Introduction
Congratulations on your purchase of the Relative Humidity and Temperature Datalogger. This
datalogger measures and stores up to 16,000 relative humidity and 16,000 temperature readings
over 0 to 100%RH and‐40 to +70°C (40 to +158°F) measurement ranges. The user can easily
configure the logging rate, high/low alarm and start
mode, and download the stored data by
pluggingthemoduleintoaPC'sUSBportand runningthesuppliedPCsoftware.RelativeHumidity,
TemperatureandDewPointdatacanthenbegraphed,printedandexportedtootherapplications.
GPP(grainsperpound)datacanbeviewedusingtheoptionalRHT10SWsoftware.Thedatalogger
issuppliedwithalonglifelithiumbattery,whichcantypicallyallowloggingforoneyear.Thismeter
is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Please visit our website (www.extech.com) to check for the latest version of
this User Guide,
ProductUpdates,andCustomerSupport.
Features
Memoryfor32,000readings(16,000temperatureand16,000humidityreadings)
DewpointindicationviasuppliedWindows
TM
Software
GrainsperpoundindicationviaoptionalWindows
TM
Software(orderRHT10SW)
Selectabledata samplingrate: 2s,5s, 10s,30s,1m,5m,10m, 30m,1hr,2hr, 3hr,6hr,12hr,
24hr
StatusIndicationviaRed/YellowLEDandGreenLED
USBInterfaceforSetupandDataDownload
UserProgrammableAlarmThresholdsforRelativeHumidityand
Temperature
Longbatterylife
RHT10 HW_SW-en-US_V5.6.2 6/16
3
Description
1. Protectivecover
2. USBconnectortoPCport
3. Startbutton
4. RHandTemperaturesensors
5. AlarmLED(red/yellow)
6. RecordLED(green)
7. Mountingclip
RHT10 HW_SW-en-US_V5.6.2 6/16
4
Operation
SystemSoftwareRequired:
Windows7,Windows8.1orWindows10
MinimumHardwareRequired:
PCorNotebookwithPentium90MHzorhigher32MBRAM;
Atleast7MBofharddiskspaceavailabletoinstallDataloggerUSBsoftware.
Recommendeddisplayresolution1024X768withHighColor(16bit).
INSTALLINGTHEGRAPHINGSOFTWAREandUSBdriver
InstallthesuppliedWindowsPCDataloggerSoftwarebyplacingthesuppliedprogramdiskinthe
PC’sCDROMdriveorbydownloadingtheprogramfromthe
http://Extech.com/instruments/software.aspwebpage.
Iftheinstallationprogramdoesnotautomaticallyopenandprovideonscreenprompts,openand
browse the CDROM drive, then find and doubleclick on the SETUP.EXE file included on the
programdisk.Followtheonscreenpromptstocompletetheinstallation.
SoftwareInstall
RunSETUP.up.exeandfollowthedefaultstoinstallheRTH10software
USBDriverInstall
1.OpentherootfolderoftheRHT10.
2.OpenthefolderDriver\DriverUSBXPRESSanddoubleclickonUSBXpressInstaller.exe.
3.FollowthedefaultchoicestoloadthisUSBXpressdriver.
Note:WhentheRHT10isconnectedtotheUSBport,thebatteryinsidethedeviceisdischargedat
ahigherratethannormal.Toconservebatterylife,donotleav etheRHT10connectedtotheUSB
portforaprolongedperiodoftime
RHT10 HW_SW-en-US_V5.6.2 6/16
5
RUNNINGTHEGRAPHINGSOFTWARE
With the RHT10 inserted in the PC’s USB port, doubleclick the Datalogger Graph icon on your
desktoptolaunchtheprogram.Theopeningscreenwillappearasshownbelow.
Clickon“Connect”toopenthemainscreen,
Clickon“About”toviewtheversionnumber,
Clickon“ViewFile”toopenanexistingfile,
Clickon“Quit”toclosetheprogram.
RHT10 HW_SW-en-US_V5.6.2 6/16
6
MENUBARDESCRIPTION
Themainmenuforthesoftwareappearsasfollows:
FromlefttorighttheMENUBARiconsrepresentDATADOWNLOAD,LOGGERSETUP,FILEOPEN,
FILE SAVEAS, FILE PRINT, VIEW ALL, and ZOOM.In addition, the FILE, VIEW, LINK, and HELP
pulldown menus are available here.All of these functions are described in the following
paragraphs.
LOGGERSETUP
Clickonthecomputericononthemenubar(2
nd
fromleft)orselectLOGGERSETfromtheLINK
pulldownmenu.TheSetup window willappearasshownbelow;descriptionsforeachfield in
theSetupwindowarelisteddirectlybelowtheillustration:
TheCURRENTTIMEfieldwillautomaticallysynchronizewiththePCdateandtimesetting.
TheMANUALandINSTANTselectbuttonsallowtheusertostartdataloggingimmediately
whentheSetupwindowisexited(INSTANT),oratalatertime(MANUAL).
TheLOGGERNAMEfieldallowstheusertoselectauniquenamefortheRHT10.
TheSAMPLEPOINTSfieldinstructsthe RHT10totakeafixednumberofreadings.
TheSAMPLERATEpulldownmenuinstructstheRHT10tologreadingsataspecificrate.
RHT10 HW_SW-en-US_V5.6.2 6/16
7
TheLED Flash Cyclecan beset by theuser:longer cycletimes help toextend the battery
life.
The ALARM SETTINGS area of the SETUP window allows the user to set high and low
temperature and humidity limits and to disable the flashing of the RHT10’s LE Ds, thereby
conservingbatteryenergy.
Once the changes have been made, click on the SETUPbutton to savethem.Pressthe DEFAULT
buttontosettheRHT10backtofactorydefaultconditions.PresstheCANCELbuttontoabortthe
setup.
Note: Anystoreddatawillbepermanentlyerasedwhentheuser
clicksonSetup.Tosavethe
databeforecreatinganewsetup,clickCancelandthendownloadthedata.
Note: Thebatterymayrun outbeforethe RHT10 has finishedcollectingaspecifiednumberof
samplepoints.Alwaysensurethattheremainingchargeinthebatteryissufficienttolast
thecompletedurationofyourlog gingexercise.Whenindoubt,werecommendthatyou
alwaysinstallafreshbatterybeforeloggingcriticaldata.
IfINSTANTwasselectedintheSETUPwindow, theRHT10beginsloggingat the instanttheSETUP
buttonispressed.IfMANUALwasselected,theRHT10
mustbemanuallystartedbypressingand
holdingthedevice’syellowbuttonforapprox.3seconds,oruntilthetwoLEDsontheRHT10flash
atthesametime.
Loggingwillcontinueattheprogrammedsamplerateuntilthenumberofreadings(samplepoints)
specifiedintheSetupwindowisreached.
The green LED flashes once per sample point and the red or yellow LED flashes when user
programmedalarmlimitsareexceeded.Fordetails,seeLEDstatusguide.
RHT10 HW_SW-en-US_V5.6.2 6/16
8
DATADOWNLOAD
TotransferthereadingsstoredintheRHT10tothePC:
1. ConnecttheRHT10tothesameUSBportusedwhentheRHT10wasinitialized.
2. Openthegraphingsoftwareprogramifitisnotstillrunning.
3. ClicktheDownloadicon(1
st
ontheleft)orselectDataDownloadfromtheLINK
pulldownmenu
4. TheWindowshownbelowwillappear.PressDOWNLOADtobegintransferring
readings
If the readings are successfully transferred, the VIEW screen and the SAVE screen (both shown
below) will appear.In the SAVE screen, name the file and save it to a convenientlocation.The
data is saved as an .rec file for use in this program only.At a later time, the user can select
SAVEASandsavethedataasanExcel,Text,orBitmapfile(explainedinnextsection).
In the screen below press VIEW to see the data graphically. If there are no readings currently
storedintheRHT10,theprogramwilllettheuserknowviaawarningscreen.
RHT10 HW_SW-en-US_V5.6.2 6/16
9
WhenVIEWispressed,theDataGraphwindowappears,
The Data Graph window presents the transferred data in an xy graph format with Date/Time
representedon the horizontalaxisandTemperature/Relative Humidity /DewPoint/Grainsper
poundGPP(withoptionalRHT10SWonly)/Alarm
Limitsrepresentedontheverticalaxis:

Tozoominandoutofdataregions,thereareseveralapproaches:
1. Usethemousetoclickanddragaboxaroundanydataareatomagnifytheselected
area.
2. Clickonthezoommagnifiericontoolonthemenubar
3. SelectVIEWALLorZOOM
OUTfromtheVIEWpulldownmenu.
In the example graph above, the Temperature is represented by the solid red line (the lines are
known as ‘traces’); the RH and Dew Point are represented by the solid yellow and green lines,
respectively. The low and high alarm values for temperature are
represented by the dashed red
traces;thecorrespondingalarmsforhumidityarerepresentedbythedashedyellowtraces.
RHT10 HW_SW-en-US_V5.6.2 6/16
10
TocustomizethegraphwindowselectSHOWTRACES,BACKGROUND,GRIDLINES,andMASK
POINTSfromtheVIEWpulldownmenu.Theseoptionsprovidethefollowingfeatures:
SHOWTRACES: Allowstheusertoselectwhichdatatoviewasatrace(Temperature&
Alarms).
BACKGROUND: Selectbackgroundcolorscheme.
GRIDLINES:
Addorremovegridlinesfromxandyaxes.
MASKPOINTS: Places dots on the traces for the actual data points, breaking up the
continuousnatureofthetrace.
FILEOPEN,FILESAVEAS
Tosave thetransferreddatainaformatotherthantheproprietary.recformat,clicktheSAVEAS
iconfromthemenubar(4
th
fromright)orselectSAVEASfromtheFILEpulldownmenu.Thedata
canbesavedinthefollowingformats:
TEXTFILE(.txt)
EXCELFILE(.xls)
BITMAPFILE(.bmp)
Toopenanexistingdatafileforviewingonthedatagraphwindow,clickontheFILEOPENiconon
themenubar(3
rd
fromleft),orselectFILEOPENfromtheFILEpulldownmenu.Whenprompted,
selecta.recfilesavedearlier.
FILEPRINT
Toprintadatagraphwindowtoalineornetworkprinter,clickontheprintericonorselectPRINT
fromtheFILEpulldownmenu.Notethatcolorgraphwindowscanbeaccuratelyprintedoncolor
printers.
Note: TheRHT10retainsdatainitsinternalmemoryuntiltheuserbeginsanewloggingsession.
Startinga new loggingsession clears allof the RHT10’sstored readings; be suretosave
previousdatabeforestartinganewloggingsession.
RHT10 HW_SW-en-US_V5.6.2 6/16
11
GRAINSPERPOUND(GPP)(RHT10SW)
GPPvaluesareobtainedonlybypurchasingandloadingtheoptionalRHT10SWsoftware.
TheoptionalsoftwarecanbepurchasedthroughanExtechdistributorusingthepartnumber
RHT10SW.
NotethatthestandardRHT10softwaremustbeuninstalledbeforetheoptionalRHT10SW
softwarecanbeinstalled.
To view GPP data, click on the VIEW menu heading and scroll down to the SHOW TRACES tab.
EnsurethattheGPPline ischecked.Ifchecked,the graphwilldisplayGPPdata.Ifunchecked,the
graphwillnotshowGPPdata.
RHT10 HW_SW-en-US_V5.6.2 6/16
12
RHT10LEDSTATUSGUIDE
LEDs Meaning Action
RecordAlarm
NoLEDsflashing
Nologginginpro cess
Or
Nobatteryfound
Or
Batterycompletelydischarge d
Fitbatteryinplace
Replacethebatteryand
downloadthedata
RecordAlarm
Greensingleflashevery10sec.*
‐Logging,noalarm**
Greendoubleflashevery10sec.*
‐Delayedstart
To start, Hold the start button
untilGreenandYellowLEDflash
RecordAlarm
Redsingleflashevery10sec. *
‐Logging,lowalarmforRH
Reddoubleashevery10sec.
‐Logging,highalarmforRH
Redsingleflashevery60sec.
‐LowBattery
Logging, if enabled, will stop
automatically. No data will be
lost.Replacebatt ery
RecordAlarm
Yellowsingleflashevery10sec.*
‐Logging,lowalarmforTEMP
Yellowdoubleashevery10sec.
‐Logging,highalarmforTEMP
Yellowsingleflashevery60sec.
‐RHT10memoryfull
Downloaddata
* Tosavepower,theRHT10’sLEDflashingcyclecanbechangedto20sor30sviathesuppliedsoftware.
** Tosavepower,alarmLEDsfortemperaturecanbedisabledviathesuppliedsoftware.
When the battery is low, all operations will be disabled automatically.NOTE: Logging
automaticallystops when the battery weakens (logged data will be retained).The supplied
softwareisrequiredtorestartlogg ingandtodownloadloggeddata.
When both temperature and relative humidity readings exceed alarm level simultaneously,
LED
statusindicationalternateseveryothercycle.
RHT10 HW_SW-en-US_V5.6.2 6/16
13
Specifications
RelativeHumidity OverallRange 0to100%
Accuracy(0to20and80to100%) ±5.0%
Accuracy(20to40and60to80%) ±3.5%
Accuracy(40to60%) ±3.0%
Temperature OverallRange‐40to158
o
F(40to70
o
C)
Accuracy(40to‐10and+40to
+70
o
C)
±2
o
C
Accuracy(10to+40
o
C) ±1
o
C
Accuracy(40to+14and104to
158
o
F)
±3.6
o
F
Accuracy(+14to104
o
F) ±1.8
o
F
DewPointTemperature OverallRange‐40to158
o
F(40to70
o
C)
Accuracy(25
o
C,40to100%RH) ±4.0
o
F(±2
o
C)
Grainsperpound 1120gpp(160g/kg)max.;AvailableonlywithoptionalRHT10SWsoftware
Loggingrate Selectablesamplinginterval:From2secondsupto24hours
Operatingtemperature 31to176
o
F(35to80
o
C)
Batterytype 3.6VLithium(1/2AA)(SAFTLS14250,TadiranTL5101orequivalent)
Batterylife 1year(typ.)depe nding onloggingrate,ambienttemperature,&useof
AlarmLEDs
Dimensions/Weight 4x1x.9”(101x25x23mm)/6oz(172g)
RHT10 HW_SW-en-US_V5.6.2 6/16
14
BatteryReplacement
It is highly recommended that the battery be replaced every 12 months, or prior to
loggingcriticaldata.
The RHT10 does not lose its stored readings when the battery weak ens or when the battery is
replaced; the data logging process will howeverbe stopped and cannot be restarted
until the
batteryhasbeenreplacedandtheloggeddatahasbeendownloadedtoPC.
Useonly3.6Vlithiumbatteries.Beforereplacingthebattery,removethedataloggerfromthePC.
Followdiagrammaticsteps1through4below.
NOTE:LeavingtheRHT10pluggedintothePCUSBportlongerthannecessary
willcausesomeof
thebatterycapacitytobelost.
WARNING:Handlelithiumbatteriescarefully,observewarningsonbatterycasing.Disposeofin
accordancewithlocalregulations.
Safety: Please dispose of batteries responsibly; never dispose of batteries in a fire, batteries may
explodeorleak.If themeterisnottobeusedfor 60 days ormore,removethebattery and store
separately.
RHT10 HW_SW-en-US_V5.6.2 6/16
15
Warranty
FLIRSystems,Inc.warrantsthisExtechInstrumentsbranddevicetobefreeofdefectsinpartsandworkmanshipforoneyear
fromdateofshipment(asixmonthlimitedwarrantyappliestosensorsandcables).Ifitshouldbecomenecessarytoreturnthe
instrumentforserviceduringorbeyondthewarranty
period,contacttheCustomerServiceDepartmentforauthorization.Visit
thewebsitewww.extech.comforcontactinformation.AReturnAuthorization(RA)numbermustbeissuedbeforeanyproductis
returned.Thesenderisresponsibleforshippingcharges,freight,insuranceandproperpackagingtopreventdamageintransit.
Thiswarrantydoesnotapplytodefectsresultingfromactionoftheusersuchasmisuse,improperwiring,operationoutsideof
specification,impropermaintenanceorrepair,orunauthorizedmodification.FLIRSystems,Inc.specifically
disclaimsanyimplied
warrantiesormerchantabilityorfitnessforaspecificpurposeandwillnotbeliableforanydirect,indirect,incidentalor
consequentialdamages.FLIR’stotalliabilityislimitedtorepairorreplacementoftheproduct.Thewarrantysetforthaboveis
inclusiveandnootherwarranty,whetherwritten
ororal,isexpressedorimplied.
Calibration,Repair,andCustomerCareServices
FLIRSystems,Inc.offersrepairandcalibrationservicesfortheExtechInstrumentsproductswe
sell.WeofferNISTtraceablecalibrationformostofourproducts.ContactourCustomerService
Departmentforinformationoncalibrationserviceavailability.Annualcalibrationsshouldbe
performedtoverifymeterperformanceandaccuracy.Technicalsupportandgeneralcustomer
serviceisalsoprovided,refertothecontactinformationprovidedbelow.
Copyright©20142016FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
www.extech.com
SupportLines:U.S.(877)2398324;International:+1(603)3247800
TechnicalSupport:Option3;Email:support@extech.com
Repair&Returns:Option4;Email:repair@extech.com
Productspecificationsaresubjecttochangewithoutnotice
Pleasevisitourwebsiteforthemostuptodateinformation
www.extech.com
FLIRCommercialSystems,Inc.,9Townsend West,Nashua,NH03063USA
ISO9001Certified
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17

Extech Instruments RHT10 User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI