Campbell Scientific CS650 and CS655 Owner's manual

Category
Measuring, testing & control
Type
Owner's manual

This manual is also suitable for

INSTRUCTION MANUAL
CS650 and CS655 Water
Content Reflectometers
Revision: 2/16
Copyright © 2011- 2016
Campbell Scientific, Inc.
Limited Warranty
Products manufactured by CSI are warranted by CSI to be free from defects
in materials and workmanship under normal use and service for twelve months
from the date of shipment unless otherwise specified in the corresponding
product manual. (Product manuals are available for review online at
www.campbellsci.com.) Products not manufactured by CSI, but that are resold
by CSI, are warranted only to the limits extended by the original manufacturer.
Batteries, fine-wire thermocouples, desiccant, and other consumables have no
warranty. CSI’s obligation under this warranty is limited to repairing or
replacing (at CSI’s option) defective Products, which shall be the sole and
exclusive remedy under this warranty. The Customer assumes all costs of
removing, reinstalling, and shipping defective Products to CSI. CSI will return
such Products by surface carrier prepaid within the continental United States of
America. To all other locations, CSI will return such Products best way CIP
(port of entry) per Incoterms ® 2010. This warranty shall not apply to any
Products which have been subjected to modification, misuse, neglect, improper
service, accidents of nature, or shipping damage. This warranty is in lieu of all
other warranties, expressed or implied. The warranty for installation services
performed by CSI such as programming to customer specifications, electrical
connections to Products manufactured by CSI, and Product specific training, is
part of CSI's product warranty. CSI EXPRESSLY DISCLAIMS AND
EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE. CSI hereby disclaims,
to the fullest extent allowed by applicable law, any and all warranties and
conditions with respect to the Products, whether express, implied or
statutory, other than those expressly provided herein.
Assistance
Products may not be returned without prior authorization. The following
contact information is for US and international customers residing in countries
served by Campbell Scientific, Inc. directly. Affiliate companies handle
repairs for customers within their territories. Please visit
www.campbellsci.com to determine which Campbell Scientific company serves
your country.
To obtain a Returned Materials Authorization (RMA), contact CAMPBELL
SCIENTIFIC, INC., phone (435) 227-9000. After an application engineer
determines the nature of the problem, an RMA number will be issued. Please
write this number clearly on the outside of the shipping container. Campbell
Scientifics shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____
815 West 1800 North
Logan, Utah 84321-1784
For all returns, the customer must fill out a Statement of Product Cleanliness
and Decontamination form and comply with the requirements specified in it.
The form is available from our website at www.campbellsci.com/repair. A
completed form must be either emailed to repair@campbellsci.com or faxed to
(435) 227-9106. Campbell Scientific is unable to process any returns until we
receive this form. If the form is not received within three days of product
receipt or is incomplete, the product will be returned to the customer at the
customers expense. Campbell Scientific reserves the right to refuse service on
products that were exposed to contaminants that may cause health or safety
concerns for our employees.
Safety
DANGER MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND
TRIPODS, TOWERS, AND ANY ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS, CROSSARMS, ENCLOSURES,
ANTENNAS, ETC. FAILURE TO PROPERLY AND COMPLETELY ASSEMBLE, INSTALL, OPERATE, USE, AND MAINTAIN TRIPODS,
TOWERS, AND ATTACHMENTS, AND FAILURE TO HEED WARNINGS, INCREASES THE RISK OF DEATH, ACCIDENT, SERIOUS
INJURY, PROPERTY DAMAGE, AND PRODUCT FAILURE. TAKE ALL REASONABLE PRECAUTIONS TO AVOID THESE HAZARDS.
CHECK WITH YOUR ORGANIZATION'S SAFETY COORDINATOR (OR POLICY) FOR PROCEDURES AND REQUIRED PROTECTIVE
EQUIPMENT PRIOR TO PERFORMING ANY WORK.
Use tripods, towers, and attachments to tripods and towers only for purposes for which they are designed. Do not exceed design
limits. Be familiar and comply with all instructions provided in product manuals. Manuals are available at www.campbellsci.com or
by telephoning (435) 227-9000 (USA). You are responsible for conformance with governing codes and regulations, including safety
regulations, and the integrity and location of structures or land to which towers, tripods, and any attachments are attached. Installation
sites should be evaluated and approved by a qualified engineer. If questions or concerns arise regarding installation, use, or
maintenance of tripods, towers, attachments, or electrical connections, consult with a licensed and qualified engineer or electrician.
General
Prior to performing site or installation work, obtain required approvals and permits. Comply
with all governing structure-height regulations, such as those of the FAA in the USA.
Use only qualified personnel for installation, use, and maintenance of tripods and towers, and
any attachments to tripods and towers. The use of licensed and qualified contractors is highly
recommended.
Read all applicable instructions carefully and understand procedures thoroughly before
beginning work.
Wear a hardhat and eye protection, and take other appropriate safety precautions while
working on or around tripods and towers.
Do not climb tripods or towers at any time, and prohibit climbing by other persons. Take
reasonable precautions to secure tripod and tower sites from trespassers.
Use only manufacturer recommended parts, materials, and tools.
Utility and Electrical
You can be killed or sustain serious bodily injury if the tripod, tower, or attachments you are
installing, constructing, using, or maintaining, or a tool, stake, or anchor, come in contact with
overhead or underground utility lines.
Maintain a distance of at least one-and-one-half times structure height, 20 feet, or the distance
required by applicable law, whichever is greater, between overhead utility lines and the
structure (tripod, tower, attachments, or tools).
Prior to performing site or installation work, inform all utility companies and have all
underground utilities marked.
Comply with all electrical codes. Electrical equipment and related grounding devices should
be installed by a licensed and qualified electrician.
Elevated Work and Weather
Exercise extreme caution when performing elevated work.
Use appropriate equipment and safety practices.
During installation and maintenance, keep tower and tripod sites clear of un-trained or non-
essential personnel. Take precautions to prevent elevated tools and objects from dropping.
Do not perform any work in inclement weather, including wind, rain, snow, lightning, etc.
Maintenance
Periodically (at least yearly) check for wear and damage, including corrosion, stress cracks,
frayed cables, loose cable clamps, cable tightness, etc. and take necessary corrective actions.
Periodically (at least yearly) check electrical ground connections.
WHILE EVERY ATTEMPT IS MADE TO EMBODY THE HIGHEST DEGREE OF SAFETY IN ALL CAMPBELL SCIENTIFIC PRODUCTS,
THE CUSTOMER ASSUMES ALL RISK FROM ANY INJURY RESULTING FROM IMPROPER INSTALLATION, USE, OR
MAINTENANCE OF TRIPODS, TOWERS, OR ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS, CROSSARMS,
ENCLOSURES, ANTENNAS, ETC.
i
Table of Contents
PDF viewers: These page numbers refer to the printed version of this document. Use the
PDF reader bookmarks tab for links to specific sections.
1. Introduction ................................................................ 1
2. Precautions ................................................................ 1
3. Initial Inspection ......................................................... 1
4. QuickStart ................................................................... 2
5. Overview ..................................................................... 4
6. Specifications ............................................................. 6
6.1 Electrical Specifications ....................................................................... 7
6.1.1 Current .......................................................................................... 7
6.2 Operational Specifications ................................................................... 8
7. Installation .................................................................. 9
7.1 Orientation and Placement ................................................................... 9
7.2 Proper Insertion .................................................................................. 10
7.3 Datalogger Wiring.............................................................................. 11
7.4 Programming ...................................................................................... 11
8. Operation .................................................................. 12
8.1 A200 and Device Configuration Utility ............................................. 12
8.1.1 Using the A200 ........................................................................... 12
8.1.1.1 Driver Installation ............................................................ 12
8.1.1.2 RS-232 Wiring ................................................................. 12
8.1.1.3 Powering the Sensor ......................................................... 13
8.1.1.4 Determining which COM Port the A200 has been
Assigned ....................................................................... 13
8.1.2 Device Configuration Utility (DevConfig) .................................. 14
8.1.2.1 Settings Editor Tab ........................................................... 15
8.1.2.2 Send OS Tab..................................................................... 17
8.1.2.3 Terminal Tab .................................................................... 17
8.2 SDI-12 Measurements ........................................................................ 19
8.2.1 Use of Multiplexers .................................................................... 20
8.3 Water Content Reflectometer Method for Measuring Volumetric
Water Content ................................................................................. 20
8.3.1 Description of Measurement Method.......................................... 20
8.3.2 The Topp Equation ..................................................................... 21
8.3.3 Electrical Conductivity ............................................................... 21
8.3.3.1 Soil Electrical Conductivity ............................................. 21
8.3.3.2 Temperature Correction of Soil Electrical
Conductivity ................................................................. 22
8.3.4 Error Sources in Water Content Reflectometer Measurement .... 22
8.3.4.1 Probe-to-Probe Variability Error ...................................... 22
Table of Contents
ii
8.3.4.2
Insertion Error ................................................................. 22
8.3.5 Temperature Dependence and Correction .................................. 22
8.3.5.1 Accurate Soil Temperature Measurement ....................... 23
8.4 Water Content Reflectometer User-Calibration ................................ 23
8.4.1 Need for Soil Specific Calibration Equation .............................. 23
8.4.2 User-Derived Calibration Equation ............................................ 23
8.4.3 Collecting Laboratory Data for Calibration ............................... 24
8.4.4 Collecting Field Data for Calibration ......................................... 26
8.4.5 Calculations ................................................................................ 28
9. Maintenance and Troubleshooting ......................... 29
10. References ................................................................ 30
Appendices
A. Importing Short Cut Code Into CRBasic Editor ... A-1
B. Example Programs ................................................. B-1
B.1 CR1000 Programs ........................................................................... B-1
B.1.1 CR1000 with One CS650 Probe ............................................... B-1
B.1.2 CR1000 with Two CS650 Probes on Same Control Port ......... B-2
B.1.3 CR1000 with 12 CS650 Probes on Multiplexer ....................... B-2
B.2 CR200X with Three CS650 Probes ................................................. B-4
C. Discussion of Soil Water Content ......................... C-1
D. SDI-12 Sensor Support .......................................... D-1
D.1 SDI-12 Command Basics ................................................................ D-1
D.2 Changing the SDI-12 Address Using Terminal Emulator and a
Datalogger .................................................................................... D-3
D.2.1 SDI-12 Transparent Mode ........................................................ D-3
D.2.2 CR200(X) Series Datalogger Example .................................... D-4
D.2.3 CR1000 Datalogger Example ................................................... D-5
Figures
5-1. CS650 (left) and CS655 Water Content Reflectometers ..................... 5
6-1. CS650 and CS655 average current drain ............................................. 8
7-1. CS650G Insertion Guide Tool ........................................................... 10
8-1. A200 Sensor-to-PC Interface ............................................................ 13
D-1. SDI-12 transparent mode on CR200(X)-series datalogger using
control port C1/SDI12 and changing SDI-12 address from 0
to 1 ............................................................................................... D-4
D-2. SDI-12 transparent mode on CR1000 datalogger using control
port 1 and changing SD1-12 address from 3 to 1 ......................... D-5
Table of Contents
iii
Tables
6-1. Size Specifications ............................................................................... 6
6-2. Relative Dielectric Permittivity Specifications .................................... 8
7-1. CS650 Wiring Code for SDI-12 ......................................................... 11
8-1. CS650 Wiring Code for RS-232 and A200 ........................................ 13
8-2. Real-Time Measurements .................................................................. 16
8-3. CS650 Terminal Commands .............................................................. 18
8-4. CS650 SDI-12 Commands ................................................................. 19
9-1. Symptom, Cause, and Solutions ......................................................... 29
B-1. CR1000 Wiring for One Probe Example Program ........................... B-1
B-2. CR1000 Wiring for Two Probe Example Program .......................... B-2
B-3. CR1000 Wiring For Multiplexer Example Program ........................ B-3
B-4. CR200(X) Wiring for Example Program ......................................... B-4
D-1. CS650 SDI-12 Command and Response Set .................................. D-1
CRBasic Examples
B-1. CR1000 with One CS650 Probe ...................................................... B-1
B-2. CR1000 with Two CS650 Probes on Same Control Port ................. B-2
B-3. CR1000 with 12 CS650 Probes on Multiplexer ............................... B-3
B-4. CR200X with Three CS650 Probes ................................................. B-4
Table of Contents
iv
1
CS650 and CS655 Water Content
Reflectometers
1. Introduction
The CS650 and CS655 are multiparameter smart sensors that use innovative
techniques to monitor soil volumetric water content, bulk electrical
conductivity, and temperature. They output an SDI-12 signal that many of our
dataloggers can measure.
The CS650 has 30 cm length rods, whereas the CS655 has 12 cm length rods.
This manual uses CS650 to reference model numbers CS650 and CS655.
Unless specifically stated otherwise, information in the manual applies equally
to both models.
This manual provides information only for CRBasic dataloggers.
It is also compatible with most of our retired Edlog dataloggers.
For Edlog datalogger support, see an older manual at
www.campbellsci.com/old-manuals or contact a Campbell
Scientific application engineer for assistance.
2. Precautions
READ AND UNDERSTAND the Safety section at the front of this
manual.
Although the CS650 is rugged, it should be handled as precision scientific
instrument.
External RF sources can affect the probe’s operation. Therefore, the probe
should be located away from significant sources of RF such as ac power
lines and motors.
3. Initial Inspection
Upon receipt of the CS650, inspect the packaging and contents for
damage. File damage claims with the shipping company.
The model number and cable length are printed on a label at the
connection end of the cable. Check this information against the shipping
documents to ensure the expected product and cable length are received.
NOTE
CS650 and CS655 Water Content Reflectometers
2
4. QuickStart
Short Cut is an easy way to program your datalogger to measure the CS650 and
assign datalogger wiring terminals. Short Cut is available as a download on
www.campbellsci.com and the ResourceDVD. It is included in installations of
LoggerNet, PC200W, PC400, or RTDAQ.
The following procedure shows using Short Cut to program the CS650.
1. Open Short Cut. Click New Program.
2. Select Datalogger Model and Scan Interval (default of 5 seconds is OK
for most applications). Click Next.
CS650 and CS655 Water Content Reflectometers
3
3. Under the Available Sensors and Devices list, select the Sensors |
Meterological | Soil Moisture | CS650/CS655 Water Content
Reflectometer. Four options are available that monitor different
parameters. In this tutorial, we’ll select CS650/CS655 Water Content
Reflectometer (VWC, EC, T, P, PA, and VR). Click to move the
selection to the Selected device window. The soil temperature defaults to
degree C, and the sensor is measured hourly. These can be changed by
clicking the Temperature or Measure sensor box and selecting one of
the other options.
CS650 and CS655 Water Content Reflectometers
4
4. After selecting the sensor, click Wiring Diagram to see how the sensor is
to be wired to the datalogger. The wiring diagram can be printed now or
after more sensors are added.
5. Select any other sensors you have, then finish the remaining Short Cut
steps to complete the program. The remaining steps are outlined in Short
Cut Help, which is accessed by clicking on Help | Contents |
Programming Steps.
6. If LoggerNet, PC400, RTDAQ, or PC200W is running on your PC, and the
PC to datalogger connection is active, you can click Finish in Short Cut
and you will be prompted to send the program just created to the
datalogger.
7. If the sensor is connected to the datalogger, as shown in the wiring
diagram in step 4, check the output of the sensor in the datalogger support
software data display to make sure it is making reasonable measurements.
5. Overview
The CS650 measures volumetric water content, electrical conductivity,
dielectric permittivity, and temperature of soils or other porous media. These
values are reported through SDI-12 communication.
CS650 and CS655 Water Content Reflectometers
5
FIGURE 5-1. CS650 (left) and CS655 Water Content Reflectometers
Volumetric water content information is derived from the probe’s sensitivity to
the dielectric permittivity of the medium surrounding the probe stainless-steel
rods. The CS650 is configured as a water content reflectometer, with the two
parallel rods forming an open-ended transmission line. A differential oscillator
circuit is connected to the rods, with an oscillator state change triggered by the
return of a reflected signal from one of the rods. The two-way travel time of the
electromagnetic waves that are induced by the oscillator on the rod varies with
changing dielectric permittivity. Water is the main contributor to the bulk
dielectric permittivity of the soil or porous media, so the travel time of the
reflected wave increases with increasing water content and decreases with
decreasing water content, hence the name water content reflectometer. The
average travel time of the reflected wave multiplied by a scaling factor of 128
is called the period average. Period average is reported in microseconds and is
considered to be the raw output of a water content reflectometer.
Electrical conductivity is determined by exciting the rods with a known non-
polarizing waveform and measuring the signal attenuation. Signal attenuation
is reported as a dimensionless voltage ratio, which is the ratio of the excitation
voltage to the measured voltage along the sensor rods when they are excited at
a fixed 100 kHz frequency. Voltage ratio ranges from 1 in non-conductive
media to about 17 in highly conductive media. Values greater than 17 are
highly unstable and indicate that the soil conditions are outside of the specified
operating range of the sensor.
Temperature is measured with a thermistor in contact with one of the rods.
It is well known that transmission line oscillators used for water content
measurements suffer from unwanted increases in oscillation period as
increasing electrical conductivity causes transmission line signal attenuation.
The CS650 handles this problem by making an electrical conductivity
measurement and then correcting the oscillator period accordingly. On-board
processing within the sensor head calculates electrical conductivity from the
signal attenuation measurement and combines the result with the oscillation
period measurement to calculate the dielectric permittivity of the media and
CS650 and CS655 Water Content Reflectometers
6
finally applies the Topp equation (Topp et al. 1980) to estimate volumetric
water content.
Probe electronics are encapsulated in the rugged epoxy probe head.
A five conductor cable including the drain or shield wire is used to provide
power and ground as well as serial communication with the CS650. The CS650
is intended to communicate with SDI-12 recorders, including Campbell
Scientific dataloggers. The orange Rx wire can be used to communicate by
means of RS-232 Tx/Rx. The A200 USB-to-Serial Module allows RS-232
serial communication between a computer and the CS650 by means of
Campbell Scientific’s Device Configuration Utility (DevConfig) software.
6. Specifications
Features:
Larger sample volume reduces error
Measurement corrected for effects of soil texture and electrical
conductivity
Estimates soil-water content for a wide range of mineral soils
Versatile sensormeasures dielectric permittivity, bulk electrical
conductivity (EC), and soil temperature
Compatible with Campbell Scientific CRBasic dataloggers: CR6,
CR200(X) series, CR800 series, CR1000, CR3000, and CR5000
TABLE 6-1 compares the size of the CS650 and CS655 reflectometers.
TABLE 6-1. Size Specifications
CS650 CS655
Rods
300 mm long
3.2 mm diameter
32 mm spacing
120 mm long
3.2 mm diameter
32 mm spacing
Probe Head
L 85 mm
W 63 mm
D 18 mm
L 85 mm
W 63 mm
D 18 mm
Probe Weight
280 g 240 g
Cable Weight
35 g m
-1
35 g m
-1
CS650 and CS655 Water Content Reflectometers
7
6.1 Electrical Specifications
Sensor Output:
SDI-12
Serial RS-232
Warmup Time:
3 s
Measurement Time:
3 ms to measure
600 ms to complete SDI-12 command
Power Supply
Requirements:
6 Vdc to 18 Vdc
Must be able to supply 45 mA @ 12 Vdc
Maximum Cable Length:
610 m (2000 ft) combined length for 1 10
sensors connected to the same datalogger
control port
Electromagnetic
Compatibility:
Πcompliant (EMC compliant performance
criteria available upon request)
Meets EN61326 requirements for protection
against electrostatic discharge and surge
External RF sources can affect CS650
measurements. CS650 circuitry should be
located away from radio transmitter aerials
and cables, or measurements ignored during
RF transmissions.
6.1.1 Current
Active (3 ms):
45 mA typical @ 12 Vdc
(80 mA @ 6 Vdc, 35 mA @ 18 Vdc)
Quiescent:
135 µA @ 12 Vdc
Average Current Drain:
I = 0.09n + [3.5 + 0.024(n-1)]n/s
I = average current in milliamps
n = number of CS650’s
s = number of seconds between
measurements
(see
FIGURE 6-1)
CS650 and CS655 Water Content Reflectometers
8
FIGURE 6-1. CS650 and CS655 average current drain
FIGURE 6-1 shows average current drain for different measurement rates and
quantities of CS650 probes. If the time between measurements is five minutes
or longer, average current drain may be approximated at 0.15 milliamps per
sensor.
6.2 Operational Specifications
TABLE 6-2 provides the operational specifications.
TABLE 6-2. Relative Dielectric Permittivity Specifications
CS650 CS655
Relative Dielectric Permittivity
Range
1 to 81
Accuracy
1 to 40:
40 to 80:
±(2% of reading + 0.6)
for solution EC
3 dS/m
±1.4 for solution EC
3 dS/m
±(3% of reading + 0.8)
for solution EC
8 dS/m
±2 for solution EC
2.8 dS/m
Precision
<0.02
CS650 and CS655 Water Content Reflectometers
9
Volumetric Water Content using Topp Equation (m
3
/m
3
)
Range
5% to 50%
Accuracy
±3% VWC typical in
mineral soils where
solution EC 3 dS/m
±3% VWC typical in
mineral soils where
solution EC 10 dS/m
Precision
<0.05%
Electrical Conductivity
Range Solution EC
0 to 3 dS/m 0 to 8 dS/m
Range Bulk EC
0 to 3 dS/m 0 to 8 dS/m
Accuracy
±(5% of reading + 0.05 dS/m)
Precision
0.5% of BEC
Temperature
Soil Measurement Range
10 to 70 °C
Operational Range
0 to 70 °C
Accuracy
±0.5 °C for probe body buried in soil
Precision
±0.02 °C
Sensing Volume*
7800 cm
3
3600 cm
3
*Sensing Volume approximately 7.5 cm radius around each probe rod and 4.5 cm
beyond the end of the rods
Accuracy specifications are based on laboratory measurements in a series of
solutions with dielectric permittivities ranging from 1 to 81 and solution electrical
conductivities ranging from 0 to 3 dS/m.
Precision describes the repeatability of a measurement. It is determined for the
CS650 by taking repeated measurements in the same material. The precision of the
CS650 is better than 0.05 % volumetric water content and 0.01 dS/m electrical
conductivity.
7. Installation
If you are programming your datalogger with Short Cut, skip Section 7.3,
Datalogger Wiring
(p. 11), and Section 7.4, Programming (p. 11). Short Cut does
this work for you. See Section 4, QuickStart
(p. 2), for a Short Cut tutorial.
7.1 Orientation and Placement
The CS650 measures the bulk dielectric permittivity, average volumetric water
content, and bulk EC along the length of the rods, which is 30 cm for the
CS650 and 12 cm for the CS655. The probe rods may be inserted vertically
into the soil surface or buried at any orientation to the surface. The probe may
be installed horizontal to the surface to detect the passing of wetting fronts or
other vertical water fluxes.
CS650 and CS655 Water Content Reflectometers
10
The sensitive volume depends on the surrounding media. In soil, the sensitive
volume extends approximately 7.5 cm (3 in) from the rods along their length
and 4.5 cm (1.8 in) beyond the end of the rods. Consequently, if the probe is
buried horizontally closer than 7.5 cm from the soil surface, it will include air
above the surface in its measurements and underestimate soil water content.
The thermistor used to measure temperature is in contact with one of the
stainless steel rods at the base of the epoxy probe body. Because of the low
thermal conductivity of stainless steel, the thermistor does not measure the
average temperature along the rod, but instead provides a point measurement of
the temperature within the epoxy. For a valid soil temperature reading, the
probe body must be in thermal equilibrium with the soil. If the probe is
installed vertically with the epoxy probe body above the surface, then the probe
body must be shielded from solar radiation and in direct contact with the soil or
media of interest.
7.2 Proper Insertion
The method used for probe installation can affect the accuracy of the
measurement. The probe rods should be kept as close to parallel as possible
when installed to maintain the design wave guide geometry. The probe is more
sensitive to permittivity close to the rods so probes inserted in a manner which
generates air voids around the rods will have reduced measurement accuracy.
In most soils, the soil structure will recover from the disturbance during probe
insertion.
In some applications, installation can be improved by using the CS650G
insertion guide tool (FIGURE 7-1). The CS650G is inserted into the soil and
then removed. This makes proper installation of the water content
reflectometer easier in dense or rocky soils.
FIGURE 7-1. CS650G Insertion Guide Tool
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56

Campbell Scientific CS650 and CS655 Owner's manual

Category
Measuring, testing & control
Type
Owner's manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI