Joerns WeCare Full-Electric Bed Owner's manual

Type
Owner's manual

Joerns WeCare Full-Electric Bed is a versatile and feature-rich bed designed to provide comfort, safety, and convenience for users. Its electric adjustability allows for effortless changes in height, backrest angle, and leg rest angle, enabling users to find the optimal position for rest, relaxation, or medical procedures. The bed's sturdy construction and 450-pound weight capacity ensure stability and support, while its floor locks and swivel locks enhance safety during transfers and use.

Joerns WeCare Full-Electric Bed is a versatile and feature-rich bed designed to provide comfort, safety, and convenience for users. Its electric adjustability allows for effortless changes in height, backrest angle, and leg rest angle, enabling users to find the optimal position for rest, relaxation, or medical procedures. The bed's sturdy construction and 450-pound weight capacity ensure stability and support, while its floor locks and swivel locks enhance safety during transfers and use.

User-Service Manual
Joerns® Bed Frames
WeCare™ Bed
To avoid injury, read user’s manual before using.
2
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
Important Precautions
Warning: Possible Injury Or Death. This product is
intended for use as an adjustable mattress support platform
that contours for resident/patient comfort with adjustable
height for resident/patient and caregiver convenience. Use
of this product in a manner for which it was not designed


Warning: Possible Injury.
that area under and near bed perimeter is free of people and

to the bed.
Warning: Possible Injury. Keep bed in lowest position



Warning: Possible Injury Or Death. Floor locks increase


transfer or resident/patient use of the bed. Failure to do so

Warning: Possible Injury Or Death. Resident/patients



or death.
Warning: Possible Injury Or Death. Do not allow children
or pets under or near the bed. Do not operate the bed while

Warning: Possible Injury Or Death. Bed safe working



450-pound safe working load. Exceeding the safe working

Warning: Possible Injury Or Death. Do not roll bed over
abrupt thresholds with resident/patient in bed. Bed is not
designed for transporting resident/patient. Transporting

Warning: Possible Fire Hazard. Use nasal mask or ½ bed

not extend below mattress support platform. Pendant should



Warning: Possible Shock Hazard. Unplug power cord from

service to the bed. Resident/patient should be removed from

Place bed on its side versus working under it. Failure to do so
Note: In case of unexpected

Warning: Possible Shock Hazard. Position the bed in a
manner that allows ease of access to the power cord and



Warning:


resident/patient. The assessment should be conducted

and federal guidelines related to the use of restraints and

Guidance for the Assessment and Implementation of Side

the U.S. Food and Drug Administration. Further information
can be obtained at the following web address: http://www.
fda.gov/MedicalDevices/ProductsandMedicalProcedures/
GeneralHospitalDevicesandSupplies/HospitalBeds/default.
htm.
Important Note: 


over the side or between the bed assist devices when the
mattress is being used.
Warning: Possible Injury Or Death. Use a mattress

that will remain centered on mattress support platform
relative to State and Federal guidelines. Joerns Healthcare
recommends the use of a mattress with minimum
dimensions of 35" width and 6" height. Length should match




Warning: Possible Injury Or Death. If a resident/patient’s




Warning: Possible Injury Or Death.

of the mattress and assist device. This gap must be small
enough to prevent resident/patient from getting his/her head
or neck caught in this location. Make sure that raising or



Warning: Possible Injury Or Death. Locate mounting

Gap between head/foot panel and assist device must be
small enough or large enough to prevent resident/patient
from getting his/her head or neck caught in this location


each so that gap between them is large enough for trunk and
3
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545



or death. Important Note: Powered air mattress surfaces


Warning: Possible Injury Or Death. Do not use an assist

resident/patient to get his/her head or neck lodged within


Warning: Possible Injury Or Death.
Joerns®


Warning: Possible Injury or Death. NEVER permit more








Warning: Possible Injury Or Death



Warning: Possible Injury Or Death. If using accessories




Caution:


Caution: Particular precaution must be considered
during exposure to strong emission sources such as High



Note:


function is selected. Do not place the bed in a location that

Warning: Use of this equipment adjacent to or stacked with
other equipment should be avoided because it could result in



Warning:

of this equipment could result in increased electromagnetic

equipment and result in improper operation.
Warning: Portable RF communications equipment (including




this equipment could result.
Caution: Particular precaution must be considered during

surgical equipment and similar devices. Do not route high


Caution: 


Joerns Healthcare recommends compliance to the regulations







mounted.








There are no known contraindications for this device.
Save These Instructions
®


and caregivers in mind.


4
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
Bed System Entrapment Information



and use.

an appropriate balance between resident/patient




representatives has addressed the potential danger

beds. These guidelines recommend dimensional
limits for critical gaps and spaces between bed






As the leading manufacturer of long-term care beds



into compliance.
Joerns Compliance Solutions
Matching the right bed components in order to meet




bed model.
Creating a Safer Care Environment


care facilities have particular exposure since serious

dementia patients.
For More Information
To learn more about compliance options with Joerns


ask for free informational publications.



consult the FDA website: http://www.fda.gov/
MedicalDevices/ProductsandMedicalProcedures/
GeneralHospitalDevicesandSupplies/HospitalBeds/
default.htm.
Table of Contents
Important Precautions .................................................................................................................................................. 2
Entrapment Information ................................................................................................................................................ 4
 ......................................................................................................................................... 5
 ...................................................................................................................................................................... 6
 ..................................................................................................................................................................... 9
Servicing ..................................................................................................................................................................... 10
Troubleshooting .......................................................................................................................................................... 11
 ..................................................................................................................................................................... 12
Maintenance and Inspection Information.................................................................................................................... 12
End of Life Disposal.................................................................................................................................................... 12
Accessories ................................................................................................................................................................ 13
 ..................................................................................................................................................................... 16
5
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
Accessories and Options

F14 Assist Device Set
F025 Assist Handle
F026 Assist Device


IV Rod
Floor Mat
Bed Transport Bar
Mattress Retainer Kit

4" Length Extension

Patient Helper and Adapter
Features and Specifications
Joerns® WeCare™ Bed
Standard Features




Grid Sleep Surface



Roll-in-low
Electrical Specifications
Power: 120V




  

Environmental Conditions
Operating Conditions:




Storage and Shipping Conditions:


Atmospheric Pressure: 500 hPa to 1060 hPa
Mechanical Specifications
Description: Dimensions:
 .............................
 ..............
 .........................
 ...........................
Maximum Back Angle ......................... 
Maximum Knee Angle.........................
Maximum Height Mattress
Support Platform................................. 26" Above Floor
Minimum Height Mattress
Support Platform.............................. 
 ............................
 ...................
Bed Identication Numbers


bed’s model and serial numbers found on the

located on bed frame under the mattress support
platform foot section.
Electrical Components Conform To:

Bed Mechanical System Aligns To:

I.D. Label Symbol Definition
Attention




support platform


Pinch Point
6
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
Assembly
Caution: 
conditions above to ensure bed electronics are within
ambient temperature.

inspect for damage.
Parts to prepare:
a. Head End Frame
b. Foot End Frame
c. 
d. 
e. Pendant
f. 

shipping.
Installation Of Bed Frame -Head and Foot
Tools required: at head screwdriver
Bed Frame Installation:

1. 

2. 



Caution: Equipment Damage. Failure to install the

to the actuators.
3. Install the back actuator and knee actuator on
the head frame and foot frame as shown using

orientation of the labels on the actuators to ensure
correct position of actuators when installed. Labels
should be visible when viewed from above the
sleep deck.
4. Insert head frame into foot frame and line up
pin hole.
5. 


Figure 6
Figure 5
Figure 3
Figure 2
Figure 1
Figure 4
7
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
6. Line up back and knee actuators and insert



 Locate back actuator and head end hi-lo cables


 


the notch on the connector is aligned with the


alignment of the notch to the control box port.

Figure 7
Figure 8
Figure 10
Figure 11
Figure 13Figure 9
Figure 12
8
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
9. 

Plug-in power cord to power source and raise the bed
frame. Note:
Install bed panels per the following instructions.
Installation Of Head and Foot Panels
Refer to Bed Panel Installation Sheet for more
Tools required: #2 Phillips screwdriver
Head Panel Installation:
1. Head panel has label on back.
2. 
3. 

4. Install head panel brackets with four screws


brackets into bed sockets.
5. 
Frame-Mounted Foot Panel Installation:
1. Foot panel does not have label on back. Foot

control was ordered.
2. 
3. 

4. Install the foot panel brackets with four screws


into bed sockets.
5. 
Final Assembly Step for Head Panel and Frame-
Mounted Foot Panel:
1. Insert head and foot panel as shown with
mounting hardware facing head of the bed

2. Tighten all screws with panels in place.
3. Panels should slip in and out of their sockets

Figure 15
Head Panel
Bracket-10.5"
Foot Panel
Bracket-6.9"

Figure 16
Figure 17
Figure 14
9
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
IV Sockets
Joerns®
standard equipment. There are two sockets at head end

are used with Joerns IV poles and most ½" IV poles.
Operation


operate the bed.
Warning: Possible Injury or Death. This bed should
not be operated (control the movements of the bed

have the cognitive skills to understand the information
in the owner’s manual or cannot understand and
perform the proper operation of the bed. This bed

Warning: Possible Injury Or Death. Do not allow
children or pets under or near the bed. Do not operate

Roll in Low Lock System
1. To keep bed from rolling when in the lowest


2. Standard caster set comes with two locking
casters placed at opposing corners of bed.
Mobile
Position
Locked
Position
Swivel
Casters
Roll in
Low Lock

 
(casters must
be locked in

Head &
Foot (in low
position

Caution: Be careful not to damage power cord when



Caution:


when transporting bed.
Caution: 
from service until repairs are performed.
Figure 18
IV Socket
Pendant Operation

Back Operation
1. Push button to raise back
2. Push button to lower back
Knee Operation
3. Push button to raise knee
4. Push button to lower knee
Auto Contour Operation
5. Push button to raise
back and knee
6. Push button to lower
back and knee
Bed Height Operation
 Push button to raise bed
 Push button to lower bed
6
5
2
1
4
3
8
7
Figure 19
Figure 20
Warning: Possible Injury. Before leaving the bed

position. Never leave an unlocked bed unattended.
Warning: Possible Injury.
in the locked position before allowing the resident to
enter or exit the bed.
10
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
Servicing
Service parts list available upon request. The actuators
and control box contain no serviceable items.


Warning: Possible Shock Hazard. Unplug

maintenance or service to control box.
Warning: Possible high temperature. Prior to

component has cooled to room temperature.
Power Cord


1. 
damaged. If power cord needs replacement due to

has been replaced.
Actuator and Control Box Information

 
 
 
 
 
Note: The actuators and control box contain no



if bed is overloaded or constrained from movement.
To Remove Control Box
1. 


2. 
3. Unplug all actuator cables and pendant control

4. Remove the three fasteners securing the control

5. 
Actuator and Control Box Troubleshooting

with the following steps:
1. 

2. Switch pendant with a new one or one from a bed

3. 

Figure 22
Figure 21
1 2 3 4 H
Power Cord Port
Figure 23
12
3
11
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
To Remove Actuator
Recommended Tools: Screwdriver, Hammer, Punch
1. 
2. Tip bed on its side or upside down to replace hi-lo
actuators.
3. Unplug actuator cable from actuator.
4. 
screwdriver to remove clip on each pin.
5. 
and punch to tap out the pins.
6. 
 
making note of the following two important items:

Back and hi-lo actuators also include washers.
Note: Ensure all hardware is re-installed in the same
locations.
To Remove AC Power Cord
Recommended Tools: Small Screwdriver, Masking
Tape, Cable Ties
1. Raise bed to maximum height if possible.
2. Unplug power cord from wall.
3. 
Tape remaining cables in their original position. It




Warning: Possible Shock Hazard.


to check condition of power cord on a regular basis.

4. 
5. 



connects to control box until plug comes out.
6. Plug replacement cord into control box until it
latches.
 
position as original routing.
 Replace cable ties along with cables in their
original position and routing to the control box.


Troubleshooting

with the following steps:
Event Possible Remedies
Bed not
responding
to pendant
control
1. 
inserted with cables secured.


2. 
connected and secure.
3. 
power source and control box
green light is illuminated
4. Thermal overload condition. This
should self-correct in 2 minutes


5. 
and clear of obstructions
6. If these steps fail to resolve the


Back sleep
surface or
knee surface
does not
raise or lower
1. Ensure the actuator pins and


2. 

the correct location.
Bed moves
across the

height
1. 
2. 

Note: Electrical damage can occur if a non-approved

To Maximize Service Life
1. Keep the bed clean.
2. 
discrepancies or problems to Joerns Healthcare

3. 





appropriate service parts.
4. Make sure each inspection includes the underside
of the bed frame and mattress support platform.

loose fasteners.
12
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
5. 




6. 



 
attachment points and ensure all related pins are


if required.
 
The recommended lubricant is Joerns-approved


9. 
excessive bending of cable. Avoid dropping or

life.
Cleaning
Caution:

Caution:

to the corrosion of steel components or hardware.

the bed with a damp cloth soaked in clean water and



of critical components. Failure to follow the above



®
well-wrung cloth. Excess water in or on electrical parts
can damage these components.
Do not submerge the pendant in liquid.


and warm water. Read and follow all label instructions
on cleaning products.
Other suggested cleaners:








cleaning with furniture polish or lemon oil. Laminate


Maintenance and Inspection
Information


basis.




cables.



End of Life Disposal


To prevent possible harm to the environment or human


material resources.
13
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
Accessories
Accessory Weights








Warning: Possible Injury Or Death. 



Baseboard Bumper
Recommended Tool: Rubber mallet
1. 
2. 


Mattress Stops
Warning: Possible Injury. 

used with mattress that compresses to less than 1.5
inches under load.
Warning: Possible Injury.
the entire width between side stops. Failure to do so

Tools Required: #2 Phillips screwdriver
1. Locate the screws at each corner of the inside
tube at the foot end of the screws mattress support

2. Remove the screws using the #2 Phillips
screwdriver; do not discard the screws.
3. Position and insert the mattress stops in the

removed in step 2.
4. Locate the screws in the holes in each corner of
the inside tube at the head end of the mattress
support platform; one per side.
5. Remove the screws using the #2 Phillips
screwdriver; do not discard the screws.
6. Position and insert the mattress stops in the holes
provided. Reinstall the screws removed in step 5.
Figure 25
Figure 24
Figure 26
Figure 27
14
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
Bed Transport Bar (Figure 28)
1. 

onto its casters. Note:
designed for use as resident/patient transport

warnings related to bed transport bar and follow all
instructions below.
2. Install bed transport bar according to supplied


3. 

Pushing up or attempting to use transport bar to


loss of control of bed.
4. 



the bed. Never use bed transport bar to move
bed on a slope that exceeds ANSI and/or ADA

Figure 28
Figure 29
Trapeze And Trapeze Adapter
Warning: Possible Injury Or Death. Improper



Warning: Possible Injury Or Death. Follow the




or death.
Warning: Possible Injury Or Death. A minimum of


lowest position. Failure to do so could damage the


Trapeze (Figure 29)
1. ®

2. 
manufacturer’s directions included with

15
Joerns® Bed Frames
WeCare™ Bed
©2023 Joerns Healthcare • 6110429 RevE • 000545
Trapeze Adapter (Figure 30)
1. 


related to this option and follow all instructions
below.
2. 


3. 

instructions.
4. 
positions to ensure proper clearance and


low operating height.
5. 
accordance with instructions from a healthcare

6. Do not exceed the maximum load condition listed
on installation instructions.
Figure 31
Figure 30
Drainage Bag Holder (Figure 31)
Warning: Possible BioHazard.
the drain bag manufacturer’s installation and use
instructions. Failure to do so could result in a bag

Warning: Possible Injury Or Death.

the bed user to function. Do not allow items attached


damaged. Exercise particular caution when the bed is
moved to its lowest position as items attached to the


Warning: Possible Injury Or Death. Make sure items
hung on this holder have proper clearance and are not
pinched or caught during bed adjustment. Failure to do

1. The drainage bag holder is designed as an
attachment point for items such as urine collection
bags. Make sure items attached to holder are

warnings related to the drainage bag holder and
follow all instructions below.
2. Install the drainage bag holder and attach items
to the holder according to supplied instructions. If

for instructions.
3. It is important to operate all bed functions through
their entire ranges to ensure that items attached
to the holder have proper clearance and are not
pinched or caught.
4. 

bed functions through their full ranges to ensure
that items attached to the holder have proper
clearance and are not pinched or caught.
5. 
and must be below bed user to function. Do not
allow items suspended from the holder to rest on



Joerns Healthcare Warranty Program
for Joerns® WeCare™ Bed Frame
www.joerns.com • email: [email protected] ©2023 Joerns Healthcare • 6110429 RevE • 000545 • MAN1
United States
2430 Whitehall Park Dr. Ste 700
Charlotte, NC 28273
(P) 800.826.0270
(F) 800.457.8827
Canada
35 Consortium Court
London, ON Canada N6E 2S8
(P) 866.546.1151
(F) 519.451.8662
United Kingdom and
Other Countries
+44 (0)344 811 1158
Netherlands
+31 (0)88 2080200
Joerns®


service.

electrical components.
Steel structural components on beds are covered


1 of
the product.


accordance with user and service instructions is not
covered.



Parts
Joerns beds contain various parts that wear from






Joerns Healthcare to be defective.




be returned for inspection at the customer’s expense.

Service

Maintenance Department with assistance from the
Joerns Healthcare Product Service Department.

the customer’s expense.


Joerns Healthcare Product Service Department can


the equipment.
1
Manufactured By:
Joerns Healthcare LLC
2100 Design Rd. Ste 100
Arlington, TX 76014
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16

Joerns WeCare Full-Electric Bed Owner's manual

Type
Owner's manual

Joerns WeCare Full-Electric Bed is a versatile and feature-rich bed designed to provide comfort, safety, and convenience for users. Its electric adjustability allows for effortless changes in height, backrest angle, and leg rest angle, enabling users to find the optimal position for rest, relaxation, or medical procedures. The bed's sturdy construction and 450-pound weight capacity ensure stability and support, while its floor locks and swivel locks enhance safety during transfers and use.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI