InstructionNumber:5697
Component Description
RESIDENTIALCOMPARATORTEST
DPDCHLORINE(FREE&COMBINED)
(.5‐5ppm)
INTHEPRESENCEOFMONOPERSULFATE
COMPONENTS
1x4028 Pipet,Calibrated0.5mL,plastic(eyedropper)w/cap
1x5697 Instruction
1x9198 SampleTube,Graduated,25mL,plasticw/cap
1xR‐0867 DeoxReagent
NOTE:Thepresenceofpotassiumperoxymonopersulfateinwaterwill
interferewithDPDchlorinetests.UseR‐0867andthe
instructionsprovidedinthiskit,incombinationwithR‐0001,
R‐0002,R‐0003,andtheResidentialSeriescomparatorfor
chlorinetodeterminefreeandcombinedchlorinewhenusing
monopersulfate.
PROCEDURE:
1. Rinseandfillsampletube(#9198)to10mLmarkwithwatertobetested.
2. Add5dropsR‐0001,5dropsR‐0002andQUICKLYswirltomix.IMMEDIATELY
add0.5mLR‐0867andQUICKLYswirltomix.Poursampleintothe
chlorine/brominecelltomarkandmatchcolorIMMEDIATELY.Savetheextra
samplewaterleftintube.Recordaspartspermillion(ppm)free
chlorine(FC).IMMEDIATELYpourbackintosampletubeandadd5drops
R‐0003.Swirltomix.Pourthissampleintothechlorine/brominecell
tomarkandmatchcolor.Again,therewillbeadditionalsamplewaterleft
intubewhichcannowbediscarded.Recordaspartspermillion(ppm)
totalchlorine(TC).
3. SubtractFCfromTC.Recordaspartspermillion(ppm)combinedchlorine.
Formula:TC‐FC=CC.
8/05
©2016TaylorTechnologies,Inc.•AllRightsReserved•ViewPrivacyPolicy