12
OPERATING GUIDELINES
C.AUTOMATEFUNCTION
The system is equipped with a 5 stage automated hot air rework
profiling system. To access and change the time and temperature
settingsoftheprofilefollowtheguidebelow:
1. Ensurethatthehotairgunfunctionswitchisdeactivated.
2. TheHotAirGunAutomateFunctionSelector(“17”fromthecontrol
panel) allows us to scroll thru thedifferentstages of the profile.
ThestagesusethesymbolsA,b,c,dandE,toindicatestages1to
5respectively.TheAutomateDisplay(“18”formthecontrolpanel)
shows the desired temperature for each stage and the Automate
Display (“16” form the control panel) shows thedurationatwhich
thesettemperatureistobemaintained.
3. NowpresstheHotAirGunAutomateFunctionSelector,thedisplay
willchangeto“###A,”and“###t”.ThesuffixAindicatesthatwe
are now adjustingthe desired temperature for stage 1.The suffix
“t”indicatesthetimeinsecondsthedesiredtemperatureistobe
maintained.
4. UsetheHotAirGunTemperatureAdjustmentButtonstoselectthe
desired temperature for this stage. Use the Hot Air Gun Airflow
Adjustment Buttons to select the desired duration. Adjust the
temperature/durationoftheotherstages.SimplypresstheHotAir
Gun Automate Function Selector to scroll to the next stage. Then
usetherespectivekeystoadjusttemperatureandduration.
5. To save the newly inputted profile into memory, repeatedly press
theHotAirGunAutomateFunctionSelectoruntilthedisplayshows
“SAUE”. Press the up button of the Hot Air Gun Temperature
AdjustmentButtontoconfirmsavingoftheprofile.
21
H.UtilizingtheSolderIronDigitalTemperatureCalibration
By default, the system is properly calibrated but for some cases
when a little adjustment of the soldering iron temperature is required
thefollowingprocedurecanbedone.
1. Turnonthesolderingironfunctionswitch.
2. Settoappropriatetemperatureyouwanttocalibrate.Placethetip
ofthesolderingirononanexternaltemperaturemeter.
3. Thereadingsontheexternaltemperaturesensorshouldbemore
orlessequaltothedisplayedtemperature.
4. Iftherearelargediscrepancyinthetemperaturereadingwecan
recalibrate the temperature setting. First write down the set
temperature of the soldering iron and the actual temperature
readingfromtheexternaltemperaturemeter.Forexample:
settemperature=350
externaltemperature=300
5. TurnofftheSolderingIronFunctionswitch(“14”fromthecontrol
panel).EnsurethattheDesolderingGunFunctionswitchisinthe
offposition(“12”fromthecontrolpanel).
6. Press and hold the UP button of the Desoldering Gun
TemperatureAdjustmentbutton(“5”fromthecontrolpanel).
7. The Soldering IronTemperatureDisplay (“11”fromthecontrol
panel).Willswitchtofourzeros“000”indicatingitisnowinthe
soldering iron calibration mode. Release the UP button of the
DesolderingGunTemperatureAdjustmentbutton
8. UsetheSolderingIronTemperatureAdjustmentbuttons(“4”from
thecontrolpanel)toincreaseordecreasethecalibrationvalues.
In our example the set temperature is 350 but the actual
temperatureis300,Thereisadifferenceof50degrees.Pressthe
upbuttonuntilwereach“050”.
9. Confirm thechangeby pressing and holding the“UP”button of
the DesolderingGun TemperatureAdjustmentbutton (“5” from
thecontrolpanel).
DIGITAL CALIBRATION