65-2515RK Oxygen Detector • 15
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 to 6 months is adequate for most applications. Unless
experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a
calibration frequency of every 3 months (quarterly).
If an application is not very demanding, for example detection in a clean, temperature
controlled environment, and calibration adjustments are minimal at calibration, then a
calibration frequency of every 6 months is adequate.
If an application is very demanding, for example if the environment is not well controlled,
then more frequent calibration than every 3 months may be necessary.
Calibration
This section describes how to calibrate the oxygen detector. It includes procedures to prepare
for calibration, set the fresh air reading, set the zero reading, and return to normal operation. It
describes calibration using a calibration kit that includes a calibration cup, calibration gas,
sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc.
recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
WARNING: The controller is not an active gas monitoring device during the calibration
procedure.
Preparing for Calibration
NOTE: Calibrating the oxygen transmitter may cause alarms. Be sure to put the controller
into its calibration program or disable external alarms before calibrating.
1. Screw the calibration cup onto the bottom of the oxygen detector.
2. Screw the regulator into the zero air calibration cylinder.
3. Use the sample tubing to connect the regulator to the calibration cup.
NOTE: If you can verify that the oxygen detector is in a fresh air environment, you do not
need to apply zero air to the detector before adjusting the zero reading.
4. Put the controller into its calibration program.
Setting the Fresh Air Reading
1. Follow the directions in the controller operator’s manual for setting the fresh air reading.
2. When the instructions call for applying zero air to the detector, turn the regulator’s on/off
knob counterclockwise to open it. Zero air will begin to flow.
3. Allow the gas to flow for two minutes.