LI-COR Pearl Macroscopic Imaging System Owner's manual

Category
Supplementary music equipment
Type
Owner's manual
CE Marking:
This product (model number 8500-01) is a CE-marked product. For conformity information, contact LI-COR
Support at http://www.licor.com/biotechsupport. Outside of the U.S., contact your local sales office or
distributor.
珍珠三部曲(型号9430/8500)小动物成像系统
Pearl Trilogy (Model 9430) and Pearl MSI (8500) Small Animal Imaging System
部件名称
Part Name
有毒有害物质或元素
Toxic and Hazardous Substances or Elements
(PB)
(Hg)
(Cd)
六价铬
(Cr(VI))
多溴联苯
(PBB)
多溴二苯醚
(PBDE)
印刷电路板
(PCBs) X O O O O O
机电配件
(Electromechanical Parts) X O O O O O
光学
(Optics) OOOOOO
电缆和电线
(Cables and Wires) X O O O O O
金属部件
(Metal Parts) X O X O O O
塑料零件
(Plastic Parts) OOOOO X
电池
(Batteries) OOOOOO
本表格依据 SJ/T 11364 的规定编制
O = 表示该有害物质在该部件所有均质材料中的含量均在 GB/T 26572 规定的限量要求以下
indicates that the content of the toxic and hazardous substance in all the Homogeneous Materials of the part is below the concentration limit
requirement as described in GB/T 26572
X = 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB/T 26572 规定的限量要求
indicates that the content of the toxic and hazardous substance in at least one Homogeneous Material of the part exceeds the concentration limit
requirement as described in GB/T 26572
Notes on Safety
LI-COR products have been designed to be safe when operated in the manner described in this manual. The
safety of this product cannot be guaranteed if the product is used in any other way than is specified in this
manual. The Pearl® Imager is intended to be used by qualified personnel. Read this entire manual before using
the Pearl Imager.
Equipment Markings:
The product is marked with this symbol when it is necessary for you to refer to the
manual or accompanying documents in order to protect against damage to the product.
The product is marked with this symbol when a hazardous voltage may be present.
Manual Markings:
WARNING Warnings must be followed carefully to avoid bodily injury.
CAUTION Cautions must be observed to avoid damaging your equipment.
NOTE Notes contain additional information and useful tips.
IMPORTANT Information of importance to prevent procedural mistakes in the operation of the
equipment or related software. Failure to comply may result in a poor experimental
outcome but will not cause bodily injury or equipment damage.
Federal Communications Commission Radio Frequency Interference Statement
WARNING: This equipment generates, uses, and can radiate radio frequency energy and if not installed in
accordance with the instruction manual, may cause interference to radio communications. It has been tested
and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC
rules, which are designed to provide a reasonable protection against such interference when operated in a
commercial environment. Operation of this equipment in a residential area is likely to cause interference in
which case the user, at his own expense, will be required to take whatever measures may be required to
correct the interference.
Page 4
Notice
The information contained in this document is subject to change without notice. LI-COR MAKES NO
WARRANTY OF ANY KIND WITH REGARD TO THIS MATERIAL, INCLUDING, BUT NOT LIMITED TO
THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
LI-COR shall not be liable for errors contained herein or for incidental or consequential damages in connection
with the furnishing, performance, or use of this material.
This document contains proprietary information that is protected by copyright. All rights are reserved. No part
of this document may be photocopied, reproduced, or translated to another language without prior written
consent of LI-COR, Inc.
The Pearl MSI is not for clinical diagnostic use.
Printing History
Publication Number: 977-19835
Revised: 2022-03-22
LI-COR Biosciences is an ISO 9001:2015 certified company. © 2022 LI-COR, Inc. Specifications subject to
change. LI-COR, Pearl, In-Cell Western, IRDye, Odyssey, and Image Studio are registered trademarks of LI-COR,
Inc. in the United States and other countries. All other trademarks belong to their respective owners. The
Odyssey Infrared Imaging Systems and IRDye reagents are covered by U.S. patents, foreign equivalents, and
patents pending.
Global Headquarters
4647 Superior Street • Lincoln, NE 68504
Toll free: 800-645-4267 • Fax: +1-402-467-0819 • [email protected] licor.com/bio
Regional Offices
LI-COR Biosciences GmbH
Siemensstraße 25A • 61352 Bad Homburg • Germany
Phone: +49 (0) 6172 17 17 771 • Fax: +49 (0) 6172 17 17 799 • [email protected]
LI-COR Biosciences UK Ltd.
St. John’s Innovation Centre • Cowley Road
Cambridge • CB4 0WS
United Kingdom • Phone: +44 (0) 1223 422104 • Fax: +44 (0) 1223 422105 • [email protected]
LI-COR Biotechnology Distributor Network: www.licor.com/bio/distributors
Page 5
Chapter 1: Installation and Safety Information
Safety Considerations 9
Laser Safety 9
Safety Interlocks 10
Placement in the Laboratory 11
Space Requirements 12
Level the Imager 12
Moving the Imager 12
Chemical Safety 12
Use of DMSO and Other Strong Solvents 12
Electrical Considerations 13
Power Cords and Disconnecting Power 13
Fuse Replacement 13
Chapter 2: System Overview
External Panels and Controls 15
Front Panel 15
Right-side Panel 16
Rear Panel 17
Imaging Drawer 18
Imaging Drawer Safety 18
Installing the Imaging Stage 19
Computer Connections and Networking 19
Optical System Description 20
References 22
Chapter 3: Getting Started
Install ImageStudio™ Software 23
Windows 7, 8, 8.1, or 10 Operating Systems 23
Mac OS X 10.9, 10.10, 10.11, or 10.12 Operating Systems 23
Start ImageStudio Software 24
Set Up a Work Area 24
Add a Work Area 24
Remove a Work Area 25
Pearl® MSI
Table of Contents
Page 6
Connect to the Instrument 25
Acquire Image 25
Setup group 26
Heater Plate group 26
Channels group 26
Camera Controls group 26
Camera group 26
Adjust Image Display 27
Choose Display 27
Adjust Display 27
Finer Adjustments 27
Compare Images by Linking Lookup Tables 27
Analyze Image 28
Set the Background Shape 29
Define the Region of Interest Using the Auto Shape Tool 29
Export Data 30
Chapter 4: Routine Maintenance
Routine Maintenance 33
Cleaning the Fan Filter 33
Desiccant Replacement 34
Pearl® Imager Installation Instructions 34
Front Panel Warning Light Indicates an Error 36
Resetting the Instrument 36
Obtaining Technical Support 36
Exporting ImageStudio Software Log Files 36
Chapter 5: Appendices
Specifications 37
Embedded Firmware Source Code 38
Page 7
Chapter 1: Installation and Safety Information
Safety Considerations
Laser Safety
The Center for Devices and Radiological Health (CDRH) was established in October, 1982, by the
U.S. Food and Drug Administration (FDA) to protect the public health in the fields of medical
devices and radiological health.
Manufacturers of products subject to performance standards under the Radiation Control for Health
and Safety Act of 1968 are required to furnish various reports to the CDRH.
The Pearl Imager is certified as a Class I laser product. This means that hazardous laser radiation is
not emitted outside the instrument. Radiation emitted inside the Pearl Imager is confined within
protective housings and external covers. A series of safety interlocks ensures that the laser beam
cannot escape during any phase of user operation.
Two lasers emit at 785 nm and the other two at 685 nm. The 785nm lasers have a peak power rating
of 1 watt. The 685nm lasers have a peak power rating of 0.5 watt. The 685 nm laser emits visible laser
radiation - direct exposure to either beam may cause eye damage. Laser radiation from each laser
uniformly illuminates the target area, which is the imaging stage (Figure 2-4) in the imaging drawer.
The CDRH implemented regulations for laser products on August 1, 1976 (CDRH radiation
performance standard 21, Code of Federal Regulations Chapter 1, Sub-chapter J). Compliance for
products marketed in the United States is mandatory. This product complies with CDRH radiation
performance standard 21 CFR Chapter 1, Sub-chapter J. Ce produit est conforme à la performance de
rayonnement CDRH chapitre CFR norme 1 Sous-chapitre J. The label that must be attached to laser
products marketed in the United States is Figure 1-1 and is located on the rear panel of the Pearl
(Figure 1-2), indicating compliance with CDRH regulations.
WARNING: Use of controls or adjustments or performance of procedures other than those specified
herein may result in hazardous radiation exposure.
AVERTISSEMENT: L'utilisation des commandes ou le réglage ou la performance des procédures
autrement qu'il a été indiqué dams le présent texte peut résulter en de dangereuses expositions aux
radiations.
Pearl® MSI
Page 9
Chapter 1: Installation and Safety Information
Figure 1-1. CDRH and IEC 60825-1:2007 regulation
compliance label. See Power Cords and Disconnecting
Power on page13 for more information about
disconnecting power before maintenance or servicing.
Figure 1-2. Location of CDRH compliance label on the
Pearl®ImagingSystem. Note: The appearance of your
instrument's back panel may vary slightly from the one
shown.
DANGER - CLASS 4 VISIBLE AND INVISIBLE LASER
RADIATION WHEN OPEN AVOID EYE OR SKIN
EXPOSURE TO DIRECT OR SCATTERED RADIATION
DANGER VARNING
DANGER - APPAREIL LASER
DE CLASSE 4. RAYONNEMENT
LASER VISIBLE ET INVISIBLE
EN CAS D’OVERTURE. EVITER
L’EXPOSITION DE L’OEIL OU DE
LA PEAU AU RAYONNEMENT
DIRECT OU DIFFUS.
VARNING – KLASS 4 SYNLIG
OCH OSYNLIG LASERSTRÅLNING
NÄR ENHETEN ÄR ÖPPEN.
UNDVIK EXPONERING AV ÖGON
OCH HUD I DIREKT ELLER
REFLEKTERAD STRÅLNING.
Figure 1-3. Class 4 laser warning label.
Safety Interlocks
Where there is a potential for class 4 laser radiation behind a given panel, the label in Figure 1-3 is
affixed to the panel. The label shown in Figure 1-3 is affixed to the Pearl Imager in two locations; one
on the rear lower access panel and one on top of the camera hood (Figure 1-4).
Figure 1-4. Location of class 4 laser warning labels located (a) on the camera hood under the top panel of the Pearl
Imager instrument enclosure, and (b) on the exterior of the rear lower access panel.
Page 10
Pearl® MSI
The Pearl Imager has two safety interlocks that prevent exposure to laser radiation when the Pearl
imaging drawer is opened during operation. Do not attempt to defeat these interlocks. The drawer
interlocks are located in the back of the instrument behind the rear lower access panel, which is also
protected by interlocks. This second set of interlocks prevents exposure to laser radiation if the rear
lower access panel is removed during operation. These interlocks are also located behind this back
access panel (Figure 1-5). Do not attempt to defeat these interlocks (Figure 1-6).
Figure 1-5. Location of laser interlock warning label behind the rear lower access panel.
DANGER - CLASS 4 VISIBLE
AND INVISIBLE LASER
RADIATION WHEN OPEN
AND INTERLOCKS
DEFEATED AVOID EYE
OR SKIN EXPOSURE TO
DIRECT OR
SCATTERED
RADIATION
VARNING – KLASS 4 SYNLIG
OCH OSYNLIG LASERSTRÅLNING
NÄR ENHETEN ÄR ÖPPEN OCH
FÖRREGLING ÅSIDOSATTS.
UNDVIK EXPONERING AV
ÖGON OCH HUD I
DIREKT ELLER
REFLEKTERAD
STRÅLNING.
DANGER - APPAREIL LASER DE
CLASSE 4. RAYONNEMENT LASER
VISIBLE ET INVISIBLE EN CAS
D’OVERTURE ET LORSQUE LE
DISPOSITIF DE SECURITE EST
NEUTRALISE. EVITER L’EXPOSITION E L’OEIL
OU DE LA PEAU AU RAYONNEMENT
DIRECT OU DIFFUS.
Figure 1-6. Laser interlock warning label.
Placement in the Laboratory
The Pearl Imaging System weighs approximately 25 kg (55 lb). The instrument should be placed on a
laboratory bench that is sufficiently sturdy to bear its weight. The bench should also be able to bear
the weight of the computer if the computer is to be installed beside the Pearl Imaging System
instrument.
Place the instrument away from sinks or other sources of water that pose a shock hazard.
Recommended operating conditions are 15-35 °C and a dew point not greater than 22 °C to prevent
condensation on the laser assemblies, which are cooled during operation.
Page 11
Chapter 1: Installation and Safety Information
Space Requirements
The Pearl Imaging System instrument requires an area approximately 41 cm W (16") x 41 cm D (16")
x 66 cm H (26"). Depth with the imaging drawer open is 63.5 cm (25”).
Level the Imager
Level the instrument to prevent the imaging drawer from scraping the lab bench or table in front of
the instrument and to keep the instrument from tilting.
The instrument can be leveled by turning the threaded metal inserts located in each of the 4 plastic
feet (Figure 1-7).
Figure 1-7. Turn feet to adjust height.
Moving the Imager
LI-COR recommends that you do not move your imager. Instead, please contact LI-COR Technical
Support for options.
LI-COR cannot be held responsible for any damage incurred while moving, transporting, or
shipping the Pearl Imager, even if an existing warranty and/or service contract applies. LI-COR
cannot be held responsible for any injury incurred while moving, transporting, or shipping
the Pearl Imager.
Chemical Safety
LI-COR Biosciences recommends that all biochemicals be handled carefully, and that safe laboratory
procedures be followed at all times. Be aware of the hazards associated with any chemical before you
begin work.
The LI-COR®Pearl Imaging System should not be used with any radioactive materials.
Use of DMSO and Other Strong Solvents
CAUTION: Strong solvents, including DMSO, will permanently discolor the imaging stage
surface. Avoid all contact. In addition, any IRDye®products in DMSO will incorporate into
the surface of the imaging stage and will be visible in images.
ATTENTION:
Les dissolvants forts, y compris DMSO, décoloreront de manière permanente la surface du lit
d'image. Évitez tout contact. En outre, tous les produits d'IRDye dans DMSO incorporeront à
la surface du lit et seront évidents dans les images.
Page 12
Pearl® MSI
Electrical Considerations
Power Cords and Disconnecting Power
The Pearl®ImagingSystem is equipped with a 3-wire grounding-type plug. This plug will only fit
into a grounding-type outlet. This is a necessary safety feature. If you are unable to insert the plug
into the outlet, you will need to replace the outlet. Do not defeat the purpose of the grounding-type plug.
Do not locate the Pearl Imaging System where the power cord will be walked on or exposed to water
or chemical spills.
The Pearl Imaging System draws a maximum of 4 amperes at 120V (see Specifications on page37). If
an extension cord is used, make sure the total of the ampere ratings on the instruments plugged into
the extension cord does not exceed the extension cord ampere rating. Also make sure the total
amperage of instruments plugged into the wall outlet does not exceed the amperage capacity for the
outlet (usually 15 or 20 amperes in the United States).
Disconnect power before maintenance or servicing.
Débrancher le pouvoir avant d'entretenir.
See Figure 1-1 on page10 for the location of these statements on the product's regulation compliance
label.
Fuse Replacement
There are no user replaceable fuses in this instrument.
Page 13
Page 14
Chapter 2: System Overview
External Panels and Controls
Front Panel
Power on/off button
Image acquisition 'in
progress' indicator (shown
unlit)
Error condition indicator
(unlit)
Imaging drawer open/close
button
Imaging drawer (closed)
Figure 2-1. Front panel features and controls of the Pearl Imager.
Power On/Off Button: Pressing this button momentarily turns the instrument on and off. Pressing
the button for 5 seconds or more cuts power to the instrument (Routine Maintenance on page33).
While the instrument is starting up or shutting down, the blue indicator light in the power button
blinks. When the blue indicator light is continuously on, the instrument is ready for operation.
Pearl® MSI
Page 15
Chapter 2: System Overview
Note: Instruments with firmware version 1.3.1 and earlier will shut down if you press the power
button, even if the image acquisition light is on. On instruments with newer firmware, the red
error indicator will illuminate if you press the power button while the image acquisition indicator
is on. Hold down the power button to force a shut down.
Green Image Acquisition Light: This light blinks during image acquisition and is continuously on
at other times, as long as communication with Image StudioSoftware is maintained. The light is
off only if communication is lost or Image Studio Software is closed.
Red Error Light: The error light illuminates to indicate that a particular process could not be
completed. See Routine Maintenance on page33.
Imaging Drawer Button: This button opens and closes the imaging drawer in normal operation.
The drawer handle on the outside of the drawer can be used to open or close the drawer in cases of
power loss or emergency.
Imaging Drawer: The imaging drawer moves the specimen into position under the camera for
imaging.
Right-side Panel
Figure 2-2. Close up of the Pearl control module showing
the power receptacle and Ethernet connectors (use only
the supplied Ethernet cable). Do not use any other
connector on this panel. Note: The appearance of your
instrument's back panel may vary slightly from the one
shown.
Page 16
Pearl® MSI
Rear Panel
Air fan inlet
Lower rear access
panel
Figure 2-3. Rear panel showing air inlet fan. Keep the inlet unblocked during operation and accessible for routine filter
cleaning (Routine Maintenance on page33).
Page 17
Chapter 2: System Overview
Imaging Drawer
Figure 2-4. The drawer front tilts forward and shifts
downward automatically when opened, making the
removable imaging stage easier to access.
Imaging Drawer Safety
Be careful not to get your fingers caught between the Pearl®ImagingSystem and the imaging drawer
when the drawer closes. The drawer is specially designed to stop if a certain level of resistance is
detected, so no harm should result, but the experience may be startling.
During an unexpected power loss, the drawer button will not function. In any case, the drawer can
be opened or closed by pulling or pushing on the drawer handle on the drawer front.
The imaging drawer should typically be opened and closed using the drawer button, but the
drawer handle can be used to manually open or close the drawer. The drawer handle is
especially useful in case of power loss or emergency.
Page 18
Pearl® MSI
Installing the Imaging Stage
1. Press the button above the drawer to open it.
2. Place the specimen on the imaging stage.
3. Place the imaging stage with the specimen in the drawer.
4. Press the button above the drawer to close it.
Computer Connections and Networking
The computer supplied with the Pearl®ImagingSystem has two Ethernet ports. During installation,
the port labeled LAN1 on the Pearl Imaging System control module (Figure 2-2) must be connected
to one of the Ethernet ports on the computer. On instruments with a serial number below 3100, the
Pearl Imaging System must be connected to the Ethernet port that is directly connected to the
computer's motherboard (see Figure 2-5).
Note: Use only the supplied Cat. 5e RJ45 Ethernet cable to connect the Pearl Imaging System and
computer.
Page 19
Chapter 2: System Overview
Figure 2-5. Ethernet
connector on the
computer motherboard
of a typical computer
(other models will vary).
When Image StudioSoftware starts, it searches for and discovers any Pearl
Imagers on the network. If only one instrument is found, Image Studio
Software establishes communication. If more than one instrument is found,
you will be presented with a list of instruments and asked which to use.
Optical System Description
Before image acquisition can be started, the imaging stage is typically locked in place inside the Pearl
Imaging System imaging drawer and the imaging drawer must be closed. An interlock prevents
operation of the laser illumination modules while the imaging drawer is open. The user can start
image acquisition via Image Studio Software, or Image Studio Software can be configured to start an
acquisition automatically each time the imaging drawer closes.
For fluorescent imaging, the Pearl Imaging System has two laser modules that work in tandem to
illuminate the imaging stage from opposite angles (Figure 2-6). Each laser module contains a 685 nm
laser, a 785 nm laser, and an LED source that produces white light. Laser illumination at near-
infrared wavelengths results in much deeper tissue penetration than white light.
During image acquisition, similar sources from each laser module are turned on simultaneously,
followed by image acquisition by the CCD detector. As each pair of similar sources are turned on and
off, the CCD acquires all three images in less than 30 seconds. A single image acquisition consists of
images for two fluorescent probes in the 700 and 800 nm channels, plus a white light image
(assuming all channels are enabled).
Page 20
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38