Omega SMSC-2 is a dual-channel signal conditioner that provides two independent, fully isolated DC outputs proportional to two DC inputs. It offers wide-ranging input and output capability for scaling and transmitting analog DC signals. The SMSC-2 accepts input voltage spans from 10mV to 100 volts and input current spans from 1mA to 100mA. It provides six configurable output ranges: 0-5V, 0-10V, 0-1mA, 4-20mA, -5 to 5V, and -10 to 10V. The SMSC-2 is ideal for eliminating ground loops, converting signal levels, and providing signal drive.

Omega SMSC-2 is a dual-channel signal conditioner that provides two independent, fully isolated DC outputs proportional to two DC inputs. It offers wide-ranging input and output capability for scaling and transmitting analog DC signals. The SMSC-2 accepts input voltage spans from 10mV to 100 volts and input current spans from 1mA to 100mA. It provides six configurable output ranges: 0-5V, 0-10V, 0-1mA, 4-20mA, -5 to 5V, and -10 to 10V. The SMSC-2 is ideal for eliminating ground loops, converting signal levels, and providing signal drive.

Specications
Input:
VoltageInput(eldcongurable):
 FullScaleRange:10mVto100V
  Impedance:>100KOhms
 Overvoltage:
400Vrms,max(Intermittent);
264Vrms,max(Continous)
CurrentInput(eldcongurable):
 FullScaleRange:1mAto100mA
  Impedance:20Ohms,typical
 Overcurrent:170mArms,max
 Overvoltage:60VDC
CommonMode(InputtoGround):
1500VDC,max
Zero Turn-Up:
50%offullscalerange
Span Turn-Down:
50%offullscalerange
Output:
VoltageOutput:
 Output:0-5V,0-10V
Drive:10mA,max.(1KOhmsmin.load
@10V)
CurrentOutput:
 Output:0-1mA,4-20mA
Compliance:
0-1mA:10V,max(10KOhmsmaxload)
4-20mA:20V,max(1KOhmsmaxload)
LED Indication (green):
InputRange:
>110%input:8Hzash
<-10%input:4Hzash
Accuracy (Including Linearity, Hysteresis):
<20mV,<2mA:±0.35%offullscale,
typical,0.5%,max
>20mV,>2mA:±0.1%offullscale,typical,
0.2%,max
Response Time:
(10-90%)200mSec.,typical
Stability (Temperature):
±0.025%offullscale/°C,typical,
±0.05%/°C,max.
Common Mode Rejection:
DCto60Hz:120dB
Isolation (Input to Output):
1500VDCbetweenchannels,input,
outputandpower
Pin Connections
1 Power(Hot)
2 SpareTermination
3 Power(Neu)
4 OutputB(+)
5 InputA(+)
6 InputA(-)
7 OutputA(+)
8 OutputA(-)
9 OutputB(-)
10 InputB(+)
11 InputB(-)
ESD Susceptibility:
Level2(4KV)
Humidity (Non-Condensing):
Operating:15to95%(@45°C)
Soak:90%for24hours(@65°C)
Temperature Range:
Operating:-15to60°C(5to140°F)
Storage:-25to70°C(-13to158°F)
Power:
Consumption:
4Wtypical,6Wmax
Standard:
selectable120/240VAC,±10%,50-
60Hz
Weight: 0.66lbs
Approvals:
ULrecognizedperstandardUL508
(FileNo.E150323/E99775).
Servicing North America:
U.S.A.:
Omega Engineering, Inc., One Omega Drive,
P.O. Box 4047, Stamford, CT 06907-0047 USA
Toll-Free: 1-800-826-6342 (USA & Canada only)
Customer Service: 1-800-622-2378
(USA & Canada only)
Engineering Service: 1-800-872-9436
(USA & Canada only)
Tel: (203) 359-1660
Fax: (203) 359-7700
e-mail: info @omega.com
For Other Locations Visit
omega.com/worldwide
omega.com info@omega.
com
The information contained in this document is
believed to be correct, but OMEGA accepts no
liability for any errors it contains, and reserves
the right to alter specifications without notice.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC., warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase.
OMEGAs WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping
time. This ensures that OMEGAs customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGAs Customer Service Department will issue an Authorized Return
(AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired
or replaced at no charge. OMEGAs WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to
mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if
the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current,
heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGAs control. Components
in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions
or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA,
either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA
MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND
ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY
DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to
this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of
the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used
in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear
installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic
WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage
whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA,
PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following
information available BEFORE contacting OMEGA.
1. Purchase order number which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to
the product.
OMEGAs policy is to make running changes, not model changes, whenever an improvement is possible. This affords customers the latest
technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© COPYRIGHT 2015 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, translated, or reduced to
any electronic medium-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
FOR NON-WARRANTY RETURNS, consult OMEGA for current repair
charges. Have the following information available BEFORE contacting
OMEGA.
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product and
3. Repair instructions and/or specific problems relative to the product.
Description
The eld congurable SMSC-2 series dual channel isolators
offerwideranginginputandoutputcapabilityforscalingand
transmittinganalogDCsignals.TheSMSC-2serieswillaccept
inputvoltagespansfrom10mVupto100volts,aswellasinput
currentspansfrom1mAto100mA.Forafullscaleoutputrange,
theinputzeroandspanpotentiometersenable50%inputzero
andspanadjustability.Forexample,the0-10Vinputrangecan
beelevatedto5-10Vorcompressedto0-5V.
TheSMSC-2offers4popularouputrangeswhicharepositive
voltagesandcurrents(e.g.0-5V,0-10V,0-1mAand4-20mA).
Themodelnumberdenestheouputchannelrangesasshown
inTable1.
The4-20mAcomplianceisapowerful20VDCperchannel.The
SMSC-2 accepts bipolar inputs and each I/O channel offers
selectablenormalorreverseactingoperation(e.g.4-20mAor
20-4mA).
EachSMSC-2isadual,three-port,industrialisolator--both
outputchannelsareopticallyisolatedfromtheirrespectiveinput
channelsupto1500VDC.ThetwoASIC-basedI/Ochannels
are independently transformer isolated from the selectable
120/240VACpowersupply.
ProvidesTwoIndependent,FullyIsolatedDCOutputsinProportiontoTwoDCInputs
 HighDensity2-ChannelPackage
 EliminatesGroundLoopswith1500VIsolation
 SixCongurableOutputRanges:0-5V,0-10V,
0-1mA,4-20mA,-5to5Vand-10to10V
 Plug-inInstallation
 Selectable120/240VACInputPower
Table1:SMSC-2SeriesOutputRanges
(Bold indicatesfactorypresetranges)
Application
TheSMSC-2eldcongurableisolatorsisusefulineliminating
groundloops,convertingsignallevelsandprovidingsignaldrive.
TheSMSC-2series'dualchanneldesignconservesinstallation
space in high density applications and offers superior cost-
benetvalueoversinglechannelisolators.Thewiderangingca-
pabilityoftheSMSC-2providesuniversalsparepartcoverage.
Diagnostic LED
TheSMSC-2seriesisequippedwithdualfunctionLEDsignal
monitors.Thegreen,top-mounted LEDindicateslinepower
andinputsignal status. Activeline power isindicatedby an
illuminatedLED.Iftheinputsignalis10%abovethefullscale
range,theLEDwillashat8Hz.Below0%,theashrateis4Hz.
Option
U
 Urethanecoatingofinternalcircuitryforprotection
fromcorrosiveatmospheres.
Conguration
EachchanneloftheSMSC-2seriescanbeindependentlyset
forawidevarietyofinputandoutputranges.
Factory Presets
ThefactorypresetstheSMSC-2inputandoutputto4-20mA(as
showninTable1andFigure1).Thesupplypoweriscongured
for120VACoperation.ForotherI/Oranges,removethefourbase
screwsandcasetoaccesstheI/Ocards.
Refer to gures 1 & 2 for conguration and program the I/O
channelsasdesired.
Replacethecoverbeforeapplyingpower.
Input
1.Positioninputjumper“W1”forCurrent(I)orVoltage(V)input.
SMSC-2
A
B
0-5V, 0-10V, 0-1mA, 4-20mA
0-5V, 0-10V, 0-1mA, 4-20mA
Model
Channel Output Selections
Current
Voltage
W1
W1
SMSC-2
DC Input, Dual Channel
Signal Conditioner
M5482/0715
INSTRUCTION
SHEET
Shop online at omega.com
SM
For latest product manuals: www.omegamanual.info
2.Set position 5 of theInput RangeSelector for Unipolar or
Bipolarinputoperation.
Note:AbipolarrangeselectionwilldoubleanyrangefromTable
2(e.g,10Vspan=±10Vbipolarspan)
3.Setposition6oftheInputRangeSelectorforNormalorRe-
verseoperation.Reverseactingproducesadecreasingoutput
withanincreasinginput.
4. Using Table 2, congure positions 1 through 4 of the Input
RangeSelectorforthedesiredmaximuminput.Roundthedesired
maximuminputvaluetothenexthighestrange(e.g.,0-70V=100V
range).
WARNING:Donotchangeswitchsettingswithpowerapplied.
Severedamagewillresult.
Output
1.FortheSMSC-2channelsAandBuseTable3tocongure
the output selector switches for one of the four(4) standard
outputranges.
Power
1.ConguretheACjumpersforeither120or240VACopera-
tion.SeeFigure3.
Calibration
1.ConnecttheinputtoacalibratedDCvoltageorcurrentsource
andapplypower.RefertoPINCONNECTIONS.Wait1hourfor
thermalstabilitybeforemonitoringthevoltage/currentoutput.
2.Setthecalibratortothedesiredminimuminputandadjustthe
Zeropotentiometerforthedesiredminimumoutput.
3.Setthecalibratortothedesiredmaximuminputandadjust
theSpanpotentiometerforthedesiredmaximumoutput.
4.Repeatsteps2and3forbestaccuracy.
I/O Card Conguration
Unipolar
Bipolar
1
2 3
4
6
5
>
>
1
2 3
4
6
5
ON
ON
>
>
Normal
Reverse
1
2 3
4
6
5
1
2 3 4
6
5
Table2:SMSC-2SeriesInputRanges
Table3:SMSC-2OutputRanges
Figure1:SMSC-2I/Ocardfactorycalibration:
4-20mAinputand4-20mAoutput
*Usejumper(W1)tocongureeithervoltageorcurrentinput.
Allunipolarrangesarezerobased.
Input Range
Selector (SW1)
Voltage* Current*
20mV 2mA
50mV 5mA
100mV 10mA
200mV 20mA
500mV 50mA
1V 100mA
2V
5V
10V
25V
50V
100V
>
>
>
1
2 3 4
>
ON
>
>
>
>
>
>
5
6
>
>
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
1
2 3
4
ON
5
6
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
0to10V
0to5V
0to1mA
4to20mA
Output Range
Selector (SW2)
Range
>
>
>
>
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
1
2 3 4
ON
5
6
W1
I/V Input Selection
Input Range
Selector (SW1)
Diagnostic
LED
Output Range
Selector (SW2)
>
ON
1
2 3
4
6
5
>
I
V
Top
Span
Zero
ON
1
2 3
4
6
5
Mounting
Allmodulesfeatureplug-ininstallation.ModelSMSC-2usesan11-pin
baseandeithermoldedsocketSKT-SM-11PorDINsocket
SKT-DR-11Pformounting.
Dimensions
Dimensionsareinmillimeters(inches)
SKT-SM-11P (Track/Surface)
SKT-DR-11P (DIN Rail)
Warning:
Donotchangewithpowerconnected!
Top View Diagrams
Warning:Donotchangeswitchsettingswithpowerapplied.
Severedamagewillresult!
Warning: ApplyingvoltagetotheinputwithW2incurrent(I)
positionwillresultindamagetotheunit.
Mark II
  • Page 1 1
  • Page 2 2

Omega SMSC-2 Owner's manual

Type
Owner's manual

Omega SMSC-2 is a dual-channel signal conditioner that provides two independent, fully isolated DC outputs proportional to two DC inputs. It offers wide-ranging input and output capability for scaling and transmitting analog DC signals. The SMSC-2 accepts input voltage spans from 10mV to 100 volts and input current spans from 1mA to 100mA. It provides six configurable output ranges: 0-5V, 0-10V, 0-1mA, 4-20mA, -5 to 5V, and -10 to 10V. The SMSC-2 is ideal for eliminating ground loops, converting signal levels, and providing signal drive.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI