RKI Instruments 65-2515RK Owner's manual

Type
Owner's manual
www.rkiinstruments.com
65-2515RK
Oxygen Detector
Operator’s Manual
Part Number: 71-0137RK
Revision: A
Released: 7/7/20
2 • 65-2515RK Oxygen Detector
WARNING
Read and understand this instruction manual before operating detec-
tor. Improper use of the detector could result in bodily harm or death.
Periodic calibration and maintenance of the detector is essential for
proper operation and correct readings. Please calibrate and maintain
this detector regularly! Frequency of calibration depends upon the
type of use you have and the sensor types. Typical calibration frequen-
cies for most applications are between 3 and 6 months, but can be
required more often or less often based on your usage.
65-2515RK Oxygen Detector 3
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in
materials, workmanship, and performance for a period of one year from date of shipment from
RKI Instruments, Inc. Any parts found defective within that period will be repaired or
replaced, at our option, free of charge. This warranty does not apply to those items which by
their nature are subject to deterioration or consumption in normal service, and which must be
cleaned, repaired, or replaced on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or
repair procedures not in accordance with the operator’s manual. This warranty indicates the
full extent of our liability, and we are not responsible for removal or replacement costs, local
repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES
AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER
OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC.
INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS,
INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR
DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR
FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors, dealers,
and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this
gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
a) Absorbent cartridges d) Batteries
b) Pump diaphragms and valves e) Filter elements
c) Fuses
4 • 65-2515RK Oxygen Detector
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Oxygen Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Junction Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Mounting the Oxygen Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Wiring the Oxygen Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Introducing Incoming Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Verifying/Setting the Fresh Air Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Replacing the Oxygen Plug-in Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Replacing the Oxygen Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Preparing for Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Setting the Fresh Air Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Setting the Zero Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Returning to Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
65-2515RK Oxygen Detector 5
Overview
This manual describes the 65-2515RK oxygen detector (internal amplifier type). This manual
also describes how to install, start up, maintain, and calibrate the detector when used with a
gas monitoring controller. A parts list at the end of this manual lists replacement parts and
accessories for the oxygen detector.
Specifications
Table 1 lists specifications for the 65-2515RK oxygen detector.
NOTE: The following symbol on the detector label is a caution to the user to refer to this
documentation for installation and operation instructions:
WARNING: When using the 65-2515RK, you must follow the instructions and warnings in
this manual to assure proper and safe operation of the 65-2515RK and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the 65-2515RK as described in this manual.
Table 1: 65-2515RK Specifications
Target Gas Oxygen
Sampling Method Diffusion
Detection Range 0 - 25.0% oxygen
Accuracy ± 0.5% O
2
Oxygen Detector Signal Output 0 mV at 0% volume oxygen nominal
18 mV at 25% oxygen nominal
Response Time 90% in 30 seconds
Area Classification Explosionproof for Class I, Groups B, C, and D
Temperature Code T6
Installation Category Installation Category 1. Signal level, special equipment or
parts of equipment, telecommunication, electronic, etc.,
with smaller transient overvoltages than Installation
Category (Overvoltage Category) II (ref. IEC 664).
Operating Temperature -20° C to 45° C
!
6 • 65-2515RK Oxygen Detector
Description
This section describes the components of the 65-2515RK. It consists of the oxygen detector
and the junction box
Oxygen Detector
The oxygen detector consists of the detector housing and the plug-in sensor.
Figure 1: Oxygen Detector Component Location
Detector Housing
The detector housing protects the sensing components within the housing. Use the mounting
threads at the top of the housing to screw the oxygen detector into the bottom conduit hub of
the junction box. Use the removable cap near the bottom of the housing to access the sensor
for maintenance or replacement. The cap protects the sensor from damage and includes a
flame arrestor which contains any sparks that may occur within the detector housing. A cap
gasket seals the interface between the housing body and cap. The flame arrestor guard protects
the flame arrestor from impact damage and is permanently bonded to the cap.
Two color coded wires extend from the top of the detector housing. Use these wires to connect
the oxygen detector to a controller via the terminal block in the junction box. The detector
housing includes two sockets installed on a circuit board. These sockets accept the plug-in
sensor’s two pins to secure the sensor within the detector housing. The circuit board with the
sockets conditions the sensor’s signal before it exits the detector housing via the two color
coded wires.
Plug-In Sensor
The plug-in sensor plugs into a circuit board in the detector housing with two pins and is
retained by the cap. Through a series of chemical and electrical reactions, the sensor produces
an electrical output that is proportional to the detection range of the oxygen detector.
Detector
Housing Body
Cap Gasket
Detector Housing
Cap
Flame Arrestor
Guard
Oxygen Sensor
65-2515RK Oxygen Detector 7
Junction Box
The junction box allows you to install the oxygen detector at a mounting site that is remote
from a controller and it protects the detector wiring connections. Two conduit hubs allow you
to mount the oxygen detector to the junction box and route the wiring from the detector to a
controller. Three spacers installed on the back of the junction box control the distance of the
junction box from a mounting surface and ensure that there is enough room to install a
calibration cup on the detector during calibration.
A terminal block within the junction box facilitates the wiring process. A cover on the front
of the junction box allows access to the interior of the junction box.
Installation
This section describes procedures to mount the oxygen detector in the monitoring
environment and wire the detector to a controller.
Mounting the Oxygen Detector
1. Select a mounting site that is representative of the monitoring environment. Consider the
following when you select the mounting site.
Select a site where the detector is not likely to be bumped or disturbed. Make sure
there is sufficient room to perform start-up, maintenance, and calibration procedures.
Select a site that is at normal breathing level.
8 • 65-2515RK Oxygen Detector
Figure 2: 65-2515RK Outline & Mounting Dimensions
NOTE: The oxygen detector in the 65-2515RK is normally provided with an Adalet
XIHSCFC3 junction box rated explosion proof for Class I, Groups B, C, and D. This
combination is shown in Figure 2 above. Any junction box with an internal volume
less than or equal to 69 cubic inches and rated explosion proof for Class I, Groups B,
C, and D may be used for this detector.
2. At the mounting site you select, hang or mount the junction box with the detector facing
down (see Figure 2).
Rubber Spacers,3X
.38
.75
3.2 Max
3/4 NPT Conduit
Hub
11/2-20Thread for
Calibration Cup
3.62
8.45 Max
2.7
65-2515RK Oxygen Detector 9
Wiring the Oxygen Detector to a Controller
WARNING: If the controller’s construction is not explosion-proof, the controller must be
installed in an ordinary location. Do non install a non-explosion-proof
controller in a classified location. Turn off the controller.
3. Turn off or unplug power to the controller.
4. Remove the junction box cover.
5. Guide a two-conductor, shielded cable or two wires in conduit through the unused conduit
hub of the junction box.
WARNING: To maintain the explosion proof classification of the oxygen detector/junction
box combination, a conduit seal must be used within 2 inches (environment
with acetic acid) or 18 inches (environment without acetic acid) of the detector
junction box conduit hub used for wiring to the amplifier. Adhere to your local
electrical code when installing the conduit seal.
6. Connect the wires to the terminals opposite the detector leads using the terminal strip in
the junction box.
CAUTION: If using shielded cable, leave the drain wire insulated and disconnected at the
detector. You will connect the opposite end of the cable’s drain wire at the
controller.
7. Reinstall the junction box cover to the junction box.
8. Route the cable or wires leading from the oxygen detector through one of the conduit hubs
at the controller housing.
CAUTION: Do not route power and detector wiring through the same conduit hub. The
power cable may disrupt the transmission of the detector signal to the controller.
9. Connect the wires to the applicable controller terminal strip as shown in Figure 3 below.
See the controller operator’s manual and the controller’s detector head specification sheet
for wiring information specific to a controller.
10 • 65-2515RK Oxygen Detector
Figure 3: Wiring the Oxygen Detector to a Controller
10. If using shielded cable, connect the cable’s drain wire to an available chassis ground at the
controller. RKI controllers typically have a ground stud that is a convenient grounding
location.
Install conduitsealwithin 18
inches (for environments
without acetic acid) or2
inches (for environments with
acetic acid)
Green (-)
White(+)
Cable
Shield
Terminal Strip
White
Green
Oxygen Detector
Controller Housing
(installed inordinary location if
device is non-explosion-proof)
Controller
Oxygen Detector
Terminals
65-2515RK Oxygen Detector 11
Start Up
This section describes procedures to start up the oxygen detector and place the detector into
normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power wiring to the controller is correct and secure. Refer to the controller
operator’s manual.
3. Turn on or plug in the incoming power to the controller, then turn on the controller.
4. Verify that the controller is on and operating properly. Refer to the controller operator’s
manual.
CAUTION: Allow the detector to warm up for 5 minutes before you continue with the next
section, “Verifying/Setting the Fresh Air Reading”.
Verifying/Setting the Fresh Air Reading
WARNING: Do not remove the detector housing cap or junction box cover while the
circuits are energized unless the area is determined to be non-hazardous. Keep
the detector housing cap and junction box cover tightly closed during
operation.
NOTE: If you can verify that the detector is in a fresh air environment (environment known to
be of normal oxygen content and free of toxic and combustible gases), it is not
necessary to apply zero air when verifying or setting the fresh air reading.
The procedure below describes applying zero air using a calibration kit that includes a
calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob.
RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
1. Screw the calibration cup onto the bottom of the oxygen detector.
2. Screw the regulator into the zero air calibration cylinder.
3. Use the sample tubing to connect the regulator to the calibration cup.
4. Turn the regulator’s on/off knob counterclockwise to open it. Gas will begin to flow.
5. Allow the gas to flow for two minutes.
12 • 65-2515RK Oxygen Detector
6. Verify a reading of 20.9% oxygen at the controller.
If the display reading is 20.9% oxygen, the oxygen detector is in normal operation and
start up is complete. Proceed to step 7 to disassemble the calibration kit.
If the display reading is not 20.9% oxygen, set the fresh air reading according to the
controller’s operator’s manual.
7. Turn the regulator’s on/off knob clockwise to close it.
8. Unscrew the regulator from the zero air calibration cylinder. For convenience, leave the
sample tubing connected to the regulator and the calibration cup.
9. Unscrew the regulator from the cylinder and remove the calibration cup from the detector.
10. Store the components of the calibration kit in a safe and convenient place.
Maintenance
This section describes maintenance procedures. It includes preventive maintenance,
troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance
of the oxygen detector. It includes daily and quarterly procedures.
Daily
Verify a display reading of 20.9% oxygen at the controller. Investigate significant changes in
the reading.
Quarterly
Calibrate the detector as described in “Calibration” on page 15.
65-2515RK Oxygen Detector 13
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for
problems you may encounter with the oxygen detector.
NOTE: This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.
Replacing the Oxygen Plug-in Sensor
1. Turn off the controller.
2. Turn off or unplug incoming power to the controller.
3. Unscrew the detector housing cap from the detector housing.
4. Unplug and remove the oxygen sensor.
5. Carefully plug the replacement sensor into the socket pattern that is located in the top
section of the detector housing.
NOTE: Match the sensor’s male pins with the two female sockets as you plug the sensor into
the sockets.
Table 2: Troubleshooting the Oxygen Detector
Condition Symptom(s) Probable Causes Recommended Action
Fail
Condition
Controller indicates
a fail condition.
The detector wiring is
disconnected or
misconnected.
The plug-in oxygen
sensor is not properly
installed.
The detector is
malfunctioning.
1. Verify that the detector wiring is
correct and secure.
2. Confirm that the plug-in oxygen
sensor inside the detector housing is
installed properly.
3. Calibrate the detector.
4. If the fail condition continues,
replace the plug-in oxygen sensor.
5. If the fail condition continues,
contact RKI for further instruction.
Slow or No
Response/
Difficult or
Unable to
Calibrate
Unable to accurately
set the fresh air or
zero reading during
calibration.
Detector requires
frequent calibration.
Note: Under
“normal”
circumstances, the
detector requires
calibration once every
three months.
Some applications may
require a more frequent
calibration schedule.
The calibration
cylinder is low, out-
dated, or defective.
The plug-in sensor is
malfunctioning.
1. Verify that the calibration cylinder
contains an adequate supply of a
fresh test sample.
2. If the calibration/response difficulties
continue, replace the plug-in sensor.
3. If the calibration/response difficulties
continue, contact RKI for further
instruction.
14 • 65-2515RK Oxygen Detector
6. Screw the detector housing cap onto the detector housing.
7. Turn on or plug in incoming power to the controller.
8. Turn on the controller.
CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with
the next step.
9. Calibrate the replacement sensor as described in “Calibration” on page 15.
Replacing the Oxygen Detector
NOTE: In most cases, it is only necessary to replace the plug-in oxygen sensor.
1. Turn off the controller.
2. Turn off or unplug incoming power to the controller.
3. Remove the junction box cover.
4. Disconnect the detector leads from the terminal block inside the junction box. Note the
position of the color-coded leads as you remove them.
5. Unscrew the detector from the junction box.
6. Guide the detector leads of the replacement detector through the bottom conduit hub of the
junction box, then screw the mounting threads of the detector into the conduit hub.
7. Wire the detector leads to the terminal block in the same position as the leads you removed
in step 4.
8. Reinstall the junction box cover.
9. Turn on or plug in incoming power to the controller.
10. Turn on the controller.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue
with the next step.
11. Calibrate the replacement detector as described in “Calibration” on page 15.
65-2515RK Oxygen Detector 15
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 to 6 months is adequate for most applications. Unless
experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a
calibration frequency of every 3 months (quarterly).
If an application is not very demanding, for example detection in a clean, temperature
controlled environment, and calibration adjustments are minimal at calibration, then a
calibration frequency of every 6 months is adequate.
If an application is very demanding, for example if the environment is not well controlled,
then more frequent calibration than every 3 months may be necessary.
Calibration
This section describes how to calibrate the oxygen detector. It includes procedures to prepare
for calibration, set the fresh air reading, set the zero reading, and return to normal operation. It
describes calibration using a calibration kit that includes a calibration cup, calibration gas,
sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc.
recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
WARNING: The controller is not an active gas monitoring device during the calibration
procedure.
Preparing for Calibration
NOTE: Calibrating the oxygen transmitter may cause alarms. Be sure to put the controller
into its calibration program or disable external alarms before calibrating.
1. Screw the calibration cup onto the bottom of the oxygen detector.
2. Screw the regulator into the zero air calibration cylinder.
3. Use the sample tubing to connect the regulator to the calibration cup.
NOTE: If you can verify that the oxygen detector is in a fresh air environment, you do not
need to apply zero air to the detector before adjusting the zero reading.
4. Put the controller into its calibration program.
Setting the Fresh Air Reading
1. Follow the directions in the controller operator’s manual for setting the fresh air reading.
2. When the instructions call for applying zero air to the detector, turn the regulator’s on/off
knob counterclockwise to open it. Zero air will begin to flow.
3. Allow the gas to flow for two minutes.
16 • 65-2515RK Oxygen Detector
4. Set the fresh air reading according to the controller operator’s manual.
5. Turn the regulator’s on/off knob clockwise to close it.
6. Unscrew the regulator from the zero air calibration cylinder. Leave the sample tubing
connected to the regulator and the calibration cup.
Setting the Zero Reading
1. Screw the regulator into the calibration gas cylinder, normally 100% nitrogen.
2. Follow the instructions in the controller’s operator’s manual for setting the oxygen zero
reading.
3. When the instructions call for exposing the detector to gas, normally 100% nitrogen, turn
the regulator’s on/off knob counterclockwise to open it. Gas will begin to flow.
4. Allow gas to flow to the detector for two minutes.
5. Set the zero reading according to the controller operator’s manual.
6. Turn the regulator’s on/off knob clockwise to close it.
7. Unscrew the regulator from the cylinder.
8. Remove the calibration cup from the detector. For convenience, leave the calibration cup
and regulator connected by the sample tubing.
Returning to Normal Operation
1. Allow about 45 seconds for the oxygen reading to increase above the decreasing alarm
point and return the controller to normal operation.
NOTE: If you do not allow the oxygen reading to increase above the decreasing alarm point,
then unwanted alarms may occur.
2. Verify that the controller display reading increases and stabilizes at 20.9% oxygen.
3. Store the components of the calibration kit in a safe and convenient place.
65-2515RK Oxygen Detector 17
Parts List
Table 3 lists replacement parts and accessories for the oxygen detector.
Table 3: Parts List
Part Number Description
06-1248RK Sample tubing (3/16 in. x 5/16 in.; specify length when ordering)
18-0400RK-01 Junction box with spacers
65-1025RK Oxygen replacement sensor, plug-in
65-2514RK Oxygen replacement detector assembly (includes plug-in sensor), CSA
classified (does not include junction box)
65-2515RK Oxygen detector/junction box, CSA classified
71-0137RK 65-2515RK Oxygen Detector Operator’s Manual (this document)
81-0076RK-01 Zero air calibration cylinder, 34 liter steel
81-0078RK Calibration cylinder, 100% nitrogen, 17 liter
81-0078RK-01 Calibration cylinder, 100% nitrogen, 34 liter steel
81-1050RK Regulator with gauge and knob, 0.5 LPM, for 17- and 34-liter steel
calibration cylinders (cylinders with external threads)
81-1117RK Calibration cup
81-F301RK-LV Calibration kit (34 liter)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17

RKI Instruments 65-2515RK Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI