Invacare Respiratory Product IRC 1740 User manual

Category
Air compressors
Type
User manual

Invacare Respiratory Product IRC 1740 is a portable medical device designed to efficiently deliver physician-prescribed medication to the bronchial lung passages. It consists of a compressor, nebulizer, air tubing, mouthpiece, and pediatric mask. With a typical operating pressure of 12.4 to 15.4 psi and a maximum pressure of 30 to 35 psi, it operates at a flow rate of 4.0 L/min. at 16 psi operating pressure. Its compact dimensions (6.5 inches wide, 8.2 inches high, and 5.5 inches deep) and lightweight (3.4 lbs) make it easy to use and store.

Invacare Respiratory Product IRC 1740 is a portable medical device designed to efficiently deliver physician-prescribed medication to the bronchial lung passages. It consists of a compressor, nebulizer, air tubing, mouthpiece, and pediatric mask. With a typical operating pressure of 12.4 to 15.4 psi and a maximum pressure of 30 to 35 psi, it operates at a flow rate of 4.0 L/min. at 16 psi operating pressure. Its compact dimensions (6.5 inches wide, 8.2 inches high, and 5.5 inches deep) and lightweight (3.4 lbs) make it easy to use and store.

Dealer: This manual MUST be given to the end user.
User:
BEFORE using this product, read this manual and
save for future reference.
For more information regarding
Invacare products, parts, and
services, please visit
www.invacare.com
Operator’s Manual
Invacare
®
Pediatric Bear
Nebulizer System
Compressor Nebulizer System
Model IRC 1740
Invacare
®
Pediatric Bear 2 Part No 1160837
WARNING
DO NOT use this product or any available optional
equipment without first completely reading and
understanding these instructions and any additional
instructional material such as owner’s manuals, service
manuals or instruction sheets supplied with this product
or optional equipment. If you are unable to understand
the warnings, cautions or instructions, contact a
healthcare professional, dealer or technical personnel
before attempting to use this equipment - otherwise,
injury or damage may occur.
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not
recommended for use with Invacare products.
WARNING
This is NOT a toy. When not administering treatment,
keep out of the reach of small children. Some of the parts
of the nebulizer are small enough for children to swallow.
Small parts may present a choking hazard.
TABLE OF CONTENTS
Part No 1160837 3 Invacare
®
Pediatric Bear
TABLE OF CONTENTS
REGISTER YOUR PRODUCT .............................. 4
SPECIAL NOTES ............................................... 5
IMPORTANT SAFEGUARDS .............................. 6
INTRODUCTION ............................................ 11
TYPICAL PRODUCT PARAMETERS ................ 12
SUPPLEMENTAL INFORMATION ..................... 14
Compressor........................................................................................14
Extension Cords................................................................................14
SECTION 1—HANDLING ................................. 15
Shipping Instructions.........................................................................15
Unpacking............................................................................................15
Inspection............................................................................................15
SECTION 2—OPERATING INSTRUCTIONS ....... 16
Setup.....................................................................................................16
Treatment Instructions for Nebulizer..........................................19
Treatment Instructions for Mask...................................................20
SECTION 3—CLEANING AND DISINFECTING ... 21
Cleaning the Outer Case.................................................................21
Cleaning the Nebulizer and Mask..................................................22
TABLE OF CONTENTS
TABLE OF CONTENTS
Invacare
®
Pediatric Bear 4 Part No 1160837
Cleaning the Compressor ...............................................................24
Cleaning the Mouthpiece.................................................................24
SECTION 4—MAINTENANCE .......................... 25
Routine Care......................................................................................25
Storage.................................................................................................25
Nebulizer Change..............................................................................26
Filter Change ......................................................................................26
SECTION 5—TROUBLESHOOTING ................... 27
LIMITED WARRANTY .................................... 29
REGISTER YOUR PRODUCT
The benefits of registering include:
1. Safeguarding your investment.
2. Ensuring long-term maintenance and servicing of your product.
3. Receiving updates with product information, maintenance tips and
industry news.
Register ONLINE at warranty.invacare.com
Please have your model number and purchase date available to complete
your registration.
Any registration information you submit will only be used by Invacare
Corporation and protected as required by applicable laws and regulations.
SPECIAL NOTES
Part No 1160837 5 Invacare
®
Pediatric Bear
SPECIAL NOTES
Signalwordsareusedinthismanualandapplytohazardsor
unsafepracticeswhichcouldresultinpersonalinjuryor
propertydamage.Refertothetablebelowfordefinitionsof
thesignalwords.
NOTICE
The information contained in this document is subject to
change without notice.
SIGNAL
WORD
MEANING
DANGER
Danger indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
WARNING
Warning indicates a potentially hazardous situation which, if
not avoided, could result in death or serious injury.
CAUTION
Caution indicates a potentially hazardous situation which, if
not avoided, may result in property damage or minor injury,
or both.
IMPORTANT SAFEGUARDS
Invacare
®
Pediatric Bear 6 Part No 1160837
IMPORTANT SAFEGUARDS
WARNING
IMPORTANT SAFEGUARDS contains important
information for the safe operation and use of this product.
CAUTION
Federallawrestrictsthisdevicetosaleorrentalbyoronthe
orderofaphysician,oranyotherpractitioner licensedbythe
lawoftheStateinwhichhe/shepracticestouseororderthe
useofthisdevice.
“Thisappliancehasapolarizedplug(onebladeiswiderthan
theother).Asasafetyfeature,thisplugwillfitinapolarized
outletonlyoneway.Iftheplugdoesnotfitfullyintheoutlet,
reversetheplug.Ifitstilldoesnotfit,contactaqualified
electrician.Donotattempttodefeatthissafetyfeature.”
WARNING
Whenusingelectricalproducts,especiallywhenchildrenare
present,basicsafetyprecautionsshouldalwaysbefollowed,
includingthefollowing:
IMPORTANT SAFEGUARDS
Part No 1160837 7 Invacare
®
Pediatric Bear
READALLINSTRUCTIONSBEFOREUSING
DANGER
FIRERISK‐DONOTusethisproductinanoxygen
enrichedenvironment(i.eoxygentent).
Riskofelectricshock.DONOTdisassemble.Refer
servicingtoqualifiedservicepersonnel.
To Reduce the Risk of Electrocution:
ALWAYSunplugthisproductIMMEDIATELYafterusing.
DONOTusewhilebathing.
DONOTplaceorstoreproductwhereitcanfallorbe
pulledintoatuborsink.
DONOTplaceinordropintowaterorotherliquid.
DONOTreachforaproductthathasfallenintowater.
UnplugIMMEDIATELY.
IMPORTANT SAFEGUARDS
Invacare
®
Pediatric Bear 8 Part No 1160837
WARNING
To Reduce the Risk of Burns,
Electrocution, Fire and Injury to Persons:
ThisproductshouldNEVERbeleftunattendedwhen
pluggedin.
Closesupervisionisnecessarywhenthisproductisusedby,
on,ornearchildrenandphysicallychallengedindividuals.
Usethisproductonlyforitsintendeduseasdescribedinthis
manualandonlywiththeprescribedmedication.Usethis
productonlyunderphysician’sdirection.
NEVERoperatethisproductwithadamagedcordorplug,if
itisnotworkingproperly,ifitwasdroppedordamaged,orif
itwasdroppedintowater.Callqualifiedservicepersonnelfor
examinationandrepair.
IMPORTANT SAFEGUARDS
Part No 1160837 9 Invacare
®
Pediatric Bear
WARNING
Keepthenebulizertubing,cord,andunitawayfromheated
orhotsurfaces,includingspaceheaters,electricblankets,
stovesandsimilarelectricalappliances.
NEVERdroporinsertobjectsintoanyopeningorhose.
NEVERusewhilesleepingordrowsy.
NEVERblocktheairopeningsoftheproductorplaceitona
softsurface,suchasabedorcouch,wheretheairopenings
maybeblocked.Keeptheairopeningsfreefromlint,hairand
thelike.Blockedopeningsmaycausetheunittoshutdown.
Contactdealerimmediately.
Thecompressornebulizersystemusesthesurroundingairto
effectivelyadministertheprescribedmedicationthroughthe
nebulizer.Theuserandadministratorshouldbeawareof
surroundingssoasnottohavecontaminatedairfiltered
throughthecompressornebulizersystem.(Examplescanbe
aircontaminatedwithpaintorcleaningfumes,weedkiller,
carexhaust,etc.)
DONOTusethecompressornebulizersysteminthese
environments.Otherwise,seriousbodilyinjurymayoccur.
Thisunitisoilless.DONOTlubricate.
IMPORTANT SAFEGUARDS
Invacare
®
Pediatric Bear 10 Part No 1160837
WARNING
Thisequipmentisnotsuitableforuseinthepresenceofa
flammableanestheticmixturewithairorwithnitrousoxide.
Whennotadministeringtreatment,keepoutofthereachof
smallchildren.Someofthepartsofthenebulizeraresmall
enoughforchildrentoswallow.Thetubingmaypresenta
strangulationhazard.
NEVERtouchtheunitwithwethands.
DONOTpullonthepowercordtounplugtheunitfroman
electricaloutlet.
DONOToverloadwalloutlets.
MaintenanceMUSTbeperformedbyaqualifiedservice
technician.
SAVE THESE INSTRUCTIONS
INTRODUCTION
Part No 1160837 11 Invacare
®
Pediatric Bear
INTRODUCTION
TheInvacarePediatricBearNebulizerSystemisaportable
medicaldevicedesignedtoefficientlydeliverphysician
prescribedmedicationtothebronchiallungpassages.The
nebulizersystemconsistsofthefollowingcomponents:
Compressor
Nebulizer
•Airtubing
•Mouthpiece
PediatricMask
TYPICAL PRODUCT PARAMETERS
Invacare
®
Pediatric Bear 12 Part No 1160837
TYPICAL PRODUCT
PARAMETERS
Electrical Requirements: 120 VAC, 60Hz, Single Phase
Running Current: 1.4 A @ 12.4-15.4 psi @ 120 volts
Sound Level: 58 dBA
Power Wattage: 68 watts @ 18 psig
Electric Motor: Class B in compliance with UL2111.
Foam Intake Filter: Polyurethane Sponge
Safety Systems: High temperature compressor shutdown.
Width: 6.5 inches
Height: 8.2 inches
Depth: 5.5 inches
Weight: 3.4 Ibs
Shipping Weight: 3.8 lbs
Regulatory Listing: In compliance with UL 60601-1.
Typical Operating Pressure: 12.4 to 15.4 psi
Maximum Pressure: 30 to 35 psi
Operating Flow Rate: 4.0 L/min. @ 16 psi operating pressure
Maximum Flow Rate: 5 to 7 L/min.
TYPICAL PRODUCT PARAMETERS
Part No 1160837 13 Invacare
®
Pediatric Bear
NOTE:Productparametersandappearancearesubjecttochange
withoutnoticeforpurposesofimprovement.
NOTE:Nebulizationrateandparticlesizearedependentonthe
characteristicsofthenebulizerchosen.
Storage Conditions: -13°F to 158°F (-25° to 70°C); 10 to 90%
RH
Operating Conditions: 50°F to 104°F (10°C to 40°C); 30 to 75%
RH
Altitude: 0 to 6000 ft (1825 m) above sea level
Mode of Operation: Intermittent Use (20 min. ON / 40 min.
OFF)
SUPPLEMENTAL INFORMATION
Invacare
®
Pediatric Bear 14 Part No 1160837
SUPPLEMENTAL
INFORMATION
Compressor
Iftheunitshutsdownfrequently,youmayhavealowvoltage
situation.
Lowvoltagecanalsobesuspectedwhen:
•Themotordoesnotgetuptofullpowerorspeed.
•Homefusesorcircuitbreakersactivatewhenstarting
compressor.
•Lightsdimorremaindimwhencompressorisstarted.
•Othermotoroperatedappliancesfailtooperateproperly
ortoomanymotoroperatedappliancesareonsame
circuit.
Extension Cords
Ifyoumustuseanextensioncord,useonlyacordhavingthe
sameorhigherelectricalratingasthedevicebeingconnected.
Makecertaintheextensioncordisingoodconditionwithout
anyexposedwiresorhazardouscontacts,andtheextension
cordwireisNOTLESSthan18gaugefor25feet.
SECTION 1—HANDLING
Part No 1160837 15 Invacare
®
Pediatric Bear
SECTION 1—HANDLING
Shipping Instructions
Theshippingcontainerhasbeendesignedtoassure
protectionofthisdevice.
Ifthisdeviceistobereshippedbycommoncarrier,itshould
bepackedinthesamecarton.
Unpacking
1. Checkforanyobviousdamagetothecartonorits
contents.Ifdamageisevident,DONOTuse.Notify
carrier/InvacareoryourlocalInvacare dealer.
2. Removealltheloosepackingfromthecarton.
3. Carefullyremoveallthecomponentsfromthecarton.
NOTE:Retaincartonandpackingmaterialsforuseinshippingor
storingtheunituntiluseisrequired.
Inspection
1. Examineexteriorfordamage.Inspectallcomponents.
SECTION 2—OPERATING INSTRUCTIONS
Invacare
®
Pediatric Bear 16 Part No 1160837
SECTION 2—OPERATING
INSTRUCTIONS
WARNING
DONOToperateunitwithoutfilter.
Avoidoperatingthisproductinadustyenvironment.
Otherwise,prematurecompressorwearcouldoccur.
NOTE:Cleananddisinfectthenebulizer,mouthpieceandmask
beforeusingthedeviceforthefirsttime.RefertoCleaningand
Disinfectingonpage 21.
Setup
NOTE:Forthisprocedure,refertoFIGURE2.1onpage 17.
1. Placetheunitonatableorotherflat,stablesurface.Besure
youcaneasilyreachthecontrolswhenyouareseated.
CAUTION
Makesurethepowerswitchisintheoff“O”position.
2. Gentlytwistthenebulizercounterclockwisetoseparateit
intotwosections.
SECTION 2—OPERATING INSTRUCTIONS
Part No 1160837 17 Invacare
®
Pediatric Bear
3. Besurethemedicationpickupinsertisinplaceinthe
bottomsectionofthenebulizer.
CAUTION
DONOTaddover6mlofmedicationtothemedication
cup.Themaximumcapacityofthenebulizermedicationcup
is6ml.
4. Addtheprescribedamountofmedicationtothecupofthe
nebulizerusinganeyedropper,premeasureddoseor
ampoule.
FIGURE 2.1 Adding Medication
5. Gentlytwistthetopandbottomsectionstogether
clockwisetoassemblethenebulizer.Besurethetwo
sectionsfitsecurely.
6. Attachoneendoftheairtubingtotheairtubingconnector
locatedonthebackofthecompressor.
Eyedropper
Nebulizer Cup
SECTION 2—OPERATING INSTRUCTIONS
Invacare
®
Pediatric Bear 18 Part No 1160837
7. Attachtheotherendoftheairtubingtothebaseofthe
nebulizer.
NOTE:Ensurenebulizertubingisfirmlyattachedtocompressor
beforeuse.
WARNING
Inthecaseofhighhumidity,condensationmayforminside
thetubing.Runtheunitfortwominutesbeforeconnecting
thenebulizertothetubing.
8. Attachthemouthpiecetothetopsectionofthenebulizer.
Refertothenebulizermanualforadditionalinstructions.
NOTE:Beforestartingacoldunit,lettheunitreachroom
temperatureinaheatedarea.
9. Withthepowerswitchoff,plugthepowercordintoan
appropriateelectricaloutlet.
CAUTION
Thisdeviceisforintermittentuse:20min.ON/40min.OFF.
Failuretooperatewithintheselimitscoulddamagethe
devicebeyondrepair.
SECTION 2—OPERATING INSTRUCTIONS
Part No 1160837 19 Invacare
®
Pediatric Bear
Treatment Instructions for Nebulizer
NOTE:Forthisprocedure,refertoFIGURE2.2.
1. Turntheuniton,pressing
theON/OFFswitchtothe
“I”position.
2. Placemouthpiece
betweenyourteeth,close
yourmouthandtiltthe
nebulizertow ardyour
mouth.
FIGURE 2.2 Using the
Nebulizer
3. Asaerosolbeginstoflow,inhaledeeply,slowly,andhold
breathfortenseconds.Thenexhaleslowlythroughthe
mouthpiece.
NOTE:Ifyouneedtorestduringatreatment,besurethenebulizer
remainsinanuprightposition;otherwise,themedicationmayspill.
4. Whentreatmentiscomplete,turntheunitoffbysetting
theswitchtothe“O”positionandunplugtheunitfrom
theelectricaloutlet.
5. Beforestoringsuppliesbetweentreatments,cleananddry
thenebulizerandaccessories.RefertoCleaningand
Disinfectingonpage 21.
SECTION 2—OPERATING INSTRUCTIONS
Invacare
®
Pediatric Bear 20 Part No 1160837
Treatment Instructions for Mask
1. Attachnebulizercuptomask.
NOTE:Forthisprocedure,refertoFIGURE2.3.
2. Placemaskonface.
3. Turntheuniton,pressing
theON/OFFswitchtothe
“I”position.
4. AsAerosolbeginsto
flow,inhaledeeply,
slowly,andholdbreath
fortenseconds.Then
exhaleslowlythrough
themask.
FIGURE 2.3 Using the Mask
NOTE:Ifyouneedtorestduringatreatment,besurethemaskand
nebulizerremaininanuprightposition;otherwise,themedication
mayspill.
5. Whentreatmentiscomplete,turntheunitoffbysetting
theswitchtothe“O”positionandunplugtheunitfrom
theelectricaloutlet.
6. Beforestoringsuppliesbetweentreatments,cleananddry
themaskandaccessories.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32

Invacare Respiratory Product IRC 1740 User manual

Category
Air compressors
Type
User manual

Invacare Respiratory Product IRC 1740 is a portable medical device designed to efficiently deliver physician-prescribed medication to the bronchial lung passages. It consists of a compressor, nebulizer, air tubing, mouthpiece, and pediatric mask. With a typical operating pressure of 12.4 to 15.4 psi and a maximum pressure of 30 to 35 psi, it operates at a flow rate of 4.0 L/min. at 16 psi operating pressure. Its compact dimensions (6.5 inches wide, 8.2 inches high, and 5.5 inches deep) and lightweight (3.4 lbs) make it easy to use and store.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI