Haier HYC-580 User manual

Type
User manual

This manual is also suitable for

HYC-940/HYC-580
Operation Manual
Read the Operation Manual carefully before using your appliance
.
Keep the Operation Manual in a safe place.
Appearance, color and la
y
out of the door ma
y
var
y
.
Pharmaceutical Refrigerator
Product
Model
Volume
External panel
Internal panel
Pharmaceutical refrigerator
HYC-940
940L
Sprayed steel sheet
Sprayed steel sheet
HYC-580
580L
Insulation
Rack
Cooling
Compressor
Condenser
Evaporator
Rrfrigerant
Defrosting
Thermostat
Over-temp warning
Low-temp warning
Door warning
Spare storage parts
Lamp
Temp recorder
Accessories
Net weight
Rechargeable battery
Styrene polyurethane foam (Freon-free)
6
Air circulation
Sealed
Finned pipes
Finned pipes
R134a
Automatic
Electronic control system
Indicator blinks for buzzer and remote warning
Indicator blinks, without deferred buzzer alarm
Door warning indicator on, and buzzer signal 10 minute later
Permanent static storage
Lamp, 15Wx2
Without
250kg
UN7.0-8(8V7.0AH)
Key: 1 set; Recording paper: 1 box; Shelf restrainer Screw: 4
Operation manual: 1; Plastic bag: 1; 9V Battery: 4
LYGX (Z) No.: 2580088
YZB/LU 0028-2004
6
178kg
Lamp, 15Wx1
LWC
930W 420W
2~8 .
2~8 .
2~8 .
HYC-940 Diagram
Inner light
Rack
Caster
s
Light
switch
Control
pannel
Door lock
3
4.Rack
There are 6 racks inside.
5.Inner light
8
10 min later
10 min
Temperature adjusting:
Press and hold both SENSOR SELECT button and CALIB CANCEL button for more than 3s until
original setting starts flashing to indicate that it enters temperature setting mode. With each press
of CALIB CANCEL button, temperature setting increases 0.1 until reaching upper limit, which
is defaulted as 8 . If press again, it returns to lower limit, which is defaulted as 2 . It repeats the
above cycle. When it reaches your desired temperature, if you don't press the button within 5s,
the temperature setting will be saved and it automatically quits the temperature setting mode (the
temperature setting range programmed at factory is 2 - 8 ). The display returns to display inside
sensor temperature or average temperature before setting is made (not flashing).
2~8 .
8
in side
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10

Haier HYC-580 User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI