Völker care bed models 2080, 3080, 3080 K built after March 2003 – Instructions for use G57 V1.0 (12.02.2009)
Use of oxygen equipment
DANGER "Risk of fire"
Do not use any oxygen equip-
ment other than that which is
delivered via nasal prongs or
masks. Do not use this bed in
a room where there is a risk of
explosion.
(Provided it has been ensured
(e.g. through information in the
instructions for use supplied with
the equipment being used) that
the use of the equipment will not
raise the O2 concentration to such
a high level that there is the risk
of explosion (even when no fault
occurs), the device can be used).
Rail spacers
When using the rail spacer, read the
separate instructions for use applica-
ble to this accessory. During techni-
cal checks, the rail spacers should be
checked to ensure they are suitable for
the size of side rail used.
Cleaning and disinfection
In order to maintain a consistent func-
tional capability, the care bed should be
cleaned, disinfected and tested at the
soonest possible opportunity following
each use, so that it can be reused im-
mediately without risk.
Risks can occur if the bed is not subject-
ed to proper cleaning/disinfection  49.
Maintenance and repair
Every individual involved with mainte-
nance and repair must at least have
read the safety regulations and the serv-
ice manual, and be trained and qualified
in accordance with Medical Devices
Operator Ordinance (MPBetreibV) §§ 4
and 6.
After maintenance work or repairs have
been carried out, a technical check
 54 must be carried out. During this
check, it must be determined that the
bed can be used in accordance with the
specifications without risk to the occu-
pant, user or third party.
Notes | General safety notes 4/4
The technical check must be carried
out at least once a year and after every
lengthy period of non-use.
Any discernible damage, such as
signs of wear and tear, loose screws
or breaks/fractures must be eliminated
immediately.
Electromagnetic / static interference
The care beds in model series 2080,
3080 and 3080 K satisfy the EMC re-
quirements in accordance with the law
on medical products (MPG). The basis
for testing is standard EN 60601-1-2.