Type 8770A Impedance Head Acceleration-Force Sensor Page 11
B08.8770Ae-04.98
6. Operation
Piezoelectric sensors are self generating transducers. While the piezoelectric
sensor does not require an external power source, it is necessary to provide
power to the internal electronic impedance converter. This section is intended
to provide the user with the information necessary to ensure accurate
measurements with Impedance Head sensors. Topics to be covered include
powering, signal conditioning, frequency response and driving long cables.
6.1 Powering
6.1.1 Using “Built In” Power Sources
Many FFT analyzers and vibration monitors are available with internal
power supplies. Often identified as "ICP" or "Constant Current Sources",
these internal power supplies are generally compatible with Kistler
Impedance Head sensors.
Before using the built-in power source, compare the measurement
instrument current source specifications with the current and voltage
specifications of the impedance to be used. If the instrument power is
within the range required by the sensor there should be no problem with
compatibility. If the user plans to drive long cables (over 430 feet/130 m)
the guidelines in Section 6.2 should be followed. The impedance head
low frequency response may be affected by the input impedance of the
measuring instrument as discussed in Section 6.4.4.
Caution: Many industrial monitors have adjustable current controls.
Exceeding the maximum current rating of any sensor may cause
permanent damage and void the warranty.
6.1.2 Kistler Couplers
Kistler offers a comprehensive line of power supply/couplers and dual mode
amplifiers. If the user is utilizing these instruments, they are designed for
compatibility with Impedance Head sensors. The chart below provides a
listing of available power supply/couplers. Please request data sheets on
specific models.
Type 5108 Passive coupler. Customer provides power.
Type 5114A Battery powered or AC-DC adapter single channel coupler. Meter monitors
bias and battery condition.
Type 5118B Battery and AC-DC Adapter powered single channel coupler with LEDs to
monitor bias. Plug-in filters available.
Type 5120 AC powered single channel coupler with AC/DC output coupling. Meter
monitors bias and offset. Factory adjustable gain and optional low pass
filters.
Type 5124A/A1
5126A/A1 Line powered, rack mounted, 12 and 6 channel couplers. A1 version offers
internally adjustable gain and optional plug-in filters.
Type 5134A 4-channel AC powered coupler and amplifier. Selectable high pass filters
and gain. Random noise generator for testing with FFT analyzer. Remote
control and sensing via RS-232C. CE certified.
Type 5350 Transducer simulator for trouble shooting or checking system performance.
It provides an ideal means to confirm transducer location or detect
inoperative cables. Connects directly to the input of any Kistler coupler or
any ICP compatible power source. Output frequencies of 100 Hz or 1000
radians/second, sine or square wave with adjustable amplitude.
Table 4: Kistler Coupler/Power Supplies and Accessories