5. Installing sensor
• Ensure all sensor hole and threads are clean
• A chamfered cable channel or cable hole prevents
damaged cables
• Check hole with tool 3.050.243 (included accessory)
8. Functional test
On completion of mold assembly, test sensor insulation and
sensitivity.
• Ensure that connector is dry
• Use cleaning spray Type 1003
8.1 Insulation test
Use insulation tester Type 5495 to check sensor insulation.
Resistance >1013 Ω.
8.2 Sensor functional check
Test the sensitivity of the installed sensor using Test Set Type
5495.
Approx. measured value: ±8 % of sensitivity as stated cali-
bration sheet or identification plate.
9. Service and repair
Piezoelectric pressure sensors are maintenance-free. Insulation
and sensitivity of the installed sensor should, however, be
checked after each molding run.
• Ensure that connector is dry
• Cover sensor hole during cleaning
• Use cleaning spray Type 1003
9.1 Insulation test
See chapter 8.1.
9.2 Sensor functional check
See chapter 8.2.
9.3 Removing sensor
Remove sensor with extraction tool Type 1358A or use a flat
ejector pin to press on the sensor front.
• Do not use a hammer on the sensor front
• Do not use a sharp center punch
9.4 Cable repair
Damaged cables cannot be repaired. Replacement cable can
be mounted in the Kistler factory.
9.5 Repairs at Kistler
Factory repairs at Kistler are arranged by the local sales office.
Information: www.kistler.com
9.6 Disposal instructions for electrical and electronic
equipment
Do not discard old electronic instruments
in municipal trash. For disposal at end of
life, please return this product to an autho-
rized local electronic waste disposal service
or contact the nearest Kistler Instrument
sales office for return instructions.
5.1 Sensor installation with spacer sleeve Type 6464A3
Install sensor into hole with extraction tool Type 1358A or
directly with spacer sleeve.
5.1.1 Machining spacer sleeve
Cut spacer sleeve so that it has an overlength of 2 … 3 mm.
Place sensor and sleeve into hole and determine exact length,
ensuring a clearance of 0,01 … 0,03 mm. Sleeve must be
ground so the end is square and planar. Remove any burrs.
For multi-cavity applications each spacer sleeve should be
numbered and allocated to a sensor.
5.1.2 Spacer sleeve and retaining plate
Install sensor and spacer sleeve. Ensure a clearance of 0,01 …
0,03 mm.
Sensor may not be preloaded when assembling the mold.
• Do not preload sensor
• Use extraction tool Type 1358A to install sensor
• For multi-cavity applications each spacer sleeve
should be numbered
Kistler Group
Eulachstrasse 22
8408 Winterthur, Switzerland
Tel. +41 52 224 11 11
Fax +41 52 224 14 14
info@kistler.com www.kistler.com
6189A_002-434e-11.17 Kis120 © 2009...2017 Kistler Group
6.1.1 Cut/grip connection
Cut the single-wire cable to length and do not strip the
insulation. Loosen the knurled connector by hand without
removing the rear section.
When using the temperature amplifier Type 2205A... strip
thermocouple cables. Attach green cable (+) to terminal and
white cable to (–) terminal.
Installation of connector Type 1839 and mounting plate
Installation of multichannel connector Type 1722A4... and
temperature amplifier Type 2205A...
This installation is for sensor Type 6189AG only: A typical
installation uses up to 4 sensors.
Assemble 4-channel connector and temperature amplifier in
adapter plate Type 5700A25 and then install in the recess.
6.1.2 Installing single-wire connector and temperature con-
nector with Mounting Plate
Install connector and mounting plate in recess. Attach cap
using one of the screws. Place caps on open connectors.
• Take care not to damage the gray single-wire cable
when stripping the green sensor cable sheath
• Use sheath stripper Type 1367 to remove green
sheathing
• Refer to Instruction leaflet 002-450
6. Cable and connector installation
• Check again that edges of cable duct are chamfered
• Use a metal plate to cover open channels
• Attach cap and place on connector
6.1 Single-wire technique
• Cables of p/T Sensors consist of one single-wire
cable for pressure and two thermocouple braids
• Single-wire cable must be completely enclosed in
the mould
• Single-wire cable may not be ducted together with
power cables
• Ensure that all contact surfaces and threads are clean
and dry
7. Installing identification plate
Rivet plate to side of mold or fit with four M2,5 screws.
p + T Sensor
Type 6189A0,4
SN 1234567
Sensitivity -6,5 pC/bar
Measuring range 2000 bar
Thermocouple Type K
Kistler Instrumente AG
Winterthur Switzerland
Made in Switzerland
25
75
__________________
_______________
_______________
_________________________
_________________________
____________ ___ ___
Kistler Instrumente AG
Winterthur Switzerland
Made in Switzerland
Zeile 1
Zeile 3 Zeile 4
Zeile 6 Zeile 7
Zeile 2 Zeile 5
Zeile 8
Beschriftungsbeispiele:
Schriftart: Helvmed
Schrifthöhe: 1,8mm
Text Laserbeschriftet
äquivalente Schriftart (z.B. Arial) zulässig
Kistler Instrumente AG
Winterthur Switzerland
Made in Switzerland
Force Sensor
Type 9223A0,2
SN 1234567
Sensitivity -4,3 pC/N
Measuring range bar
Measuring pin
Ø
mm
Kistler Instrumente AG
Winterthur Switzerland
Made in Switzerland
Pressure Sensor
Type 6183CCGQ100
SN 1234567
Sensitivity -2,273 pC/bar
Measuring range 2000 bar
ohne Einheitsempfindlichkeit
Pressure Sensor
Type 6157BAE
SN 1234567
Sensitivity -9,4 pC/bar
Measuring range 2000 bar
Kistler Instrumente AG
Winterthur Switzerland
Made in Switzerland
mit Einheitsempfindlichkeit -9,4 pC/bar
Typenschild 75x25x0,5 mit Beschriftung
A4
c
1 / 1
1:1
measure. analyze. innovate.
Erstmals verwendet
Erste Proj.-Nr.
Werkstoff
Änderung
Datum
Massstab
Ersatz für
Kopie Datum
100010525 27.01.2016
gez.
gepr.
ges.
26.01.2016
Beo
Hs
27.01.2016
Hoh
Hs
27.01.2016
100.239.050
Zeichnungs-Nr.
Material-Nr.
KIWAG-SWX_A4h
Ver.
Bl.
9223A__
551456213.0255
55147987
interne und externe Fertigung
Das Urheberrecht an dieser Zeichnung, die dem Empfänger
persönlich anvertraut wird, verbleibt unserer Firma.
Ohne unsere schriftliche Genehmigung darf die Zeichnung
weder kopiert noch vervielfältigt, noch an Drittpersonen
mitgeteilt oder zugänglich gemacht werden.
27.01.2016
100.239.050 - c - YDI - Mat.-Zeichnung - interne und externe Fertigung - Freigegeben --- 55147987 --- 14-Sep-2017 09:48 (CET) -
[email protected]