IFM SA5000 Operating instructions

Type
Operating instructions

This manual is also suitable for

Operating instructions
Flow sensor
SAxxxx
11458803 / 0103 / 2023
GB
SAxxxx Flow sensor
2
Contents
1 Preliminary note ............................................................. 4
1.1 Symbols used.......................................................... 4
1.2 Warnings used ......................................................... 4
2 Safety instructions............................................................ 5
3 Intended use................................................................ 6
3.1 Application area ........................................................ 6
3.2 Restriction of the application area........................................... 6
4 Function ................................................................... 7
4.1 Output OUT1 selection options............................................. 7
4.2 Output OUT2 selection options............................................. 7
4.3 IO-Link ............................................................... 7
5 Installation.................................................................. 8
5.1 Process connection...................................................... 8
5.2 Interference............................................................ 8
5.3 Installation position...................................................... 9
5.3.1 Immersion depth.................................................... 9
5.3.2 Recommended mounting position....................................... 9
5.3.3 Conditionally possible installation position. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.3.4 Impermissible installation position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.3.5 Orientation ........................................................ 10
6 Electrical connection.......................................................... 11
7 Operating and display elements................................................. 12
8 Menu...................................................................... 13
8.1 Menu overview......................................................... 13
8.2 Main menu and submenus ................................................ 13
8.3 Available parameters .................................................... 16
9 Set-up..................................................................... 18
10 Parameter setting............................................................ 19
10.1 Parameter setting via the unit keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
10.2 Parameter setting via IO-Link.............................................. 19
10.3 Basic settings.......................................................... 20
10.3.1 Operating mode .................................................... 20
10.3.2 Flow adjustment.................................................... 21
10.3.3 Internal pipe diameter................................................ 23
10.3.4 Standard unit of measurement ......................................... 23
10.3.5 Process value for OUT2 .............................................. 24
10.4 Output configuration..................................................... 24
10.4.1 Switching signal .................................................... 24
10.4.2 Frequency signal.................................................... 26
10.4.3 Analogue signal .................................................... 28
10.5 Application configuration.................................................. 30
10.5.1 Medium........................................................... 30
10.5.2 Output polarity of the switching outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
10.5.3 Damping.......................................................... 30
10.5.4 Error behaviour of the outputs.......................................... 31
10.5.5 Switching delay..................................................... 31
10.5.6 Calibration......................................................... 32
10.5.7 Lock / unlock....................................................... 32
10.5.8 Device reset ....................................................... 33
10.6 Display ............................................................... 34
10.6.1 Display standard process value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
10.6.2 Display colour setting ................................................ 34
10.6.3 Display update rate.................................................. 35
10.6.4 Display rotation..................................................... 35
10.6.5 Switch off the display ................................................ 36
Flow sensor SAxxxx
3
10.7 Diagnosis ............................................................. 36
10.7.1 Memory........................................................... 36
11 Operation .................................................................. 38
11.1 Process value display.................................................... 38
11.2 Reading the parameter setting ............................................. 38
12 Troubleshooting ............................................................. 39
12.1 Error messages......................................................... 39
12.2 Warning messages...................................................... 39
13 Maintenance, repair and disposal................................................ 41
14 Factory settings.............................................................. 42
SAxxxx Flow sensor
4
1 Preliminary note
You will find instructions, technical data, approvals and further information using the QR code on the
device / packaging or at documentation.ifm.com.
1.1 Symbols used
Requirement
Instructions
Reaction, result
[...] Designation of keys, buttons or indications
Cross-reference
Important note
Non-compliance may result in malfunction or interference.
Information
Supplementary note
1.2 Warnings used
CAUTION
Warning of personal injury
wSlight reversible injuries may result.
Flow sensor SAxxxx
5
2 Safety instructions
The unit described is a subcomponent for integration into a system.
The system architect is responsible for the safety of the system.
The system architect undertakes to perform a risk assessment and to create documentation in
accordance with legal and normative requirements to be provided to the operator and user of
the system. This documentation must contain all necessary information and safety instructions
for the operator, the user and, if applicable, for any service personnel authorised by the
architect of the system.
Read this document before setting up the product and keep it during the entire service life.
The product must be suitable for the corresponding applications and environmental conditions
without any restrictions.
Only use the product for its intended purpose (Ò Intended use).
Only use the product for permissible media.
If the operating instructions or the technical data are not adhered to, personal injury and/or damage
to property may occur.
The manufacturer assumes no liability or warranty for any consequences caused by tampering with
the product or incorrect use by the operator.
Installation, electrical connection, set-up, operation and maintenance of the product must be
carried out by qualified personnel authorised by the machine operator.
Protect units and cables against damage.
SAxxxx Flow sensor
6
3 Intended use
The unit monitors liquids and gases.
It detects the process categories flow and medium temperature.
3.1 Application area
• Air
• Water
Glycol solutions (reference medium: 35% ethylene glycol solution)
Low-viscosity oils (viscosity: ≤ 40 mm²/s at 40°C /≤ 40 cSt at 104°F)
High-viscosity oils (viscosity: ≥ 40 mm²/s at 40°C / ≥ 40 cSt at 104°F)
Electromagnetic compatibility (EMC):
This is a class A product. This product may cause radio interference in domestic areas:
uIf required, take appropriate EMC screening measures.
3.2 Restriction of the application area
Rapid temperature changes:
Due to the thermal operating principle, the sensor tip placed in the medium must assume the
temperature of the medium in order to achieve the specified accuracy. Rapid and uneven
temperature changes can cause temporary misinterpretation.
Temperature stratification:
There must be no temperature stratification in the medium to be measured. Temperature
differences within the cross section lead to misinterpretation of the measuring signal.
Phase mixtures:
The medium to be measured must be completely mixed. Different thermal conductivities due to
phase differences lead to misinterpretation of the measuring signal.
Medium selection:
The sensor has different characteristics that can be selected via the [MEdI] parameter. Note: For all
media listed in the "Note on media" (Ò data sheet), only the repeatability specified in the data
sheet applies.
Deposits:
Deposits lead to incorrect measurements of the sensor due to their insulating effect. If there is a
risk of biological or additive deposits, refer to the maintenance recommendation in the
"Maintenance, repair and disposal" section.
Flow sensor SAxxxx
7
4 Function
The unit detects flow based on the calorimetric measuring principle.
As additional process value the unit detects the medium temperature.
The unit can be operated in SIO mode (standard input-output ) or in IO-Link mode.
The unit displays the current process values.
The unit has many self-diagnostic options.
The unit generates two output signals according to the parameter setting.
4.1 Output OUT1 selection options
Switching signal flow
Frequency signal flow
• IO-Link
4.2 Output OUT2 selection options
Switching signal flow
Switching signal temperature
Analogue signal flow
Analogue signal temperature
Frequency signal flow
Frequency signal temperature
Input for external teach signal (remote adjustment)
4.3 IO-Link
IO-Link is a communication system for connecting intelligent sensors and actuators to automation
systems. IO-Link is standardised in the IEC 61131-9 standard.
General information on IO-Link at io-link.ifm
Input Output Device Description (IODD) with all parameters, process data and detailed
descriptions of the device at documentation.ifm.com
IO-Link offers the following advantages:
Interference-free transmission of all data and process values
Parameter setting in the running process or presetting outside the application
Parameters for identifying the connected devices in the system
Additional parameters and diagnostic functions
Automatic backup and restore of parameter sets in case of device replacement (data storage)
Logging of parameter sets, process values and events
Device description file (IODD - Input Output Device Description) for easy project planning
Standardised electrical connection
Remote maintenance
SAxxxx Flow sensor
8
5 Installation
CAUTION
If the medium temperature is above 50 °C (122 °F), parts of the housing can increase in
temperature to over 65 °C (149 °F).
wRisk of burns.
uProtect the housing against contact with flammable substances and unintentional
contact.
uEnsure that the system is free of pressure during installation.
uEnsure that no media can leak at the mounting location during installation.
5.1 Process connection
Using process adapters the unit can be adapted to different process connections. Adapters have to be
ordered separately as accessories.
Information about the available mounting accessories at www.ifm.com.
A correct fit of the unit and ingress resistance of the connection are only ensured using ifm
adapters.
uObserve the instructions of the mounting accessories.
uUse a lubricating paste which is suitable and approved for the application. Grease the
threads of the process connection, adapter and sensor. Ensure no grease is applied to the
sensor tip.
uTake the tightening torques for sensor and fixing elements into account.
The following tightening torques apply to ifm sensors:
Types M18x1.5 and G1/2: 25 Nm.
Type 1/2" NPT: 100 Nm.
5.2 Interference
Structures in the pipe, bends, valves, reducing pieces and the like affect the function of the unit.
uAdhere to the distances between sensor and interference.
S S
5...10 x D 3...5 x D
Fig.1: Inlet and outlet pipe lengths
D: Outside diameter of the pipe
S: Interference
Flow sensor SAxxxx
9
5.3 Installation position
5.3.1 Immersion depth
15 mm
Fig.2: Immersion depth
The sensor tip must be completely surrounded by the medium.
Recommended immersion depths:
With an internal pipe diameter (diA) < 120 mm: ~ 15 mm
With an internal pipe diameter (diA) ≥ 120 mm: ~ ⅛ diA
uIn case of strong leverage on the measuring probe, e.g. due to high-viscosity or strongly
flowing media, do not exceed the specified immersion depths.
5.3.2 Recommended mounting position
Fig.3: Recommended mounting position
For horizontal pipes: mounting from the side.
For vertical pipes: mounting in the rising pipe.
Place the sensor tip as centrally as possible in the pipe.
Minimum immersion depth 15 mm.
5.3.3 Conditionally possible installation position
Fig.4: Conditionally possible installation position
For horizontal pipes, if the pipe is free from build-up: mounting from below.
For horizontal pipes, if the pipe is completely filled with medium: mounting from the top.
SAxxxx Flow sensor
10
5.3.4 Impermissible installation position
Fig.5: Impermissible installation position
The sensor tip must not be in contact with the pipe wall.
Do not mount in downpipes that are open at the bottom.
5.3.5 Orientation
2
1
Fig.6: Flow direction
uTo achieve the optimum measuring accuracy, mount the sensor in a way that the flow goes to the
larger of the two key surfaces (1):
For units with an external thread, a bore hole in the key surface (2) indicates the flow direction.
For easier readability of the display the sensor housing can be rotated by 345° with regard to the
process connection.
Do not go beyond the end stop.
Flow sensor SAxxxx
11
6 Electrical connection
The unit must be connected by a qualified electrician.
Observe the national and international regulations for the installation of electrical equipment.
Voltage supply according to SELV, PELV.
uDisconnect power.
uConnect the unit as follows:
L
1 BN
2 WH
4 BK
3 BU
OUT1/IO-Link
OUT2
L+
43
2 1
Fig.7: Wiring diagram (colours to DINEN60947-5-2)
BK: black BN: brown
BU: blue WH: white
Pin Connection
1 L+
3 L-
4 (OUT1) Switching signal flow
Frequency signal flow
• IO-Link
2 (OUT2) Switching signal flow
Switching signal temperature
Analogue signal flow
Analogue signal temperature
Frequency signal flow
Frequency signal temperature
Input for external teach signal (remote adjustment)
1
L
1
BN
2
WH
4
BK
3
BU
L+
2
L
BN
WH
BK
BU
L+
4
L
BN
BK
WH
BU
L+
3
L
1
BN
2
WH
4
BK
3
BU
L+
Fig.8: Circuit examples
1: 2 x positive switching
2: 2 x negative switching
3: 1 x positive switching / 1 x analogue
4: 1 x negative switching / 1 x analogue
SAxxxx Flow sensor
12
7 Operating and display elements
13
4
2
5
Fig.9: Operating and display
elements
1: Switching status LED for OUT1
2: Switching status LED for OUT2
3: Process value in the indicated unit of measurement *
4: 4-digit alphanumeric display
5: Keys for changing views and parameter setting
* %, m/s, l/min, m3/h, °C, x103 (SAxx10: %, fps, gpm, cfm, °F, x103)
Flow sensor SAxxxx
13
8 Menu
The figures in which the menus are displayed show the parameters that can be set on the unit by key
input. These parameters and other functions are also available via the IO-Link interface.
8.1 Menu overview
Use the operating keys to navigate from the process value display to the main menu and from there to
the submenus.
Process value display
1: Main menu
2: Initialisation menu [INI]
3: Extended functions
[EF]
4: Basic settings [CFG]
5: Min/max memory [MEM]
6: Display settings [DIS]
Fig.10: Menu overview
8.2 Main menu and submenus
The displayed parameters change when the factory setting is changed. The following menu
displays show the maximum available parameters.
SAxxxx Flow sensor
14
EF EF
rP1
SP1 FH1
FL1
FEP1
FrP1
FrP2
FSP2
t.HGH
t.LOW
INI
INI
FEP2
FSP2FH2
FL2rP2
SP2
ASP2
AEP2
Process value display
Fig.11: Main menu
MEdI
CGA 60...140 %
15...400
diA
ModE REL LIQU GAS
H2O
GLYC
OIL1 OIL2 AIR
INI
Main menu
Fig.12: Initialisation menu [INI]
EF
rES
CFG CFG
- - - -
DIS
MEM MEM
DIS
Main menu
Fig.13: Extended functions [EF] menu
Flow sensor SAxxxx
15
ou2
ou1 Hno Hnc Fno Fnc FRQ
dS1
dr1
dS2
uni
P-n
dAP
CFG
0,0...5,0 s
FOU1
FOU2
OFFOn OU
OFFOn OU
PnP nPn
m3/hl/min m/s
Hno Hnc Fno Fnc Itch FRQ
dr2
SEL2 FLOW TEMP
0,0...60 s
0,0...60 s
0,0...60 s
0,0...60 s
EFEF
Fig.14: Basic settings [CFG] menu
* For SAxx10 devices: cfm, gpm, fps
MEM
Lo.F
Hi.F
Lo.T
Hi.T
0,0 l/min
0,0 l/min
0,0 °C
0,0 °C
EFEF
Fig.15: Min/max memory [MEM] menu
* For SAxx10 devices: gpm and °F
SAxxxx Flow sensor
16
DIS
coLr
SELd FLOW TEMP
GrEnrEd
diS d2 d3d1
rd1 rd2 rd3 OFF
r1ou G1ou
r2ou G2ou
EFEF
Fig.16: Display settings [DIS] menu
8.3 Available parameters
The table shows the parameters as they appear in the device menu. The designation of the
parameters in the parameter setting software may differ.
Parameter Explanation
AEPx Analogue end point for OUTx = process value at which the output signal is 20 mA.
ASPx Analogue start point for OUTx = process value at which the output signal is 4 mA.
CGA Calibration factor in % for adapting the measured value curve to the application
colr Colour configuration of the display
dAP Damping constant in seconds (63 % rise time τ).
diA Inner diameter of the pipe
diS Update rate and orientation of the display
drx Switch-off delay for switching output OUTx in seconds
dSx Switch-on delay for switching output OUTx in seconds
FEPx Frequency end point for OUTx = Upper measured value at which the frequency signal set under
FrPx is output.
FHx Upper limit for switching signal OUTx with window function
FLx Lower limit for switching signal OUTx with window function
FOUx Behaviour of output OUTx in case of an error
FrPx Frequency signal that is output when the upper measured value (MEW or FEPx) is reached.
FSPx Frequency start point for OUTx = Lower measured value from which a frequency signal is output
(only for temperature measurement).
Hi.F Maximum flow value measured (flow volume or flow velocity)
Hi.T Maximum temperature value measured
Lo.F Minimum flow value (flow volume or flow velocity)
Lo.T Minimum measured temperature value
MEdi Selection of the medium to be monitored
ModE Operating mode (REL, LIQU, GAS)
oux Output configuration for output OUTx (e.g. switching output with hysteresis function)
P-n Output polarity for the switching outputs
P-n Output logic
rES Reset to factory settings
rPx Reset point for switching output OUTx with hysteresis function
SELd Standard process value of the display
SELx Process value for output OUTx
Flow sensor SAxxxx
17
Parameter Explanation
SPx Switch point for switching output OUTx with hysteresis function
tch Teach signal for flow adjustment
t.HGH High flow adjustment = flow adjustment to a maximum flow in the installation
t.LOW Low flow adjustment = flow adjustment to a minimum flow in the installation
uni Standard unit of measurement for flow
SAxxxx Flow sensor
18
9 Set-up
After power on and expiry of the power-on delay time, the unit is in the normal operating mode. It
carries out its measurement and evaluation functions and generates output signals according to the
set parameters.
During the power-on delay time, the outputs are in the following status according to the set
parameters:
ON with normally open function (Hno / Fno)
OFF with normally closed function (Hnc / Fnc)
OFF for frequency output (FRQ)
20 mA for current output (I)
Flow sensor SAxxxx
19
10 Parameter setting
Parameter setting can be carried out via the IO-Link interface or via the operating elements on the
unit.
Parameters can be set before installation or during operation.
If you change parameters during operation, this will influence the function of the plant.
uEnsure that there will be no malfunctions in your plant.
During parameter setting the unit remains in the operating mode. It continues to monitor with the
existing parameter until the parameter setting has been completed.
Depending on the parameter setting, the parameters available in the menu may change.
10.1 Parameter setting via the unit keys
CAUTION
If the medium temperature is above 50 °C (122 °F), parts of the housing can increase in
temperature to over 65 °C (149 °F).
wRisk of burns
uDo not touch the device with your hands.
uUse another object (e.g. a ballpoint pen) to carry out settings on the unit.
Parameter setting process in general:
Intent Action
Change from the process value display to the main menu [●]
Change to the submenu Use [▼] to navigate to the sub-
menu (e.g. EF), then [●]
Selection of the desired parameter ▲ or [▼]
Change to setting mode [●]
Changing the parameter value ▲ or [▼] > 1 s
Acceptance of the set parameter [●]
Exit parameter setting without saving [▲] + [▼]
Return to next higher menu
(Repeat several times to reach process value display)
[▲] + [▼]
Return to the process value display > 30 seconds (timeout)
10.2 Parameter setting via IO-Link
Requirements for parameter setting via the IO-Link interface:
üA suitable parameter setting software, e.g. ifm moneo|configure
üThe Input Output Device Description (IODD) for the device, see documentation.ifm.com
üOne IO-Link master
uConnect the IO-Link master to a parameter setting software.
uSet the port of the master to the IO-Link operating mode.
uConnect the device to a free port of the IO-Link master.
wThe unit switches to IO-Link mode.
SAxxxx Flow sensor
20
uChange parameter settings in the software.
uWrite parameter settings to the unit.
Notes on parameter setting Ò Manual of the parameter setting software
10.3 Basic settings
uStart by making the following settings in the order listed, as all further parameter settings are based
on this selection:
1. Operating mode (Ò/20).
Further settings depending on the operating mode:
[REL]: Flow adjustment (Ò/21).
[GAS] and [LIQU]: Internal pipe diameter (Ò/23) and Standard unit of measurement (Ò/23).
2. Process value for OUT2 (Ò/24)
10.3.1 Operating mode
The device provides three selectable operating modes for flow measurement. The operating mode
must be selected at the beginning of the parameter setting.
The selected operating mode has no effect on the temperature measurement, only absolute values in
°C (°F) are indicated.
Operating mode [ModE] Medium Display unit
REL liquids, air % of the taught range
LIQU liquids m/s; l/min m3/h
(fps, gpm, cfm)
GAS air m/s; l/min m3/h
(fps, gpm, cfm)
Tab.1: Operating modes for flow measurement
A change of the operating mode leads to a restart of the device. The parameter settings are
saved in the respective operating mode, i.e. when the operating mode is changed, the settings
are not lost.
10.3.1.1 Parameter setting via the device keys: Operating mode
uOpen the [INI] menu.
uSelect [ModE] and define the operating mode: [REL], [GAS], [LIQU].
wThe device carries out a reboot.
If you select [ModE] = [REL]:
uParameter setting via the device keys: Flow adjustment (Ò/21).
If you select [ModE] = [LIQU] or [GAS]:
uParameter setting via the device keys: Internal pipe diameter (Ò/23).
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42

IFM SA5000 Operating instructions

Type
Operating instructions
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI