Enraf-Nonius CT-7 User manual

Type
User manual
DBe sure to read this manual before use.
Keep this manual in a safe place for future reference.
Combination Therapy Equipment
PHYSISYSTEM
Model
CT-7
Instruction manual
For Professional Use
About This Manual
zKeep this manual in a safe place where it can be referred to whenever necessary.
zIf this manual becomes lost or damaged, contact your OG Wellness Technologies
distributor immediately.
zFollow the instructions and procedures specied in this manual.
zDo not operate, care for or provide maintenance for the product except as specied in
this manual.
zThe contents of this manual and caution labeling on the product may not cover all
potential dangers associated with the product.
OG Wellness is unable to foresee dangers arising from any use of the product other than its intended use.
zKeep safety considerations in mind when using, cleaning or inspecting the product,
including, but not limited to the contents of this manual and caution labeling on the
product.
zThe product you have purchased may dier in some details from the illustrations in
this manual. Parts of the illustrations in this manual may be omitted or simplied to aid
explanation.
zThe specication and design of the product, as well as the contents of this manual are
subject to improvement and change without notice.
zFor inquiries or further information, contact your OG Wellness Technologies distributor.
zGreat care has been taken in the preparation of this manual. In the event of any error or
omission, or if any of the contents are unclear, please contact your OG Wellness
Technologies distributor.
zOG Wellness Technologies holds the copyright of this manual.
zIt is prohibited to copy, edit or translate all or any part of this manual without the prior
written consent of OG Wellness Technologies.
zDo not copy the contents of this manual via any electronic reading or scanning device
including, but not limited to, image scanners.
zCompany names and product names described in this manual are trademarks or
registered trademarks of their respective companies in Japan and other countries.
Table of Contents
1. Foreword ------------------------------------------------------------------------------------------------------ 4
1. 1 IntendedUser/Operator ------------------------------------------------------------------------------------ 4
1. 2 ProductLiability ------------------------------------------------------------------------------------------------ 4
2. Product Description --------------------------------------------------------------------------------------- 4
3. Precautionary Instructions ------------------------------------------------------------------------------ 5
4. Intended Use ------------------------------------------------------------------------------------------------- 6
4. 1 Electrotherapy -------------------------------------------------------------------------------------------------- 6
4.1.1 Painmanagement --------------------------------------------------------------------------------------------6
4.1.2 Musclestimulation --------------------------------------------------------------------------------------------7
4.1.3 Description of current waveforms ------------------------------------------------------------------------- 8
4.1.4 Illustrationsofcurrentwaveforms ----------------------------------------------------------------------- 16
4. 2 UltrasoundTherapy ------------------------------------------------------------------------------------------21
4.2.1 Indications - Ultrasoundtherapy ------------------------------------------------------------------------ 21
4.2.2 Contra-indications - Ultrasoundtherapy -------------------------------------------------------------- 21
4.2.3 Precautionsandwarnings - Ultrasoundtherapy --------------------------------------------------- 21
4.2.4 Relevanthazards - Ultrasoundtherapy--------------------------------------------------------------- 21
4.2.5 Potentialadverseeects - Ultrasoundtherapy ----------------------------------------------------- 21
4.2.6 Parameters - Ultrasoundtherapy ---------------------------------------------------------------------- 22
4. 3 CombinationTherapy ---------------------------------------------------------------------------------------22
5. Package Contents ---------------------------------------------------------------------------------------- 22
6. Installation -------------------------------------------------------------------------------------------------- 24
6. 1 SystemswithoutaVacuumUnit --------------------------------------------------------------------------24
6. 2 SystemswithaVacuumUnit ------------------------------------------------------------------------------24
6. 3 ConnectiontoMainsSupply ------------------------------------------------------------------------------- 24
6. 4 DisconnectionfromMainsSupply -----------------------------------------------------------------------24
6. 5 PlacingtheBattery -------------------------------------------------------------------------------------------25
6. 6 OperationfromBattery -------------------------------------------------------------------------------------- 25
7. Application Information -------------------------------------------------------------------------------- 26
7. 1 Electrotherapy -------------------------------------------------------------------------------------------------26
7.1.1 Beforetreatment -------------------------------------------------------------------------------------------- 26
7.1.2 Duringtreatment -------------------------------------------------------------------------------------------- 26
7.1.3 Endingtreatment -------------------------------------------------------------------------------------------- 26
7.1.4 After treatment ----------------------------------------------------------------------------------------------- 26
7.1.5 Rubberelectrodes ------------------------------------------------------------------------------------------ 27
7.1.6 Vacuumelectrodes ----------------------------------------------------------------------------------------- 27
7.1.7 Self-adhesiveelectrodes ---------------------------------------------------------------------------------- 27
7.1.8 Electrolyticeects ------------------------------------------------------------------------------------------- 28
7.1.9 Current density ---------------------------------------------------------------------------------------------- 28
7.1.10 Connection and disconnection reactions ------------------------------------------------------------- 28
7. 2 Ultrasound ------------------------------------------------------------------------------------------------------28
7.2.1 Contactcontrol ----------------------------------------------------------------------------------------------- 28
7.2.2 Thecontactmedium ---------------------------------------------------------------------------------------- 29
7.2.3 Beforetreatment -------------------------------------------------------------------------------------------- 29
7.2.4 Startingtreatment ------------------------------------------------------------------------------------------- 29
7.2.5 Duringtreatment -------------------------------------------------------------------------------------------- 29
7.2.6 Aftertreatment ----------------------------------------------------------------------------------------------- 29
7. 3 Vacuum ---------------------------------------------------------------------------------------------------------30
8. Operating Instructions ---------------------------------------------------------------------------------- 30
8. 1 OperatorControls --------------------------------------------------------------------------------------------30
8. 2 BasicOperation -----------------------------------------------------------------------------------------------32
8.2.1 Turningonthedevice -------------------------------------------------------------------------------------- 32
8.2.2 Displayscreen ----------------------------------------------------------------------------------------------- 32
8.2.3 Navigation ----------------------------------------------------------------------------------------------------- 33
8.2.4 Shuttingdevicedown -------------------------------------------------------------------------------------- 42
8.2.5 Operatingdetails -------------------------------------------------------------------------------------------- 43
9. Maintenance and Troubleshooting ----------------------------------------------------------------- 45
9. 1 CleaningandDisinfection ----------------------------------------------------------------------------------45
9.1.1 Cleaning ------------------------------------------------------------------------------------------------------- 45
9.1.2 Disinfection --------------------------------------------------------------------------------------------------- 46
9. 2 Troubleshooting ----------------------------------------------------------------------------------------------- 46
9.2.1 Errorcode ----------------------------------------------------------------------------------------------------- 46
9.2.2 Warningmessage ------------------------------------------------------------------------------------------- 46
9. 3 Maintenance ---------------------------------------------------------------------------------------------------48
9.3.1 Optimizecontactcontrolultrasoundapplicator ------------------------------------------------------ 48
9.3.2 Back-upandrestorefavorites---------------------------------------------------------------------------- 48
9.3.3 To restore your favorites----------------------------------------------------------------------------------- 48
9.3.4 Firmware update -------------------------------------------------------------------------------------------- 48
9. 4 TechnicalMaintenance--------------------------------------------------------------------------------------49
9. 5 Disposal ---------------------------------------------------------------------------------------------------------49
10. Specications --------------------------------------------------------------------------------------------- 50
10. 1 TechnicalData ------------------------------------------------------------------------------------------------50
10. 2 SafetyandPerformanceStandards ---------------------------------------------------------------------50
10. 3 DescriptionofSymbols -------------------------------------------------------------------------------------51
10. 4 EMCDetails ---------------------------------------------------------------------------------------------------52
11. Consumable Parts --------------------------------------------------------------------------------------- 56
12. Warranty and Disclaimer ------------------------------------------------------------------------------- 57
13. Supplementary Information --------------------------------------------------------------------------- 58
4
1. Foreword
1. 1 Intended User / Operator
Thismanualhasbeenwrittenfortheownersandoperatorsofthisproduct.Itcontainsgeneral
instructionsonoperation,precautionarypractices,maintenanceandpartsinformation.Inorderto
maximizetheuse,eciencyandlifespanofyourunit,pleasereadthismanualthoroughlyandbecome
familiar with the controls as well as the accessories before operating the unit.
This product is intended to be used, and shall only be used, by or under the supervision of professional
usersintheeldofphysicaltherapyandrehabilitation.
Thisproductisprescriptionequipment.Usebyanypersonsotherthanphysiciansisprohibited.
Specicationsputforthinthismanualwereineectatthetimeofpublication.However,owingtoOG
WellnessTechnologiesherpolicyofcontinualimprovement,changestothesespecicationsmaybe
made at any time without obligation on the part of OG Wellness Technologies.
1. 2 Product Liability
Tothemaximumextentpermittedbyapplicablelaw,innoeventwillEnraf-Noniusoritssuppliersor
resellers be liable for any indirect, special, incidental or consequential damages arising from the use of
orinabilitytousetheproduct,including,withoutlimitation,damagesforlossofgoodwill,workand
productivity, computer failure or malfunction, or any and all other commercial damages or losses, even
if advised of the possibility thereof, and regardless of the legal or equitable theory (contract, tort or
otherwise)uponwhichtheclaimisbased.Inanycase,Enraf-Nonius’sentireliabilityunderany
provision of this agreement shall not exceed in the aggregate the sum of the fees paid for this product
andfeesforsupportoftheproductreceivedbyEnraf-Noniusunderaseparatesupportagreement(if
any),withtheexceptionofdeathorpersonalinjurycausedbythenegligenceofEnraf-Noniustothe
extent applicable law prohibits the limitation of damages in such cases.
Enraf-Noniuscannotbeheldliableforanyconsequenceresultingfromincorrectinformationprovided
byitspersonnel,orerrorsincorporatedinthismanualand/orotheraccompanyingdocumentation
(including commercial documentation).
Theopposingparty(product’suseroritsrepresentative)shalldisclaimEnraf-Noniusfromallclaims
arising from third parties, whatever nature or whatever relationship to the opposing party.
2. Product Description
Combination Therapy Equipment CT-7:
This product is a combination device, combining the functions of the electrotherapy and the ultrasound
therapy in a single device.
The simultaneous application of ultrasound and electrotherapy (combination therapy) is also possible.
The remaining electrotherapy channel can then be used independently.
Vacuum unit for CT-7 (CT7-001):
ElectrotherapycanbeperformedusingtheoptionalVacuumunit.Electrotherapycanbeapplied
throughstandardorvacuumelectrodes.WithvacuumelectrodestheVacuumunitforCT-7generates
the vacuum through which the vacuum electrodes are attached to the patient. The unit is placed
beneathCombinationTherapyEquipmentCT-7,fromwhichitspowerisderivedandthroughwhichitis
also operated.
StatUS for CT-7 (CT7-005):
StatUStherapy(StationaryUltraSoundtherapy)canbeperformedusingtheoptionalStatUSforCT-7.
WhenperformingStatUStreatment,readtheinstructionmanualaccompanyingtheStatUSforCT-7
(CT7-005).
ElectrotherapyandStatUStherapycannotbeusedincombination.
StatUSforCT-7(CT7-005)isnotdesignedforuseincombinationwiththeVacuumUnitforCT-7
(CT7-001).
5
3. Precautionary Instructions
Inthissection,generalWarningsandPrecautionsarelisted,thatyoushouldbeawareofwhenusingthis
product.Shouldtheuseofthisdeviceleadtoanundesirableeventsuchasdeathor serious injury to the
user,themanufacturerandthecompetentauthorityoftheMemberStateMUSTbeinformedwithoutdelay.
Seealsochapter4. 1 and 4. 2forWarningsandPrecautionsthatareapplicationspecic.
D WARNING:
•This product should be used only under the continued supervision of a physician or licensed
practitioner.
•Makecertainthatthisproductiselectricallygroundedbyconnectingonlytoagroundedelectrical
service receptacle conforming to the applicable national and local electrical codes.
•Do not operate the unit in an environment of short-wave or micro-wave diathermy.
•Thisproductisnotsuitableforuseinthepresenceofammableanestheticsmixturewithair,oxygen,
or nitrous oxide.
•Thisproductshouldbekeptoutofthereachofchildren.
•Thisproductisnotsuitableforuseinthepresenceofammableanestheticmixturewithair,oxygen,
or nitrous oxide.
•Useofcontrolsoradjustmentsorperformanceofproceduresotherthanthosespeciedhereinmay
resultinhazardousexposuretoultrasonicenergy.
•Useofaccessories,transducersandcablesotherthanthosespeciedorprovidedbythe
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
•Caremustbetakenwhenoperatingthisequipmentaroundotherequipment.Potential
electromagneticorotherinterferencecouldoccurtothisortotheotherequipment.Trytominimize
this interference by not using other equipment in conjunction with it.
•Itisrecommendedthatthisproductshouldnotbeusedafterextremetemperatureuctuations.
•This product should not be used when any sort of mechanical damage is noticed.
•Do not operate the unit in an environment of short-wave or micro-wave diathermy.
D CAUTION:
•Alwaysdeterminethedosagebasedonthepatient’sperceptionofheat.Anysensationgreaterthan
mild warmth could cause a burn.
•Do not operate this product when connected to any other devices.
•Donotusethisdeviceinsocalled“WetRooms”(hydrotherapyrooms)
•Thisunitshouldbeoperatedintemperaturesbetween10°Cand40°C(50°Fand104°F),witha
RelativeHumidityrangingfrom10-90%(non-condensing).
•Do not expose the unit to direct sunlight, heat radiated from a heat radiator, excessive amounts of
dust,moisture,vibrationsandmechanicalshocks.
•Inthecaseofingressofliquids,unplugtheunitfromthemainssupplyandhaveitcheckedbyan
authorizedperson(seetheparagraphontechnicalmaintenance).
•Beforeadministeringanytreatmenttoapatient,youshouldbecomeacquaintedwiththeoperating
procedures for each mode of treatment available, as well as the indications, contraindications,
warnings and precautions.
•Handleultrasoundtreatmentheadwithcare.Inappropriatehandlingoftheultrasoundtreatmenthead
mayadverselyaectitscharacteristics.
•Inspectultrasoundtreatmentheadforcracksandothermechanicaldefectswhichmayallowthe
ingressofconductiveuidbeforeeachuse.
6
4. Intended Use
This product is designed to only be used by or under the supervision of persons using the medical device
inthecourseoftheirworkandintheframeworkofaprofessionalhealthcareactivity,whounderstandthe
benetsandlimitationsofelectrotherapyandultrasoundtherapy.
4. 1 Electrotherapy
This product is intended to be used to treat nerve disorders or muscle disorders or pain, by providing
transcutaneous stimulation to mitigate of pain or musculer atrophy.
4.1.1 Pain management
PainManagementistheuseofelectricalstimulationforpainrelief.
4.1.1.1 Indications - Pain management
Symptomaticreliefofchronic,intractablepain.Managementofpainassociatedwithpost-traumaticor
postoperative conditions.
4.1.1.2 Contra-indications - Pain management
•This product should not be used for symptomatic pain relief unless etiology is established or unless a
pain syndrome has been diagnosed.
•Thisproductshouldnotbeusedonpatientswithdemand-typecardiacpacemakers.
•This product should not be used over cancerous lesions.
•Electrodeplacementsthatapplycurrenttothecarotidsinusregion(anteriorneck)mustbeavoided.
•Electrodeplacementsthatapplycurrenttranscerebrally(throughthehead)mustbeavoided.
•Electrodeplacementsthatapplycurrenttransthoracically(theintroductionofelectricalcurrentintothe
heart may cause cardiac arrhythmias) must be avoided.
4.1.1.3 Warnings - Pain management
•BenetsofTENScurrentshavenotbeenestablishedforpainofcentralorigin.
•Thisproductistobeusedasasymptomatictreatmentforpainandhasnocurativevalue.Patients
should be cautioned and their activities regulated if pain that would otherwise serve as a protective
mechanism is suppressed.
•Thelong-termeectsofchronicelectricalstimulationareunknown.
•Safetyhasnotbeenestablishedfortheuseoftherapeuticelectricalstimulationduringpregnancy.
•Stimulationshouldnotbeappliedoverswollen,infected,orinamedareasofskineruptionse.g.,
phlebitis, thrombophlebitis, varicose veins, etc.
•Seealsochapter3,PrecautionaryInstructions,forgeneralWarningsandPrecautions.
4.1.1.4 Precautions - Pain management
•Isolatedcasesofskinrashmayoccuratthesiteofelectrodeplacementfollowinglong-term
applications. The irritation may be reduced by use of an alternate conductive medium or an alternative
electrode placement.
•Eectivenessofthistreatmentisdependentuponpatientselection.
•Seealsochapter3,PrecautionaryInstructions,forgeneralWarningsandPrecautions.
4.1.1.5 Adverse eects - Pain management
Skinirritationandburnsbeneaththeelectrodeshavebeenreportedwiththeuseoftherapeutic
electrical stimulation.
4.1.1.6 Current waveforms - Pain management
For pain management the following current waveforms are recommended.
4.1.3.1, 4.1.3.2, 4.1.3.3, 4.1.3.5, 4.1.3.6, 4.1.3.8, 4.1.3.9.2
7
4.1.2 Muscle stimulation
MuscleStimulationistheuseofelectricalstimulationtotreatmuscledisorders.
4.1.2.1 Indications - Muscle stimulation
•Relaxationofmusclespasms
•Preventionorretardationofdisuseatrophy
•Increasinglocalbloodcirculation
•Musclere-education
•Immediatepost-surgicalstimulationofcalfmusclestopreventvenousthrombosis
•Maintainingorincreasingrangeofmotion
•Dysphagia
4.1.2.2 Contra-indications - Muscle stimulation
•Thisproductshouldnotbeusedonpatientswithdemand-typecardiacpacemakers.
•This product should not be used over cancerous lesions.
•Electrodeplacementsthatapplycurrenttothesinuscarotidregion(anteriorneck)mustbeavoided.(1)
•Electrodeplacementsthatapplycurrenttranscerebrally(throughthehead)mustbeavoided.
•Electrodeplacementsthatapplycurrenttransthoracically(theintroductionofelectricalcurrentintothe
heart may cause cardiac arrhythmias) must be avoided.
(1)
However,theapplicationareaforthetreatmentofDysphagiaisdistantenoughfromthesinuscarotis
area,whenthetherapistfollowstheguidelinesasdescribedinthetherapybook“Dysphagia(by
H.C.A.BogaardtSLP,PhD)”
4.1.2.3 Warnings - Muscle stimulation
•Thelong-termeectsofchronicelectricalstimulationareunknown.
•Safetyhasnotbeenestablishedfortheuseoftherapeuticelectricalstimulationduringpregnancy.
•Stimulationshouldnotbeappliedoverswollen,infected,orinamedareasofskineruptionse.g.,
phlebitis, thrombophlebitis, varicose veins, etc.
•Seealsochapter3,PrecautionaryInstructions,forgeneralWarningsandPrecautions.
4.1.2.4 Precautions - Muscle stimulation
•Adequateprecautionsshouldbetakenwhentreatingindividualswithsuspectedordiagnosedheart
problems, or epilepsy.
•Caution should be used when there is a tendency to hemorrhage following acute trauma or fracture.
•Caution should be used following recent surgical procedures when muscle contraction may disrupt
the healing process.
•Caution should be used over the menstruating uterus.
•Cautionshouldbeusedoverareasoftheskinwhichlacknormalsensation.
•Somepatientsmayexperienceskinirritationorhypersensitivityduetoelectricalstimulationor
electrical conductive medium. The irritation can usually be reduced by using an alternative conductive
medium, or alternate electrode placement.
•Seealsochapter3,PrecautionaryInstructions,forgeneralWarningsandPrecautions.
4.1.2.5 Adverse eects - Muscle stimulation
Skinirritationandetchingbeneaththeelectrodeshavebeenreportedwiththeuseoftherapeutic
electrical stimulation.
4.1.2.6 Current waveforms - Muscle stimulation
•For muscle stimulation the following current waveforms are recommended.
4.1.3.2.1, 4.1.3.3, 4.1.3.4, 4.1.3.6, 4.1.3.9.1
These waveforms are often applied in combination with a surge program, which consists of a sequence
of exercise and rest periods. Two options are available here:
•Reciprocalapplication,wherestimulationalternatesbetweenagonistsandantagonists.Thisis
accomplished through asynchronous stimulation over two current channels with an appropriate delay
between the two channels.
•Co-contract application, where two channels operate synchronously to co-contract agonist and
antagonistordierentsectionsofalargermusclegroup.
Seeparagraph8.2.5.6 for the available parameters.
8
4.1.3 Description of current waveforms
4.1.3.1 4-pole interferential currents
With the interferential current type a medium frequency carrier frequency is used to pass the low
frequencystimulation(beat)frequencythroughtheskin.Therelativelylowresistanceoftheskintothe
carrier frequency contributes to the patient comfort that is often associated with this current type.
InterferentialcurrentsareallACcurrentswithoutanyresidualDCcomponents.Severalvariationsof
theinterferentialcurrenttypeareknown,ofwhichthefollowingareavailableinthisproduct.
4.1.3.1.1 Classical interferential
With this therapy method four electrodes are used and two non modulated currents are generated. The
frequencyofonechannelisxedatthecarrierfrequency,whiletheotherchannelhasavariable
frequency,basedontheBeatfrequencyandFrequencyModulationsettings.Interferenceoccurswhere
the two currents intersect in the tissue. The modulation depth (which determines the current amplitude
ofthestimulation)dependsonthedirectionofthecurrents,andcanvaryfrom0to100%.100%
modulation depth only occurs at the diagonals (and hence at the intersection) of the two currents. This
isofcourseatheoreticalsituation,basedontheassumptionthatthetissueishomogeneous.Inreality,
the tissue is heterogeneous, so that the current balance between the two channels has to be used to
obtainthe100%modulationdepth(Fig.1).Thecurrentbalancecanalsobeusedtocompensatefor
dierencesinsensationoccurringundertheelectrodes.
Modulationdepthisonly100%atthediagonal
4.1.3.1.2 Isoplanar vector
Theisoplanarvectortechniqueisintendedtoincreasetheareawhereeectivestimulationoccurs.
Amplitude modulation occurs in the equipment and a special phase relation between the two channels
ensuresa100%modulationdepthbetweenthefourelectrodesinallpositions.
Modulationdepthis100%overtheentiretreatmentarea.
Theadvantageofthismethodisthatthepositioningofthefourelectrodestoeectivelytreatthe
aectedtissueislesscritical.ThesensationoftheIsoplanarvectormodeissoftandequallydivided
over the treatment area.
Fig.1
9
4.1.3.1.3 Dipole vector manual
With the dipole vector technique the currents from the two electrode pairs are vectorially summed in
thetissue.Theeectisthatstimulationonlyoccursintothedirectionoftheresultingvector,whichcan
be adjusted over a range of 360º. Amplitude modulation occurs in the equipment and the modulation
depthis100%.
Stimulationwith100%modulationdepthonlyoccursintothe
direction of the vector.
The advantage of this method is that the direction of the stimulation
can be adjusted electronically after positioning the electrodes.
4.1.3.1.4 Dipole vector automatic
With the automatic dipole technique the dipole vector described above is rotated at an adjustable
speed.Ifthecurrentamplitudeisincreasedexceedingthemotorialthreshold,thetissuewillcontract
and relax rhythmically. The automatic dipole vector current is ideally suited for areas where mechanical
pressure (massage) is not desirable.
Parameters:
•Carrier frequency, expressedinkHz,isthebasefrequencyofthealternatingcurrent.
•Beat frequency, expressedinHz,denesthechannelfrequencydierenceinclassicalinterferential
mode and the rate at which the amplitude is internally modulated in the vector modes.
•Frequency Modulation, expressedinHz,denesavariablefrequencyrangethatissummedtothe
Beatfrequencyi.ewhentheBeatfrequencyissetto80HzandtheFrequencymodulationissetto40
Hz,thenalfrequencywillvaryfrom80to120Hz.Frequencymodulationisoftenusedtoprevent
accommodation to stimulation or to improve patient tolerance.
•Modulation Programdenesthetimeandsequenceinwhichthefrequencywillsweepthroughthe
Frequency modulation range.
•Balancedenesthedierenceincurrentamplitudebetweenthetwochannels.Onlyavailablein
classical interferential mode.
•Vector position adjustmentdenestheangleofthedipolevectorwithrespecttothepositionofthe
electrodes.
•Rotation Speed, expressedinsecond,denesthetimeelapsedduringonerevolutionofthevectorin
automatic dipole vector mode.
4.1.3.2 Biphasic pulsed currents (TENS)
4.1.3.2.1 Asymmetrical and alternating asymmetrical
TheasymmetricalbiphasicpulsedcurrentwaveformisoftenusedinTENS(TranscutaneousElectrical
NerveStimulation)applications.Thiswaveformischaracterizedbyvariablephasedurationand
variablepulsefrequency.Itstypicalamplitude,duration,andrateofriseanddecayareunequalfor
each phase with respect to the baseline. The waveform is fully balanced, i.e. the phase charges of
eachphaseareequal.Seeg4.1.4.3.1 for a graphical representation.
A variation to the standard biphasic asymmetrical pulsed current is the alternating one, in which the
successivepulsephasesalternatewithrespecttothebaseline.Seeg4.1.4.3.2 for details. This
waveform is also fully balanced.
To prevent accommodation to stimulation or to improve patient tolerance, the pulse frequency can be
variedthroughfrequencymodulation.Severalfrequencymodulationprogramsareavailable.Seeg
4.1.4.7 for details.
10
Parameters:
•Phase Duration, expressed in µs, is the elapsed time from the beginning to the end of the initial pulse
phase.Seeg4.1.4.3.1 for details.
•Pulse Frequency, expressedinHzorpps(pulsespersecond),denestherepetitionrateofthe
TENSpulses.Seeg4.1.4.3.1 for details.
•Frequency Modulation, expressedinHz,denesavariablefrequencyrangethatissummedtothe
Pulsefrequencyi.ewhenthePulsefrequencyissetto80HzandtheFrequencymodulationissetto
40Hz,thenalfrequencywillvaryfrom80to120Hz.
•Modulation Program, denesthetimeandsequenceinwhichthefrequencywillsweepthroughthe
Frequencymodulationrange.SeefortheavailableModulationprograms4.1.4.7.
4.1.3.2.2 Burst asymmetrical and burst alternating asymmetrical
The burst biphasic and burst biphasic alternating asymmetrical pulsed currents are variations to their
nonburstcounterparts,inwhichthecontinuoustrainofpulsesisinterruptedbypulsepauses.Seeg
4.1.4.3.3andg4.1.4.4.4 for details. A burst frequency can be set for treating chronic pains, where the
useofcontinuousstimulationwithalowpulsefrequencywouldbetoopainful.Eachburstlastsfor
100msandtheburstratecanbeadjustedseparately.WiththismilderTENSwaveformitiseasierto
exceed the motorial threshold stimulus.
Parameters:
•Phase Duration, expressed in µs, is the elapsed time from the beginning to the end of the initial pulse
phase.Seeg4.1.4.3.3 for details.
•Pulse Frequency, expressedinHzorpps(pulsespersecond),denestherepetitionrateofthe
TENSpulses.Seeg4.1.4.3.3 for details.
•Burst Frequency, expressedinHz,denestherepetitionrateofburstsofpulses.Aburstconsistsof
atrainofpulses.Eachburstlastsfor100msandthenumberofpulsesinaburstdependsonthe
selectedPulsefrequencyi.eataPulseFrequencyof100Hz,10pulsesareavailableineachburst.
Seeg4.1.4.3.4 for details.
4.1.3.2.3 Symmetrical
TENScurrentpulsescanalsobeusedformusclestimulationapplications.Oftenthesymmetrical
biphasicpulsedcurrentwaveformisused.Seeg4.1.4.3.5 for a graphical representation. The
speciedphasedurationappliestobothpulsephases,whichdoublestheamountofavailableenergy
with respect to the asymmetrical pulsed current waveform. This waveform is fully balanced (no residual
DC components are present).
Parameters:
•Phase Duration, expressed in µs, is the elapsed time from the beginning to the end of a pulse phase.
Thephasedurationappliestoeachpulsephase.Seeg4.1.4.3.5 for details.
•Pulse Frequency, expressedinHzorpps(pulsespersecond),denestherepetitionrateofthe
TENSpulses.Seeg4.1.4.3.5 for details.
•Frequency Modulation, expressedinHz,denesavariablefrequencyrangethatissummedtothe
Beatfrequencyi.ewhentheBeatfrequencyissetto80HzandtheFrequencymodulationissetto40
Hz,thenalfrequencywillvaryfrom80to120Hz.
•Modulation Program denesthetimeandsequenceinwhichthefrequencywillsweepthroughthe
Frequencymodulationrange.SeefortheavailableModulationprogramsg4.1.4.7.
•Surge Programcanbeusedtoadjustrepeatedsequencesofcontractionandrestperiods.See
paragraph 8.2.5.6 for details.
11
4.1.3.2.4 Burst symmetrical
The burst biphasic symmetrical pulsed current is a variation to its non burst counterpart, in which the
continuoustrainofpulsesisinterruptedbypulsepauses.Seeg4.1.4.3.6 for details. A burst frequency
can be set for treating chronic pains, where the use of continuous stimulation with a low pulse
frequencywouldbetoopainful.Eachburstlastsfor100msandtheburstratecanseparatelybe
adjusted.WiththismilderTENSwaveformitiseasiertoexceedthemotorialthresholdstimulus.
Parameters:
•Phase Duration, expressed in µs, is the elapsed time from the beginning to the end of the initial pulse
phase.Seeg4.1.4.3.5 for details.
•Pulse Frequency, expressedinHzorpps(pulsespersecond),denestherepetitionrateofthe
TENSpulses.Seeg4.1.4.3.5 for details.
•Burst Frequency, expressedinHz,denestherepetitionrateofburstsofpulses.Aburstconsistsof
atrainofpulses.Eachburstlastsfor100msandthenumberofpulsesinaburstdependsonthe
selectedPulsefrequencyi.eataPulseFrequencyof100Hz,10pulsesareavailableineachburst.
Seeg4.1.4.3.6 for details.
4.1.3.3 Premodulated
AswithInterferentialcurrents,amediumcarrierfrequencyisusedtopassthelowfrequency
stimulation(beat)frequencythroughtheskin.Seeg4.1.4.1forthecurrentwaveform.‘Premodulated’
implies that amplitude modulation occurs in the equipment, allowing it to be applied with a single
electrode pair.
ThePremodulatedalternatingcurrentisoftenusedwheretheobjectiveistostrengthenthemuscleand
changethedistributionofmusclebers(twitchspeed).TheBeatfrequencyisusedtoaectthemuscle
berdistribution.Theoptimumcarrierfrequencyforthispurposevariesbetween2000and4000Hz.
AtalowBeatfrequency(uptoabout20Hz)themusclebecomes‘red’,whileatahigherBeat
frequency(uptoabout150Hz)themusclebecomes‘white’.Thiscanbeusedtoincreasethe
explosive release of energy in high-jumpers, provided that is supplemented by functional exercises.
ThemostcomfortabletetaniccontractionsareobtainedataBeatfrequencybetween40and80Hz.
MusclestimulationisnormallyappliedwithaSurgeprogram,allowingthemusclestorestbetween
exercise cycles.
Parameters:
•Carrier Frequency, expressedinkHz,isthebasefrequencyofthealternatingcurrent.
•Beat Frequency, expressedinHz,denestherateatwhichtheamplitudeisinternallymodulated.
•Frequency Modulation, expressedinHz,denesavariablefrequencyrangethatissummedtothe
Beatfrequencyi.e.whentheBeatFrequencyissetto80HzandtheFrequencymodulationissetto
40Hz,thenalfrequencywillvaryfrom80to120Hz.
•Modulation Programdenesthetimeandsequenceinwhichthefrequencywillsweepthroughthe
Frequencymodulationrange.SeefortheavailableModulationprogramsg4.1.4.7.
•Surge Programcanbeusedtoadjustrepeatedsequencesofcontractionandrestperiods.See
paragraph 8.2.5.6 for details.
12
4.1.3.4 Russian stimulation
Thiscurrenttypeisanintermittentalternatingcurrentwithacarrierfrequencyaround2500Hz.Seeg
4.1.4.2forthecurrentwaveform.RussianStimulationwasrstusedbyKots,alecturerinsports
medicineattheMoscowStateAcademy.Kotsuseditformusclestrengtheninginprosthesiologyandin
thetrainingofRussiancosmonauts.Withthistechniquetheelectrostimulationisappliedbothto
individualmusclesandtogroups(eitherdirectlyorviathenerve).Indirectstimulation,afrequencyof
2500Hzwasfoundtoproducethelargestcontraction,whiletheoptimumfrequencyinindirect
stimulationwas1000Hz.
Aspecicfeatureofthistypeofmusclestimulationisthatthealternatingcurrentisinterrupted50times
persecond.Thisresultsinapulsetrain,comparabletothe‘burst’inTENS.Thetotaldurationofthe
pulsetrainis20ms,givingaphaseduration/phaseintervalratioof1:1.KotsusesaBurstfrequencyof
50Hz,approximatelyinthemiddleofthefrequencyspectrumusedtoproducetetaniccontraction(40
to80Hz).Inadditiontothe1:1ratio,Kotsalsodescribesaphaseduration/phaseintervalratioof1:5.
The amplitude should be increased until a powerful contraction is produced (from the motor stimulation
leveluptothelimitoftolerance).AswithallmusclestimulationapplicationsaSurgeprogramcanbe
used, allowing the muscles to rest between exercise cycles.
Parameters:
•Carrier Frequency, expressedinkHz,isthebasefrequencyofthealternatingcurrent.
•Burst Frequency, expressedinHz,denestherepetitionrateofthebursts.
•Burst / Interval Ratio, denestheratiooftheburstlengthtotheintervalbetweenthebursts.The
sum of the burst and interval duration is the reciprocal of the burst frequency i.e. with a burst
frequencysetat50Hzandaburst/intervalratioof1:5,theburstdurationwillbe20*1/6=3.3ms
andtheintervaldurationwillbe20*5/6=16.7ms.
•Surge Program canbeusedtoadjustrepeatedsequencesofcontractionandrestperiods.See
paragraph 8.2.5.6 for details.
4.1.3.5 Micro current
MicroCurrentisamonophasicrectangularwaveformwithmanuallyselectableoralternatingpolarity.
Seeg4.1.4.5.4foragraphicalrepresentation.ManytherapistspreferMicroCurrenttherapybecause
of the low current amplitudes used. Alternating polarity can be used to average out the DC component,
thereby reducing the formation of electrolysis by-products.
Parameters:
•Frequency, expressedinHz,isthenumberofcyclespersecond.
•Alternation mode deneswhetherthepolarityofthewaveisautomaticallyalternatingornot.
•Alternation Period, expressedinsecond,denesthepolarityreversaltiminginthealternatingmode.
•Surge Program canbeusedtoadjustrepeatedsequencesofcontractionandrestperiods.See
paragraph 8.2.5.6fordetails.Surgeprogramsareonlyavailableinthenonalternatingmode.
13
4.1.3.6 High voltage
Thiscurrenttypehasatwinpeakmonophasicwaveformwithaxeddurationof64µsbetweenthetwo
voltagepeaks.TheamplitudeisadjustedinvoltsratherthaninmA.Theshortrisetimeandshort
durationofeachvoltagepeak(approximately7µs)iswellsuitedtonervestimulationandecient
discrimination between sensory, motor and pain responses. The very short pulse duration of high
voltage creates a stimulation which is quite comfortable, and one which most patients can tolerate.
The very short pulse duration followed by a very long interpulse interval eliminates the formation of any
appreciablechemicalorthermaleectsinthetissue.Highvoltageisusedforstimulatingnervesand
muscles,causingmusclecontractions.Examplesforclinicalusearetotreatacuteorchronicpain,
edemaabsorptionandulcerhealing.Musclecontractionormotorresponseofisolatedmusclegroups,
supercialordeep,canbeeasilyandcomfortablystimulated.Therelativecomfortanddepthof
penetrationmaybethekeyfortheusefulnessofhighvoltagestimulationinclinicalconditionssuchas
tendontransplants,jointmobilizationandmusclere-education.
Parameters:
•Pulse Frequency, expressedinHzorpps(pulsespersecond),denestherepetitionrateofthetwin
pulses.See4.1.4.5.3 for details.
•Frequency Modulation, expressedin%,denesavariablefrequencyrangethatissubtractedfrom
thePulsefrequencyi.e.whenthePulsefrequencyissetto80HzandtheFrequencymodulationis
setto50%,thenalfrequencywillvaryfrom40–80Hz.
•Modulation Program denesthetimeandsequenceinwhichthefrequencywillsweepthroughthe
Frequencymodulationrange.SeefortheavailableModulationprogramsg4.1.4.7.
•Alternation Modedeneswhetherthepolarityofthepulsesisautomaticallyalternatingornot.
•Alternation Period, expressedins,denesthepolarityreversaltiminginthealternatingmode.
•Surge Program canbeusedtoadjustrepeatedsequencesofcontractionandrestperiods.See
paragraph 8.2.5.6fordetails.Surgeprogramsareonlyavailableinthenonalternatingmode.
4.1.3.7 Diadynamic currents
TheDiadynamicCurrentsthatwereintroducedbyBernard(2) are mainly used for pain reduction and the
improvement of blood circulation.
Bernardusestheterm‘DiadynamicCurrent’torefertoamonophase(MF–MonophaséFixe)or
double-phase(DF–DiphaséFixe)rectiedalternatingcurrent.Thefrequencywasdirectlyderived
from the mains supply, resulting in sinusoidal pulses with a duration of 10ms. This phase time of 10ms
willmainlydepolarizethickbers.Stimulationofthinberscanonlybeobtainedathighercurrent
amplitudes.
(2)Bernard,PierreD.
Lathérapiediadynamique,
Paris,Editions‘Physio’,1962.
D CAUTION:
•Diadynamic currents are monophasic currents that produce electrolysis by-products. These by-
products can result in burns beneath the electrodes. Always use properly moistened sponge and
electrodecombinationstoabsorbtheseby-productsduringtreatment.Seeparagraph7.1.8 for details.
The following variations are available:
4.1.3.7.1 MF (Monophasé Fixe)
Isasinglephaserectiedsinusoidalcurrentwithafrequencyof50Hz.Seeg4.1.4.6.1fordetails.MF
is a vibrating waveform that easily induces contractions.
14
4.1.3.7.2 DF (Diphasé Fixe)
Isadualphaserectiedsinusoidalcurrentwithafrequencyof100Hz.Seeg4.1.4.6.2 for details. DF
isusuallyexperiencedasaslightvibration.Itisapleasantwaveformthatisoftenusedasan
introductiontoCPorLP.
4.1.3.7.3 LP (Longues Périodes)
IsaslowalternationbetweensixsecondsofMFcurrentandasix-secondDFcurrent.IntheDFphase
theintervalsbetweentheMFpulsesarelledwithadditionalpulseswithgraduallyincreasingand
decreasingamplitude.Seeg4.1.4.6.3fordetails.LPissmootherthanCP.
4.1.3.7.4 CP (Courtes Périodes)
IsarapidalternationbetweenonesecondofMFcurrentandonesecondofDFcurrent.Seeg
4.1.4.6.4fordetails.CPhasastrongresorbingeect.
4.1.3.7.5 CPid
IsidenticaltoCP,exceptthatthecurrentamplitudeduringtheMFphaseis12.5%lowerthanduring
theDFphase.Seeg4.1.4.6.5fordetails.Normallyalowerfrequencyisexperiencedtobemore
aggressivethanahigherfrequency.CPidpreventsthisdierenceinsensation.
Parameters:
•Surge Program canbeusedtoadjustrepeatedsequencesofcontractionandrestperiods.See
paragraph 8.2.5.6fordetails.SurgeprogramsareonlyavailablewithMFandDF.
4.1.3.8 Galvanic current
4.1.3.8.1 Continuous galvanic current
Galvaniccurrentworkswhencombinedwiththecorrectionized/electricallychargedsolutions,(i.e.
theyareionscarryingeitherapositiveornegativeelectricalcharge,orwillionizewithelectricity).
Thismakesitpossibletoinuencetheskin'sabilitytoabsorbserumsintotheintracellularspacesinthe
dermis. The absorption process is called iontophoresis because the electrical currents literally carry
ionsintothetissuesbetweenthecells(seeg4.1.4.5.2 for details).
D CAUTION:
The continuous galvanic current is a monophasic current that produces electrolysis by-products. These
by-products can result in burns beneath the electrodes. Always use properly moistened sponge and
electrodecombinationstoabsorbtheseby-productsduringtreatment.Seeparagraph7.1.8 for details.
4.1.3.8.2 Interrupted galvanic current
The medium frequency interrupted galvanic current is a monophasic rectangular waveform with a pulse
frequencyof8000Hzandadutycycleof90%.Seeg4.1.4.5.1 for details. As opposed to direct
galvanic current, the pulsed waveform provides increased patient comfort.
D CAUTION:
•The interrupted galvanic current is a monophasic current that produces electrolysis by-products.
These by-products can result in etching beneath the electrodes. Always use properly moistened
spongeandelectrodecombinationstoabsorbtheseby-productsduringtreatment.Seeparagraph
7.1.8 for details.
15
4.1.3.9 Faradic current
4.1.3.9.1 Faradic Rectangular or Triangular Pulsed Current
Faradic currents are often used for muscle stimulation applications that are based on prior diagnostics.
Seeg4.1.4.4.2 and 4.1.4.4.3 for the current waveforms. The diagnostic objective is to obtain
information on the sensitivity of the neuromuscular apparatus to electrical stimulation. This gives an
indication of the degree of denervation of the muscle tissue. With this technique the relationship
between the current amplitude and phase duration of a rectangular and triangular pulse is plotted in a
strength/durationcurve.Thestrength/durationcurveisrecordedbyobservingthecurrentamplitude
required at various phase duration values (ranging from 0.01 to 1000ms) that produce a just
perceptible (i.e. just visible or palpable) contraction of a muscle or muscle group. The values observed
canbeplottedongraphpaperwithalogarithmicscale.Inthecaseofreducedorabsentsensitivityto
electricalstimulation,thestrength/durationcurvegivesanindicationofthecurrentwaveform,phase
duration and current amplitude of the electrical stimulus to be used in any therapy that may be applied.
D CAUTION:
Faradic currents are monophasic currents that produce electrolysis by-products. These by-products
can result in etching beneath the electrodes. Always use properly moistened sponge and electrode
combinationstoabsorbtheseby-productsduringtreatment.Seeparagraph7.1.8 for details.
Parameters:
•Phase Duration, expressed in ms or s, is the elapsed time from the beginning to the end of the pulse
phase.Seeg4.1.4.4.2 and 4.1.4.4.3 for details.
•Pulse Frequency, expressedinHzorpps(pulsespersecond),denestherepetitionrateofthe
currentpulses.Seeg4.1.4.4.2 and 4.1.4.4.3 for details.
•Surge Programcanbeusedtoadjustrepeatedsequencesofcontractionandrestperiods.See
paragraph 8.2.5.6 for details.
4.1.3.9.2 Träbert, 2 – 5 Current
The2-5or‘Ultra-Reiz’currentwasintroducedbyTräbert.(1)Itisoftenusedtotreatheadachesandneck
pain.The2-5currentisafaradicrectangularpulsedcurrentwithaphasedurationof2msandaphase
intervalof5ms.Thesesettingsarethedefaultsettingsforthefaradicrectangularcurrentwaveformand
resultinapulsefrequencyofapproximately143Hz.Träbertoerednoexplanationforthechoiceof
theseparameters.Nevertheless,manyspecialistshaveadoptedthetherapyanditisstillappliedwith
success.Seeg4.1.4.4.1 for a graphical representation.
(1) Träbert,H.
Ultra-Reizstrom,einneuestherapeutischesPhänomen,
Elektromedizin2,1957(7).
D CAUTION:
Faradic currents are monophasic currents that produce electrolysis by-products. These by-products
canresultinburnsbeneaththeelectrodes.Alwaysuseproperlymoistenedsponge/electrode
combinationstoabsorbtheseby-productsduringtreatment.Seeparagraph7.1.8 for details.
Parameters:
•Phase Duration, expressed in ms or s, is the elapsed time from the beginning to the end of the pulse
phase.Thedefaultsettingis2ms.Seeg4.1.4.4.1 for details.
•Phase Interval, expressed in ms or s, is the elapsed time between successive pulse phases. The
defaultsettingis5ms.Seeg4.1.4.4.1 for details.
16
4.1.4 Illustrations of current waveforms
4.1.4.1 Premodulated / Isoplanar vector / Dipole vector
fb Beatfrequency
4.1.4.2 Russian stimulation
fBBurstfrequency
4.1.4.3 Biphasic pulsed current TENS
4.1.4.3.1 Asymmetrical
tPhaseduration
fpPulsefrequency
4.1.4.3.2 Asymmetrical alternating
tPhaseduration
fpPulsefrequency
4.1.4.3.3 Burst asymmetrical
fBBurstfrequency
4.1.4.3.4 Burst asymmetrical alternating
fBBurstfrequency
100mA max.
1/fb
17
4.1.4.3.5 Symmetrical
tPhaseduration
tiPhaseinterval
fpPulsefrequency
4.1.4.3.6 Burst symmetrical
fBBurstfrequency
4.1.4.4 Faradic current
4.1.4.4.1 Träbert, 2 – 5 Current
tpPhaseduration:2ms
tiPhaseinterval:5ms
4.1.4.4.2 Rectangular pulsed current (ms)
tpPhaseduration
tiPhaseinterval
4.1.4.4.3 Triangular pulsed current (ms)
tpPhaseduration
tiPhaseinterval
ti
ti
18
4.1.4.4.4 Rectangular pulsed current (Hz)
tpPhaseduration
fpPulsefrequency
4.1.4.4.5 Triangular pulsed current (Hz)
tpPhaseduration
fpPulsefrequency
4.1.4.5 Galvanic current
4.1.4.5.1 Galvanic interrupted
fCarrierfrequency-8kHz
xed
Dutycycle-90%xed
4.1.4.5.2 Galvanic continuous
4.1.4.5.3 High voltage
tPeakinterval-64µsxed
fpPulsefrequency
4.1.4.5.4 Micro current
f Frequency
1/
1/
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64

Enraf-Nonius CT-7 User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI