INSTRUCTION MANUAL
Read and understand all of the instructions
and safety information in this manual before
operating or servicing this tool.
English ............ 1–18
Français ......... 19–36
Italiano........... 37–54
Deutsch .......... 55–72
Español .......... 73–90
Português...... 91–108
Nederlands ... 109–126
2
Description
The Greenlee DM-60 Digital Multimeter is a hand-held testing device
with the following measurement capabilities: AC and DC voltage, AC
and DC current, frequency, and resistance. It also checks diodes and
verifies continuity.
Safety
Safety is essential in the use and maintenance of Greenlee tools and
equipment. This instruction manual and any markings on the tool
provide information for avoiding hazards and unsafe practices related
to the use of this tool. Observe all of the safety information provided.
Purpose of This Manual
This instruction manual is intended to familiarize all personnel with the
safe operation and maintenance procedures for the Greenlee DM-60
Digital Multimeter.
Keep this manual available to all personnel. Replacement manuals are
available upon request at no charge.
All specifications are nominal and may change as design improvements
occur. Greenlee Textron Inc. shall not be liable for damages resulting from
misapplication or misuse of its products.
® Registered: The color green for electrical test instruments is a registered
trademark of Greenlee Textron Inc.
KEEP THIS MANUAL
Do not discard this product or throw away!
For recycling information, go to www.greenlee.com.
DM-60
3
Important Safety Information
This symbol is used to call your attention to hazards or unsafe
practices which could result in an injury or property damage. The
signal word, defined below, indicates the severity of the hazard.
The message after the signal word provides information for
preventing or avoiding the hazard.
SAFETY ALERT SYMBOL
Immediate hazards which, if not avoided, WILL result in severe
injury or death.
Hazards which, if not avoided, COULD result in severe injury or
death.
Hazards or unsafe practices which, if not avoided, MAY result in
injury or property damage.
Read and understand this material before
operating or servicing this equipment. Failure
to understand how to safely operate this tool
could result in an accident causing serious
injury or death.
4
Important Safety Information
Electric shock hazard:
Contact with live circuits could result in severe
injury or death.
Electric shock and fire hazard:
Do not expose this unit to rain or moisture.
Do not use the unit if it is wet or damaged.
Use test leads or accessories that are appropriate for the
application. Refer to category and voltage rating of the test lead
or accessory.
Inspect the test leads or accessory before use. They must be
clean and dry, and the insulation must be in good condition.
Use this unit for the manufacturer’s intended purpose only, as
described in this manual. Any other use can impair the
protection provided by the unit.
Failure to observe these warnings could result in severe injury or
death.
Electric shock hazard:
Do not apply more than the rated voltage between any two input
terminals, or between any input terminal and earth ground.
Do not contact the test lead tips or any uninsulated portion of the
accessory.
Failure to observe these warnings could result in severe injury or
death.
DM-60
5
Important Safety Information
Electric shock hazard:
Do not operate with the case open.
Before opening the case, remove the test leads from the circuit
and shut off the unit.
Failure to observe these warnings could result in severe injury or
death.
Electric shock hazard:
The fuses are an integral part of the overvoltage protection. When
fuse replacement is necessary, refer to “Specifications” for the
correct type, size and capacity. Using any other type of fuse will
void the overvoltage protection rating of the unit.
Failure to observe this warning could result in severe injury or
death.
Electric shock hazard:
Unless measuring voltage, current or frequency, shut off and
lock out power. Make sure that all capacitors are discharged.
Voltage must not be present.
Using this unit near equipment that generates electromagnetic
interference can result in unstable or inaccurate readings.
Failure to observe these warnings could result in severe injury or
death.
6
Important Safety Information
Set the selector and connect the test leads so that they correspond
to the intended measurement. Incorrect settings or connections
can result in a blown fuse.
Electric shock hazard:
Do not attempt to repair this unit. It contains no user-serviceable
parts.
Do not expose the unit to extremes in temperature or high
humidity. Refer to “Specifications.”
Failure to observe these precautions may result in injury and can
damage the unit.
Electric shock hazard:
Do not change the measurement function while the test leads are
connected to a component or circuit.
Failure to observe this precaution may result in injury and can
damage the unit.
DM-60
7
Identification
5
1
2
3
4
6 7 8
9
14
17
18
15,16
10
11
12
13
1. LCD display
2. HOLD button
3. MAX button
4. Selector
5. Positive input terminal for
high current measurements
6. Positive input terminal for
low current measurements
7. Negative, common (COM),
or ground input terminal for
all measurements
8. Positive input terminal for
all measurements except
current
Display Icons
9. MAX function is
enabled.
10. HOLD function is
enabled.
11. Polarity indicator
12. Low battery indicator
13.
O.L
Overload indicator
14. 600 V AC or DC range
selected
Symbols on the Unit
Warning—Read the instruction manual
Risk of electric shock
Double insulation
Battery
Fuse
Recycle product in accordance with manufacturer’s
directions
15. Diode
16. Continuity
17. kHz Kilohertz
(frequency, 10
3
cycles per second)
18. MHz Megahertz
(frequency, 10
6
cycles per second)
8
Using the Features
HOLD Button Press momentarily to hold the present value on the
display.
will appear on the display.
Press again to return to normal mode.
MAX Button Press momentarily to hold the maximum value on the
display.
will appear on the display. The highest value will remain
on the display, which will update when the meter measures a new
maximum.
Press again to return to normal mode.
Operation
Electric shock hazard:
Contact with live circuits could result in severe
injury or death.
1. Refer to the Settings Table. Set the selector to the proper setting
and connect the test leads to the meter. Start with the highest
measurement range.
2. Refer to “Typical Measurements” for specific measurement
instructions.
3. Test the unit on a known functioning circuit or component.
If the unit does not function as expected on a known functioning
circuit, replace the battery and/or fuses.
If the unit still does not function as expected, send the unit to
Greenlee for repair. Refer to the instructions under the Warranty.
4. Take the reading from the circuit or component to be tested. If the
resolution is not satisfactory, remove the meter from the circuit
and change to the next lower range.
DM-60
9
Operation (cont’d)
Set the Connect Connect
To measure selector to the red the black
this value: this symbol: lead to: lead to:
Continuity* VHz COM
Current (AC)
10A Max COM
10 A range
Current (AC)
µA mA COM
other ranges
Current (DC)
10A Max COM
10 A range
Current (DC)
µA mA COM
other ranges
Diode VHz COM
Frequency 10MHz VHz COM
Resistance VHz COM
Voltage (AC) VHz COM
Voltage (DC) VHz COM
*Tone and
symbol on display indicate continuity. Threshold is
approximately 25 .
Settings Table
10
Current Measurement
Voltage Measurement
Typical Measurements
DM-60
11
Typical Measurements
Resistance
Measurement
Continuity
Check
12
Diode Measurement
Typical Measurements
Forward Bias
Reverse Bias
DM-60
13
Accuracy
Refer to “Specifications” for operating conditions and temperature
coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a
fixed amount) at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 80% relative
humidity.
AC Voltage
Frequency Input
Range Accuracy Range Impedance
199.9 mV ± (1.2% + 0.3 mV)
1.999 V ± (1.0% + 0.003 V)
19.99 V ± (1.0% + 0.03 V) 40 to 400 Hz 10 M
199.9 V ± (1.0% + 0.3 V)
600 V ± (1.2% + 3 V)
DC Current
Range Accuracy Burden Voltage
199.9 µA ± (1.0% + 0.2 µA)
1.999 mA ± (1.0% + 0.002 mA)
19.99 mA ± (1.0% + 0.02 mA) < 0.2 V
199.9 mA ± (1.2% + 0.2 mA)
10.00 A* ± (2.0% + 0.03 A)
* Duty Cycle: 60 seconds on, 5 minutes off
AC Current
Burden Frequency
Range Accuracy Voltage Range
199.9 µA ± (1.0% + 0.3 µA)
1.999 mA ± (1.0% + 0.003 mA)
19.99 mA ± (1.0% + 0.03 mA) < 0.2 V 40 to 400 Hz
199.9 mA ± (1.5% + 0.4 mA)
10.00 A* ± (2.5% + 0.04 A)
* Duty Cycle: 60 seconds on, 5 minutes off
14
Accuracy (cont’d)
Frequency
Range Accuracy
1.999 kHz ± (0.5% + 0.003 kHz)
19.99 kHz ± (0.5% + 0.03 kHz)
199.9 kHz ± (0.5% + 0.3 kHz)
1.999 MHz ± (0.5% + 0.003 MHz)
15.00 MHz ± (0.5% + 0.03 MHz)
Sensitivity:
10 Hz to 1 MHz: 0.5 Volts RMS
1 MHz to 10 MHz: 1 Volt RMS
Resistance
Range Accuracy Typical Open Circuit Voltage
199.9 ± (0.8% + 0.5 ) 3.2 VDC
1.999 k ± (0.8% + 0.002 k)
19.99 k ± (0.8% + 0.02 k)
199.9 k ± (0.8% + 0.2 k)> 0.7 VDC
1.999 M ± (0.8% + 0.002 M)
20.00 M ± (1% + 0.05 M)
Diode Test
Open Circuit Voltage: 3.45 VDC (maximum)
Range: 0 mV to 2000 mV
Test Current (Typical): 1.0 mA
Continuity
Tone on: Circuit resistance is 25
Resolution: 1
Open Circuit Voltage: 3.45 VDC (maximum)
DC Voltage
Range Accuracy Input Impedance
199.9 mV ± (0.5% + 0.2 mV)
1.999 V ± (0.5% + 0.002 V)
19.99 V ± (0.5% + 0.02 V) 10 M
199.9 V ± (0.5% + 0.2 V)
600 V ± (0.8% + 1 V)
DM-60
15
Specifications
Display: 3-1/2–digit LCD (1999 maximum count)
Polarity: Automatic
Sampling Rate: 2.5 per second
Overload Protections:
Volts: 600 VAC RMS and 600 VDC
Amps:
A: 10 A/250 V F fuse, interrupting rating 200 A, 1/4" x 1-1/4"
mA: 0.315 A/250 V F fuse, interrupting rating 1500 A,
5 mm x 20 mm
Measurement Categories:
V Terminal: Category III, 300 VAC and 300 VDC; Category II,
600 VAC and 600 VDC
mA Terminal: Category III, 250 VAC and 250 VDC
A Terminal: Category III, 250 VAC and 250 VDC
Operating Conditions: 0 °C to 40 °C (32 °F to 104 °F),
0 to 80% relative humidity
Altitude: 2000 m (6500’) maximum
Indoor use only
Pollution Degree: 2
Storage Conditions: –10 °C to 50 °C (14 °F to 122 °F),
0 to 70% relative humidity
Remove battery
Battery: 9-Volt battery (NEDA 1604, JIS 006P or IEC 6F22)
16
Measurement Categories
These definitions were derived from the international safety standard
for insulation coordination as it applies to measurement, control, and
laboratory equipment. These measurement categories are explained in
more detail by the International Electrotechnical Commission; refer to
either of their publications: IEC 61010-1 or IEC 60664.
Measurement Category I
Signal level. Electronic and telecommunication equipment, or parts
thereof. Some examples include transient-protected electronic circuits
inside photocopiers and modems.
Measurement Category II
Local level. Appliances, portable equipment, and the circuits they are
plugged into. Some examples include light fixtures, televisions, and
long branch circuits.
Measurement Category III
Distribution level. Permanently installed machines and the circuits they
are hard-wired to. Some examples include conveyor systems and the
main circuit breaker panels of a building’s electrical system.
Measurement Category IV
Primary supply level. Overhead lines and other cable systems. Some
examples include cables, meters, transformers, and other exterior
equipment owned by the power utility.
Statement of Conformity
Greenlee Textron Inc. is certified in accordance with ISO 9000 (2000)
for our Quality Management Systems.
The instrument enclosed has been checked and/or calibrated using
equipment that is traceable to the National Institute for Standards and
Technology (NIST).
DM-60
17
1. Disconnect the unit from the circuit. Turn the unit OFF.
2. Remove the screw from the battery compartment cover.
3. Remove the battery compartment cover.
4. Replace the battery (observe polarity).
5. Replace the cover and screw.
Maintenance
Battery Replacement
Electric shock hazard:
Do not attempt to repair this unit. It contains no user-serviceable
parts.
Do not expose the unit to extremes in temperature or high
humidity. Refer to “Specifications.”
Failure to observe these precautions may result in injury and can
damage the unit.
Electric shock hazard:
Do not operate with the case open.
Before opening the case, remove the test leads from the circuit
and shut off the unit.
Failure to observe these warnings could result in severe injury or
death.
18
1. Disconnect the unit from the circuit. Turn the unit OFF.
2. Remove the screw from the battery compartment cover. Remove
the battery compartment cover.
3. Remove the screws from the back cover. Remove the back cover.
4. Replace the fuse(s).
5. Replace the covers and screws.
Cleaning
Periodically wipe the case with a damp cloth and mild detergent; do
not use abrasives or solvents.
Electric shock hazard:
The fuses are an integral part of the overvoltage protection. When
fuse replacement is necessary, refer to “Specifications” for the
correct type, size and capacity. Using any other type of fuse will
void the overvoltage protection rating of the unit.
Failure to observe this warning could result in severe injury or
death.
Maintenance (cont’d)
Fuse Replacement
Lifetime Limited Warranty
Greenlee Textron Inc. warrants to the original purchaser of these
goods for use that these products will be free from defects in
workmanship and material for their useful life, excepting normal
wear and abuse. This warranty is subject to the same terms and
conditions contained in Greenlee Textron Inc.’s standard one-year
limited warranty.
122
Nauwkeurigheid (vervolg)
Frequentie
Bereik Nauwkeurigheid
1,999 kHz ± (0,5% + 0,003 kHz)
19,99 kHz ± (0,5% + 0,03 kHz)
199,9 kHz ± (0,5% + 0,3 kHz)
1,999 MHz ± (0,5% + 0,003 MHz)
15,00 MHz ± (0,5% + 0,03 MHz)
Gevoeligheid:
10 Hz tot 1 MHz: 0,5 Volt RMS
1 MHz tot 10 MHz: 1 Volt RMS
Weerstand
Bereik Nauwkeurigheid Typische nullastspanning
199,9 ± (0,8% + 0,5 ) 3,2 V DC
1,999 k ± (0,8% + 0,002 k)
19,99 k ± (0,8% + 0,02 k)
199,9 k ± (0,8% + 0,2 k)> 0,7 V DC
1,999 M ± (0,8% + 0,002 M)
20,00 M ± (1% + 0,05 M)
Diodetest
Nullastspanning: 3,45 V DC (maximum)
Bereik: 0 mV tot 2000 mV
Teststroom (typisch): 1,0 mA
Continuïteit
Toon aan: circuitweerstand is 25
Resolutie: 1
Nullastspanning: 3,45 V DC (maximum)
Gelijkstroomspanning
Bereik Nauwkeurigheid Ingangsimpedantie
199,9 mV ± (0,5% + 0,2 mV)
1,999 V ± (0,5% + 0,002 V)
19,99 V ± (0,5% + 0,02 V) 10 M
199,9 V ± (0,5% + 0,2 V)
600 V ± (0,8% + 1 V)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128

Greenlee DM-60 Digital Multimeter (Europe) User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI