PL4PL4
PH4
SH4
SL4
PL5PL5
PH5
SH5
SL5
PL6PL6
PH6
SH6
SL6
PL7PL7
PH7
SH7
SL7
SL4
SL1
SL0
SH3
PH1
SL6
PL2
PH6
PH2
SL2
PL3
PH3
SH4
SH1
PL1
PH7
PH4
PL4
SL7
SL5
PL6
PL0
SH6
SH5
SH0
PL7
SL3
PH0
SH7
PH5
PL5
SH2
SL0
SL1
SL2
SL3
SL4
SL5
SL6
SL7
R3
10K
R1
3.9K
R6
3.3K
R5
12K
R2
1K
GND
TEST-POINT
GND
I+
I-
V+
V-
S
V+
I+
V-
I-
S
V+
I+
V-
I-
S
V+
I+
V-
I-
S
TB1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
U1
LM335
I+
I-
V+
V-
S
V+
I+
V-
I-
S
V+
I+
V-
I-
S
V+
I+
V-
I-
S
TB2
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
P501
2
6
10
14
18
22
26
30
4
8
12
16
20
24
28
32
1
5
9
13
17
21
25
29
3
7
11
15
19
23
27
31
36
37
38
33
34
39
40
SH0
SH1
SH2
SH3
SH4
SH5
SH6
SH7
SL0
SL1
SL2
SL3
SL4
SL5
SL6
SL7
PH0
PH1
PH2
PH3
PH4
PH5
PH6
PH7
PL0
PL1
PL2
PL3
PL4
PL5
PL6
PL7
+12V
TREF
GND
GND
GND
SHLD
SHLD
R4
1K-POT
1 3
2
PWB
0400095 RevA
CH0
CH1
CH2
CH5
CH6
CH7
RN1
1M-SIP
CH4
CH3
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4

Sensoray 7409TB Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI