2 • Installation
Page 12 FocusJet Nebulizer Installation Guide
2.3 Gas Pressure Setting for the Nebulizer
NOTICE Operating the nebulizer with a gas pressure other than the
pressure specified for this nebulizer can cause a loss in performance of
the nebulizer and the detector. After installation of the nebulizer, set the
nebulizer gas pressure to the pressure that is specified for your
nebulizer. The nebulizer gas pressure is defined on the gas pressure
certificate that is shipped with the nebulizer.
Before operating the detector, verify and if required adjust the gas
pressure setting for the nebulizer. The nebulizer must only be operated
with the gas pressure setting stated on the certificate that was shipped
with the nebulizer.
Preparations
1. Connect the power cord to the power-inlet connector on the
detector.
2. Turn on the detector with its main power switch.
Follow these steps
1. Check if the gas pressure on the nebulizer certificate corresponds to
the pressure set in the detector. For instructions, refer to the
Operating Manual for the detector.
♦ If the gas pressure corresponds to the gas pressure on the
nebulizer certificate, continue with step 3, page 13.
♦ If the gas pressure deviates from the pressure on the nebulizer
certificate, adjust the gas pressure. See next step.
2. Adjust the nebulizer gas pressure to the pressure that is specified on
the nebulizer certificate. Monitor the gas pressure as it reaches the
setpoint.
♦ Vanquish VH-/VF-D20 detectors:
In Chromeleon, on the Command window for the detector, set
the nebulizer gas pressure (Gas Regulator Pressure Nominal).
♦ Corona Veo RS detectors:
On the detector display, navigate to the Gas Regulator screen.
Set nebulizer gas pressure under Setpoint.