eM860T True RMS Multimeter
User Manual
Version 1.4. Revised 3/19/2015.
© 2015 ennoLogic™. All rights reserved.
http://ennologic.com
TABLE OF CONTENTS
1. GENERAL INSTRUCTIONS ................................................. 1
1.1 Precautionary Safety Measures .......................................... 1
1.1.1 Preliminary ............................................................... 2
1.1.2 During Use ............................................................... 4
1.2 Battery Installation or Replacement ................................... 7
1.3 Symbols ............................................................................... 9
1.4 Instructions ......................................................................... 9
2. INSTRUMENT DESCRIPTION ........................................... 11
2.1 Main User Elements .......................................................... 11
2.2 LCD Display ....................................................................... 13
2.3 Keypad .............................................................................. 15
2.3.1 SELECT ................................................................... 15
2.3.2 HOLD/BACKLIGHT ............................................. 15
2.3.3 RANGE ................................................................... 15
2.3.4 REL ....................................................................... 16
2.3.5 Hz % ........................................................................ 16
2.3.6 MAX/MIN .............................................................. 16
3. FUNCTION DESCRIPTION ................................................ 17
3.1 General Functions ............................................................. 17
3.1.1 DATA HOLD mode .............................................. 17
3.1.2 Manual ranging and Autorange mode ............... 17
3.1.3 Battery Saver .......................................................... 18
3.1.4 Relative Measurement Mode ............................... 19
3.1.5 TRUE RMS Measurement .................................... 19
3.2 Measurement Functions ................................................... 20
3.2.1 AC and DC Voltage measurement ..................... 20
3.2.2 Resistance Measurement ...................................... 23
3.2.3 Capacitance Measurement ................................... 26
3.2.4 Continuity Check .................................................. 28
3.2.5 Diode Test .............................................................. 31
3.2.6 Frequency and Duty Cycle Measurement ......... 34
3.2.7 Temperature Measurement ................................. 35
3.2.8 Current Measurement .......................................... 36
3.2.9 NCV (Non-Contact Voltage Detect) ................... 38
4. TECHNICAL SPECIFICATIONS ........................................ 39
4.1 General specifications ....................................................... 39
4.2 Measurement Specifications ............................................ 40
4.2.1 Voltage .................................................................... 40
4.2.2 Frequency ............................................................... 42
4.2.3 Resistance ............................................................... 42
4.2.4 Diode Test .............................................................. 43
4.2.5 Continuity Check .................................................. 43
4.2.6 Capacitance ............................................................ 43
4.2.7 Temperature ........................................................... 44
4.2.8 Current .................................................................... 44
5. MAINTENANCE .................................................................. 45
5.1 General Maintenance ....................................................... 46
5.2 Fuse replacement ............................................................. 46
6. ACCESSORIES ....................................................................... 47
1
1. GENERAL INSTRUCTIONS
This instrument complies with IEC 61010-1: 2001, CAT
1000V and CAT 600V overvoltage standards. See
Specifications.
To get the best service from this instrument, read this user's
manual carefully and respect the detailed safety
precautions.
International symbols used on the Meter and in this manual
are explained in chapter 1.2.
1.1 Precautionary Safety Measures
Important Note: Limited Liability
Cascadia Innovations is the exclusive distributor of all
ennoLogic™ products. Except as explicitly stated,
Cascadia Innovations is not liable for direct, indirect,
incidental, or other types of damages arising out of, or
resulting from the use of this product. By using the
eM860T you agree to hold ennoLogic™ and Cascadia
Innovations harmless for any and all consequences of the
use of this product or application of data from the use of
this instrument.
2
1.1.1 Preliminary
As the possibilities of high transient over-voltages occurring
in today’s power systems increase, more stringent safety
standards are set for electrical test equipment. Transients on
electrical systems (power grid, feeder or branch circuits) will
trigger a series of incidents that may result in serious
personal injury. To protect you against transients, safety
must be built into the test equipment.
Overvoltage
category In brief Examples
CAT I Electronic
• Protected electronic
equipment.
• Equipment connected to
(source) circuits in
which measures are
taken to limit transient
over-voltages to an
appropriately low level.
• Any high-voltage, low-
energy source derived
from a high winding
resistance transformer,
such as the high-voltage
section of a copier.
3
CAT II
Single-phase
receptacle
connected
loads
• Appliance, portable
tools, and other household
and similar loads.
• Outlet and long branch
circuits.
• Outlets at more than 10
meters (30 feet) from CAT
III source.
• Outlets at more that 20
meters (60 feet) from CAT
IV source.
CAT III
Three-phase
distribution,
including
single-phase
commercial
lighting
• Equipment in fixed
installations, such as
switchgear and polyphase
motors.
• Bus and feeder in
industrial plants.
• Feeders and short
branch circuits,
distribution panel devices.
• Lighting systems in
larger buildings.
• Appliance outlets with
short connections to
service entrance.
4
CAT VI
Three-phase at
utility
connection,
any outdoor
conductors
• Refers to the “origin of
installation”; i.e., where
low-voltage connection
is made to utility power.
• Electricity meters,
primary overcurrent
protection equipment.
• Outside and service
entrance, service drop
from pole to building,
run between meter and
panel.
• Overhead line to
detached building,
underground line to well
pump.
When using this Multimeter, the user must observe all
normal safety rules concerning:
protection against the dangers of electric current.
protection of the Multimeter against misuse.
For your own safety, only use the test probes supplied with
the instrument. Before use, check that they are in good
condition.
1.1.2 During Use
5
If the meter is used near noise generating equipment, be
aware that display may become unstable or indicate large
errors.
Do not use the meter or test leads if they look damaged.
Use the meter only as specified in this manual; otherwise,
the protection provided by the meter may be impaired.
Use extreme caution when working around bare
conductors or bus bars.
Do not operate the meter around explosive gas, vapor, or
dust.
Verify a Meter's operation by measuring a known
voltage. Do not use the Meter if it operates abnormally.
Protection may be impaired. When in doubt, have the
Meter serviced.
Use the proper terminals, function, and range for your
measurements.
When the range of the value to be measured is unknown,
check that the range initially set on the multimeter is the
highest possible or, wherever possible, choose the
autoranging mode.
To avoid damage to the instrument, do not exceed the
maximum limits of the input values shown in the
technical specification tables.
When the multimeter is linked to measurement circuits,
do not touch unused terminals.
Caution when working with voltages above 60Vdc or 30V
AC rms. Such voltages pose a shock hazard.
6
When using the probes, keep your fingers behind the
finger guards.
When making connections, connect the common test lead
before connecting the live test lead; when disconnecting,
disconnect the live test lead before disconnecting the
common test lead.
Before changing functions, disconnect the test leads from
the circuit under test.
For all DC functions, including manual or auto-ranging,
to avoid the risk of shock due to possible improper
reading, verify the presence of any AC voltages by first
using the AC function. Then select a DC voltage range
equal to or greater than the AC range.
Disconnect circuit power and discharge all high-voltage
capacitors before testing resistance, continuity, diodes, or
capacitance.
Never perform resistance or continuity measurements on
live circuits.
Before measuring current, check the meter's fuse and turn
off power to the circuit before connecting the meter to
the circuit.
In TV repair work, or when carrying out measurements
on power switching circuits, remember that high
amplitude voltage pulses at the test points can damage
the multimeter. Use of a TV filter will attenuate any such
pulses.
7
Use just one 6F22 battery, properly installed in the
Meter's battery case, to power the Meter.
Replace the battery as soon as the battery indicator
() appears. A low battery might produce false
readings that can lead to electric shock and personal
injury.
Do not measure voltages above 1000V in Category III, or
600V in Category installations.
When in REL mode, the “REL” symbol is displayed.
Caution must be used because hazardous voltage may be
present.
Do not operate the Meter with the case (or part of the
case) removed.
1.2 Battery Installation or Replacement
To avoid false readings, which could lead to
possible electric shock or personal injury, replace
the battery as soon as the battery indicator
() appears.
Before installing or replacing the battery,
disconnect test leads and/or any connectors from
any circuit under test, turn the meter off and remove
test leads from the input terminals.
8
1. Set rotary switch to the OFF position.
2. Disconnect test leads and/or any connectors from the
terminals.
3. Remove the red protective rubber casing by carefully
peeling it away from the black plastic enclosure.
4. Use a screwdriver to unscrew the two screws on the
battery cover, on the back of the unit near the bottom.
5. Remove the battery cover from the meter.
6. If you are replacing the battery: remove the used battery
7. Install a new 9V battery (6F22).
8. Replace the battery cover and tighten the screws.
9. Attach protective rubber casing.
9
1.3 Symbols
Symbols used in this manual and on the instrument:
Caution: refer to the instruction manual.
Incorrect use may result in damage to the device
or its components.
~ AC (Alternating Current)
DC (Direct Current)
AC or DC
Earth ground
Double insulated
Fuse
Conforms to European Union directives
1.4 Instructions
Remove test leads from the Meter before opening the
Meter case or battery cover.
When servicing the Meter, use only specified replacement
parts.
Before opening up the instrument, always disconnect
from all sources of electric current and make sure you are
10
not charged with static electricity, which may destroy
internal components.
Any adjustment, maintenance or repair work carried out
on the meter while it is live should be carried out only by
appropriately qualified personnel, after having taken into
account the instructions in this manual.
A "qualified person" is someone who is familiar with the
installation, construction and operation of the equipment
and the hazards involved. He is trained and authorized to
energize and de-energize circuits and equipment in
accordance with established practices.
When the instrument is opened up, remember that some
internal capacitors can retain a dangerous potential even
after the instrument is switched off.
If any faults or abnormalities are observed, take the
instrument out of service and ensure that it cannot be
used until it has been checked out.
If the meter is not going to be used for a long time, take
out the battery and do not store the meter in high
temperature or high humidity environments.
11
2. INSTRUMENT DESCRIPTION
2.1 Main User Elements
The front panel is shown in Figure 2-1, and its user
elements are described below:
LCD display: Used for displaying measurement results
and various symbols.
Keypad: Measurement function keys.
Rotary switch: Used for selecting measurement
functions.
V
Hz
:
Connection terminal for the red test lead during
voltage, resistance, capacitance, frequency,
temperature, diode and continuity measurements.
uA/mA:
Connection terminal for the red test lead during µA
and mA measurements.
A:
Connection terminal for the red test lead during 6A
and 10A measurements.
COM:
Connection terminal for the black test lead (common
reference.)
12
Figure 2-1: Meter Front Panel
13
2.2 LCD Display
Figure 2-2: LCD Display
The LCD screen is shown in Figure 2-2 above, and the
meanings of the symbols are explained in Table 1 below:
No.
Symbol Meaning
1 Indicates negative readings
2
Indicator for AC voltage or
current
14
3
Indicator for DC voltage or
current
4 AUTO
The meter is in the Autorange
mode in which the meter
automatically selects the range
with the best resolution.
5 NCV No-contact AC Voltage detect
6 H The meter is in Data Hold
mode.
7 REL The meter is in Relative
Measurement mode.
8 MAX Display maximum data
9 MIN Display minimum data
10 Low battery indication
11 The meter is in Continuity
Check mode.
12 The meter is in Diode Test
mode.
13 %°C°F KM
Hz
nμmFAV Measurement units
14
This symbol means that the
input is too large for the
Table 1: LCD Display Symbols
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56

ennoLogic eM860T True RMS User manual

Category
Measuring, testing & control
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI