Minebea Intec WE1 option for PR 5900 Owner's manual

Type
Owner's manual

Minebea Intec WE1 option for PR 5900

The Minebea Intec WE1 option for PR 5900 is a device that allows you to connect load cells or weighing platforms located in a hazardous area (zone 1 or 21) to a PR 5900 indicator located in a safe area. This device ensures that no electrical power exceeding the maximum permissible limits is transmitted into the hazardous area, even in the event of a circuitry fault.

Some possible use cases for this device include:

  • Weighing hazardous materials in a laboratory or manufacturing environment
  • Monitoring the weight of a tank or silo in a hazardous area
  • Controlling the filling or emptying of a hazardous material container

Minebea Intec WE1 option for PR 5900

The Minebea Intec WE1 option for PR 5900 is a device that allows you to connect load cells or weighing platforms located in a hazardous area (zone 1 or 21) to a PR 5900 indicator located in a safe area. This device ensures that no electrical power exceeding the maximum permissible limits is transmitted into the hazardous area, even in the event of a circuitry fault.

Some possible use cases for this device include:

  • Weighing hazardous materials in a laboratory or manufacturing environment
  • Monitoring the weight of a tank or silo in a hazardous area
  • Controlling the filling or emptying of a hazardous material container
WE1 option for PR 5900
Additional information
Translation of original additional information 9499 059 59100 Edition 3.4.0 03/02/2023
Minebea Intec GmbH, Meiendorfer Str. 205 A, 22145 Hamburg, Germany
Phone: +49.40.67960.303 Fax: +49.40.67960.383
Foreword
Must be followed!
Any information in this document is subject to change without notice and does not represent a commitment on
the part of Minebea Intec unless legally prescribed. This product should only be operated/installed by trained
and qualied personnel. In correspondence concerning this product, the type, name, and release number/serial
number as well as all license numbers relating to the product have to be cited.
Note
This document is partially protected by copyright. It may not be changed or copied, and it may not be used
without purchasing or written permission from the copyright owner (Minebea Intec). The use of this product
constitutes acceptance by you of the abovementioned provisions.
Table of contents
1 Safety instructions...............................................................................................................................2
1.1 General notes..............................................................................................................................................................2
1.2 Description...................................................................................................................................................................2
1.3 Marking for Ex Zones.................................................................................................................................................2
2 Installation........................................................................................................................................... 3
2.1 General information...................................................................................................................................................3
2.2 Weighing electronics board for zones 1 and 21...................................................................................................... 3
2.3 Display..........................................................................................................................................................................4
2.4 Connection within the Ex area..................................................................................................................................5
2.5 WE1 load cell supply voltage.....................................................................................................................................8
2.6 WE1 potential-free load cell supply voltage...........................................................................................................8
2.7 Potential equalization................................................................................................................................................9
2.8 Screens....................................................................................................................................................................... 10
3 Repairs/Maintenance/Solder work.....................................................................................................11
3.1 Repairs......................................................................................................................................................................... 11
3.2 Maintenance............................................................................................................................................................... 11
3.3 Soldering work...........................................................................................................................................................11
4 Technical data.....................................................................................................................................12
4.1 General data...............................................................................................................................................................12
4.2 Weighing electronics board WE1.............................................................................................................................12
4.3 Connecting load cells/junction boxes/weighing platforms ...............................................................................13
4.4 Load cell supply ........................................................................................................................................................ 14
4.5 Connecting cables .................................................................................................................................................... 14
4.6 Documentation on the CD included ...................................................................................................................... 14
5 Appendix ............................................................................................................................................15
5.1 Certicates/control drawing/safety instructions.................................................................................................15
WE1 option for PR 5900 Table of contents
Minebea Intec EN-1
1 Safety instructions
1.1 General notes
CAUTION
Warning of personal injury.
Observe the safety instructions listed in the PR 5900 installation manual.
1.2 Description
The WE1 option is used to establish an intrinsically safe interface between a PR 5900
installed in a safe area (or with Y2 option in zone 2/22) and load cells or weighing
platforms located in a zone 1 or 21 Ex area as dened by European Directive 2014/34/EU
and related harmonized European standards (see also Chapter 5.1).
The device is designed for weighing applications requiring intrinsically safe interfacing.
The PR 5900 with WE1 option ensures that no electrical power exceeding the maximum
permissible limits is transmitted into the Ex area, even in the event of a circuitry fault.
The PR 5900 with WE1 option provides for intrinsically safe connection to
-a voltage output for the supply of load cells or weighing platforms located in zone 1
or 21,
-a measuring voltage input from zone 1 or 21, and
-a sense voltage input from zone 1 or 21.
This is achieved by means of double current and voltage limiters in the supply circuit as
well as voltage limiters in the analog electronics (weighing electronics board) of the
PR 5900 with WE1 option.
1.3 Marking for Ex Zones
Zone Marking Certicate no.
1 II (2) G [Ex ib Gb] IIC/IIB
[Ex ib Gb] IIC/IIB DEKRA 13ATEX0186X
IECEx DEK 13.0065X
21 II (2) D [Ex ib Db] IIIC
[Ex ib Db] IIIC DEKRA 13ATEX0186X
IECEx DEK 13.0065X
Note:
See also Chapter 5.1.
NOTICE
Installation in Ex areas
The Ex safety instructions provided in the appendix must be observed when
installing in Ex areas.
WE1 option for PR 5900 1 Safety instructions
EN-2 Minebea Intec
2 Installation
2.1 General information
Installation should be performed by an authorized specialist, in accordance with
applicable laws, rules and regulations, ordinances, and standards. In particular, the
EN 60079-14 standard must be taken into account.
All cables leading to and from the PR 5900 have to be securely installed.
Any connections that are not being used (cable glands) must be sealed o with approved
locking screws in order to guarantee appropriate IP protection.
WARNING
Working on the device while it is switched on may have life-threatening
consequences.
Disconnect the device from the power supply.
Live parts may be exposed when tools are used to open covers or remove parts.
Please note that capacitors in the device may still carry an electrical charge even
after disconnecting the device from all voltage sources.
Do not remove any existing safety equipment/covers.
2.2 Weighing electronics board for zones 1 and 21
Note:
The weighing electronics board cannot be replaced retroactively as this option and the
basic unit are permanently assigned during production.
This assignment is regulated by the BIOS.
If these components do not match, an error message will appear and the device will not
start.
2 Installation WE1 option for PR 5900
Minebea Intec EN-3
No. Description
1 Connection for Ex connection cable PR 6136/.
2 CAL switch
3 Ex partition wall
2.3 Display
When installed correctly, the Maxxis displays an Ex symbol in the status display next to
weighing point A.
WE1 option for PR 5900 2 Installation
EN-4 Minebea Intec
2.4 Connection within the Ex area
Note:
Components are only depicted schematically on the connection diagram. The
connections are specically described in the PR 5900 installation manual.
Example with PR 1625/60 and external remote display
2 Installation WE1 option for PR 5900
Minebea Intec EN-5
Load cell
Potential equalization line
Ex zone
No Ex zone
Screen
RS-422/485 interface PR1604/00; solder link X3: closed; solder link X7: closed
Factory setting, see also Chapter 2.5
Screen clamping rail
PR 5900 with WE1 option
UDC = 24 V or UAC = 230 V
power supply
external remote display
WE1 option for PR 5900 2 Installation
EN-6 Minebea Intec
Example with remote terminal PR 5900/60
2 Installation WE1 option for PR 5900
Minebea Intec EN-7
Load cell
Potential equalization line
Ex zone
No Ex zone
Screen
Power supply unit YPSC01-X
Factory setting, see also Chapter 2.5
Screen clamping rail
PR 5900 with WE1 + CX1 option
UDC = 24 V or UAC = 230 V
With cable lengths >50 m, the cable screen may only be connected at one end.
Note:
The regulations for installation in the Ex zone must be complied with.
2.5 WE1 load cell supply voltage
The load cell supply voltage cannot be changed retroactively.
The desired voltage must be specied when ordering, see Chapters 4.2 and 4.4.
2.6 WE1 potential-free load cell supply voltage
The device with WE1 option will be delivered with an intrinsically safe circuit that is
galvanically connected with the equipotential bonding conductor (PE).
At distances >50 m a potential-free load cell supply voltage is required. This is done as
follows:
When jumper "J250" is open, the intrinsically safe circuit is isolated from the
housing (PE) at a minimum test voltage of 500 V.
1. Cut jumper "J250" on the weighing electronics board.
2. Turn the remaining wire leads upward and cut them just above the circuit board.
Make sure that both contacts are separated by a minimum distance of 2 mm and do
not touch other potentials.
WE1 option for PR 5900 2 Installation
EN-8 Minebea Intec
2.7 Potential equalization
In Ex areas, equipotential bonding must be established to avoid any compensatory
currents between the individual conducting system components.
Legend
No. Component Cross-section
(1) Conductor for additional potential equalization ≥4.0 mm2Cu
(2) Conductor for potential equalization ≥6.0 mm2Cu
(3) Conductor for potential equalization ≥10.0 mm2Cu
(4) Main conductor for potential equalization ≥10.0 mm2Cu
(5) Earthing conductor ≥10.0 mm2Cu
(6) Load cell bypass ≥10.0 mm2Cu
(A) Equipotential bonding conductor (busbar)
(B) Main equipotential busbar
(E) To main earth (electrode)
2 Installation WE1 option for PR 5900
Minebea Intec EN-9
2.8 Screens
WARNING
Improper connection of the connecting cable screens can cause severe personal
injury and damage to systems!
At distances > 50 m, the screen may only be connected at one end.
Connect the connecting cable screen to the PR 5900, see installation manual.
Open bridge "J250" on the WE1 weighing electronics board; see Chapter 2.6.
For load cells where the screen is ush against the housing and when using junction
box PR 6130/64 Sa, connect the screen of the connecting cable and open the bridge
above the sense terminals on the board of the junction box, see PR 6130/64 Sa
installation manual.
Do not connect the connecting cable screen in the case where load cell screens are
ush against the housing and when using junction box PR 6130/65 S; see also
Chapter 2.4.
WE1 option for PR 5900 2 Installation
EN-10 Minebea Intec
3 Repairs/Maintenance/Solder work
3.1 Repairs
Repairs are subject to inspection and must be carried out at Minebea Intec.
In case of defect or malfunction, please contact your local Minebea Intec dealer or service
center for repair.
When returning the device for repair, please include a precise and complete description of
the problem.
3.2 Maintenance
Maintenance work may only be carried out by a trained technician with expert knowledge
of the hazards involved and the required precautions.
3.3 Soldering work
Soldering work on the weighing electronics board is not permitted.
3 Repairs/Maintenance/Solder work WE1 option for PR 5900
Minebea Intec EN-11
4 Technical data
4.1 General data
Note:
For further technical data, refer to the PR 5900 installation manual.
Connection for the Ex area Via 4 mm2terminals
Connection values See Chapter 5.1
Equipotential bonding conductor (PE) Via terminal, see Chapter 2.6
4.2 Weighing electronics board WE1
The weighing electronics board is inserted in the "WP B" slot. The CAL switch "B" (1) is
located on the board. Adjustment data and parameters are saved to the EAROM (non-
volatile memory) of the weighing electronics board.
Note:
The voltage (UDC = 7.2 V or UDC = 12 V) must be specied when ordering!
7.2 V:
The total resistance (RLC) of the load cells is between 80 and 150 , see also
Chapter 4.4.
12 V:
The total resistance (RLC) of the load cells is >150 , see also Chapter 4.4.
WE1 option for PR 5900 4 Technical data
EN-12 Minebea Intec
Technical data
Description Data
Connection 7-pin plug (male), screen connection on housing and
to GNDA
Number of channels 1× 6wire or 4wire
Load cell type SG cells, 6 or 4wire connection possible
Load cell supply See Note
Capacity (number of load
cells) See Chapter 4.4
Sense input UDC = ±6 V, with monitoring
Sense voltage monito-
ring Sense voltage below 8 V, switchable to <8 V
Measurement input UDC = 0…36 mV, symmetrical to 0
Dead load suppression UDC = max. 36 mV (dead load + range)
Accuracy* 0.8 µV/e corr. to 4.8 mV for 6,000 e
Class III, according to OIML R76/EN45501
Min. measuring signal
(OIML)* 3,000 e: ≥0.20 mV/V
6,000 e: ≥0.40 mV/V
Max. resolution 7.5 million internal counts at 3 mV/V
Linearity* <0.003%
Zero point stability error
(Tk0)* <0.05 µV/K RTI; 0.004%/10 K at 1 mV/V
SPAN stability error
(Tkspan)* < ±4.0 ppm/K
Cable type 6wire with screen for entire cable and screen for measure-
ment cables, e.g. PR 6136/..
Cable length system-dependent
* with a measurement time of 160 ms and a 12 V supply
4.3 Connecting load cells/junction boxes/weighing platforms
The weighing electronics board WE1 is connected using a 7-pin plug.
Note:
For connection values, see Chapter 5.1.
4 Technical data WE1 option for PR 5900
Minebea Intec EN-13
4.4 Load cell supply
UDC = ±6.0 V (nominal)/≤96 mA Max. 4 load cells of 650 each, connected
in parallel (= 162.5 )
Supply voltage
UDC = ±3.6 V (nominal)/≤96 mA Max. 8 load cells of 650 each, connected
in parallel (= 81.3 )
or
4 load cells with 350 each, connected in
parallel (= 87.5 )
Tolerance range UDC = ±5.8…±6.2 V;
UDC = ±3.4…±3.8 V
Max. current ≤100 mA short circuit
Measuring voltage Vmeas Max. 36 mV (dierential)
Sense voltage Vsense ±6 V nominal (with ref. to ground)
±3.6 V nominal (with ref. to ground)
4.5 Connecting cables
Length of the connecting cable between junction box and instrument
Cable type PR 6136, PR 6136A max. 200 m – length of the load cell cable
Length of the connecting cable between weighing platform and instrument
Cable type LiYCY max. 200 m
4.6 Documentation on the CD included
The documents and manuals listed in the appendix (see Chapter 5.1) can be found on the
PR 5900 CD.
WE1 option for PR 5900 4 Technical data
EN-14 Minebea Intec
5 Appendix
5.1 Certicates/control drawing/safety instructions
Ser. no. Description Document no.
1 EC-Type Examination Certicate DEKRA 13 ATEX 0186 X
2 Certicate of Conformity IECEx DEK 13.0065 X
3 Safety instructions 940515900000-460
4 Control drawing 2004137
The documents listed in the table can be found on the PR 5900 CD.
5 Appendix WE1 option for PR 5900
Minebea Intec EN-15
Published by
Minebea Intec GmbH | Meiendorfer Strasse 205 A | 22145 Hamburg, Germany
Phone: +49.40.67960.303 | Email: [email protected]
www.minebea-intec.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18

Minebea Intec WE1 option for PR 5900 Owner's manual

Type
Owner's manual

Minebea Intec WE1 option for PR 5900

The Minebea Intec WE1 option for PR 5900 is a device that allows you to connect load cells or weighing platforms located in a hazardous area (zone 1 or 21) to a PR 5900 indicator located in a safe area. This device ensures that no electrical power exceeding the maximum permissible limits is transmitted into the hazardous area, even in the event of a circuitry fault.

Some possible use cases for this device include:

  • Weighing hazardous materials in a laboratory or manufacturing environment
  • Monitoring the weight of a tank or silo in a hazardous area
  • Controlling the filling or emptying of a hazardous material container