Hollywog Wireless Remote Control Pain Relief for Lower Back, WiTouch User manual

  • Hello! I am an AI chatbot trained to assist you with the Hollywog Wireless Remote Control Pain Relief for Lower Back User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Hollywog
2830 Amnicola Hwy
Chattanooga, TN 37406 USA
+1 423 305 7778
www.hollywog.com
® and ©2013 Hollywog, LLC. All Rights Reserved
Patents Pending
Patente Pendiente
Brevets en cours d’homologation
Zum Patent angemeldet
User Manual
Manual de Usuario
Manuel d’utilisation
Benutzerhandbuch
What is Hollywog
®
WiTouch
®
?
Congratulations on your purchase of the Hollywog WiTouch device. Hollywog WiTouch is the first wireless remote controlled
pain relief device incorporating TENS technology to specifically target back pain. The thin and flexible design perfectly contours
the back for maximum surface contact. The advanced electronics design maximizes energy use, providing over 150 30-minute
treatment sessions per battery life. This innovative device is safe, drug-free, easy to use, discreet & comfortable to wear, and
most importantly allows you to control your pain to maintain an active lifestyle.
What is TENS?
TENS (Transcutaneous Electrical Nerve Stimulation) is a non-invasive, drug-free method of controlling pain by the transmission of
electrical stimulation to the body’s nerve or muscle fibers through the skin. Pain relief or suppression is achieved by:
•Suppressingthetransmissionofpaininnerves(GateControlTheory)and
•Increasingthedischargeofthebody’sendorphins,whichreducesthesensitivitytopaininthecentralnervoussystem
(Opiate Release Theory)
The Hollywog WiTouch’s exclusive and patented waveform incorporates BOTH clinical theories of TENS to provide pain relief.
The Hollywog WiTouch provides a safe, easy to apply, pain management system for self-treatment when needed.
INDICATIONS FOR USE
The Hollywog WiTouch is indicated for temporary relief of pain associated with sore and aching muscles in the lower back due to
strain from exercise or normal household and work activities.
CONTRAINDICATION
Do not use this device if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device.
Such use could cause electric shock, burns, electrical interference, or death.
WARNINGS
• Donotusethisdeviceacrossorthroughyourchestbecausetheelectricalcurrentsintroducedintothechestmaycauserhythm
disturbances to your heart, which may be lethal.
• Donotusethisdeviceifyouaresusceptibletorhythmdisturbancestotheheartunlessunderthedirectionofyourphysician.
• Donotusethisdeviceoveryourneckbecausethiscouldcauseseveremusclespasmsresultinginclosureofyourairway,
difficulty in breathing, or adverse effects on heart rhythm or blood pressure.
• Consultwithyourphysicianbeforeusingthisdeviceifyouareinthecareofaphysician.
• Consultwithyourphysicianbeforeusingthisdeviceifyouhavehadmedicalorphysicaltreatmentforyourpain.
• Stopusingthisdeviceandconsultyourphysicianifyourpaindoesnotimprove,becomesmorethanmild,orcontinuesformore
than five days.
• Donotusethisdevicewhiledriving,operatingmachinery,orduringanyactivityinwhichelectricalstimulationcanputyouat
risk of injury.
• Donotusethisdeviceoveropenwoundsorrashes,oroverswollen,red,infected,orinamedareasorskineruptions(e.g.,
phlebitis, thrombophlebitis, varicose veins);
• Donotusethisdeviceover,orinproximityto,cancerouslesions.
• Donotusethisdeviceonchildrenbecauseithasnotbeenevaluatedforpediatricuse.
• Donotusethisdeviceinthepresenceofelectronicmonitoringequipment(e.g.,cardiacmonitors,ECGalarms),whichmaynot
operate properly when the electrical stimulation device is in use.
• Donotusethisdevicewheninthebathorshower.
• Donotusethisdevicewhilesleeping.
• Donotusethisdeviceonabnormalskin,orskinthatisnotintact,clean,orhealthy.
PRECAUTIONS
• DONOTSTARTstimulationofthedevicepriortoapplicationofthedevicetotheback.
• Keepthisdeviceoutofthereachofchildren.
• Thesafetyofnervestimulationhasnotbeenestablishedduringpregnancy;thereforedonotusethisdeviceifyouare
pregnant, or suspect that you are pregnant, unless under the direction of your physician.
• Thisdeviceshouldnotbeappliedonoracrossyourheadsincetheeffectsofstimulationofthebrainareunknown.
• Thisdeviceisforsymptomatictreatmentand,assuch,suppressesthesensationofpainthatwouldotherwiseserveasa
protective mechanism.
• Ifyouhavesuspectedordiagnosedheartdisease,youshouldfollowprecautionsrecommendedbyyourphysician.
• Ifyouhavesuspectedordiagnosedepilepsy,youshouldfollowprecautionsrecommendedbyyourphysician.
• Usethisdevicewithcautionifyouhaveatendencytobleedinternally,suchasfollowinganinjuryorfracture.
• Consultwithyourphysicianpriortousingthisdeviceafterarecentsurgicalprocedure,becausestimulationmaydisruptthe
healing process.
• Donotusethisdeviceforpainofcentralorigin,includingheadache.
• Thisdevicedoesnotprovidecurativevalue.
• Thelong-termeffectsofnervestimulationareunknown.
• Youmayexperienceskinirritationorhypersensitivityduetotheelectricalstimulationoradhesivemedium(gelpads).
• Usethisdevicewithcautionifstimulationisappliedoverthemenstruatingorpregnantuterus.
• Usethisdevicewithcautionifstimulationisappliedoverareasofskinthatlacknormalsensation.
• Usethisdeviceonlywiththegelpadsandaccessoriesrecommendedbythemanufacturer.
•Donotremovethisdevicefromyourskinduringstimulationtreatment.
• Donotplaceyournger,oranyobject,betweenornearyourskinandtheadhesivegelpadswiththestimulationactivated.
• Thisdeviceisnottobeusedinthepresenceofammableoranesthesiagasses.
ADVERSE REACTIONS
• Isolatedcasesofskinirritationorburnsmayoccurduetoelectricalstimulationoradhesivemedium(gelpads).
• Stopusingthedeviceandconsultwithyourphysicianifyouexperienceadversereactionsfromuseofthisdevice.
CONTENTS
• 1HollywogWiTouchdevice
• 1HollywogWiTouchremotecontrol
• 1PackageHollywogWiTouchgelpads(5pair)
• 2AAAalkalinebatteries
• 1CR2032lithiumbattery(installedinremote)
• 1Size#0Phillipsheadscrewdrivertool
• 1StorageBox
• 1UserManual
• 1QuickStartGuide
PREPARE your Hollywog WiTouch
1. Prepare the treatment area by washing and drying the skin thoroughly. Excessive hair should be trimmed (NOT shaved), and
the treatment area should be void of oils and/or lotions. DO NOT apply the main device over broken skin.
2. Remove battery cover from the device with the included screwdriver tool and insert 2 AAA batteries noting correct +/- polarity.
Replace the battery cover.
3. Press and hold the ON/OFF button on the WiTouch device for one (1) second. The LED will begin blinking green once every
two(2seconds).YourWiTouchdeviceisnowONandreadyforuse.
4. Remove clear battery insulation tab from the remote control to activate battery. Discard clear tab.
5. Remove two (2) WiTouch gel pads from the package and separate the pads at the perforation in the liner.
6. Remove the blue liners from one (1) gel pad by peeling away the folded edges of the liners. Discard the blue liners.
DO NOT remove the green protective liner during this step.
7. Apply the gel pad to one (1) black electrode area on the bottom of the Hollywog WiTouch device. Align the shape of the gel
pad to match the shape of the black electrode area. Firmly press the gel onto the electrode across the entire surface of the gel
pad to ensure good adhesion.
8. Repeat steps 6 - 7 for the second gel pad and electrode area.
9. YourHollywogWiTouchisnowreadyforuse.DONOTactivatestimulationofthedevicepriortoapplicationofthedevice
to the back.
OPERATE your Hollywog WiTouch
1. Remove the green protective liners from both gel pads on the device by slowly peeling the liner from one inside corner towards
the outside corner of the electrode area. DO NOT discard the protective liners as the protective liners will be reapplied to the
gel pads for storage after use. Take care to not touch or allow objects to come in contact with the gel pad as this may interfere
with the adhesion properties of the gel pad.
2. Placethedeviceonyourbackovertheareaofbackpain,aligningthecenterofthedeviceoverthespine.Makesurethegel
padsmakegoodcontactwiththeskin.DONOTapplythegelpads/electrodesdirectlyoverthespine.Iftheareaofbackpain
cannotbeaccessed,requestassistancefromanotherpersonforapplication.
3. OntheHollywogWiTouchremotecontrol,presstheSTART/STOPTREATMENTbuttonmomentarily.TheLEDwillblink
green one (1) time. The Hollywog WiTouch electrical stimulation treatment has now started. On the WiTouch device,
the LED will now begin blinking green very rapidly for the duration of the treatment.
4. PresstheINCREASEINTENSITY(+)buttontoincreasethestimulationtoacomfortablelevel.Todecreasethestimulation
intensityanytimeduringtreatment,presstheDECREASEINTENTISY(-)button.
5. Treatment will automatically stop after 30 minutes. Treatment can be stopped at any time by pressing the START/STOP
TREATMENTbuttonontheremotecontrol.DONOTremovethedeviceuntiltreatmenthasstopped.
6. CareshouldbetakentonotinadvertentlydepresstheON/OFFbuttononthedevicewhenbeingworn.IftheON/OFFbuttonis
depressed during use, the treatment will stop. To restart treatment, press and hold the ON/OFF button for 1 second, then
initiate the treatment with the remote control. The 30 minute treatment will start over from the beginning.
7. The Hollywog WiTouch device can be left in position without removal for multiple treatments during the day. To initiate
additional treatments, repeat step 3 above when treatment is desired. NOTE:Itisrecommendedtowaitaminimumof30
minutes between treatments.
8. The Hollywog WiTouch device will automatically turn off after two (2) hours of inactivity. To turn on the device, press
and hold the ON/OFF button on the WiTouch device for one (1) second. The LED will begin blinking green once every two (2
seconds).OntheHollywogWiTouchremotecontrol,presstheSTART/STOPTREATMENTbuttonmomentarily.TheLEDwill
blink green one (1) time. The Hollywog WiTouch electrical stimulation treatment has now started. On the WiTouch
device, the LED will now begin blinking green very rapidly for the duration of the treatment..
STORE your Hollywog WiTouch
1. When you have completed treatments and ready to store the Hollywog WiTouch, you can remove the device from your
back. Slowly peel the WiTouch device away from the skin, grasping the edge of the device and gel pad simultaneously to
ensure gel pad does not come off the device and stay on the skin.
2. Applythegreenprotectivelinerstothegelpads.Makesurethegelpadiscompletelycovered.
3. Press and hold the ON/OFF button on the device for 1 second to turn off the device. The LED will will stop blinking green.
4. Place the Hollywog WiTouch device and remote control in the storage box.
5. IftheHollywogWiTouchisnottobeusedforalongperiodoftime,removethe2AAAbatteriesfromtheHollywog
WiTouch battery compartment to prolong battery life.
6. To turn on the device, press and hold the ON/OFF button on the WiTouch device for one (1) second. The LED will begin
blinking green once every two (2 seconds). On the Hollywog WiTouch remote control, press the START/STOP
TREATMENTbuttonmomentarily.TheLEDwillblinkgreenone(1)time.TheHollywogWiTouchelectricalstimulation
treatment has now started. On the WiTouch device, the LED will now begin blinking green very rapidly for the duration of
the treatment.
REPLACE Hollywog WiTouch Gel Pads
HollywogWiTouchGelPadsaresinglepersonuse.Thegelpadswilllastforapproximately2-5applications.Thenumberof
applications will vary person-to-person depending upon skin type, oils and PH levels. When the gel pads no longer adhere completely,
it is time to replace the gel pads. To purchase replacement gel pads for your WiTouch device contact your original purchase retailer.
1. Slowly peel the gel pad away from the black electrode area of the Hollywog WiTouch device. Be sure to remove all gel
residues from the electrode area.
2. Remove two (2) gel pads from the package and separate the pads at the perforation in the liner.
3. Remove the blue liners from one (1) gel pad by peeling away the folded edges of the liners. Discard the blue liners. DO NOT
remove the green protective liner at this time.
4. Align the shape of one (1) gel pad to match the shape of the black electrode area. Apply the gel pad to the black electrode
area, firmly pressing the gel onto the electrode. Press across the entire surface of the gel pad to ensure good adhesion.
5. Repeat step 3 and 4 for the second gel pad and electrode area.
6. Follow instructions for operation or storage of your Hollywog WiTouch
MAINTAIN your Hollywog WiTouch
• TheHollywogWiTouchLEDindicatorwillalertwhenbatteriesrequirereplacement:
1. Hollywog WiTouch device:
 •
When pressing ON/OFF button to turn on device: LED indicator blinks yellow every two (2) seconds when the device is ON.
 •
During stimulation treatment: LED indicator will blink yellow rapidly during the treatment.
2. Hollywog WiTouch remote control: LED indicator blinks yellow once when the START/STOP button is pressed.
•ReturntheHollywogWiTouchtoitsstorageboxwhennotinusetoensurethedeviceisprotectedduringstorageand
transport.
•TheHollywogWiTouchshouldbeoperated,transportedandstoredattemperaturesbetween50°Fand104°F(10°C
and40°C),withrelativehumiditybetween30%-85%.Allvalueshave+/-10%tolerance.
•Consultyourlocalauthoritiesforproperbatterydisposal.
•IftheHollywogWiTouchisnotworkingproperly,stopusingimmediatelyandcontactHollywogCustomerService.
•Donotdisassembleormodifythisdevice.
CLEAN your Hollywog WiTouch
Hollywog WiTouch Electrodes are single person use. Use a damp cloth and mild soap to gently wipe clean. Remove any soap residue
fromcleaning.Damagetothedeviceand/orremotecontrolwilloccurifsubmergedinwaterorotherliquids.WiTouchismanufactured
with water detection technology to detect exposure to water that may cause damage to the electronics and void the product warranty.
TECHNICAL SPECIFICATIONS
Channels: Single channel
Waveform:Asymmetricbiphasicsquarepulse.
Pulse Amplitude: 0 ~110 mA = 0 ~ 55 volts, adjustable (at 500 ohm load)
PulseFrequency:(Hz)5-120
Pulse Width: (μs) 120 - 240
Timer Control: (mins) 30
Power Supply:
Device:Two(2)AAAbatteries(InternallyPowered)
RemoteControl:One(1)CR2032Lithiumbattery(InternallyPowered)
Size (D x W x H): 0.7” x 7.5” x 3.5” (18 mm x 191 mm x 90 mm)
Weight (including battery): 4.8 oz (136 g)
SafetyStandards:IEC/EN/UL60601-1,IEC/EN60601-1-2,IEC60601-2-10,CAN/CSAC22.2No.601.1,EN300200/EN301498-3
MadeinUSA
LED Indicator Guide
Event
Low battery - device
Low battery - remote control
Device transition from OFF to ON
Device transition from ON to OFF
Device transition from ON to START
STIMULATION mode
On remote control, press of
START/STOP TREATMENT button to
either START or STOP treatment
Device at maximum intensity and
+/- button pressed on remote or device
Remote control communication error
LED Color
Yellow
Yellow
Green
None
Green
Green
Yellow
Yellow
LED Sequence
One (1) blink every two (2)
seconds when the device is ON.
Rapid blinking during the
stimulation treatment.
One (1) blink every time the
START/STOP button is pressed.
One (1) blink every two (2)
seconds.
Green blinking stops.
Rapid blinking during treatment.
One (1) blink when START/STOP
button is pressed.
Two (2) blinks each time button
is pressed.
One (1) rapid blink, followed by
two (2) rapid blinks.
This device complies with part 15 of the
FCC Rules. Operation is subject to the
following two conditions: (1) This device
may not cause harmful interference,
and (2) this device must accept any
interference received, including
interference that may cause undesired
operation.
RF:2.4GHztransceiver
Device:
FCCID:N8L-HWOGTENS11
IC:10346A-HWOGTENS11
Remote Control:
FCCID:N8L-HWOGTENS12
IC:10346A-HWOGTENS12
0120
European
Conformity
European
Directive
2002/96/EC
85%
30%
Atmospheric
Pressure Limitations
Temperature
Limitations
Humidity Limitations
Limitations of Use and Storage
MDSS GmbH
Schiffgraben 41
30175 Hannover, Germany
European Authorized Representative:
TROUBLESHOOTING
YourHollywogWiTouchdevicecomesfromthefactorypairedwiththeremote.YouDONOThavetopairthedeviceandremote
priortouse.Intherareinstancetheremoteanddevicebecomeunpairedandwirelesscommunicationbetweenthemstop,follow
these steps to pair the units:
1. Remove the two (2) AAA batteries from the device battery compartment
2. On the remote control, press and hold both the + and – buttons for five (5) seconds. The LED indicator will blink green,
one (1) blink every second.
3.Insertthetwo(2)AAAbatteriesintothedevicebatterycompartment.
The device and the remote control will pair automatically, and the remote control LED indicator will blink green five (5) times.
WARRANTY
Hollywog, LLC (“Company”) warrants the Hollywog WiTouch (“Product”) is free of defects in materials and workmanship. This
warrantyshallremainineffectfor1year(12months)fromthedateofpurchase.IftheseProductsfailtofunctionduringthe1year
warranty period due to a defect in material or workmanship, at the Company’s Option, Company or the selling dealer will replace
the respective Product without charge within a period of thirty (30) days from the date on which the Product is returned to the
Company or the dealer.
This Warranty Does Not Cover:
HollywogGelPads
Any defects, damage, loss, malfunction or failure in the Product caused by product misuse, alteration, disassembly or any use
thatisinconsistentwiththeProduct’sInstructionManual.
COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
¿Que es Hollywog
®
WiTouch
®
?
Felicitaciones por su compra del dispositivo Hollywog
®
WiTouch. El dispositivo Hollywog WiTouch es el primer dispositivo
controladoporremotoinalámbricoparaelaliviodedoloryqueincorporalatecnologíadeTENSparaapuntarespecícamente
el dolor de espalda. El diseño delgado y flexible perfectamente perfila la espalda para el contacto superficial máximo. El diseño
electrónicoavanzadoutilizaalmaximizolaenergía,produciendomásde150tratamientosde30minutosporcargadebatería.
Este dispositivo innovador es seguro, sin medicina, fácil de usar, discreto y cómodo para llevar puesto, y más importante, permite
queustedcontrolesudolor,paramantenerunestilodevidaactiva.
¿Qué es TENS?
TENS (Estimulación Trascutánea Eléctrica del Nervio) es un no invasivo, sin medicina de controlar el dolor por la transmisión del
estímuloeléctricoalasbrasdemúsculoonerviodelcuerpoporlapiel.Elaliviodedolorolasupresiónesadquiridapor:
•Supresióndelatransmisióndedolorennervios(TeoríadeControlGate)y
•Aumentoeldescargodeendornasdelcuerpo,quereducelasensibilidaddedolorenelsistemanerviosocentral(Teoría
Liberación de Opiatos)
LaformadeondaexclusivaypatentadadeHollywogWiTouchincorporalasDOSteoríasclínicasdeTENSparaproporcionar,alivio
de dolor. El Hollywog WiTouch proporciona un sistema seguro, fácil de aplicar, mantenimiento de dolor para tratamiento personal
cuando necesario.
INDICACIONES PARA USO
HollywogWiTouchestáindicadoparaelaliviotemporaldeldolorasociadoconmúsculosdoloridosenlapartebajadelaespalda
debido a tensiones de ejercicios, actividades caseras o de empleo.
CONTRAINDICACIÓN
Nouseestedispositivosiustedtieneunmarcapasocardíaco,undesbriladorimplantado,uotrodispositivometálicoo
electrónicoimplantado.Talusopodríacausarsacudidaeléctrica,quemaduras,interferenciaeléctrica,omuerte.
ADVERTENCIAS
• Estedispositivonosedebeusaratravésoporsupechoporquelascorrienteseléctricasintroducidasenelpechopueden
causarperturbacionesderitmoasucorazón,quepuedeserletal.
• Estedispositivonosedebeusarsiustednoessusceptibleaperturbacionesderitmoalcorazónamenosqueestebajola
dirección de su médico.
• Estedispositivonosedebeusarsobresucuelloporqueestepodríacausarespasmosdemúsculoseverosquecausanelcierre
desuvíaaérea,dicultadenrespiración,oefectosadversosenritmodecorazónotensiónarterial.
• Consultesumédicoantesdeusarestedispositivosiustedestábajoelcuidadodeunmédico.
• Consulteconsumédicoantesdeusarestedispositivosiustedhatenidoeltratamientomédicoofísicoporsudolor.
• Dejedeusarestedispositivoyconsulteasumédicosisudolornomejora,sinosehacemássuave,osiguecondolormásde
cincodías.
• Estedispositivonosedebeusarmientrasconduce,trabajaconmaquinariadeoperaciones,odurantecualquieractividaden
queunestímuloeléctricopuedaponerloenpeligro.
• Estedispositivonosedebeusarsobreheridasabiertasoerupciones,osobreáreasaumentadas,rojas,infectadasoinamadas
o erupciones de piel (p.ej, Flebitis, Tromboflebitis, venas varicosas);
• Estedispositivonosedebeusar,oenproximidada,lesionescancerosas.
• Estedispositivonosedebeusarenniñosporquenohasidoevaluadoparaelusopediátrico.
• Estedispositivonosedebeusarenlapresenciadelequipodemonitoreselectrónico(p.ej,monitorescardíacos,alarmasde
ECG),nofuncionancorrectamentemientrasseutilizaeldispositivodeestímuloeléctrico.
Manual de Usuario
/