Leica M205 User manual

Category
Microscopes
Type
User manual

This manual is also suitable for

Leica M165 FC
Leica M205 FA
User Manual
Leica M-series User Manual 2
General Instructions
Safety concept
Before using your microscope for the first time,
please read the "Safety concept" brochure
included with your instrument. It contains addi-
tional information about handling and care.
Use in clean rooms
The Leica M series can be used in clean rooms
without any problems.
Cleaning
Do not use any unsuitable cleaning agents,
ρ
chemicals or techniques for cleaning.
Never use chemicals to clean colored
ρ
surfaces or accessories with rubberized
parts. This could damage the surfaces,
and specimens could be contaminated by
abraded particles.
In most cases, we can provide special solu-
ρ
tions on request. Some products can be
modified, and we can offer other accesso-
ries for use in clean rooms.
Servicing
Repairs may only be carried out by Leica
ρ
Microsystems-trained service technicians.
Only original Leica Microsystems spare
parts may be used.
Responsibilities of person in charge of
instrument
Ensure that the Leica stereomicroscope
ρ
is operated, maintained and repaired by
authorized and trained personnel only.
Leica M-series User Manual 3
Important Safety Notes
User Manual
The individual modules of the Leica M stereo-
microscopy series include an interactive
CD-ROM with all relevant user manuals in
20 other languages. Keep it in a safe place,
and readily accessible to the user. User manu-
als and updates are also available for you
to download and print from our web site at
www.stereomicroscopy.com.
This User Manual describes the special func-
tions of the individual modules of the Leica M
stereomicroscopy series and contains impor-
tant instructions for their operational safety,
maintenance, and accessories.
The "Safety concept" booklet contains addi-
tional safety information regarding the service
work, requirements and the handling of the
stereomicroscope, accessories and electrical
accessories as well as general safety instruc-
tions.
You can combine individual system articles
with articles from external suppliers (e.g. cold
light sources, etc.). Please read the user manual
and the safety requirements of the supplier.
Before installing, operating or using the instru-
ments, read the user manuals listed above.
In particular, please observe all safety instruc-
tions.
To maintain the unit in its original condition
and to ensure safe operation, the user must
follow the instructions and warnings contained
in these user manuals.
Special manuals
Special manuals are provided for a number of
accessories:
User Manual M2-216-1 for lamp housing
ρ
with high-pressure mercury burner.
User Manual for ebq 100 isolated supply
ρ
unit for 100W high-pressure mercury
burners.
User Manual M2-267-1 for the motorized
ρ
focus system.
User Manual M2-166-2 for FluoCombi III™.
ρ
Leica M-series User Manual 4
Symbols Used
Warning of a danger
•
This symbol indicates especially impor-
tant information that must be read and
complied with.
Failure to comply can cause the following:
Hazards to personnel
ρ
Functional disturbances or damaged instru-
ρ
ments
Warning of hazardous electrical voltage
This symbol indicates especially impor-
tant information that is mandatory to
read and observe.
Failure to comply can cause the following:
Hazards to personnel
ρ
Functional disturbances or damaged instru-
ρ
ments
Danger due to hot surface
This symbol warns against touching
accessible hot surfaces, e.g. those of
light bulbs.
Important information
This symbol indicates additional infor-
mation or explanations that intend to
provide clarity.
Explanatory notes
This symbol within the text stands for addi-
ρ
tional information and explanations.
Figures
(1) Numbers in parentheses within the
descriptions relate to the figures and the
items within those figures.
Leica M-series User Manual 5
Safety Instructions
Description
The individual modules fulfill the highest
requirements for observation and documenta-
tion of Leica stereomicroscopes of the M series.
Intended use
Refer to "Safety Concept" booklet
Non-intended use
Refer to "Safety Concept" booklet
Never use M series microscopes or their compo-
nents for surgical procedures (e.g. on the eye)
unless they are specifically intended for that
purpose.
The instruments and accessories described in
this operating manual have been tested for
safety and potential hazards. The responsible
Leica affiliate must be consulted whenever
the instrument is altered, modified or used in
conjunction with non-Leica components that
are outside of the scope of this manual.
Unauthorized alterations to the instrument or
noncompliant use shall void all rights to any
warranty claims.
Place of use
Refer to "Safety Concept" booklet
Electrical components must be placed at
ρ
least 10 cm away from the wall and from
flammable substances.
Avoid large temperature uctuations,
ρ
direct sunlight and vibrations. These condi-
tions can distort measurements and micro-
graphic images.
In warm and warm-damp climatic zones,
ρ
the individual components require special
care in order to prevent the build-up of
fungus.
Responsibilities of person in charge of
instrument
Refer to "Safety Concept" booklet
Ensure that:
The M series stereomicroscopes and
ρ
accessories are operated, maintained and
repaired by authorized and trained person-
nel only.
All operators have read, understood and
ρ
observe this User Manual, and particularly
the safety instructions.
Leica M-series User Manual 6
Safety Instructions (cont'd.)
Repairs, service work
Refer to "Safety Concept" booklet
Only original Leica Microsystems spare
ρ
parts may be used.
Before opening the instruments, switch off
ρ
the power and unplug the power cable.
Touching the live circuit can cause
injury.
Transport
Use the original packaging for shipping or
ρ
transporting the individual modules of the
Leica M stereomicroscopy series and the
accessory components.
In order to prevent damage from vibrations,
ρ
disassemble all moving parts that (accord-
ing to the user manual) can be assembled
and disassembled by the customer and
pack them separately.
Integration in third-party products
Refer to "Safety Concept" booklet
Disposal
Refer to "Safety Concept" booklet
Legal requirements
Refer to "Safety Concept" booklet
EC Declaration of Conformity
Refer to "Safety Concept" booklet
ρ
Health risks
•
Workplaces with stereomicroscopes
facilitate and improve the viewing task,
but they also impose high demands on the eyes
and holding muscles of the user. Depending on
the duration of uninterrupted work, asthenopia
and musculoskeletal problems may occur. For
this reason, appropriate measures for reduction
of the workload must be taken:
Optimal arrangement of workplace, work
ρ
assignments and work flow (changing tasks
frequently).
Thorough training of the personnel, giving
ρ
consideration to ergonomic and organiza-
tional aspects.
The ergonomic design and construction
ρ
of the Leica M stereomicroscopy series are
intended to reduce the exertion of the user
to a minimum.
Leica M-series User Manual 7
Safety Instructions (cont'd.)
•
Direct contact with eyepieces is a poten-
tial transmission method for bacterial
and viral infections of the eye.
The risk can be kept to a minimum by using
personal eyepieces for each individual or
detachable eyecups.
Light sources: Safety regulations
Protective measures of the manufacturer:
UV protection screen in front of the speci-
ρ
men plane prevents the user from looking
directly into the UV rays.
Dummy filter carriers in the free positions
ρ
of the rapid filter changer prevent direct UV
radiation from reaching the eyes.
UV filters are installed in the observation
ρ
beam paths to protect the eyes.
The stray-light protection on the lamp
ρ
housing prevents irradiation of the hands.
Warning
•
UV radiation could damage the eyes.
Therefore:
Never look into the light spot on the speci-
ρ
men plane without a UV protection screen.
Never look into the eyepieces if no excita-
ρ
tion filter is in the beam path.
Fill empty filter positions with dummy filter
ρ
carriers (e.g. M165 FC).
Do not select a white, strongly reflective
ρ
background for the specimen.
Leica M-series User Manual 8
Safety Instructions (cont'd.)
Supply unit
Always unplug the supply unit from the power
supply:
When installing and disassembling the
ρ
lamp housing
Before opening the lamp housing
ρ
When replacing the high-pressure mercury
ρ
lamp and other parts, such as the heat-
absorbing filter or the collector
During maintenance work on the supply
ρ
unit
Lamp housing
•
Never open the lamp housing while the
lamp is switched on. Risk of explosion,
UV exposure, blinding!
•
Before opening the lamp housing, allow
it to cool off for at least 15 minutes.
Danger of explosion!
•
Never cover the air duct on the lamp
housing. Danger of fire!
Mercury lamp
Follow the user manual and safety instruc-
ρ
tions provided by the lamp manufacturer,
and, in particular, the section on how to
proceed if a lamp breaks and releases
mercury.
For transport, remove the mercury lamp,
ρ
transport it in its original packaging and
protect moving parts in the lamp housing
using the transport anchors.
When it has reached the end of its rated life
ρ
(follow the manufacturer's specifications
and the minute meter on the supply unit).
To minimize the risk of explosion, replace a
ρ
discolored mercury lamp promptly.
Leica Microsystems assumes no liability
ρ
for damage caused by exploding, incor-
rectly installed or improperly used mercury
lamps.
Leica M-series User Manual 9
Contents
General Instructions 2
Important Safety Notes 3
Symbols Used 4
Safety Instructions 5
The Leica M Series for Fluorescence
Congratulations! 13
A Step Towards Infinity 14
The Electronics: Comfort, Convenience and Safety 15
The Modular Design: Everything is Relative 16
Maximum Compatibility 17
On We Go 18
Assembly
Assembling the Focusing Column for MDG Bases 20
Assembling the Focusing Column with an Incident-light Base 21
Motorized Focus: Restricting the Travel Path 22
Optics Carrier 23
Tube 24
Eyepieces 25
Objective 26
UV Protection Screen – Assembly 27
Objective Nosepiece – Assembly 28
Objective Nosepiece – Adjusting Parfocality 29
Transmitted-light Base TL ST 30
FluoCombi III™ – Preparations 31
FluoCombi III™ – Assembly 32
FluoCombi III™ – Adjusting the Parfocality 34
FluoCombi III™ – Centering the Objectives 36
FluoCombi III™ – Filter and UV Protection Screen 37
Transmitted-light Base TL BFDF: Before First Use 38
Transmitted-light Base TL BFDF 39
TL RC™ / TL RCI™ 40
IsoPro™ Manual Mechanical Stage: Assembly 41
Leica IsoPro™ Motorized Mechanical Stage: Assembly 44
Cables: Connections 48
Cables: Cable Duct 49
Cables: Diagram 50
Leica LED5000 MCI™ 51
Leica LED5000 MCI™: Alternative Assembly 52
Leica LED5000 RL 53
Leica EL6000 – Safety Notes 54
Leica EL6000 – Preparations 57
Leica EL6000 – Connection to the Fluorescence Lamp Housing 59
Leica EL6000 – Replacing the Lamp 61
Leica EL6000 – Connection to the Fluorescence Lamp Housing (cont'd.) 62
Leica M-series User Manual 10
Quick Start Guide
The Fastest Route to Success 64
Overview of an M Series Microscope 65
The Correct Interpupillary Distance 66
Using the Eyepieces 67
Focusing 68
Adjusting the Resistance of the Focus Drive 69
Changing Magnification (Zoom) 70
Ratchet Steps and Magnification Levels 71
Parfocality: More Comfort and Convenience for Your Work 72
Iris Diaphragm 73
Eyepieces
Magnification Factors of the Eyepieces 75
Health Notes 76
Dioptric Correction 77
Dioptric Correction and Parfocality 78
Graticules 79
Photography & Video
Photography & Video 81
Photo Tubes and C-mounts 82
Trinocular Video/Phototube 50% 83
Trinocular Video/Phototube 100% 84
Microscope Carrier
The Objective Nosepiece 86
Objectives andOptical Accessories
The Different Types of Objectives 88
Bases
Leica TL ST Transmitted-light Base: Controls 90
Leica TL ST Transmitted-light Base: Operation 91
Leica TL ST Transmitted-light Base: Changing Bulbs 92
Leica TL BFDF Transmitted-light Base: Controls 93
Leica TL BFDF Transmitted-light Base: Operation 94
Leica TL RC™ / TL RCI™: Controls 95
Leica TL RCI™: The Deflection Mirror 96
Leica TL RCI™: Color Intensity and Temperature 97
Leica TL RC™ / TL RCI™: Operation 98
Leica TL RCI™: Methods in Transmitted Light 99
Leica TL RCI™: Relief Images 100
Using Filters 102
Leica IsoPro™ (Non-motorized): Controls 103
Leica IsoPro™ (Motorized): Controls 104
Leica M-series User Manual 11
System Illumination
Leica LED5000 MCI™ 106
Leica LED5000 RL 108
Leica EL6000 – About the Instrument 109
Leica EL 6000 – Operation 111
Leica EL 6000 – Troubleshooting 112
Cleaning and Maintenance 113
Fluorescence
Filter Changer 115
FIM – Fluorescence Intensity Manager 116
About Fluorescence Microscopy 117
Rapid Filter Changers and Filter Types 118
Simple filter holders 119
Equipping the Filter Changer 120
Observation without Fluorescence 121
Double Iris Aperture 122
Commissioning the Fluorescence System 123
Accessories
Leica SmartMove 125
Leica SmartTouch™ 126
Leica PSC Controller 127
Dimensional Drawings
Leica M165 FC 129
Leica M205 FA 132
Leica LED5000 MCI 135
Leica LED5000 RL 136
MATS TL 137
Specications
Leica M165 FC / Leica M205 FA 139
Leica EL6000 – Specifications 143
Leica TL ST Transmitted-light Base 145
Leica TL BFDF Transmitted-light Base 146
Leica TL RC™ / TL RCI™ 147
Leica IsoPro™ Motorized XY Stage 148
Appendix
Calculating the Total Magnification and Field of View Diameter 150
Care, Maintenance, Contact Persons 151
Leica M-series User Manual 12
The Leica M Series
for Fluorescence
Leica M-series User Manual 13
Congratulations!
Congratulations on obtaining your new Leica M series stereomicroscope.
We are convinced it will exceed your expectations, as never before have
we applied our decades of experience in the areas of optics, mechanical
engineering and ergonomics in such an uncompromising manner.
The Leica M series embodies all the qualities you associate with the name
Leica Microsystems: excellent objectives, high-quality engineering, and
reliability. Furthermore, the modular design ensures that the M series
adapts perfectly to your needs—no matter which accessories you require
for your tasks.
The entire imaging system, including the zoom, objective and ErgoTube®,
is apochromatically corrected with much technological effort. Contrast,
sharpness, richness in detail, resolution, image and color fidelity are opti-
mum. In addition, the patented illumination beam path guarantees at every
zoom level that light utilization is at a maximum and that fluorescence images
are intensely luminous on a jet black background.
Though the reliability and robustness of Leica stereomicroscopes is
legendary, like any high-tech product, the Leica M series requires a certain
degree of care and attention. Therefore, we recommend that you read
this manual. It contains all the information you need regarding operation,
safety and maintenance. Simply observing a few guidelines will ensure
that even after years of intensive use, your stereomicroscope will continue
to work as smoothly and reliably as on the very first day.
We wish you the best of success in your work—after all, you are now
equipped with the best tool!
Leica M-series User Manual 14
A Step Towards Infinity
Ever since their introduction by Horatio S.
Greenough, stereomicroscopes have worked
according to the optical principles based
primarily on Ernst Abbe's research. For over a
century, ingenious optics designers and engi-
neers have worked to push magnification, reso-
lution and image fidelity to the limit permitted
by optics.
In doing so, they have always been constrained
by the interrelation between three factors: the
higher a microscope's resolution, the lower the
available working distance. If one increases the
distance of the optical axes, the three-dimen-
sional image seen by the observer becomes
distorted—a sphere becomes an ellipse, a flat
surface curves towards the observer.
Limits are made to be broken
The Leica M205 FA is the world's first fluores-
cence stereomicroscope with a zoom range
of 20.5:1. This accomplishment, however, was
not enough for Leica's engineers. With the new
FusionOptic™ in the Leica M205 FA, they have
succeeded in going yet another step beyond
previous limits. In addition to the increase in
magnification, the resolution, too, has been
increased to up to 1050 lp / mm, which corre-
sponds to a resolved structure size of 472 nm.
Of course, this increase in performance benefits
your everyday work. Set up your specimens on
the microscope table with comfortable freedom
of movement and discover details in stereomi-
croscopy that you could never see before.
* Leica M205 FA with 1.0× planapochromat and 10× eyepieces
Leica M-series User Manual 15
The Electronics: Comfort, Convenience and Safety for Your Experiments
Never before have electronics been used as
extensively in a Leica series as in the new M
series lineup. Optics carrier, stand, base and
illuminator are all connected using electrical
contacts—which provides a number of advan-
tages.
Reliability for your experiments
Exact reproducibility of results is an increas-
ingly common demand, particularly in research
and development. The continuous encoding
captures parameters such as the magnifica-
tion, the illumination, the position of the iris
diaphragm and more and transmits the infor-
mation to the Leica LAS software. Thus you
always know the conditions under which an
image was acquired.
Fewer cables
A large part of the cables have been routed in
the interior of the column. Data are also trans-
ferred via the interface between the column,
the optics carrier and the carrier. The direct
result is that you need fewer cables—this not
only makes your workstation neater and more
comfortable, it even makes it appear larger.
Contacts not only transmit data, but also supply
the power.
Leica Application Suite (LAS) evaluates the
transmitted data and can restore the test situa-
tion later.
Leica M-series User Manual 16
Have a special request? Let us know!
Leica Microsystems enjoys an exceptional repu-
tation when it comes to devising customer-
specific solutions. If you have a special request
that cannot be met with standard parts, contact
your Leica consultant. We have a solution for
every problem.
The Modular Design: Everything is Relative
The Leica M series provides maximum flexibil-
ity in choosing equipment, thanks primarily to
the modular configuration and the compatibil-
ity that Leica has painstakingly maintained for
decades. The optics carriers, eyepieces, bases
and more can be combined in any way you
choose, allowing you to create the microscope
that best suits your needs.
Despite this, you will notice that the controls
and individual components do not differ signif-
icantly. Whichever configuration you choose,
you will quickly feel right at home.
Leica M-series User Manual 17
Leica engineers were careful to ensure that the
new Leica M series—like its predecessors—
remains compatible with existing series. This
means that objectives, bases, tubes and so on
can be reused.
Objectives
All new objectives of the Leica M series are
parfocal, meaning that when used with the
objective nosepiece, they can be replaced while
keeping the specimen in perfect focus.
If you prefer, you can continue to use the previ-
ous Leica objective series. In this case, parfocal-
ity is no longer guaranteed.
Tubes
The interface between the optics carrier and the
tube has remained the same, so existing tubes
fit the new M series. The new tubes are designed
for eyepieces with field number 23, while the
predecessor models were only designed for field
number 21, resulting in a smaller object field.
Eyepieces
However, the new M-series eyepieces have an
audible and tangible click to provide immedi-
ate feedback in case of accidental adjustment.
Maximum Compatibility
Leica M-series User Manual 18
On We Go
If your new Leica microscope has already been
assembled and commissioned by your Leica
consultant, click here to skip through the instal-
lation instructions and go directly to the Quick
Start Guide on Page 63.
If, on the other hand, you are assembling
the microscope yourself, continue with the
"Assembly" chapter, which begins on Page 19.
For everything you need to know about the
correct use of fluorescence-related parts, refer to
Page 114.
Leica M-series User Manual 19
Assembly
Leica M-series User Manual 20
The first step is to connect the focusing column
of the M series to the corresponding base.
Tools used
Hex socket screwdriver, 3 mm
ρ
Assembling the column adapter
1. Securely install the column adapter on the
column using the four included screws.
Assembling the focusing column
2. Securely screw the focusing column to the
base using the six included screws.
Assembling the Focusing Column for MDG Bases
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152

Leica M205 User manual

Category
Microscopes
Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI