Top Membrane Filter Replacement -
1. Remove the Top Filter Cap from the Membrane Dryer assembly by
turning the cap counter clockwise.
2. Remove the Top Membrane Filter and discard.
3. Install the replacement Top Membrane Filter by carefully inserting the
filter over the nozzle.
4. Remove the Top Cover O-ring from the Top Filter Cap and discard.
5. Install the replacement Top Cover O-ring onto the Top Filter Cap.
6. Secure the Top Filter Cap to the Membrane Dryer assembly by turning
the cap clockwise until tight.
Bottom Membrane Filter Replacement -
1. Remove the Filter Bowl from the Membrane Dryer assembly by
carefully turning the bowl counter clockwise.
2. Remove the Bottom Membrane Filter from the Membrane Dryer
assembly by pushing the filter in while turning a quarter turn counter
clockwise. Discard the filter.
3. Install the replacement Bottom Membrane Filter into the Membrane
Dryer as follows:
a. Align the four filter grooves with the four locking tabs inside the
Membrane Dryer.
b. Push in on the filter while turning a quarter turn clockwise.
4. Remove the Filter Bowl O-ring from the bottom threads of the
Membrane Dryer assembly and discard o-ring.
5. Install replacement Filter Bowl O-ring onto the Membrane Dryer.
6. Make sure the Filter Bowl interior is empty of any moisture. Drain and
towel dry as necessary.
7. Install the dry Filter Bowl onto the Membrane Dryer by turning the bowl
clockwise until tight.
Compressor Air Intake Filter Replacement -
1. Using a slight twisting motion, carefully pull and remove the Compressor
Air Intake Filter with the sound reducer attached from the associated
compressor head. Discard the filter.
2. Install the replacement Compressor Air Intake Filter onto the sound
reducer.
Important: Use care when installing intake filter. The filter is made
of paper and can be damaged
3. Install the Compressor Air Intake Filter and sound reducer by care-
fully inserting the filter into the compressor head.
Operational Check -
Checking that the replacement AirStar compressor units with a Membrane Dryer Membrane Dryer are properly installed and the
AirStar under repair may be returned to service is easily accomplished by operating the compressor as normal and verifying
that the unit reaches approximately 115 PSI of pressure and then shuts off automatically.
Sound
Reducer
Compressor Air
Intake Filter
Compressor
Head
Membrane
Dryer
Assembly
Membrane
Dryer
Assembly
Top
Membrane
Filter
Top Cover O-ring
Filter Bowl
Filter Bowl
O-ring
Bottom
Membrane Filter
Bottom
Membrane
Filter
© Air Techniques, Inc. Copyright 2012 • P/N 87365, Rev. D • December 2016