General GPIB Information 1-6 MG369xC GPIB Operation
MG369xC SCPI PM PN: 10370-10375 Rev. E 1-9
1-6 MG369xC GPIB Operation
All Series MG369xC Synthesized Signal Generator functions, settings, and operating modes (except for
power on/standby) are controllable using commands sent from an external controller via the GPIB. When in
the remote (GPIB) mode, the signal generator functions as both a listener and a talker. The GPIB interface
function capability of the MG369xC is listed in Table 1-2 on page 1-7.
Setting GPIB Operating Parameters
The MG369xC leaves the factory with the GPIB address value set to 5 and the data delimiting terminator set
to carriage return and line feed (CR/LF). A different address value can be entered from the front panel using
the Configure GPIB menu. Using this same menu, the data delimiting terminator can be changed to carriage
return (CR) only. Refer to the Series MG369xC Synthesized Signal Generator Operation Manual for the
procedure.
Selecting the Interface Language
Series MG369xC Synthesized Signal Generators with the SCPI option can be remotely operated using one of
two external interface languages—Native or SCPI. The Native interface language uses a set of MG369xC GPIB
product specific commands to control the instrument; the SCPI interface language uses a set of the Standard
Commands for Programmable Instruments (SCPI) commands to control the unit. Selecting which of these
external interface languages is to be used can be made from the front panel using the Configure GPIB menu.
Refer to Section 2-4 “SCPI Interface Language Selection” on page 2-6.
Response to GPIB Interface Function Messages
Table 1-3 lists the GPIB interface function messages that the MG369xC will recognize and respond to. With
the exception of the Device Clear and Selected Device Clear messages, these messages affect only the operation
of the MG369xC GPIB interface. The signal generator’s response for each message is indicated.
Interface function messages are transmitted on the GPIB data lines and interface management lines as either
unaddressed or addressed commands. The manner in which these messages are invoked in programs is
implementation dependent. For programming information, refer to the documentation included with the GPIB
interface used for the external controller.
Table 1-3. MG369xC Response to GPIB Interface Function Messages
Interface Function Message
Addressed
Command MG369xC Response
Device Clear (DCL)
Selected Device Clear (SDC)
No
Yes
Resets the MG369xC to its default state. (Equivalent to
sending the *RST command.)
Go To Local (GTL) Yes Returns the MG369xC to local (front panel) control.
Group Execute Trigger (GET) Yes Executes a string of commands, if programmed.
Interface Clear (IFC) No Stops the MG369xC GPIB interface from listening or talking.
(The front panel controls are not cleared.)
Local Lockout (LLO) No Disables the front panel menu RETURN TO LOCAL soft key.
Remote Enable (REN) No Places the MG369xC under remote (GPIB) control when it has
been addressed to listen.
Serial-Poll Enable (SPE) No Outputs the serial-poll status byte.
Serial-Poll Disable (SPD) No Disables the serial-poll function.
Parallel-Poll Configure (PPC) Yes Responds to a parallel-poll message (PPOLL) by setting
assigned data bus line to the logical state (1,0) that indicates
its correct SRQ status.
Parallel-Poll Unconfigure (PPU) No Disables the parallel-poll function.