19
ENGLISH
SPECIFICATION AND DIMENSION OF EACH PART
Specifications are subject to change without notice.
The symbol ~ means alternating current, and the symbol means direct current.
Specifications
Detector
Item Specifications
Model 10HQ701G
Sensor Type a-Si TFT
Scintillator Type CsI:TI
Total Pixel Matrix 2048 x 1792pixels
Total Pixel Area 286.72mm x 250.88mm
Pixel Pitch 140um
Effective Pixel Matrix 2008 x 1752pixels
A/D Conversion 16bits
Data transmission 802.11 a/b/g/n/ac Wireless LAN Standard, 150Mbps
Wired Gigabit Ethernet Standard, 500Mbps
Cycle time Typ. 4.5Sec (Wired)
Typ. 5Sec (Wireless)
Image Transmission Typ. 1.5Sec (Wired)
Typ. 2Sec (Wireless)
Image Save Stores up to 200 images
Semi-dynamic mode 5fps @ 140um (full FOV)
Energy Range 40kVp ~ 150kVp
MTF Typ. 84% @ 0.5lp/mm
DQE Typ. 66% @ 0.1lp/mm
Item Specifications
Size (Width x Height x Depth) (mm) 333 x 282 x 15.6
Weight (kg) Typ. 1.9
Window Materials Carbon Fibre
Trigger Mode Manual Mode
Auto Mode (Auto Exposure Detection)
Power Consumption Typ.24W
Wireless Standard:
802.11 a/b/g/n/ac compliance
Peak Mode: 867Mbps
Frequency: 2.4GHz / 5GHz
Bandwidth: 20MHz / 40MHz / 80MHz
MIMO: 2 x 2
Rating 24V 2.1A
Applied part Type: BF
Location: The front side of the Detector (Effective area only)
NOTE
• Maximum wireless signal rate derived from IEEE standard specifications. Actual data throughput will vary. Network
conditions and environmental factors, including volume of network traffic, building materials and construction,
and network overhead, lower actual data throughput rate.
• Recommended Maximum operable distance: 2m (From the Access Point)
• Wireless antennas: The module adopts the latest 802.11ac technology. The transmitter of the module is powered by
host equipment (Detector). The antennas are 2 printed-dipole antennas.
• Wireless module: 802.11 a/b/g/n/ac USB2.0 module is implemented. It supports 2T2R (2 transmit 2 receive) MIMO
technology, which delivers throughput up to 300Mbps.
• Images can be saved by the X-ray generator while the power of the detector is turned on without connecting to a
PC. To produce images, X-ray is irradiated at intervals of more than 10 seconds. Check and load the saved images
from LG Acquisition Workstation Software.