Invacare PLATINUM 10L is an oxygen concentrator designed to provide supplemental oxygen to patients with respiratory disorders. It uses a molecular sieve and pressure swing adsorption methodology to produce oxygen gas output. The oxygen concentration level of the output gas ranges from 87% to 95.6%. The oxygen is delivered to the patient through the use of a nasal cannula. The device is not intended to sustain or support life. The PLATINUM 10L is capable of operation by the patient in a home environment or in an institutional environment.

Invacare PLATINUM 10L is an oxygen concentrator designed to provide supplemental oxygen to patients with respiratory disorders. It uses a molecular sieve and pressure swing adsorption methodology to produce oxygen gas output. The oxygen concentration level of the output gas ranges from 87% to 95.6%. The oxygen is delivered to the patient through the use of a nasal cannula. The device is not intended to sustain or support life. The PLATINUM 10L is capable of operation by the patient in a home environment or in an institutional environment.

Invacare®Platinum™10L
OxygenConcentrator
IRC10LX
enHomeFill®SystemCompatible
UserManual
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.
©2016Invacare®Corporation
Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout
priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™and®.Alltrademarksare
ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
MakingLife’sExperiencesPossibleisatrademarkintheU.S.A.
Pine-solandLestoilareregisteredtrademarksofTheCloroxCompany.
DawnisaregisteredtrademarkofTheProctorandGambleCompany.
Contents
1General........................................4
1.1Symbols.....................................4
1.2IntendedUse..................................5
1.3IndicationsForUse.............................6
1.4Description...................................6
1.5Contraindications..............................6
2Safety..........................................7
2.1LabelLocations................................7
2.2GeneralGuidelines.............................9
2.3RadioFrequencyInterference......................12
3Components....................................13
3.1ComponentIdentification.........................13
3.2PneumaticDiagram.............................14
4Accessories.....................................15
4.1OptionalAccessories............................15
5Setup..........................................16
5.1Unpacking....................................16
5.2Inspection....................................16
5.3Storage......................................16
6Usage..........................................17
6.1Introduction..................................17
6.2SelectaLocation...............................17
6.3SettingUptheConcentrator......................18
6.3.1SettingUptheHumidifier......................18
6.3.2TurningtheConcentratorOn...................20
6.3.3Flowrate..................................21
6.3.4UnitswithoutSensO
2
........................21
7Maintenance....................................22
7.1ServiceLife...................................22
7.2CleaningtheCabinet............................22
7.3CleaningtheCabinetFilter........................23
7.4HumidifierCleaningandThermicDisinfection..........24
7.5CleaningandDisinfectionBetweenPatients............24
7.6PreventiveMaintenanceChecklist...................25
8AfterUse.......................................26
8.1RecyclingInformation............................26
8.2WearandTear................................26
9Troubleshooting.................................27
9.1Troubleshooting...............................27
10Technicaldata..................................28
10.1Specifications.................................28
11Warranty......................................31
11.1NorthAmericanLimitedWarranty.................31
Invacare®Platinum™10LOxygenConcentrator
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe
practiceswhichcouldresultinpersonalinjuryorpropertydamage.
Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious
injury.
WARNING!
Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious
injury.
CAUTION!
Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage
orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor
efficient,trouble-freeuse.
Refertothefollowingtableforthemeaningofsymbols
markedontheequipmentand/orpackaging.
SymbolsinDocumentation
GeneralWarningSign
Thebackgroundcolorinsidethe
triangleisyellowonproductlabels.
Readmanual
Thecolorofthesymbolbackground
isblueonproductlabels.
NoSmoking
Thecolorofthecirclewithdiagonal
barisredonproductlabels.
NoOpenFlame
Thecolorofthecirclewithdiagonal
barisredonproductlabels.
ClassII,DoubleInsulated
Protectedagainstsolidforeignobjectsof12.5
mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops.
IndoorUseONLY
KeepDry
4
1118353-M-00
General
-20F
MIN
150F
MAX
TransportandStorageTemperature
15%
MIN
95%
MAX
TransportandStorageHumidity
AlternatingCurrent
TypeBFequipment
Recycle
DONOTdisposeofinhouseholdwaste
ElectricalHazard
Manufacturer
Callsupplier
SymbolsonProduct
Unitrunning
Unitnotrunning
LXIndicators
SymbolO
2
PurityIndicatorLights(LED)
I/O
SYSTEMOKAYGREENIndicatorLight
SYSTEMFAILURE
REDIndicatorLight
ContinuousAudibleAlarm
SieveGARD™
CompressorShutdown
Callaqualifiedtechnician.
1.2IntendedUse
TheintendedfunctionanduseoftheInvacare®Platinum™Oxygen
Concentratoristoprovidesupplementaloxygentopatientswith
respiratorydisorders,byseparatingnitrogenfromroomair,byway
ofamolecularsieve.Itisnotintendedtosustainorsupportlife.
DANGER!
RiskofInjuryorDeath
Thisproductistobeusedasanoxygensupplementand
isnotintendedtobelife-supportingorlife-sustaining.
ONLYusethisproductifthepatientiscapableof
spontaneousbreath,abletoinhaleandexhalewithout
theuseofamachine.
DONOTuseinparallelorserieswithotheroxygen
concentratorsoroxygentherapydevices.
1118353-M-005
Invacare®Platinum™10LOxygenConcentrator
DANGER!
RiskofInjuryorDeath
Dependingontheirmedicalcondition,patientson
flowratesgreaterthan5L/minmaybeatanincreased
riskforseriousinjuryordeathintheeventoffailure.
ALWAYSdiscussthisincreasedriskwithyourhealth
caresupplierBEFOREusingthisproductifyouare
prescribedaflowrategreaterthan5L/min.
WARNING!
RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduseand
specificationshasnotbeentestedandmayleadto
productdamage,lossofproductfunction,orpersonal
injury.
DONOTusethisproductinanywayotherthan
describedintheSpecificationsandIntendedUse
sectionsofthismanual.
DANGER!
RiskofInjuryorDeath
WhileInvacarestrivestoproducethebestoxygen
concentratorinthemarkettoday,thisoxygen
concentratorcanfailtoproduceoxygenduetopower
failureordevicemalfunction.
ALWAYShaveabackupsourceofoxygenreadily
available.
Intheeventtheconcentratorfailstoproduceoxygen,
theconcentratorwillbrieflyalarmsignalingthepatient
toswitchtotheirbackupsourceofoxygen.Referto
Troubleshootingformoredetail.
1.3IndicationsForUse
TheintendedfunctionanduseoftheInvacare®Platinum™Oxygen
Concentratoristoprovidesupplementaloxygentopatientswith
respiratorydisorders,byseparatingnitrogenfromroomair,byway
ofamolecularsieve.Itisnotintendedtosustainorsupportlife.
1.4Description
TheInvacarePlatinumconcentratorisusedbypatientswith
respiratorydisorderswhorequiresupplementaloxygen.Thedevice
isnotintendedtosustainorsupportlife.
Theoxygenconcentrationleveloftheoutputgasrangesfrom87%
to95.6%.Theoxygenisdeliveredtothepatientthroughtheuse
ofanasalcannula.
TheInvacarePlatinumconcentratorusesamolecularsieveand
pressureswingadsorptionmethodologytoproducetheoxygen
gasoutput.Ambientairentersthedevice,isfilteredandthen
compressed.Thiscompressedairisthendirectedtowardoneof
twonitrogenadsorbingsievebeds.
Concentratedoxygenexitstheoppositeendoftheactivesievebed
andisdirectedintoanoxygenreservoirwhereitisdeliveredtothe
patient.
TheInvacarePlatinumconcentratoriscapableofoperationbythe
patientinahomeenvironmentorinaninstitutionalenvironment.
Deviceoperatesatanominal120VAC/60Hertzsupply.
Serviceinformationwillbeavailableuponrequesttoqualified
technicalpersonnelonly.
1.5Contraindications
Therearenoknowncontraindications.
61118353-M-00
Safety
2Safety
2.1LabelLocations
A
SerialNumberLabelislocatedontheresonator
intakeassembly.
B
C
SpecificationandDoubleInsulationLabelislocated
onthebackoftheconcentratoratthebase
1118353-M-007
Invacare®Platinum™10LOxygenConcentrator
IRC10LX
D
81118353-M-00
Safety
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyandoperationofthe
concentratortheseinstructionsMUSTbefollowed.
DANGER!
RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath,injuryor
damage.Thissectioncontainsimportantinformationfor
thesafeoperationanduseofthisproduct.
DONOTusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand
understandingtheseinstructionsandanyadditional
instructionalmaterialsuchasusermanuals,service
manualsorinstructionsheetssuppliedwiththis
productoroptionalequipment.
Ifyouareunabletounderstandthewarnings,cautions
orinstructions,contactahealthcareprofessional,
dealerortechnicalpersonnelbeforeattemptingto
usethisequipment.
CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproductisnot
workingcorrectly,contactatechnicianorInvacare
forrepair.
THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER!
RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,greasy
substancesandothercombustiblesareeasilyignited
andburnwithgreatintensityinoxygenenrichedairand
whenincontactwithoxygenunderpressure.Toavoid
fire,death,injuryordamage:
DONOTSMOKEwhileusingthisdevice.
DONOTusenearOPENFLAMEorIGNITION
SOURCES.
DONOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
NOSMOKINGsignsshouldbeprominentlydisplayed.
Avoidcreationofanysparknearoxygenequipment.
Thisincludessparksfromstaticelectricitycreatedby
anytypeoffriction.
Keepallmatches,lightedcigarettes,electronic
cigarettesorothersourcesofignitionoutofthe
roominwhichthisconcentratorislocatedandaway
fromwhereoxygenisbeingdelivered.
Keeptheoxygentubing,cord,andconcentratorout
fromundersuchitemsasblankets,bedcoverings,
chaircushions,clothing,andawayfromheatedorhot
surfacesincludingspaceheaters,stoves,andsimilar
electricalappliances.
CAUTION!
Federal(statutory)lawrestrictsthisdevicesaletoor
ontheorderofamedicalpractitionerlicensedbya
governmentalagencywherehe/shepractices.
ONLYalicensedmedicalpractitionermayorderthe
purchaseoruseofthisdevice.
1118353-M-009
Invacare®Platinum™10LOxygenConcentrator
DANGER!
RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,deathor
injurytopersons:
DONOTdisassemble.Referservicingtoqualified
servicepersonnel.Therearenouserserviceable
parts.
Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the
concentratormustbelocatedinanotherroomat
least7ft(2.1m)fromthebath.
DONOTcomeincontactwiththeconcentrator
whilewet.
DONOTplaceorstoreconcentratorwhereitcan
dropintowaterorotherliquid.
DONOTreachforconcentratorthathasfalleninto
water.UnplugIMMEDIATELY.
DONOTusefrayedordamagedACpowercords.
WARNING!
RiskOfInjuryOrDamage
Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare
accessories.Accessoriesdesignedbyother
manufacturershavenotbeentestedbyInvacareand
arenotrecommendedforusewithInvacareproducts.
Therearemanydifferenttypesofhumidifiers,oxygen
tubing,cannulasandmasksthatcanbeusedwiththis
device.Youshouldcontactyourlocalhomecare
supplierforrecommendationsonwhichofthese
deviceswillbebestforyou.Theyshouldalsogiveyou
adviceontheproperusage,maintenance,andcleaning.
DANGER!
RiskofInjuryorDeath
Toavoidchokingoringestionofchemicalsfromairway
contamination:
DONOTusetheconcentratorinthepresenceof
pollutants,smoke,fumes,flammableanesthetics,
cleaningagents,orchemicalvapors.
WARNING!
RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
Closelysupervisewhenthisconcentratorisusedby
ornearchildrenorimpairedindividuals.
Monitorpatientsusingthisdevicewhoareunableto
hearorseealarmsorcommunicatediscomfort.
WARNING!
RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing
entanglement:
Keepchildrenandpetsawayfromnasalcannulaand
tubing.
Closesupervisionisnecessarywhenthenasalcannula
isusedbyornearchildrenand/orimpairedpersons.
WARNING!
RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
Replacethenasalcannulaonaregularbasis.Check
withyourequipmentsupplierorphysicianto
determinehowoftenthecannulashouldbereplaced.
DONOTsharecannulasbetweenpatients.
101118353-M-00
Safety
WARNING!
RiskofInjury
Achangeinaltitudemayaffecttotaloxygenavailableto
you.Topreventoxygendeprivation:
Consultyourphysicianbeforetravelingtohigheror
loweraltitudestodetermineifyourflowsettings
shouldbechanged.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamagefromcordmisuse:
DONOTmoveorrelocateconcentratorbypulling
onthecord.
DONOTuseextensioncordswithACpowercord
provided.
Properlystoreandpositionelectricalcordsand/or
tubingtopreventatrippinghazard.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
NEVERleaveconcentratorunattendedwhenplugged
in.
Makesureconcentratorisoffwhennotinuse.
WARNING!
RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned
tominimizeroutinepreventivemaintenance.Toprevent
injuryordamage:
Onlyprofessionalsofthehealthcarefieldor
personsfullyconversantwiththisprocesssuchas
factorytrainedpersonnelshouldperformpreventive
maintenanceorperformanceadjustmentsonthe
oxygenconcentrator,exceptfortasksdescribedin
thismanual.
UsersshouldcontactyourdealerorInvacarefor
service.
CAUTION!
RiskofDamage
Topreventdamagefromliquidingress:
Iftheconcentratorisnotworkingproperly,ifit
hasbeendroppedordamaged,ordroppedinto
water,callequipmentsupplier/qualifiedtechnicianfor
examinationandrepair.
NEVERdroporinsertanyobjectorliquidintoany
opening.
ForIndoorUseONLY.
1118353-M-00
11
Invacare®Platinum™10LOxygenConcentrator
2.3RadioFrequencyInterference
WARNING!
RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamagefrom
interferencewithwirelessequipment:
Keepconcentratoratleast9.8ft(3.0m)awayfrom
wirelesscommunicationequipmentsuchaswireless
homenetworkdevices,mobilephones,cordless
phonesandbasestations,walkie-talkies,etc.
ThisequipmenthasbeentestedandfoundtocomplywithEMClimits
specifiedbyIEC/EN60601-1-2.Theselimitsaredesignedtoprovide
areasonableprotectionagainstelectromagneticinterferenceina
typicalmedicalinstallation.
Otherdevicesmayexperienceinterferencefromeventhelowlevels
ofelectromagneticemissionspermittedbytheabovestandards.
Todetermineiftheemissionsfromtheconcentratorarecausing
theinterference,turntheconcentratorOff.Iftheinterference
withtheotherdevice(s)stops,thentheconcentratoriscausingthe
interference.Insuchrarecases,interferencemaybereducedor
correctedbyoneofthefollowingmeasures:
Reposition,relocate,orincreasetheseparationbetweenthe
equipment.
Connecttheequipmentintoanoutletonacircuitdifferentfrom
thattowhichtheotherdevice(s)isconnected.
12
1118353-M-00
Components
3Components
3.1ComponentIdentification
A
OxygenOutlet
B
Flowmeter
C
FaultandPowerIndicatorLights
D
PowerSwitch
E
CircuitBreaker
F
ElapsedTimeMeter
G
CabinetFilter
RearView
H
PowerCord
I
HomeFillOutletFitting
Accessories(NotShown):HomeFIllhomeoxygen
compressorIOH200
TheHomeFilloutletfittingIistobeusedonlyfor
fillingoxygencylinderswiththeHomeFillhomeoxygen
compressor.Theoutletfittingdoesnotaffectconcentrator
performance.RefertotheHomeFIllusermanual,part
number1100873,forconnectionandoperatinginstructions.
Whennotinuse,theplugprovidedwiththeconcentrator
shouldbeinsertedintotheoutletfitting.Formore
informationabouttheHomeFill,contactyourInvacare
dealer.
1118353-M-0013
Invacare®Platinum™10LOxygenConcentrator
3.2PneumaticDiagram
14
1118353-M-00
Accessories
4Accessories
4.1OptionalAccessories
Therearemanydifferenttypesofhumidifiers,oxygentubing,cannulas
andmasksthatcanbeusedwiththisdevice.Youshouldcontactyour
localhomecareproviderforrecommendationsonwhichofthese
deviceswillbebestforyou.Theyshouldalsogiveyouadviceonthe
properusage,maintenance,andcleaning.
Thesupplyaccessories(nasalcannula,mask,tubing,humidifier,etc.)
usedtodeliveroxygentothepatientneedtoincludeameansto
reducethepropagationoffireintheaccessoriesforthesafetyofthe
patientandothers.Ifacommerciallyavailable,fire-activatedflow
stopdeviceisusedintheaccessoriessetup,itshouldbeplacedas
closetothepatientaspossible.
StandardAdultNasalCannula,7ft(2.1m)M3120
StandardHumidifier-InvacarePart#1155719
HomeFillHomeOxygenCompressor-IOH200
OxygenSupplyTubing7ft(2.1m)-MS4107
OxygenSupplyTubing21ft(6.3m)-MS4121
OxygenTubingConnector-MS4301
1118353-M-0015
Invacare®Platinum™10LOxygenConcentrator
5Setup
5.1Unpacking
1.Checkforanyobviousdamagetothecartonoritscontents.If
damageisevident,notifythecarrier,oryourlocaldealer.
2.Removeallloosepackingfromthecarton.
3.Carefullyremoveallthecomponentsfromthecarton.
Unlesstheoxygenconcentratoristobeused
IMMEDIATELY,leaveconcentratorinitspackagingfor
storageuntiluseisrequired.
5.2Inspection
1.Inspect/examineexterioroftheconcentratorfornicks,dents,
cracks,scratchesorotherdamage.
2.Inspectallcomponents.
5.3Storage
1.Storetherepackagedoxygenconcentratorinadryarea.Refer
toSpecificationsforstoragetemperatureparameters.
2.DONOTplaceobjectsontopofrepackagedconcentrator.
161118353-M-00
Usage
6Usage
6.1Introduction
Youroxygenconcentratorisintendedforindividualuseindoors.
Itisanelectronicallyoperateddevicethatseparatesoxygenfrom
roomair.Itprovideshighconcentrationofoxygendirectlytoyou
throughanasalcannula.Clinicalstudieshavedocumentedthat
oxygenconcentratorsaretherapeuticallyequivalenttoothertypesof
oxygendeliverysystems.
Yoursupplierwillshowyouhowtouseyouroxygenconcentrator.
He/Sheshouldbecontactedwithanyquestionsorproblems
regardingyouroxygenconcentrator.Thisusermanualwilltellyou
aboutyourconcentratorandwillserveasareferenceasyouuse
yourconcentrator.
6.2SelectaLocation
WARNING!
RiskofInjury
Toavoidinjuryduringtherapy:
TheoxygenconcentratorMUSTbeplacedonalevel
surfaceforuse.
DONOTrelocatetheoxygenconcentratorwhilein
use.
WARNING!
RiskofInjuryorDamage
Toavoidinjuryordamagefromairbornepollutants
and/orfumesandforoptimalperformance:
Locateandpositiontheconcentratorinawell
ventilatedspacesothattheairintakeandtheair
exhaustsarenotobstructed.
NEVERblocktheairopeningsoftheconcentrator
orplaceitonasoftsurface,suchasabedorcouch,
wheretheairopeningmaybeblocked.
Keeptheopeningsfreefromlint,hairandsimilar
foreignitems.
Keepconcentratoratleast12in(30.5cm)awayfrom
walls,draperiesandfurniture.
Avoiduseinpresenceofpollutants,smokeorfumes,
flammableanesthetics,cleaningagentsorchemical
vapors.
Placeconcentratorinawellventilatedareatoavoid
airbornepollutantsand/orfumes.
DONOTuseinacloset.
Youmayselectaroominyourhousewhereusingyouroxygen
concentratorwouldbemostconvenient.Yourconcentratorcanbe
easilyrolledfromroomtoroomonitscasters.
Youroxygenconcentratorwillperformbestwhenoperatedunder
theconditionsoutlinedinSpecifications.
Usageinenvironmentsotherthanthosedescribedmayresultinthe
needforincreasedequipmentmaintenance.Theairintakeofthe
unitshouldbelocatedinawellventilatedareatoavoidairborne
pollutantsand/orfumes.
1118353-M-0017
Invacare®Platinum™10LOxygenConcentrator
6.3SettingUptheConcentrator
1.Pluginpowercordtoanelectricaloutlet.
2.ConnectHumidifier(ifsoprescribed).
6.3.1SettingUptheHumidifier
WARNING!
RiskofInjuryorDamage
Toavoidburnsfromsteamorhotwater,inhalationof
waterand/orwaterdamagetotheconcentrator:
DONOTfillhumidifierbottlewithhotwater.Allow
boiledwatertocooltoroomtemperaturebefore
filling.
DONOToverfillhumidifier.
Replacethehumidifiercapandsecurelytighten.
Confirmthatthecapisnotcross-threadedonthe
humidifierbottle.
DONOTreversetheoxygeninputandoutput
connections.Waterfromthehumidifierbottlewill
travelthroughthecannulabacktothepatientifinput
andoutputconnectionsarereversed.
Whenusingtubingconnectionslongerthan7ft
(2.1m)inlength,positionthehumidifierasclose
tothepatientaspossibletoallowformaximum
humidificationoutput.
WARNING!
RiskofInjuryorDamage
Failuretoproperlyinstallthehumidifierbottleorother
accessoriestotheconcentratorwillimpacttheflowof
oxygen.
Tocheckforproperoperationoftheoxygen
concentratorandaccessories,placetheendofthe
nasalcannulaunderthesurfaceofahalf-fullcupof
wateronlywhentheoxygenisflowingandlookfor
bubblesindicatingproperoperation.Wipeanyexcess
wateroffthenasalprongswhenfinishedbeforeplacing
onthepatient.
Iftherearenobubbles,checkallconnections
(includinghumidifierbottleandotheraccessories,if
applicable)andrepeat.Contactyourdealerorservice
supplierimmediatelyifbubblesstilldonotappear.
181118353-M-00
Usage
HumidifierBottle
withCap
HumidifierBottlewithoutCap
1.RemovecapAfrombottleB.
2.Fillhumidifierbottlewithboiledtapwaterorbottledwater
tothelevelindicatedbythemanufacturer.Boiltapwaterfor
approximatelytenminutesandcooltoroomtemperatureprior
touse.
CAUTION!
RiskofDamage
Confirmthehumidifiercapisnotcross-threaded
onthehumidifierbottle.
3.InsertaflatheadscrewdriverintheplategrooveConthetop
edgeofthefilteraccessdoorDandgentlyprythefilteraccess
dooroff.
4.PullupandremovethehumidifierbottleadapterE(nextto
inletfilter).
WARNING!
RiskofInjury
Toavoidimpactingtheflowofoxygen:
Installthehumidifierbottleand/orother
accessoriesproperly.
1118353-M-0019
Invacare®Platinum™10LOxygenConcentrator
5.AttachthehumidifierbottleadapterEtothehumidifier
bottleBbyturningthewingnutGonthehumidifierbottle
counterclockwiseuntilitissecurelyfastened.
6.WhenplacingthehumidifierbottleinthecompartmentFon
theconcentrator,attachthehumidifierbottle/adaptertubeHto
theoxygenoutletconnectorIontheconcentrator.
Whenusingtubingconnectionslongerthan7ft(2.1m)
inlength,positionthehumidifierasclosetothepatient
aspossibletoallowformaximumhumidificationoutput.
7.Attachthepatient’snasalcannulasupplytubetothehumidifier
bottleoutletJ.
8.Afterassembly,ensurethatoxygenisflowingthroughthenasal
cannula.
9.Replacethefilteraccessdooronthesideofconcentratorby
snappingitbackintotheplategroove.
6.3.2TurningtheConcentratorOn
1.PresspowerswitchAtoOnposition.
Allthepanellightsandtheaudiblealarmwillcome
onforonesecond,confirmingthattheindicatorsare
functioningproperly.
201118353-M-00
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36

Invacare PLATINUM 10L User manual

Category
Oxygen Equipment
Type
User manual
This manual is also suitable for

Invacare PLATINUM 10L is an oxygen concentrator designed to provide supplemental oxygen to patients with respiratory disorders. It uses a molecular sieve and pressure swing adsorption methodology to produce oxygen gas output. The oxygen concentration level of the output gas ranges from 87% to 95.6%. The oxygen is delivered to the patient through the use of a nasal cannula. The device is not intended to sustain or support life. The PLATINUM 10L is capable of operation by the patient in a home environment or in an institutional environment.