Extech Instruments RF41 User manual

  • Hello! I am an AI chatbot trained to assist you with the Extech Instruments RF41 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
RF41-en-US_V1.3 3/15
User'sGuide
Battery/CoolantBrixRefractometer(ATC)
ModelRF41
Introduction
CongratulationsonyourpurchaseofthisExtechRefractometer.
Precisionopticalinstrumentsshouldbehandledgently;avoid
touchingtheopticalsurface.Carefuluseoftheseinstrumentswill
provideyearsofreliableservice.
Specifications
Range 0to‐50ºCPropyleneGlycolfreezepoint
0to‐50ºCEthyleneGlycolfreezepoint
1.15to1.30specificgravitybatteryacid
0to‐40ºCcleaningfluidfreezepoint
Resolution 1ºC
Dimensions 6.3x1.6x1.2"(160x40x30mm);
Weight 6.4oz.(180g)
Description
1. Eyepiece
2. MirrorTube
3. Adjustmentscrew
4. CoverPlate
5. Prism
Scale
Operation
Thisinstrumentmeasurestherefractiveindexofasampleand
displaystheresultsinºCorSpecificgravity.
1.ZeroAdjustment
Covertheprismwithafewdropsofdistilledwater.Closethe
coverplateandviewthelight/darkboundary(shadowline).If
necessary,turntheadjustmentscrewuntiltheshadowline
aligns
withthescale’szeroline.Thezerocalibrationshouldbe
performedatambienttemperatureandshouldberepeatedevery
30minutesifthetemperaturechanges.Afterthezeroadjustment,
cleantheprismwithasoftcloth.
2.SamplePreparationandReading
Covertheprismwithafewdropsofdistilled
waterfromthe
includedvial.Closethecoverplatesothattheliquidspreads
acrosstheentiresurfaceoftheprismwithoutairbubblesordry
spots.Allowthesampleto
remainontheprismfor
approximately30seconds.
Whileholdingtheinstrument
underalightsource,look
throughthe
eyepiece.Thebrix
concentrationisdeterminedby
theintersectionofthe
boundaryofthelightanddark
fieldsonthescale.Ifthescale
appearsoutoffocus,adjustbyrotatingtheeyepieceuntilthe
scaleisinfocus.Theinstrumentalsofeaturesaneyeguardto
preventstraylightfrom
enteringtheeyepieceandcausing
reflections.
Itmaybenecessarytoadjustthepositionofthelightsourceto
maximizethecontrastoftheshadowline.Undernormal
conditions,optimalcontrastisobtainedbyholdingtheinstrument
underneathandperpendiculartoalightsource.
Onceareadinghasbeentaken,wipe
drywithacleancloth(donot
washorrinse)andplacetheinstrumentinthesuppliedplastic
case.Storetheinstrumentinasafe,dryenvironment.
3.AutomaticTemperatureCompensation(ATC)
Temperaturecompensationrelievestheuseroftheresponsibility
tomeasuretemperatureandapplyacorrectionfactorwhen
taking
readings.Thisrefractometermakesthiscorrection
automatically.Whenambienttemperaturevariesfrom20°C,
readingsareautomaticallyadjustedtocompensatefor
temperaturevariancebetween10°Cto30°C.
Warranty
FLIRSystems,Inc.warrantsthisExtechInstrumentsbranddevicetobefreeof
defectsinpartsandworkmanshipforoneyearfromdateofshipment(asixmonth
limitedwarrantyappliestosensorsandcables).Ifitshouldbecomenecessaryto
returntheinstrumentforserviceduringorbeyondthewarranty
period,contact
theCustomerServiceDepartmentforauthorization.Visitthewebsite
www.extech.comforcontactinformation.AReturnAuthorization(RA)number
mustbeissuedbeforeanyproductisreturned.Thesenderisresponsiblefor
shippingcharges,freight,insuranceandproperpackagingtopreventdamagein
transit.Thiswarrantydoesnotapply
todefectsresultingfromactionoftheuser
suchasmisuse,improperwiring,operationoutsideofspecification,improper
maintenanceorrepair,orunauthorizedmodification.FLIRSystems,Inc.specifically
disclaimsanyimpliedwarrantiesormerchantabilityorfitnessforaspecific
purposeandwillnotbeliableforanydirect,indirect,incidentalor
consequential
damages.FLIR’stotalliabilityislimitedtorepairorreplacementoftheproduct.
Thewarrantysetforthaboveisinclusiveandnootherwarranty,whetherwritten
ororal,isexpressedorimplied.
Calibration,Repair,andCustomerCare
Services
FLIRSystems,Inc.offersrepairandcalibrationservicesforthe
ExtechInstrumentsproductswesell.NISTcertificationformost
productsisalsoprovided.CalltheCustomerService
Departmentforinformationoncalibrationservicesavailablefor
thisproduct.Annualcalibrationsshouldbeperformedtoverify
meterperformanceandaccuracy.Technicalsupportand
generalcustomerserviceisalsoprovided,refertothecontact
informationprovidedbelow.
Copyright©20132015FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinany
form
www.extech.com
XX
12
3
4
5
SupportLines:U.S.(877)4398324;International:+1(603)3247800
TechnicalSupport:Option3;Email:support@extech.com
Repair&Returns:Option4;Email:[email protected]
Productspecificationsaresubjecttochangewithoutnotice
Pleasevisitourwebsiteforthemostuptodateinformation
www.extech.com
FLIRCommercialSystems,Inc.,9TownsendWest,Nashua,NH03063USA
ISO9001Certified
/