4 -Digit programmable Multimeter HM8012
Valid at 23 °C after a 30 minute warm-up period
DC voltage
Measurement ranges: 500 mV, 5 V, 50 V, 500 V, 600 V
Resolution: 10 μV, 100 μV, 1m V, 10m V, 100m V
Accuracy:
5 V, 500 V, 600 V: ±(0.05 % of reading +0.002% of full scale)
500 mV, 50 V: ±(0.05 % of reading + 0.004% of full scale)
Overload protection:
V/Ω/T°/dB/ to COM and to chassis:
850 Vpat max. 60 Hz or 600 VDC
COM against chassis: 250 Vrms at max. 60 Hz or 250 VDC
Input resistance:
50 V, 500 V, 600 V: 10 MΩ II 90 pF
500 mV, 5 V: › 1 GΩ II 90 pF
Input current: 10 A
CMRR1): ≥100 dB (50/60Hz ± 0.5 %)
SMRR2):≥60 dB (50/60 Hz ± 0.5 %)
dB Mode
Accuracy: ±(0.02dB+2 digits) (display › -38.7 dBm)
Resolution: 0.01 dB above 18 % of rating
DC current
Measurement ranges: 500 μA, 5 mA, 50 mA, 500 mA, 10 A
Resolution: 10 nA, 100 nA, 1 μA, 10 μA, 1 mA
Accuracy:
0.5-500 mA: ±(0.2 % of reading + 0.004% of full scale)
10 A: ±(0.3 % of reading + 0.004% of full scale)
Voltage drop:
10 A range: 0.2 V max.
500 mA range: 2.5 V max.
other ranges: 0.7 V max.
AC voltage
Measurement ranges: 500 mV, 5 V, 50 V, 500 V, 600 V
Resolution: 10 μV, 100 mV, 1 mV, 10 mV, 100 mV
Accuracy 0.5-50 V:
40 Hz-5 kHz: ± (0.4 % of reading + 0.07% of full scale)
20 Hz-20 kHz: ± (1 % of reading + 0.07% of full scale)
Accuracy 500 V and 600 V:
40 Hz-1 kHz: ± (0.4 % of reading + 0.07% of full scale)
20 Hz-1 kHz: ± (1 % of reading + 0.07% of full scale)
Overload protection:
V/Ω/T°/dB/ to COM and to chassis:
850 Vpat max. 60 Hz or 600 VDC
COM against chassis: 250 Vrms at max. 60 Hz or 250 VDC
Input impedance
AC mode: 1 MΩ II 90 pF
AC + DC mode: 10 MΩ II 90 pF
Bandwidth at –3 dB: 80 kHz typical
dB mode: 20 Hz - 20 kHz
Accuracy
–23.8 dBm to 59.8 dBm: ±0.2 dBm
Resolution: 0.01 dB above 9 mV
CMRR1):≥60 dB (50/60 Hz ± 0.5 %)
Crest factor: 7 max.
34
AC current
Measurement ranges: 500 μA, 5 mA, 50 mA, 500 mA, 10 A
Resolution: 10 nA, 100 nA, 1 μA, 10 μA, 1 mA
Accuracy:
0.5 - 500 mA: ±(0.7 % of reading + 0.07% of f.s.)40 Hz - 5 kHz
10 A: ±(1 % of reading + 0.07% of full scale)
AC + DC measurements
As shown for AC + 25 digits
Resistance
Measurement ranges: 500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 50 MΩ
Resolution: 10 mΩ, 100 mΩ, 1 Ω, 10 Ω, 100 Ω, 1 kΩ
Accuracy:
500 Ω to 500 kΩ: ±(0.05 % of reading + 0.004% of f.s.+50 mΩ)
5 MΩ to 50 MΩ: ±(0.3 % of reading + 0.004% of full scale)
Input protection max. 300 Vrms
Measurement current: 500 Ω-5 kΩ range: 1 mA
50 kΩ range: 100 μA
500 kΩ range: 10 μA
5-50 MΩ range: 100 nA
Measurement voltage: 10 V typical for open inputs, depending on the
value of resistance to be measured. Negative polarity of measurement
voltage is across common terminal.
Temperature
2-wire resistance measurement with linearization for PT100 sensors as
per standard EN60751
Range: –200° C to +500° C
Resolution: 0.1° C
Measurement current: approx. 1 mA
Display: in ° C, ° F
Accuracy: ± 0.1° C from – 200°C to + 200° C
± 0.2° C from 200°C to 500° C
(exception: sensor tolerance)
Temperature coefficient: (reference 23°C)
V = 500 mV, 50 V 30 ppm/° C
600 V range 80 ppm/° C
other ranges 20 ppm/° C
V ~ 600 V range 80 ppm/° C
other ranges 50 ppm/° C
mA all ranges 200 ppm/° C
mA~all ranges 300 ppm/° C
Ω 5 MΩ, 50 MΩ ranges 200 ppm/° C
other ranges 50 ppm/° C
Miscellaneous
Power supply (from mainframe):
+ 5 V 300 mA
~26 V 140 mA
Operating temperature: + 10° C to + 40° C
Max. relative humidity: 80 % (without condensation)
Dimensions ( W x H x D) (without 22-pole flat plug):
135 x 68 x 228 mm
Weight: approx. 0.5 kg
1) Common mode rejection ratio 2) Serial mode rejection ratio
Accessories supplied: Operator’s Manual, HZ15 PVC test leads, HM8012: +
Software CD and interface cable
Optional accessories: HZ10 Silicone test leads, HZ812 PT100 Temperature Sensor