  • Hello! I am an AI chatbot trained to assist you with the Taylor Technologies K-1520 Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
InstructionNumber:5697
Component Description
RESIDENTIALCOMPARATORTEST
DPDCHLORINE(FREE&COMBINED)
(.5‐5ppm)
INTHEPRESENCEOFMONOPERSULFATE
COMPONENTS
1x4028 Pipet,Calibrated0.5mL,plastic(eyedropper)w/cap
1x5697 Instruction
1x9198 SampleTube,Graduated,25mL,plasticw/cap
1xR‐0867 DeoxReagent
NOTE:Thepresenceofpotassiumperoxymonopersulfateinwaterwill
interferewithDPDchlorinetests.UseR‐0867andthe
instructionsprovidedinthiskit,incombinationwithR‐0001,
R‐0002,R‐0003,andtheResidentialSeriescomparatorfor
chlorinetodeterminefreeandcombinedchlorinewhenusing
monopersulfate.
PROCEDURE:
1. Rinseandfillsampletube(#9198)to10mLmarkwithwatertobetested.
2. Add5dropsR‐0001,5dropsR‐0002andQUICKLYswirltomix.IMMEDIATELY
add0.5mLR‐0867andQUICKLYswirltomix.Poursampleintothe
chlorine/brominecelltomarkandmatchcolorIMMEDIATELY.Savetheextra
samplewaterleftintube.Recordaspartspermillion(ppm)free
chlorine(FC).IMMEDIATELYpourbackintosampletubeandadd5drops
R‐0003.Swirltomix.Pourthissampleintothechlorine/brominecell
tomarkandmatchcolor.Again,therewillbeadditionalsamplewaterleft
intubewhichcannowbediscarded.Recordaspartspermillion(ppm)
totalchlorine(TC).
3. SubtractFCfromTC.Recordaspartspermillion(ppm)combinedchlorine.
Formula:TC‐FC=CC.
8/05