Rosemount T1055 Clarity II Turbidimeter Owner's manual

Type
Owner's manual
Clarity II™ Turbidimeter
Turbidity Measurement System
Instruction Manual
PN 51-T1055/rev.L
October 2008
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Your purchase from Rosemount Analytical, Inc. has
resulted in one of the finest instruments available for
your particular application. These instruments have
been designed, and tested to meet many national and
international standards. Experience indicates that its
performance is directly related to the quality of the
installation and knowledge of the user in operating and
maintaining the instrument. To ensure their continued
operation to the design specifications, personnel
should read this manual thoroughly before proceeding
with installation, commissioning, operation, and main-
tenance of this instrument. If this equipment is used in
a manner not specified by the manufacturer, the pro-
tection provided by it against hazards may be impaired.
Failure to follow the proper instructions may cause
any one of the following situations to occur: Loss of
life; personal injury; property damage; damage to
this instrument; and warranty invalidation.
Ensure that you have received the correct model
and options from your purchase order. Verify that
this manual covers your model and options. If not,
call 1-800-854-8257 or 949-757-8500 to request
correct manual.
For clarification of instructions, contact your
Rosemount representative.
Follow all warnings, cautions, and instructions
marked on and supplied with the product.
Use only qualified personnel to install, operate,
update, program and maintain the product.
Educate your personnel in the proper installation,
operation, and maintenance of the product.
Install equipment as specified in the Installation
section of this manual. Follow appropriate local and
national codes. Only connect the product to electri-
cal and pressure sources specified in this manual.
Use only factory documented components for repair.
Tampering or unauthorized substitution of parts and
procedures can affect the performance and cause
unsafe operation of your process.
All equipment doors must be closed and protective
covers must be in place unless qualified personnel
are performing maintenance.
If this equipment is used in a manner not specified
by the manufacturer, the protection provided by it
against hazards may be impaired.
Equipment protected throughout by double insulation.
Installation of cable connections and servicing of this
product require access to shock hazard voltage levels.
Main power and relay contacts wired to separate power
source must be disconnected before servicing.
Do not operate or energize instrument with case open!
Signal wiring connected in this box must be rated at
least 240 V.
Non-metallic cable strain reliefs do not provide grounding
between conduit connections! Use grounding type bush-
ings and jumper wires.
Unused cable conduit entries must be securely sealed by
non-flammable closures to provide enclosure integrity in
compliance with personal safety and environmental pro-
tection requirements. Unused conduit openings must be
sealed with NEMA 4X or IP65 conduit plugs to maintain
the ingress protection rating (NEMA 4X).
Electrical installation must be in accordance with the
National Electrical Code (ANSI/NFPA-70) and/or any
other applicable national or local codes.
Operate only with front and rear panels fastened and in
place over terminal area.
Safety and performance require that this instrument be
connected and properly grounded through a three-wire
power source.
Proper relay use and configuration is the responsibility
of the user.
This product generates, uses, and can radiate radio fre-
quency energy and thus can cause radio communication
interference. Improper installation, or operation, may
increase such interference. As temporarily permitted by reg-
ulation, this unit has not been tested for compliance within
the limits of Class A computing devices, pursuant to
Subpart J of Part 15, of FCC Rules, which are designed to
provide reasonable protection against such interference.
Operation of this equipment in a residential area may cause
interference, in which case the user at his own expense, will
be required to take whatever measures may be required to
correct the interference.
This product is not intended for use in the light
industrial, residential or commercial environments
per the instrument’s certification to EN50081-2.
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2008
WARNING
RISK OF ELECTRICAL SHOCK
CAUTION
WARNING
C
QUICK START GUIDE
FOR CLARITY II TURBIDIMETER
1. Refer to Section 2.0 for installation instructions.
2. The sensor cable is pre-wired to a plug that inserts into a receiving socket in the analyzer. The cable also passes
through a strain relief fitting. To install the cable…
a. Remove the wrenching nut from the strain relief fitting.
b. Insert the plug through the hole in the bottom of the enclosure nearest the sensor socket. Seat the fitting in the
hole.
c. Slide the wrenching nut over the plug and screw it onto the fitting.
d. Loosen the cable nut so the cable slides easily.
e. Insert the plug into the appropriate receptacle on the circuit board.
f. Adjust the cable slack in the enclosure and tighten the cable nut. For the wall/pipe mount version, be sure to
leave sufficient cable in the enclosure to avoid stress on the cable and connections.
g. Plug the cable into the back of the sensor.
h. Place the sensor in either the measuring chamber or the calibration cup. The sensor must be in a dark place
when power is first applied to the analyzer.
3. Make power, alarm, and output connections as shown in the drawing below.
CONTINUED ON THE FOLLOWING PAGE
Power, Alarm, and Output Connections
Panel Mount version
The release clip on the sensor plug
faces the top of the analyzer enclosure.
Wall/Pipe Mount version
The release clip on the sensor
plug faces the user.
D
6. Choose the desired language. Move the cursor to >> and press ENTER to show
more choices.
7. Choose the number of sensors. This screen will be displayed only for dual input
analyzers.
8. Choose Turbidity or TSS (total suspended solids). If you choose TSS you must
enter a calibration curve. Refer to Section 6.5.
9. Choose units for turbidity (NTU, FTU, FNU) or TSS (ppm, mg/L, none).
10. If you have a dual input analyzer the screen at left appears. Repeat steps 8 and
9 for the second sensor.
11. The main display appears. The initial turbidity reading will be 0.000 NTU. Over
the next 60 seconds the reading will gradually reach a final value. The error in
the displayed value may be as great as 20%. For best results, the sensor must
be calibrated. See Section 6.0.
The outputs and alarms are assigned to default values. To change settings,
refer to Section 5.0, Programming the Analyzer. To reinstall factory settings and
return to Quick Start, see Section 5.9.
# of sensors?
One
Two
Sensor2 is for:
Turbidity
TSS
Sensor1 is for:
Turbidity
TSS
Units?
NTU
FTU FNU
Units?
ppm
mg/L none
4. Once connections are secured and verified, apply power to the analyzer.
5. When the analyzer is powered up for the first time Quick Start screens appear. Using Quick Start is easy.
a. A blinking field shows the position of the cursor.
b. Use the or key to move the cursor left or right. Use the or key to increase or decrease the value of a
digit. Use the or key to move the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT also returns the
display to the language selection screen.
English
Franais
Espaol >>
QUICK REFERENCE GUIDE
MENU TREE FOR TURBIDITY/TSS MEASUREMENTS
About This Document
This manual contains instructions for installation and operation of the Clarity II Model
T1055 Turbidimeter.
The following list provides notes concerning all revisions of this document.
Rev. Level
Date Notes
A 5/04 This is the CD-launch version containing only installation
information.
B 6/04 This is the electronic launch version, which added more detail
instructions for programming and troubleshooting.
C 11/04 This is the initial full release of the product manual. The manual
has been reformatted to reflect the Emerson documentation
style and updated to reflect any changes in the product offering.
D 1/05 Updated ordering matrix, added lamp calibration section,
revised information screens section.
E 3/05 Revised panel mount drawing.
F 7/05 Revised text on pp. 32 & 34; revised text & figure on page 46.
G 10/05 Revised text on pp 9 & 47 and drawings on pp. 10 & 48 to
show new molded debubbler; revised debubbler specifications.
Addition of agency-required warnings to pp. 12, 43, 45, & 48.
H 1/06 Revised Analyzer Enclosure Specifications - page 2.
I 5/06 Added FM and CSA Non-Incendive approval ratings to
Specifications - Analyzer, page 2.
Added FM and CSA Non-Incendive installation control drawings
to section 3.0 - Wiring, pp. 12-15.
Added wetted materials to debubbler specifications, page 2.
Changed part numbers for replacement lamp boards, page 50.
J 7/06 Added troubleshooting advice for “SN Warning”, pp. 11, 36, 38,
40, 51, 52.
K 4/07
L 10/08 PN number changed on page 50 for #9 on replacement parts.
i
MODEL CLARITY II TABLE OF CONTENTS
MODEL CLARITY II TURBIDIMETER
TABLE OF CONTENTS
Section Title Page
1.0 DESCRIPTION AND SPECIFICATIONS .............................................................................. 1
1.1 Features and Applications ...................................................................................................... 1
1.2 Specifications ......................................................................................................................... 2
2.0 INSTALLATION...................................................................................................................... 3
2.1 Unpacking and Inspection ...................................................................................................... 3
2.2 Installation — Analyzer........................................................................................................... 3
2.3 Installation — Flow Chamber and Debubbler......................................................................... 7
2.4 Installation — Sensor ............................................................................................................. 9
2.5 Sample Point .......................................................................................................................... 9
3.0 WIRING .................................................................................................................................. 10
3.1 Preparing Conduit Openings .................................................................................................. 10
3.2 Power, Alarm, Output, and Sensor Connections ................................................................... 10
3.3 Non-Incendive Installation Control Drawings ......................................................................... 12
4.0 DISPLAY AND OPERATION ................................................................................................. 16
4.1 Display.................................................................................................................................... 16
4.2 Keypad ................................................................................................................................... 16
4.3 Programming and Calibrating the Solu Comp II - Tutorial...................................................... 17
4.4 Security .................................................................................................................................. 18
4.5 Using Hold.............................................................................................................................. 18
5.0 PROGRAMMING THE ANALYZER ....................................................................................... 19
5.1 General................................................................................................................................... 19
5.2 Changing StartUp Settings ..................................................................................................... 19
5.3 Configuring and Ranging the Outputs .................................................................................... 22
5.4 Configuring Alarms and Assigning Setpoints.......................................................................... 23
5.5 Choosing Turbidity or Total Suspended Solids....................................................................... 26
5.6 Choosing Single Sensor or Dual Sensor Input....................................................................... 29
5.7 Setting a Security Code.......................................................................................................... 30
5.8 Noise Rejection ...................................................................................................................... 30
5.9 Resetting Factory Default Settings......................................................................................... 31
5.10 Selecting a Default Screen, Language, and Screen Contrast................................................ 32
6.0 CALIBRATION....................................................................................................................... 34
6.1 Introduction............................................................................................................................. 34
6.2 Calibrating Against a User-Prepared Standard ...................................................................... 34
6.3 Calibrating Against a Commercial Standard........................................................................... 37
6.4 Calibrating the Turbidity Sensor Against a Grab Sample ....................................................... 39
6.5 Lamp Calibration .................................................................................................................... 41
6.6 Using the Dry Check Standard ............................................................................................... 42
6.7 Entering a Turbidity to TSS Conversion Equation .................................................................. 43
6.8 Calibrating Current Outputs.................................................................................................... 45
7.0 MAINTENANCE .................................................................................................................... 46
7.1 SoluComp II Analyzer............................................................................................................. 46
7.2 Sensor .................................................................................................................................... 48
7.3 Debubbler and Measuring Chamber ...................................................................................... 49
7.4 List of Replacement Parts ...................................................................................................... 50
8.0 TROUBLESHOOTING........................................................................................................... 51
8.1 Overview ................................................................................................................................ 51
8.2 Troubleshooting Using Fault Codes ....................................................................................... 51
8.3 Troubleshooting Calibration Problems ................................................................................... 53
8.4 Troubleshooting Other Problems............................................................................................ 54
8.5 Information Screens ............................................................................................................... 56
9.0 RETURN OF MATERIAL ...................................................................................................... 58
MODEL CLARITY II TABLE OF CONTENTS
LIST OF FIGURES
Number Title Page
2-1 Panel Mount Installation ........................................................................................... 4
2-2 Pipe Mount Installation ............................................................................................. 5
2-3 Surface Mount Installation........................................................................................ 6
2-4 Opening the Supporting Clamps .............................................................................. 7
2-5 Debubbler and Flow Chamber ................................................................................. 7
2-6 Sensor ...................................................................................................................... 9
2-7 Sampling for Turbidity............................................................................................... 9
3-1 Removing the Knockouts ......................................................................................... 10
3-2 Wiring Diagram for Model T1055-10 analyzer (panel mount version)...................... 10
3-3 Wiring Diagram for Model T1055-11 analyzer (wall/pipe mount version) ................. 10
3-4 Non-Incendive Field Wiring (FM) T1055-10 ............................................................. 12
3-5 Non-Incendive Field Wiring (FM) T1055-11.............................................................. 13
3-6 Non-Incendive Field Wiring (CSA) T1055-10 ........................................................... 14
3-7 Non-Incendive Field Wiring (CSA) T1055-11 ........................................................... 15
4-1 Displays During Normal Operation........................................................................... 16
4-2 Solu Comp II Keypad ............................................................................................... 16
5-1 Assigning Outputs 1 and 2 in a dual input instrument ............................................. 17
5-2 High Alarm Logic ..................................................................................................... 19
5-3 Low Alarm Logic ....................................................................................................... 19
5-4 Turbidity Sensor — General ..................................................................................... 22
5-5 Turbidity Sensor — EPA 180.1 ................................................................................. 22
5-6 Turbidity Sensor — ISO 7027................................................................................... 23
6-1 Calibration Against a User-Prepared Standard ........................................................ 30
6-2 Converting Turbidity to TSS ..................................................................................... 39
6-3 Lowest Turbidity (TSS) ............................................................................................. 39
7-1 Exploded View of Solu Comp II (Panel Mount Version) ........................................... 42
7-2 Exploded View of Solu Comp II (Pipe/Surface Mount Version)................................ 43
7-3 Replacing the Lamp/LED Board............................................................................... 44
LIST OF TABLES
Number Title Page
5-1 Default Settings ........................................................................................................ 20
7-1 Replacement Parts for Solu Comp II (Panel Mount Version) ................................... 46
7-2 Replacement Parts for Solu Comp II (Pipe/Surface Mount Version)........................ 47
APPENDIX
APPENDIX .............................................................................................................. 59
ii
1
MODEL CLARITY II SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0.
DESCRIPTION AND SPECIFICATIONS
COMPLETE SYSTEM includes single or dual input analyzer, sensor(s), and debub-
bler assembly
CHOOSE U.S. EPA METHOD 180.1 or ISO METHOD 7027 compliant sensors
RANGE 0-200 NTU
RESOLUTION 0.001 NTU
FULL FEATURED ANALYZER with fully scalable analog outputs and optional fully
programmable alarms
INTUITIVE, USER-FRIENDLY MENU in six languages makes setup and calibration
easy
Clarity II is a trademark of Emerson Process Management.
FEATURES AND APPLICATIONS
The Clarity II turbidimeter is intended for the determi-
nation of turbidity in water. Low stray light, high stabili-
ty, efficient bubble rejection, and a display resolution of
0.001 NTU make Clarity II ideal for monitoring the tur-
bidity of filtered drinking water. Because it measures
turbidity as high as 200 NTU, Clarity II is also suitable
for most raw waters. The Clarity II turbidimeter can be
used in applications other than drinking water treat-
ment. Examples are monitoring wastewater dis-
charges, condensate returns, and clarifiers.
Both USEPA 180.1 and ISO 7027-compliant sensors
are available. USEPA 180.1 sensors use a visible light
source. ISO 7027 sensors use a near infrared LED.
For regulatory monitoring in the United States, USEPA
180.1 sensors must be used. Regulatory agencies in
other countries may have different requirements.
The Clarity II turbidimeter consists of an analyzer,
which accepts either one or two sensors, the sensors
themselves, and a debubbler/measuring chamber and
cable for each sensor. The cable plugs into the sensor
and the analyzer, making setup fast and easy. Sensors
can be located as far as 50 ft (15.2 m) away from the
analyzer.
The Clarity II turbidimeter incorporates the popular
and easy to use Solu Comp II analyzer. Menu flows
and prompts are so intuitive that a manual is practical-
ly not needed. Analog outputs are fully scalable. An
optional alarm board with three relays is also avail-
able. Alarms are fully programmable for high/low logic
and dead band. To simplify programming, the analyzer
automatically detects whether an EPA 180.1 or ISO
7027 sensor is being used.
Clarity II is available in an optional configuration in
which the analyzer, sensor(s), and debubbling flow
cell(s) are mounted on a single back plate. The sensor
cables are pre-wired to the analyzer, so setup is
exceptionally fast and easy. All the user does is mount
the unit on a wall, bring in power and sample, and pro-
vide a drain. To order this option, consult the factory.
A dry check is also available to periodically confirm
Clarity II operation.
2
MODEL CLARITY II SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.2 SPECIFICATIONS — ANALYZER
Enclosure: ABS (panel mount), polycarbonate (pipe/wall
mount); NEMA 4X/CSA 4 (IP65)
Dimensions:
Panel mount version: 6.10 X 6.10 X 3.72 in (155 X 155 X
94.5 mm)
Pipe/ Wall mount version: 6.23 X 6.23 X 3.23 in (158 X 158 X
82 mm)
Conduit openings: Accepts PG 13.5 or ½-in fittings.
Display: Two line, 16-character back lit display. Character
height 4.8 mm. Display can be customized to meet individ-
ual requirements.
Security Code: 3-digit code prevents accidental or unautho-
rized changes in instrument settings and
calibration.
Languages: English, German, Spanish, Italian, French,
Portuguese
Units: Turbidity (NTU, FTU, or FNU); total suspended solids
(mg/L, ppm, or no units)
Display resolution-turbidity: 4 digits; decimal point moves
from x.xxx to xxx.x
Display resolution-TSS: 4 digits; decimal point moves from
x.xxx to xxxx
Calibration methods: user-prepared standard, commercially
prepared standard, or grab sample. For total suspended
solids user must provide a linear calibration equation.
Ambient temperature and humidity: 0 to 50°C, (32 to 122°F);
RH 10 to 90% (non-condensing)
Power: 85 to 265 Vac, 47.5 to 65.0 Hz. Maximum current with-
out option -60 alarm board: 1.0 amp, with option -60 alarm
board: 1.3 amp.
Equipment protected by double insulation.
Hazardous Location:
Class I, Div. 2, Groups A, B, C, & D: T3C Tamb 0°-50°C
Suitable for use in Class II and III, Division 2, Groups E, F
and G. Enclosure Type 4/4X
Install in accordance with control drawing
no. 1400311 or 1400312 (FM).
Install in accordance with control drawing
no. 1400313 or 1400314 (CSA).
(Ordinary Location only)
Inputs: Choice of single or dual input
RFI/EMI: EN-61326
LVD: EN-61010-1
Outputs: Single input analyzer has single output. Dual input
analyzer has dual outputs. Outputs are 0-20 mA or 4-20
mA isolated. Maximum load is 600 ohms. Output dampen-
ing with 5 sec time constant is user-selectable.
Output Accuracy: 0.05 mA
Alarms: Optional alarm relay board includes three alarms.
Alarm 3 can be configured as a fault alarm in-stead of a
process alarm. Each relay can be configured independent-
ly. Alarm logic (low or high) and dead band are user-pro-
grammable.
Relays: Form C, single pole, double throw, epoxy sealed.
Alarm Board Ratings:
Field wiring terminals: removable terminal blocks for power,
analog outputs, and sensors
SPECIFICATIONS — SENSOR
Method: EPA 180.1 or ISO 7027 (using 860 nm LED source).
Must be specified when ordering.
Incandescent lamp life: two years
LED life: five years
Wetted materials: Delrin
®1
, glass, EPDM
Accuracy after calibration at 20.0 NTU:
0 - 1 NTU: ±2% of reading or ±0.015 NTU, whichever is
greater.
0 - 20 NTU: ±2% of reading
Cable: 20 ft (6.1 m) or 50 ft (15.2 m). Maximum 50 ft
(15.2 m).
Maximum Pressure: 30 psig (308 kPa abs)
Temperature: 40 - 95°F (5 - 35°C)
Sensor Body: IP65 when cable is connected.
1
Delrin is a registered trademark of DuPont Performance Elastomers.
SPECIFICATIONS — DEBUBBLER AND
FLOW CHAMBER
Dimensions: 18.1 in. x 4.1 in. diam. (460 mm x 104 mm diam.)
(approx.)
Wetted materials: ABS, EPDM, polypropylene, nylon,
Kynar
®2
, Delrin
Inlet: compression fitting accepts 1/4 in. OD tubing; fitting can
be removed to provide 1/4 in. FNPT
Drain: barbed fitting accepts 3/8 in. ID tubing; fitting can be
removed to provide 1/4 in. FNPT. Must drain to atmos-
phere.
Sample temperature: 40 - 95°F (5 - 35°C)
Minimum inlet pressure : 3.5 psig (125 kPa abs). 3.5 psig will
provide about 250 mL/min sample flow.
Maximum inlet pressure: 30 psig (308 kPa abs). Do not block
drain tube.
Recommended sample flow: 250 - 750 mL/min
Response Time: The table shows the time in minutes to per-
cent of final value following a step change in turbidity.
SPECIFICATIONS — MISCELLANEOUS
Weight/shipping weight:
Sensor: 1 lb/2 lb (0.5 kg/1.0 kg)
Analyzer: 2 lb/3 lb (1.0 kg/1.5 kg)
Debubbler: 3 lb/4 lb (1.5 kg/2.0 kg)
(rounded to the nearest lb or 0.5 kg)
2
Kynar is a registerd trademark of Elf Atochem North America, Inc.
Specifications subject to change without notice.
Resistive Inductive
115 Vac 5.0 A 3.0 A
230 Vac 5.0 A 1.5 A
% response
following step
change
response time (minutes)
4 gph
(250 mL/min)
12 gph
(750 mL/min)
10 2.0 0.5
50 2.5 1.0
90 4.5 2.5
99 7.0 4.0
SECTION 2.0.
INSTALLATION
2.1 UNPACKING AND INSPECTION
The Clarity II Turbidimeter is a complete system for the determination of turbidity in drinking water. The system
consists of the analyzer, sensor(s), cable(s), and flow chamber/debubbler(s). Consult the table to verify that you
have received the parts for the option you ordered.
(1)
The analyzer model number is printed on a label attached to the side of the instrument. For example, if you
ordered a panel mount, dual input analyzer, with alarm board, the label should read: T1055-10-22-60.
2.2 INSTALLATION — ANALYZER
2.2.1 General Information
1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme tem-
peratures.
2. Install the analyzer in an area where vibrations and electromagnetic and radio frequency interference are min-
imized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy
access to the analyzer.
4. Do not run AC power and relay wiring through the top conduit openings. Keep AC power and relay
wiring separate from other wiring in the analyzer after installation.
5. The analyzer is suitable for panel, pipe, or surface mounting. Refer to the table below.
6. See Section 3.1 for removal of conduit knockouts.
7. To reduce the likelihood of stress on wiring connections, remove the hinged front panel (-11 models) from the
base during wiring installation. Allow sufficient wire length to avoid stress on conductors.
Item Model/part number
(1)
Analyzer-panel mount
(1)
T1055-[10]-[ ]
Analyzer-pipe/wall mount T1055-[11]-[ ]
Analyzer-single input T1055-[ ]-[21]
Analyzer-dual input T1055-[ ]-[22]
Analyzer- alarm board input T1055-[ ]-[ ]-[60]
Sensor-EPA standard 8-0108-0002-EPA
Sensor-ISO standard 8-0108-0003-ISO
Cable-3 ft (0.9 m) 24138-00
Cable-20 ft (6.1 m) 24097-00
Cable-50 ft (15.2 m) 24098-00
Calibration cup 24101-00
Molded chamber/debubbler 24170-00
Type of Mounting Section
Panel 2.2.2
Pipe 2.2.3
Surface 2.2.4
MODEL CLARITY II SECTION 2.0
INSTALLATION
3
4
MODEL CLARITY II SECTION 2.0
INSTALLATION
FIGURE 2-1. Panel Mount Installation
Access to the wiring terminals is through the rear cover. Four screws hold the cover in place.
2.2.2 Panel Mounting.
MILLIMETER
INCH
FIGURE 2-2. Pipe Mount Installation
The front panel is hinged at the bottom. The panel swings down for access to the wiring terminals.
2.2.3 Pipe Mounting.
MILLIMETER
INCH
MODEL CLARITY II SECTION 2.0
INSTALLATION
5
6
MODEL CLARITY II SECTION 2.0
INSTALLATION
FIGURE 2-3. Surface Mount Installation
The front panel is hinged at the bottom. The panel swings down for access to the wiring terminals.
2.2.4 Surface Mounting.
MILLIMETER
INCH
2.3 INSTALLATION — DEBUBBLER ASSEMBLY
See Figure 2-4 for installation.
Connect the sample line to the inlet fitting. The fitting accepts 1/4-inch OD tubing. See Section 2.6 for recom-
mended installation of the sample port.
Attach a piece of 3/8 inch ID soft tubing to the drain fitting. The debubbler must drain to atmosphere.
NOTE
During operation, the debubbler is under pressure. A 0.040 inch (1 mm) orifice in the outlet pro-
vides the pressure. Back pressure helps prevent outgassing, which can lead to bubbles accu-
mulating on the sensor face resulting in erroneous readings. DO NOT EXCEED 30 psig (308
kPa abs) inlet pressure.
The amount of pressure in the debubbler can be estimated from the flow rate. See Table 2-1.
To control and monitor sample flow, a valved rotameter with fittings is available (PN 24103-00). Attach the
rotameter to the debubbler outlet. The rotameter can also be used to increase back pressure on the debubbler if
additional pressure is needed to prevent outgassing.
MODEL CLARITY II SECTION 2.0
INSTALLATION
TABLE 2-1. Approximate debubbler pressure
as a function of flow (0.040 inch outlet orifice)
gph psig
21
43
68
814
10 21
11 26
12 31
——
mL/min kPa abs
100 110
200 120
300 140
400 160
500 190
600 240
700 280
800 340
7
WARNING
Before removing the sensor be absolutely certain the
process pressure is reduced to 0 psig and the
process temperature is at a safe level.
CAUTION
8
FIGURE 2-4. Debubbler and Flow Chamber
INCH
MILLIMETER
MODEL CLARITY II SECTION 2.0
INSTALLATION
9
2.4 INSTALLATION — SENSOR
Unscrew the nut on the side of the debubbler. Insert the sensor in the mouth of the measuring chamber. Be sure
the pin on the debubbler lines up with the hole in the sensor. Replace the nut. Remove the protective cap from
the sensor and screw the cable onto the receptacle. The plug and receptacle are keyed for proper alignment.
The sensor is rated to IP65 when properly connected to the cable. To prevent possible water damage to the
connector contacts, be sure the cable receptacle and the connector on the back of the sensor are dry when
connecting or disconnecting the cable.
2.5 SAMPLE POINT
Locate the sample tap to minimize pickup of sediment or air. See Figure 2-6. If possible, install a sampling port
that extends one or two inches (25 - 50 mm) into the pipe. Use ¼ inch OD rigid plastic tubing. Avoid soft plastic
tubing if possible. To reduce sample lag time, install the debubbler and flow chamber as close to the sample tap
as possible.
FIGURE 2-6. Sampling for Turbidity
FIGURE 2-5. Sensor
INCH
MILLIMETER
DWG. NO. REV.
40T105501 A
MODEL CLARITY II SECTION 2.0
INSTALLATION
1010
MODEL CLARITY II SECTION 3.0
WIRING
SECTION 3.0.
WIRING
3.1 PREPARING CONDUIT OPENINGS
The analyzer enclosure has five conduit openings. Two are
open and three are knockouts.
Conduit openings accept 1/2-inch conduit fittings or PG 13.5
cable glands. To keep the case watertight, block unused
openings with NEMA 4X or IP65 conduit plugs.
NOTE
Use watertight fittings and hubs that comply with the
requirements of UL514B. Connect the conduit hub to the
conduit before attaching the fitting to the analyzer
(UL508-26.16).
Figure 3-1 shows how to remove the knockouts. The knock-
out grooves are on the outside of the case. Place the screw-
driver blade on the inside of the case and align it approxi-
mately along the groove. Rap the screwdriver sharply with a
hammer until the groove cracks. Move the screwdriver to an uncracked portion of the groove and continue the
process until the knockout falls out. Use a small knife blade to remove the flash from the inside of the hole.
3.2 POWER, ALARM, OUTPUT, AND SENSOR CONNECTIONS
3.2.1 General Information
The analyzer is available in two mounting configurations.The positions of the power, alarm, output, and sensor ter-
minal blocks are different in each. See Figure 3-2 (panel mount) or Figure 3-3 (pipe/wall mount).
To reduce the likelihood of stress on wiring connections, keep the hinged front panel (-11 option only) attached to
the back cover while installing wiring. Be sure there is sufficient cable within the analyzer enclosure to avoid stress
on conductors and connections.
For ease of wiring, connections for power, outputs, and alarms are removable.
FIGURE 3-1. Removing the Knockouts
FIGURE 3-2. Wiring Diagram for Model T1055-10
analyzer (panel mount version)
The release clip on the sensor plug faces
the top of the analyzer enclosure.
FIGURE 3-3. Wiring Diagram for Model T1055-11
analyzer (wall/pipe mount version)
The release clip on the sensor plug faces the user.
3.2.2 Sensor
The sensor cable is pre-wired to a plug that inserts into a receiving socket in the analyzer. See Figures 3-2 and
3-3 for the locations of the sockets. If you are using a single input analyzer, be sure to plug the sensor into the
SENSOR 1 receptacle. The cable also passes through a strain relief fitting. To install the cable…
1. Remove the wrenching nut from the strain relief fitting.
2. Insert the plug through the hole in the bottom of the enclosure nearest the sensor socket. Seat the fitting in the
hole.
3. Slide the wrenching nut over the cable plug and screw it onto the fitting.
4. Loosen the cable nut so the cable slides easily.
5. Insert the plug into the appropriate receptacle. To remove the plug, squeeze the release clip and pull straight
out.
6. Adjust the cable slack in the enclosure and tighten the cable nut. Be sure to allow sufficient slack to avoid plac-
ing stress on the cable and connections.
7. Plug the cable into the back of the sensor. The sensor is rated to IP65 when properly connected to the cable.
To prevent possible water damage to the connector contacts, be sure the cable receptacle and the connec-
tor on the back of the sensor are dry when connecting or disconnecting the cable.
8. Place the sensor in either the measuring chamber or the calibration cup. The sensor must be in a dark place
when power is first appled to the analyzer.
Note: If “S1 Warning” appears, check sensor cable connection and confirm sample water flow at debubbler
drain outlet.
3.2.3 Power, Alarm, and Ouput
AC wiring should be 14 gauge or greater. Provide a switch or breaker to disconnect the analyzer from the main
power supply. Install the switch or breaker near the analyzer and label it as the disconnecting device for the
analyzer.
Keep sensor and output signal wiring separate from power wiring. Do not run sensor and power wiring in the same
conduit or close together in a cable tray.
For best EMI/RFI protection use shielded output signal cable enclosed in an earth-grounded metal conduit.
Connect the shield to earth ground.
Do not apply power to the analyzer until all connections are secured and verified.
3.2.4 Non-Incendive Installation
The sensor cable is rated as Non-Incendive field wiring by FM and CSA.
The AC power wiring, analog output wiring, and alarm wiring (optional configuration) is not
rated as Non-Incendive
and must be installed in metal conduit. Refer to the FM and CSA control drawings herein, Figures 3-4 through 3-7, for
proper installation in hazardous areas.
MODEL CLARITY II SECTION 3.0
WIRING
11
WARNING
Explosion hazard. Do not disconnect equipment
when a flammable or combustible atmosphere is
present.
WARNING
Exposure to some chemicals may degrade the
sealing properties of materials used in the following
device: PN AZ8-1CH-6DSEA (Zettler Inc.)
AC connections and grounding must be in compli-
ance with UL 508 or local electrical code. DO NOT
apply power to the analyzer until all electrical
connections are verified and secure.
WARNING
RISK OF ELECTRICAL SHOCK
12
B
C
1
2
3
4
3
4
A
B
C
D
C
1400311
12
NOTES: UNLESS OTHERWISE SPECIFIED
A
DRAWN
CHECKED
ENG APVD
B. JOHNSON
APPROVALS
TITLE
SCALE: WEIGHT:
SHEET 1 OF 1
DATE
ROSEMOUNT
ANALYTICAL
Emerson
PROCESS MANAGEMEN T
FINISH
MATERIAL
DIMENSIONS ARE IN INCHES
REMOVE BURRS & SHARP EDGES
MACHINE FILLET RADII .020 MAX
NOMINAL SURFACE FINISH: 125
ANGLES – 1/2.
.XX – .03 .XXX – .010
THIS FILE CREATED USING
SOLID EDGE
SIZE
C
DWG NO
REV
12/7/04
J. FLOCK
J. FLOCK
1/5/05
1/5/05
This document contains information proprietary to
Rosemount Analytical, and is not to be made available
to those who may compete with Rosemount Analytical.
REV
ECO NORELEASE DATE
C9160
AUG 1, 2005
LTR ECO DESCRIPTION
BY
DATE
REVISION
CHECKED/APPROVED
NON INCENDIVE FIELD
WIRING INSTALLATION (FM)
T1055-10
1400311
C
NONE
THIS DOCUMENT IS
CERTIFIED BY
REV
REV
REV
REV
REV
REV
REVISIONS NOT PERMITTED
W/O AGENCY APPROVAL
FM C
SENSOR CABLE
IS SHIELDED
METAL CONDUIT
SENSOR #2
CLARITY II
TURBIDITY
SENSOR #1
CLARITY II
TURBIDITY
POWER SUPPLY
3
(OPTIONAL)
115/230
VAC
SENSOR CABLE
IS SHIELDED
METAL CONDUIT
ALARM
WIRING (VAC)
(OPTIONAL)
UNCLASSIFIED
AREA
HAZARDOUS AREA
T1055-10
METAL CONDUIT
ANALOG OUTPUT
UNCLASSIFIED
AREA
!
WARNING
FOR USE WITH NON-FLAMMABLE
PROCESS MEDIA ONLY
5
5
CLASS I, DIV. 2, GPS A-D, 0 - 50  C
CLASS II, III, DIV 2, GPS E-G
6 NON INCENDIVE FIELD WIRING IS ALLOWED.
5 MAX CABLE LENGTH IS 50 FEET.
4. NO REVISION TO DRAWING WITHOUT FM APPROVAL.
3 GROUND CONNECTION MAY BE MADE IN HAZARDOUS AREA.
2. SEAL REQUIRED AT EACH CONDUIT ENTRANCE.
1. INSTALLATION MUST CONFORM TO THE
NEC (ANSI-NFPA 70)
6
6
MODEL CLARITY II SECTION 3.0
WIRING
FIGURE 3-4. Non-Incendive Field Wiring (FM) T1055-10
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68

Rosemount T1055 Clarity II Turbidimeter Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI