UV Lighting DC-LD-25 User guide

Type
User guide

UV Lighting DC-LD-25 is a pulsed-xenon UV disinfection device designed to eliminate pathogens and microorganisms using high-intensity ultraviolet (UV-C) light. Its cordless and portable nature allows for easy maneuverability and disinfection of hard-to-reach areas in hospitals, clinics, laboratories, and other settings. With a single cycle disinfection time of 5 minutes, it efficiently disinfects surfaces and objects, including those made of plastic, metal, and fabric, making it ideal for use in various healthcare and commercial applications.

UV Lighting DC-LD-25 is a pulsed-xenon UV disinfection device designed to eliminate pathogens and microorganisms using high-intensity ultraviolet (UV-C) light. Its cordless and portable nature allows for easy maneuverability and disinfection of hard-to-reach areas in hospitals, clinics, laboratories, and other settings. With a single cycle disinfection time of 5 minutes, it efficiently disinfects surfaces and objects, including those made of plastic, metal, and fabric, making it ideal for use in various healthcare and commercial applications.

Pulsed-Xenon UV Dosimeter
User Guide
Anatomy:
Irradiance Level
Indicator
3M Adhesive
Protective Strip
UV Dosimeter
15
mJ/cm2
25
mJ/cm2
Date Time
Operator Location
Check mark will appear when exposed to 5mJ/cm2 of UV
PULSED XENON
UV DOSIMETER INSTRUCTIONS
PULSED XENON
The UV dose indicators are designed to provide a visual indication
of applied UV energy.
1) Place the unexposed card(s) at desired location in the area.
Cards can be used in horizontal or vertical positions. Always
make sure the color changing area faces towards the UV source.
2) As the center yellow indicator is exposed to UV, it will gradually
change colors based on exposure or UV dosage. For accurate
results, record the exposure level immediately after treatment.
3) If your results don’t match the target dosage shade exactly, for
recording purposes use the lesser value.
For more information including a
short list of common pathogens
and the dosimeter reading that
would indicate that energy
exceeding a lethal dose for that
pathogen has been delivered,
please refer to uvlighting.com/
dosimeter
Dosimeter Card is not intended to reflect exact UV dosage measurements,
but is designed as visual reference for UV treated areas confirming
treatment has occurred with UV light. This dosimeter is intended as
a visual reference of accumulated UV dose only, and does not determine
level of cleanliness. A card that has been pre-exposed to light should be
disposed of and not used.
For more information, visit uvlighting.com/dosimeter
QR Code for
Dosimeter
Webpage.
More detailed
instructions and
further guidance
on understanding
results.
Quick
Instructions
Irradiance
Level Color
Reference
Testing Record
Journal
1) First identify desired disinfection areas within your space. Once identied ensure
that your pulsed-Xenon product has a direct light path to your chosen disinfection
area. The examples below show a pulsed-Xenon unit directed towards the desk
and bed as disinfection targets within a room.
Figure 1.
Pulsed-Xenon unit angled towards desk. Ideal placement for dosimeter is circled.
2 3
Figure 2. Figure 4.
Figure 5.
Figure 3.
Pulsed-Xenon unit angled towards bed. Ideal placement for dosimeter is circled. Correct – Chair has been moved no longer obstructing light. Dosimeter on side of
ling cabinet has been removed as well.
Three dosimeter cards prepared for testing on bed.
Incorrect - Objects obstructing light and improper placement of dosimeter card.
2) Prepare selected surfaces in step one for testing. Remove anything that may obstruct
the light pathway to the target surface. Avoid shadowing and indirect exposure to
dosimeter card surface.
3) Remove the 3M protective strip from the top of the back of the cards to expose
adhesive. Place cards adhesive side down on target surfaces selected in step one.
a. At minimum place one card per target surface.
b. To understand irradiance over time we recommend three cards per target
surface.
Dosimeter on side of filing
cabinet will receive indirect light
and provide low readings.
Chair will obstruct light from
pulsed-Xenon unit, casting shadow
over cards and lower readings.
4 5
Figure 6.
Figure 7.
Three dosimeter cards prepared for testing on desk.
Example photo of multiple cards tested at one location for easy reference of irradiance
over time.
4) With cards adhered to target surfaces, power on pulsed-Xenon unit and step
out of the room.
5) Once testing is complete, and cards are retrieved, mark the following: Date, Test
Time Elapsed, Operator, and Location on irradiated cards. Take pictures of the
cards for records as chemical reaction will reverse over time changing results.
Do not reuse cards.
a. If testing one card, re-enter room after disinfection cycle has completed and
retrieve card.
b. If testing multiple cards re-enter room after disinfection cycle has completed
and retrieve one card. Leave the other cards and recommence the disinfection
cycle. Repeat until all cards/disinfection cycles are complete.
a. If multiple dosimeter cards are tested, it is recommended to take a photo
of all tested cards for easy reference of irradiance over time.
15
mJ/cm2
25
mJ/cm2
Date Time
Operator Location
Check mark will appear when exposed to 5mJ/cm2 of UV
UV Dosimeter
PULSED XENON
15
mJ/cm2
25
mJ/cm2
Date Time
Operator Location
Check mark will appear when exposed to 5mJ/cm2 of UV
UV Dosimeter
PULSED XENON
15
mJ/cm2
25
mJ/cm2
Date Time
Operator Location
Check mark will appear when exposed to 5mJ/cm2 of UV
UV Dosimeter
PULSED XENON
John Doe OR-204
14:18
07/24
John Doe OR-204
14:22
07/24
John Doe OR-204
14:36
07/24
  • Page 1 1
  • Page 2 2
  • Page 3 3

UV Lighting DC-LD-25 User guide

Type
User guide

UV Lighting DC-LD-25 is a pulsed-xenon UV disinfection device designed to eliminate pathogens and microorganisms using high-intensity ultraviolet (UV-C) light. Its cordless and portable nature allows for easy maneuverability and disinfection of hard-to-reach areas in hospitals, clinics, laboratories, and other settings. With a single cycle disinfection time of 5 minutes, it efficiently disinfects surfaces and objects, including those made of plastic, metal, and fabric, making it ideal for use in various healthcare and commercial applications.