4 T-901/T-903/T-905 Instruction Bulletin
3. Set master pressure at the extreme opposite
value of that used for Step 1.
4. Turn the lower adjusting screw until the control
line value of Step 2 is obtained. Make a note of
the dial setting at this pressure.
5. The number of F degrees rotation of the lower
dial should equal the degree F setting on the
readjusting range dial. If not, follow Steps 6, 7,
and 8.
6. Loosen the readjusting dial and tighten it at the
value obtained in Step 5.
7. Turn the readjusting dial and post to the original
desired value.
8. Repeat Steps 1 through 5 until the readjusting
dial setting equals the actual readjusting range
obtained from the readings.
Adjusting the Set Point
Referring to the example above, the set point of the
submaster thermostat is 140 F when the master
pressure is 0 psi.
1. Set the master pressure to its minimum value,
e.g. 0 psi.
2. Set the sensitivity selector at a position
determined by experience, or about 1/3 of the
distance up from the inner pivots. The sensitivity
should be as high as possible without producing
undue “hunting.”
3. Check the temperature at the measuring
element.
4. Turn the lower adjusting screw until the mid-point
control pressure of the controlled device
operating range is obtained. For example, 6 psi
is the mid-point of a valve having an operating
range of 4.8 psi.
5. Loosen the lower dial and turn it to indicate the
temperature found at the measuring element.
Tighten the dial in this position.
6. Turn the set point adjusting screw until the dial
setting corresponds to the set point determined
by the master pressure.
Referring to the example, this would be 140°F at
0 psi master pressure.
Adjusting the Stops
The procedure for adjusting the stops is
shown by another example.
MASTER PRESSURE TEMPERATURE
5 psi 60°F
10 psi 53°F
This means that 5 psi change in master pressure
must readjust the temperature 60 - 53 = 7 degrees.
In order to use the readjusting dial which is calibrated
for 15 psi maximum master pressure it is necessary
to know how much a 15 psi change in master
pressure will readjust the instrument. Using the
above example, 15/5 x 7 = 21 degrees. The
readjusting dial is then set at 21 degrees. (Refer to
the procedure for Checking the Range of Remote
Adjustment.) Proceed as follows:
1. Adjust the master pressure to 10 psi.
2. Turn the set point adjusting screw until the
control line pressure is at some intermediate
value, for example, 7 psi. Make sure that the
stop screws are not touching the stop plate.
3. Turn the stop screw for the upper pressure
clockwise until it just touches the plate. (This will
be indicated by a change in control air pressure.)
4. Turn the same stop screw counterclockwise until
the former intermediate value is obtained.
5. Repeat this procedure for the other stop screw.
6. Adjust the master pressure to 5 psi.
7. Turn the set point adjusting screw until the
control line pressure is at some intermediate
value as in Step 2.
8. Adjust the stop screws for the lower pressure
limit by repeating the procedure described for the
stop screw for the upper pressure limit, Steps 2
through 5.
Controls Group
507 E. Michigan Street
P.O. Box 423 Printed in U.S.A.
Milwaukee, WI 53202