Operating instructions
1. Preparation
1-1. Connecting the probe
Insert the probe plug ⑨ of the probe ⑩ into the probe connector ⑤ of the main unit.
Align the connector keyway to the plug guide and lightly push in until it makes a clicking sound to fix.
1-2. Power switch ON
Insert the plug of the accessorized AC adaptor into AC adaptor connector ⑥ of the back face of AC adaptor.
Connect AC adaptor to the socket(AC100~240V).
Switch Power source ON to emit a 「beeping」 sound ④ and indicate 「UUU」と「μm」 on display.
Leave the state as it is for about 3 min to wait for the stability of the circuit.
1-3. How to load weight
Place the accessorized 100 g weight ⑬ on the weight guide of the probe ⑩.When this weight is always in use,
fix the weight with double-sided adhesive tape. The measurements can be taken under the perfect pressure
condition by burdening the weight on the arm of the probe.
2.Adjustments
2-1. Zeroing
Make sure that the HOLD button ③ of the main unit is OFF (green lamp on the upper side of the
HOLD button is OFF), and press the probe ⑩ on the accessorized Zero plate (standard ferrous plate for test)
(placing 100 g weight ⑬ on the probe ⑩ and turn Zero adjustment dial ① so that a numeric value on the Digital
Display ④ can become 「0 (zero)」. It is OK that a measurement value become in the neighborhood of 「0」.
when pressing the probe ⑩ plural times on Zero plate.
2-2. Calibration standard
Press an accessorized standard plate on the Zero plate and place the aforementioned probe ⑩ on the
standard plate.
Turn standard calibration dial ② so that a numerical value indicated on Digital display ④ can become a numerical
value displayed on Thickness standard. (Zeroing/Standard calibration dials are equipped with a lock function. A
press-down of the pick on the upper right activates a lock function)
Repeat the above mentioned operation 2-1, 2-2 a few times and adjust so that both can obtain stable displays.
3. Measurement
3-1. Take measurement after completing the abovementioned adjustments. Place measuring objects (screen, resist,
film, sheet, foils, paper, etc.) on the adjusted Zero plate and furthermore, place the probe ⑩ on that object.
The numerical value on Digital display ④ is the thickness of the measured object.
3-2. Measurement with a hold function
After adjusting, press Hold button ③(green light is ON), when measuring, a numerical value
is held/displayed with a 「beep」 sound. The display does no change if the probe is released.
A press again of the probe hold/displays a new measured value with a 「beep」 sound.
※When adjusting, never fail to press this button and make sure that Hold is OFF(green light is OFF).
◆ Do not press strongly like smashing to measuring objects by the probe.
Or, it may cause of the breakdown of the probe or damage measuring objects.
◆ Scrape by the probe the surface of the measuring objects.
Or, it may quicken the abrasion of electrodes and damage measuring objects.
-5-