Bard DUET Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Bard DUET Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
CARE AND MAINTENANCE
The care and maintenance of the catheter requires well trained, skilled personnel following a detailed protocol. The protocol should include a
directive that the catheter is not to be used for any purpose other than the prescribed therapy.
ACCESSING CATHETER, CAP CHANGES, DRESSING CHANGES
3
• Experiencedpersonnel
• Useaseptictechnique
•Properhandhygiene
•Cleanglovestoaccesscatheterandremovedressingandsterileglovesfordressingchanges
•Surgicalmask(1forthepatientand1forthehealthcareprofessional)
• Catheterexitsiteshouldbeexaminedforsignsofinfectionanddressingsshouldbechangedateachdialysistreatment.
• CatheterLuer-lockConnectorswithendcapsattachedshouldbesoakedfor3to5minutesinpovidoneiodineandthenallowedtodry
before separation.
• Carefullyremovethedressingandinspecttheexitsiteforinammation,swellingandtenderness.Notifyphysicianimmediatelyifsignsof
infection are present.
EXIT SITE CLEANING
9
• Useaseptictechnique(asoutlinedabove)
• Cleantheexitsiteateachdialysistreatmentwithchlorhexidinegluconateunlesscontraindicated.Applyantisepticpermanufacturer’s
recommendations.Allowtoairdrycompletely.
• Covertheexitsitewithsterile,transparent,semipermeabledressingorperhospitalprotocol.
RECOMMENDED CLEANING SOLUTIONS
Catheter Luer-lock Connectors/End Caps:
•Povidoneiodine(allowconnectors/endcapstosoakfor3to5
minutes)
3
WARNING:Alcoholshouldnotbeusedtolock,soakordeclotpolyure-
thaneDialysisCathetersbecausealcoholisknowntodegradepoly-
urethanecathetersovertimewithrepeatedandprolongedexposure.
Hand cleaner solutions are not intended to be used for disinfecting
ourdialysiscatheterLuer-lockConnectors.
Exit Site:
•Chlorhexidinegluconate2%solution(preferred)
3,7,8,9,10
•Chlorhexidinegluconate4%solution
•Diluteaqueoussodiumhypochlorite
•0.55%sodiumhypochloritesolution
•Povidoneiodine
•Hydrogenperoxide
•Chlorhexidinepatches
•Bacitracinzincointmentsinpetrolatumbases
WARNING:AcetoneandPEG-containingointmentscancausefailure
of this device and should not be used with polyurethane catheters.
Chlorhexidinepatchesorbacitracinzincointments(e.g.Polysporin*
ointment)arethepreferredalternative.
POST DIALYSIS
Useaseptictechnique(asoutlinedabove).
1. Flusharterialandvenouslumenswithaminimumof10mLofsterilesaline.
WARNING: Toavoiddamagetovesselsandviscus,infusionpressuresmustnotexceed25psi(172kPa).Theuseofa10mLorlargersyringeis
recommended because smaller syringes generate more pressure than larger syringes.
2. Injectheparinsolutionintoboththearterialandvenouslumensofthecatheter.Theappropriateheparinsolutionconcentrationandushing
frequencyshouldbebasedonhospitalprotocol.Heparinsolutionof1,000to5,000units/mLhasbeenfoundtobeeectiveformaintaining
thepatencyofhemodialysisandapheresiscatheters.Wheninjectingheparinsolution,injectquicklyandclampextensionwhileunderposi-
tivepressure.Heparinsolutionvolumetolockeachlumenmustbeequaltotheprimingvolumeofeachlumen.Primingvolumesaremarked
on each lumen.
3. CleancatheterLuer-lockconnectorsperhospitalprotocol.Attachsterileendcapstoboththearterialandthevenousclampingextension
pieces.
WARNING: Topreventsystemicheparinizationofthepatient,theheparinsolutionmustbeaspiratedoutofbothlumensimmediatelyprior
tousingthecatheter.Inmostinstances,nofurtherheparinsolutioninjectionisnecessaryfor48-72hours,providedthecatheterhasnot
beenaspiratedorushed.
CATHETER REMOVAL
Evaluatethecatheterroutinelyandpromptlyremoveanynonessentialcatheter
9
perphysiciansorders.Afterremovingthecatheter,apply
manualpressuretothepuncturesitefor10-15minutesuntilnosignsofbleedingarepresent.Thenapplysterile,transparent,semipermeable
dressingordressingperhospitalprotocolforaminimumof8hours.Followhospitalprotocolregardingbedrestaftercatheterremoval.
DISPOSAL
Afteruse,thecathetersandaccessoriesmaybeapotentialbiohazard.Handleanddisposeofinaccordancewithacceptedmedical
practice and all applicable laws and regulations.
TROUBLESHOOTING
PATIENT WITH FEVER
Patientwithfeverandchillsfollowingtheproceduremaybeindicativeofcatheter-relatedsepsis.Ifthisresults,removalofthecathetermaybe
indicated.
INSUFFICIENT FLOW
Excessiveforcemustnotbeusedtoushanobstructedlumen.Insufficientbloodowmaybecausedbyanoccludedtipresultingfromaclot
orbycontactingthewallofthevein.Ifmanipulationofthecatheterorreversingarterialandvenouslinesdoesnothelp,thenthephysicianmay
attempttodissolvetheclotwithathrombolyticagent(e.g.TPA,Catho*Activase*thrombolytic).Physiciandiscretionadvised.
CATHETER EXCHANGE
Donotroutinelyreplacedialysiscatheterstopreventcatheter-relatedinfections
9
.Itmaybecomenecessarytoexchangetheindwellingcatheter
duetoapersistentriseinpressuresordecreaseofowrateswhichcannotberectiedthroughtroubleshooting.Catheterexchangesshould
be performed under strict aseptic conditions in which the physician should wear a cap, mask, sterile gown, sterile gloves, and use a large sterile
drape to cover the patient.
REFERENCES
1. Aitken,D.R.andMinton,J.P.“ThePinch-OSign:AWarningofImpendingProblemswithPermanentSubclavianCatheters”,AmericanJournal
ofSurgery,Vol.148,Nov.1984,pp.633-638.
2. Mickley,V.,“Centralvenouscatheters:manyquestions:fewanswers”,NephrolDialTransplant,(2002)17:1368-1373.
3. NationalKidneyFoundationK/DOQIGUIDELINES2006.
4. Sulek,CA.,Blas,ML.,Lobato,EB,Arandomizedstudyofleftversusrightinternaljugularveincannulationinadults.JClinAnesth.2000
Mar;12(2):142-5
5. Tan,P.L.,Gibson,M.,“CentralVenousCatheters:theroleofradiology”,ClinRad.2006,61:13-22
6. Octavio,Bella,Colemenares,Garcia,andFlores;“RightVersesLeftInternalVeinCatheterizationforHemodialysis:ComplicationsandImpact
onIpsilateralAccessCreation”;ArticalOrgans;2004,28(8):728-733.
7. TheInstituteforHealthcareImprovement,“How-to-Guide:PreventCentralLineInfections,2006.
8. TheJointCommissionHospitalAccreditationOrganization,NationalPatientSafetyGoals,2009.
9. CenterforDiseaseControlandPrevention,“GuidelinesforthePreventionofIntravascularCatheter-RelatedInfections,MorbidityandMortal-
ityWeeklyReport,Aug.9,2002,51(RR-10),1-32.
10.TheSocietyforHealthcareEpidemiologyofAmerica,“StrategiestoPreventCentralLine-AssociatedBloodstreamInfectionsinAcuteCare
Hospitals,InfectionControlandHospitalEpidemiology,Oct.2008,29(S1):S22-S30.
Otherreferencesavailableuponrequest
MRConditional
Non-clinicaltestinghasdemonstratedthedeviceisMRConditional.Itcanbescannedsafelyunder:
• staticmagneticeldof3Teslaorless
• spatialgradienteldof360Gauss/cmorless
• maximumspecicabsorptionrate(SAR)or4W/kgfor15minutesofscanning.
• Innon-clinicaltesting,thedeviceproducedatemperatureriseoflessthan1.4°Catamaximumspecicabsorptionrate
(SAR)of4W/kgfor15minutesofMRscanningina3TeslaSiemensTriowithsoftwareversionB15.
ForMinimalImageArtifact
• MRimagequalitymaybecompromisediftheareaofinterestisinthesameareaorrelativelyclosetothepositionoftheextensionlegs.
Therefore,itmaybenecessarytooptimizeMRimagingparametersforpresenceofthemetallicportionoftheextensionlegsormoveexten-
sion legs away from area of interest.
Anissuedorrevisiondatefortheseinstructionsisincludedfortheusersinformation.Intheeventtwoyearshaveelapsedbetweenthisdateand
product use, the user should contact Bard Access Systems, Inc. to see if additional product information is available.
Revisiondate:March2010.
*Bard, Duet and StatLockaretrademarksand/orregisteredtrademarksofC.R.Bard,Inc.
Allothertrademarksarethepropertyoftheirrespectiveowners.
©2010C. R. Bard, Inc. Allrightsreserved.
Bard Access Systems, Inc.
605North5600West
SaltLakeCity,UT84116U.S.A.
1-801-522-5000
CustomerService:800-545-0890
ClinicalInformation:800-443-3385
www.bardaccess.com
InstructionsForUse
For
Duet*
Long-TermHemodialysisCatheterSystem
Access Systems
DESCRIPTION
The Duet* Long-Term Hemodialysis Catheter System is made of Carbothane* radiopaque polyurethane. The catheter system is comprised of
twosingle10Frenchlumensthatallowforowratesashighas400mL/min.Eachcathetercomeswithanextensionlegandaretentioncufor
tissueingrowthtoanchorthecatheter.Thearterial(red)andvenous(blue)lumensaredistinguishedbythecolorcodingoftheprintingonthe
catheterandthecolor-codedLuer-lockconnector.
STERILIZATIONBYETHYLENEOXIDE.
Sterile,
Non-Pyrogenic
Unless Package is
Damaged or Opened.
Do not use if package is damaged
MRConditional 
Non-pyrogenic
Sterilize
This product does
notcontainDEHP
This product and packaging do not
containnaturalrubberlatex.
INDICATIONS FOR USE
The Duet*Long-termHemodialysisCatheterSystemisindicatedforuseinattainingshort-termorlong-termvascularaccessforhemodialysis,
hemoperfusion,orapheresistherapy.Thecathetersmaybeinsertedpercutaneouslyintotheinternaljugularvein,subclavian,orexternaljugular
vein.
CONTRAINDICATION
Thisdeviceiscontraindicatedforpatientsexhibitingsevere,uncontrolledthrombocytopeniaorcoagulopathy.
WARNINGS
First Rib
Subclavian Vein
Clavicle
Vertebra
Internal Jugular Vei
Superior Vena Cava
Sternum
Pinch-off Area
Infraclavicular Fossa
Axillary Vein
WARNING:Ifthesubclavianveinisused,percutaneousinsertionofthecatheter
shouldbemadeintotheaxillary-subclavianveinatthejunctionoftheouterand
mid-thirdsoftheclaviclelateraltothethoracicoutlet.Thecathetershouldnotbe
insertedintothesubclavianveinmediallybecausesuchplacementcanleadtocom-
pressionofthecatheterbetweentherstribandclavicleandcanleadtodamageor
fractureandembolizationofthecatheter.
1
Fluoroscopicorradiographicconrmation
of catheter tip placement should be helpful in demonstrating that the catheter is not
beingpinchedbytherstribandclavicle.
1
•Useofthesubclavianveinforcatheterplacementmayresultin
subclavianveinstenosis.Subclavianveinstenosismaypreventfuture
useoftheipsilateralextremityforpermanentaccess.Useofthejugular
vein is preferred.
•Alcoholoralcohol-containingantiseptics(suchaschlorhexidine)maybeusedto
cleanthecatheter/skinsite;however,careshouldbetakentoavoidprolongedorexcessivecontactwiththesolutions(s).
• AcetoneandPEG-containingointmentscancausefailureofthisdeviceandshouldnotbeusedwithpolyurethanecatheters.Chlorhexidine
patchesorbacitracinzincointments(e.g.,Polysporin* ointment)arethepreferredalternative.
• FollowUniversalPrecautionswheninsertingandmaintainingthisdevice.
• Cardiacarrhythmiasmayresultiftheguidewireisallowedtopassintotherightatrium.
• Donottunnelthroughmuscle.
• Clampingthecatheterrepeatedlyinthesamespotcouldweakenthetubing.Changethepositionoftheextensionclampsregularlyto
prolongthelifeoftheextensiontubing.AvoidclampingnearorontheLuer-lockconnectors,asthismaycausetubingfatigueandpossible
disconnection.Extensiontubingmaydeveloptearsorcutsifsubjectedtoexcessivepullingorcontactwithroughedges.Examinetubingfor
damage both prior to and after each treatment.
• Cathetersshouldbeimplantedcarefullytoavoidanysharporacuteangleswhichcouldcompromisetheopeningofthecatheterlumens.
• Topreventairembolismand/orbloodloss,placethumbovertheexposedoriceofthesheathintroducer.
• Toavoiddamagetovesselsandviscus,infusionpressuresshouldnotexceed25psi(172kPa).Theuseofa10mLorlargersyringeisrecom-
mended because smaller syringes generate more pressure than larger syringes. Note:Athreepound(13.3Newton)forceontheplungerof
a3mLsyringegeneratespressureinexcessof30psi(206kPa)whereasthesamethreepound(13.3Newton)forceontheplungerofa10mL
syringegenerateslessthan15psi(103kPa)ofpressure.
• AccessoriesandcomponentsusedinconjunctionwiththiscathetershouldincorporateLuer-lockconnectors.
• Theheparinsolutionmustbeaspiratedoutofbothlumensimmediatelypriortousingthecathetertopreventsystemicheparinizationofthe
patient.
• Failuretoclampextensionswhennotinusemayleadtoairembolism,bleeding,andpossibleocclusions.
• Intherareeventofaleak,thecathetershouldbeclampedimmediately.Necessaryremedialactionmustbetakenpriortoresumingdialysis,
hemoperfusion, or apheresis procedure.
• Donotresterilizethecatheterorcomponentsbyanymethod.Themanufacturerwillnotbeliableforanydamagescausedbyreuseofthe
catheter or accessories.
• Cannulationoftheleftinternaljugularveinisreportedlyassociatedwithahigherincidenceofcomplicationscomparedtocatheterplace-
mentintherightinternaljugularvein.
4
• Donotadvanceguidewireorcatheterifunusualresistanceisencountered.
• TheblueCompressionSleeveisanecessarycomponentoftheExtensionLegAssembly.AlwaysvisuallyconrmthattheCompressionSleeve
remainsintheFemaleConnectorduringassembly.
• Alcoholshouldnotbeusedtolock,soakordeclotpolyurethaneDialysisCathetersbecausealcoholisknowntodegradepolyurethane
cathetersovertimewithrepeatedandprolongedexposure.
• Reusemaycreateariskofpatientinfectionofuserinfection,compromisethestructuralintegrityofthedevice,leadtodevicefailureand/or
leadtoinjury,illnessordeathofthepatient.
• TheSlideClamp,ThumbClamp,andPlugareprovidedforuseduringcatheterplacementonly.Donotreuse.
• Notintendedforpediatricorneonataluse.
• IntendedforSingleUse.DONOTREUSE.Reuseand/orrepackagingmaycreateariskofpatientoruserinfection,compromisethestructural
integrityand/oressentialmaterialanddesigncharacteristicsofthedevice,whichmayleadtodevicefailure,and/orleadtoinjury,illnessor
death of the patient.
CAUTIONS
• Repeatedovertighteningofbloodlines,syringesandcapswillreduceLuer-lockconnectorlifeandcouldleadtopotentialconnectorfailure.
Incaseofdamage,clampthecatheterbetweenthepatientandthedamagedareawithasmooth-edged,atraumaticclamp.
• Sterileandnon-pyrogeniconlyifpackagingisnotopened,damagedorbroken.
• Readtheinstructionsforusecarefullybeforeusingthisdevice.
• Federal(USA)lawrestrictsthisdevicetosalebyorontheorderofaphysician.
• CareshouldbetakenNOTtoforcethedilatorsheathintroducerintothevesselduringinsertionasvesseldamageincludingperforationcould
result.
• Leftsidedcatheterplacementmayprovideuniquechallengesduetotherightanglesformedbytheinnominateveinandattheleftbrachio-
cephalicjunctionwiththeSVC.
2,5
• Ensurethatthesheathintroducerisonlytornexternally.Cathetermayneedtobefurtherpushedintothevesselassheathintroduceristorn.
• Foroptimalproductperformance,donotinsertanyportionofthecuintothevein.
• Careshouldbetakennottoadvancethesplitsheathtoofarintovesselasapotentialkinkwouldcreateanimpassetothecatheter.
• PlacementandcareoftheDuet*Cathetersshouldbeperformedbypersonsknowledgeableoftherisksinvolvedandqualiedintheproce-
dures.
• Donotcreateasharpbendincathetertunnelasthismaycausekinkingandimpactow.
• Iftheguidewiremustbewithdrawnwhiletheneedleisinserted,removeboththeneedleandguidewireasaunittopreventtheneedlefrom
damaging or shearing the guidewire.
• Cuttingthecatheteranywherebutthepre-printedmarkswillresultintheinabilitytoreadthecatheterlengthandprimingvolumes.
• Cathetertipsmustbestaggeredby4cmtominimizerecirculation.
• BeforeattemptingtheinsertionofDuet*Catheters,ensurethatyouarefamiliarwiththefollowingcomplicationsandtheiremergencytreat-
ment should any of them occur.
POSSIBLE COMPLICATIONS
These and other complications are well documented in medical literature and should be carefully considered before placing the catheter.
Theuseofanindwellingcentralvenouscatheterprovidesanimportantmeansofvenousaccessforcriticallyillpatients;however,thepotential
existsforseriouscomplicationsincludingthefollowing:
• AirEmbolism
• Bleeding
• BrachialPlexusInjury
• CardiacArrhythmia
• CatheterorCuErosionthroughtheSkin
• CatheterEmbolism
• CatheterOcclusion
• CatheterOcclusion,DamageorBreakage
duetoCompressionbetweentheClavicle
andFirstRiborKinking
• Catheter-relatedSepsis
• CardiacTamponade
• Endocarditis
• ExitSiteInfection
• ExitSiteNecrosis
• Extravasation
• FibrinSheathFormation
• Hemothorax
• Hemotoma
• Hydrothorax
• Inammation,Necrosisorscarringofskin
over implant area
• IntoleranceReactiontoImplantedDevice
• LacerationofVesselsorViscus
• Mediastinalwidening
• PerforationofVesselsorViscus
• Pneumothorax
• RetroperitonealBleed
• RisksNormallyAssociatedwithLocaland
GeneralAnesthesia,Surgery,andPost-
OperativeRecovery
• ThoracicDuctInjury
• Thromboembolism
• VenousThrombosis
• VentricularThrombosis
• VesselErosion
0724304/1003R
F. Assembly of Catheter and Extensions
31.(Asapplicable)usingascalpelorscissors,cutcathetersquarelyatdesiredpre-printedmarktoproduceaclean,smoothsurface(Figure5).
Ensureatleast6cmofcatheterextendfromexitsitetoallowforconnection.
35 cm/2.0ml
35 cm/x.xml
40 cm/2.2ml
45 cm/2.4ml
Figure 5
CAUTION: Cuttingthecatheteranywherebutthepre-printedmarkswillresultintheinabilitytoreadthecatheterlengthandpriming
volumes.
Figure 6
Female Connector
Figure 7
Blue
Compression
Sleeve
Female Connector
WARNING: TheblueCompressionSleeveisanecessarycomponentoftheExtensionLegAssem-
bly.AlwaysvisuallyconrmthattheCompressionSleeveremainsintheFemaleConnectorduring
assembly(Figure7).
32.HoldunclampedExtensionLegAssemblyandattachsterile10mLsyringeandprimewithnormalsterilesaline,leavingsyringeattached
(Figure8).WARNING: Toavoiddamagetovesselsandviscus,infusionpressuresshouldnotexceed25psi(172kPa).Theuseofa10mLor
larger syringe is recommended because smaller syringes generate more pressure than larger syringes.
Figure 8
33.With catheters cut per Step 31, slip Female
Connector over catheter visually conrming
blue Compression Sleeve remains inside
Female Connector (Figure9).
Note:Attachtheappropriatecolor-coded
(blue-venous,red-arterial)ExtensionLegassembly.
Figure 9
34.SlideproximalendofcatheterovermetalcannulaontheExtensionLegBody,ensuringthatcatheterisattachedentirelengthofmetal
cannulaandnometalisvisible(Figure10).
Figure 10
35.SlideFemaleConnectortowardsExtensionLegBodyandassemblethetwotogetheruntilthetwoarefullyseated(Figure11).
WARNING: TheblueCompressionSleeveisanecessarycomponentoftheExtensionLegAssembly.AlwaysvisuallyconrmthattheCom-
pressionSleeveremainsintheFemaleConnectorduringassembly(Figure7andFigure11).
Figure 11
Blue Compression Sleeve must be present during assembly.
36.AfterExtensionLegAssemblyisoncatheter,ensurecatheterlengthandprimingvolumeprintingisvisible.
37.Graspingtheconnectorinonehand,andthecathetertubingintheother,gentlytugonthejointtotestthesecurityoftheconnector.Ifthe
connectorpullsoutofcatheter,repeattheattachmentprocedure.Aconnectionfailuremaybeduetoone,oracombinationofthefollowing:
•TheExtensionLegConnectormetalcannulaisnotfullyinsertedintothecatheter.
•Missingbluecompressionsleeveinfemaleconnector.
38.Withthe10mLsyringeattachedtotheExtensionLegAssembly,removeSlideClampfromcatheter.
WARNING:TheSlideClamp,ThumbClampandPlugareprovidedforuseduringcatheterplacementonly.Donotreuse.
39.Verifycatheterfunctionbyaspiratingtoensureadequatebloodow.Notethatextensionclampsmustbeunclampedtoaspirate.
40.Onceowissatisfactory,ushcatheterwithaminimumof10mLofsterilesaline.
WARNING:Toavoiddamagetovesselsandviscus,infusionpressuresmustnotexceed25psi(172kPa).Theuseofa10mLorlargersyringeis
recommended because smaller syringes generate more pressure than larger syringes.
41.Injectheparinsolutionof1,000to5,000units/mLinamountsequaltotheprimingvolumeasdenotedonthecatheter.Injectquicklyand
clampextensionwhileunderpositivepressure.Attachasterileendcap.WARNING:Failuretoclampextensionswhennotinusemayleadto
air embolism, bleeding, and possible occlusions.
42.Repeatsteps31-41forsecondcatheter(Figure12).
Figure 12
Cu
G. Placement Verication and Securement
43.Afterinsertion,conrmplacementofthecatheterusingx-rayoruoroscopy.
44.Foradditionalsecurity,suturetheinsertionsite,orifpreferred,useaStatLock*
CatheterStabilizationDeviceSecurementdevicetoanchorthecatheter.
45.Managetheexitsiteperhospitalprotocol.
46.Dressthecatheterperhospitalprotocol.
WARNING:AcetoneandPEG-containingointmentscancausefailureofthisdevice
andshouldnotbeusedwithpolyurethanecatheters.Chlorhexidinepatchesor
bacitracinzincointments(e.g.,Polysporin* ointment)arethepreferredalterna-
tive.
47.Recordindwellingcatheterlengthandinsertionsiteonpatient’schart.
PERFORMANCE GUIDELINES
Priming Volumes
Refertoindividualcathetersforprimingvolumeinformationprintedatpre-denedcentimetermarkings.
Assuggestedby In Vitro data
Flow Rate vs Lumen Pressure
Venous
Arterial
35 cm Length (400 ml/min)
50 cm Length (350 ml/min)
199 mmHg
229 mmHg
-208 mmHg
-235 mmHg
Note: Reverse ow will result in higher recirculation
INSERTION TECHNIQUE
Forpercutanenousplacement,thecatheterisinsertedthroughasheathintroducerintothesuperiorvenacavaviatheinternaljugularvein
3,6
(preferred),externaljugularvein,orsubclavianvein.ThepatientshouldbeplacedinTrendelenburgpositionwiththeheadturnedtotheop-
posite side of the entry site.
CATHETERS MUST BE INSERTED UNDER STRICT ASEPTIC CONDITIONS.
WARNING:Cannulationoftheleftinternaljugularveinwasreportedlyassociatedwithahigherincidenceofcomplicationscomparedtocath-
eterplacementintherightinternaljugularvein.
4
CAUTION:Leftsidedcatheterplacementmayprovideuniquechallengesduetotherightanglesformedbytheinnominateveinandattheleft
brachiocephalicjunctionwiththeSVC.
2,5
A. Sterile Field and Skin Preparation
1. Provideasterileeldthroughouttheprocedure.Usesterilegloves,masks,caps,sterilegowns,andusealargesteriledrapetocoverthe
patient.Ifhairremovalisneededuseclippersordepilatories.
2. Preparetheaccesssiteusingstandardsurgicaltechniqueanddrapethepreppedareawithsteriletowels.
3. (Ifapplicable)administerlocalanaesthesiatotheinsertionsite.
B. Pre-ush the Catheters
Slide Clamp
Thump Clamp
Plug
4. Irrigate,prime,andclampbothcatheterlumenswithheparinsolutionoraccordingtohospitalproto-
col.Withcathetersfullofuid,clampeachcatheterusingSlideClamporThumbClampbetweenthe
proximalendofthecatheterandthe50cmmarkingoruseplug.
C. Perform Venipuncture
5. Inserttheintroducerneedlewithanattachedsyringetothedesiredlocationusingultrasoundguidance(preferred).Aspirategentlyasthe
insertion is being made.
6. Whentheveinhasbeenentered,removethesyringeleavingtheneedleinplaceandoccludeendofneedlewiththumb.
7. InserttheexibleJendofthestandardguidewirethroughtheintroducerneedlehubthatwillpermitpassageofa0.038in.(0.97mm)guide-
wire.Advancethestandardguidewiretothedesiredlocationinvessel.
WARNING:Cardiacarrhythmiasmayresultiftheguidewireisallowedtopassintotherightatrium.
WARNING: Do not advance guidewire or catheter if unusual resistance is encountered.
8. Removetheneedlewhileholdingtheguidewireinplace.Wipetheexposedguidewirecleanandsecureitinplace.
CAUTION:Iftheguidewiremustbewithdrawnwhiletheneedleisinserted,removeboththeneedleandguidewireasaunittopreventthe
needle from damaging or shearing the guidewire.
9. Repeatsteps5-8abovetointroducesecondneedleandguidewireintosametargetveinapproximately1-2cmadjacenttotherstincision
site.
10.Aftersecondwirehasbeenplaced,makeasmallincisionateachvenousinsertionsitetoremoveanydermalbridgeandeaseinsertion.
D. Insert Sheath and Advance Catheter
11.Advancethedilatorsheathintroducerovertheexposedguidewireintothevessel.
CAUTION:CareshouldbetakenNOTtoforcethedilatorsheathintroducerintothevesselduringinsertionasvesseldamageincluding
perforation could result.
WARNING:Cardiacarrhythmiasmayresultiftheguidewireisallowedtopassintotherightatrium.
12.Withdrawthedilatorandguidewire,leavingthesheathintroducerinplace.
CAUTION:Careshouldbetakennottoadvancethesplitsheathtoofarintovesselasapotentialkinkwouldcreateanimpassetothecath-
eter.
WARNING:Topreventairembolismand/orbloodloss,placethumbovertheexposedoriceofthesheathintroducer.
13.Removethumbandfeeddistalsectionofcatheterintothesheathintroduceruntiltipiscorrectlypositioned.Thedepthmarkingsinonecm.
increments may be used to determine insertion length.
WARNING:Ensurethecatheterislledwithheparinizedsaline,isclampedandisfreeofairbubblesbeforeinsertingitintothevein.
WARNING: Do not advance guidewire or catheter if unusual resistance is encountered.
14.Withthecatheteradvanced,peelawaythesheathintroducerbygrippingtheT”handleandbreakingitapartwithadownwardandoutward
motion to initiate separation and withdrawal of the sheath introducer.
CAUTION:Ensurethatthesheathintroducerisonlytornexternally.Cathetermayneedtobefurtherpushedintothevesselassheathintro-
ducer is torn.
CAUTION:Foroptimalproductperformance,donotinsertanyportionofthecuintothevein.
15.Repeatsteps11-14aboveforsecondcatheter.
16.Verifycatheterpositionunderuoroscopyandmakenecessaryadjustments.The venous (infusion) tip should be located at the level of
the caval atrial junction or into the right atrium and approximately 4 cm past the arterial (withdrawal) catheter tip.
CAUTION:Cathetertipsmustbestaggeredby4cmtominimizerecirculation.
E. Subcutaneous Tunneling of Catheters and Cu Placement
Tu nnel Tr act
Catheter
Exit Site
Catheter
Insertion
Site
Figure 1
17.Identifydesiredtunnellocationandexitsitetoinsureagentlearc.Ensure
thattheexposedproximalportionofcatheterislongenoughtocutthe
cathetertothedesiredlengthandtoreadlengths/primingvolumes.
18.Makenoteofthedesiredlocationatwhichthecuwillresideinthetunnel.
19.(Ifapplicable)administerlocalanaesthesiatotunneltract.
20.Makeanincisionatthecatheterexitsite(Figure1).
21.Usingthemetaltunnelerprovided,createasubcutaneoustunnelfromthe
insertionsitetoemergeatthecatheterexitsite(Figure2b).
WARNING: Do not tunnel through muscle.
22.Oncemetaltunneleremergesfromcatheterexitsite,placetaperedend
of tunnel slide dilator over tip of tunneler at insertion site. While holding
endofmetaltunneler,expandtunneltractfrominsertionsitetohalfway
throughthetunneltodesiredculocationorwithin2cmofthecatheter
exitsiteusingaback-and-forthmotion(Figure3aand3b).
Figure 2bFigure 2a
Tunneler
Tunnel Slide Dilator
Figure 2c
23.Removetunnelslidedilatorwhileholdingmetaltunnelerinplace.Ifused,removeplug.
24.Attachcathetertotunnelersothatcatheter’sproximalendslidesoverbarbedendofmetaltunneler.
Tunneler
Tunnel
Slide
Dilator
Figure 3a
Figure 3b
25.RemoveSlideClamporThumbClamp. 
Figure 4
Cu
WARNING:TheSlideClamp,ThumbClampandPlugareprovidedforuse
during catheter placement only. Do not reuse.
26.Pullmetaltunnelercarefullyuntilcatheteremergesfromcatheterexitsite.
The catheter should not be forced through the tunnel.
CAUTION: Do not create a sharp bend in catheter tunnel as this may cause
kinkingandimpactow.
27.EngageSlideClampbetweendesiredcutlengthandcatheterexitsiteto
prevent blood loss and air embolism.
28.Removetunnelerbycuttingcathetersquarelyatdesiredpre-printedmarkto
produceaclean,smoothsurface(Figure5).
CAUTION:Cuttingthecatheteranywherebutthepre-printedmarkswillresult
in the inability to read the catheter length and priming volumes.
29.Rechecktippositioningusingx-rayoruoroscopy.
30.Repeatsteps17-29forsecondcatheter(Figure4).
/