  • Hello! I've reviewed the instruction manual for the Hanna Instruments HI98203 Salinity Meter. This document covers the device's features, calibration procedure, and maintenance guidelines, including battery replacement and junction renewal. It also explains how to use the chart to convert the reading to g/L of NaCl. I'm ready to help you with any questions you might have about this device.
  • How often should I calibrate the meter?
    What should I do if the readings are erroneous?
    How many times can the cloth junction be pulled out?
    What type of batteries does the meter use?
    What is the warranty period for the HI98203?
SALINTEST
HI98203
Salinity Concentration
Pocket-sized Meter
INSTRUCTION MANUAL
Care & Maintenance
To improve accuracy, immerse the meter in HI7081
solution for a few minutes at least once a week.
Erroneous readings after accurate calibrations may
indicate a contaminated or clogged junction. Pull
out 2 mm (1/8”) of the cloth junction to renew the
electrode reference. Cut the cloth leaving at least
2 mm (1/8”) over the reference compartment and
recalibrate the meter.
The cloth junction can be pulled out approximately
12 times.
PULL OUT
Battery Replacement
When the display fades or the meter switches off, the
batteries need replacing.
To replace the batteries:
1. With the meter off, slide off the battery compartment
cover.
2. Remove the old batteries.
3. Insert 4 new 1.5V Alkaline batteries in the battery
compartment while paying attention to the correct
polarity.
4. Replace the battery compartment cover.
SLIDE
OFF
Note: Do not mix old and new Alkaline batteries. Only
use specified battery type. Dispose of old batteries in
accordance with local regulations.
Accessories
HI7081L 30 g/L NaCl standard solution, 500 mL
HI7081M 30 g/L NaCl standard solution, 250 mL
HI7061M General purpose cleaning solution, 230 mL
HI731326 Calibration screwdriver (20 pcs.)
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895, USA.
IST98203 08/21
Certification
All Hanna Instruments conform to the CE European
Directives.
Disposal of Electrical & Electronic Equipment. The
product should not be treated as household waste. Instead,
hand it over to the appropriate collection point for the
recycling of electrical and electronic equipment, which will
conserve natural resources.
Disposal of waste batteries. This product contains
batteries, do not dispose of them with other household
waste. Hand it over to the appropriate collection point for
recycling.
Ensuring proper product and battery disposal prevents
potential negative consequences for the environment and
human health. For more information, contact your city,
your local household waste disposal service, or the place
of purchase.
Recommendations for Users
Before using this meter, make sure that it is entirely suitable
for your specific application and for the environment
in which it is used. Avoid touching the electrode at all
times. Any variation introduced by the user to the supplied
equipment may degrade the meter’s performance. For your
and the meter’s safety do not use or store the meter in
hazardous environments.
Warranty
HI98203 is warranted for a period of one year against
defects in workmanship and materials when used for its
intended purpose and maintained according to instructions.
This warranty is limited to repair or replacement free of
charge. Damage due to accidents, misuse, tampering, or
lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments office.
If under warranty, report the model number, date of
purchase, serial number and the nature of the problem.
If the repair is not covered by the warranty, you will be
notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments
office, first obtain a Returned Goods Authorization (RGA)
number from the Technical Service department and then
send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete
protection.
Hanna Instruments reserves the right to modify the
design, construction, or appearance of its products
without advance notice.
Preliminary Examination
Remove the meter and accessories from the packing
material and examine it carefully. If you require any
further information, please contact Hanna Instruments
technical support team at [email protected].
Each HI98203 is supplied with:
• Protective cap
• Calibration screwdriver
• 1.5V Alkaline battery (4 pcs.)
• Quality certificate and Instruction manual
Note: Save all packing material until you are sure that
the meter works correctly. Any damaged or defective
item must be returned in its original packing material
with the supplied accessories.
General Description & Intended Use
HI98203 is designed to help users monitor the
concentration of sodium chloride in aquaculture
applications and salt concentration in marine fish tanks
and aquariums.
The meter has a one-point calibration through a
trimmer on the side and is equipped with a sodium
ion-selective glass electrode that determines the activity
of sodium ions in solution. When dissolved, NaCl ionizes
to form Na+ and Cl ions. Measurement of Na+ ion
concentration is an indicator of the concentration of NaCl.
The double junction reference ensures highly-stable
readings.
Specifications
Range 0.00 to 1.00 pNaCl
(58.4 to 5.84 g/L NaCl)
Resolution 0.01 pNaCl
Accuracy ±0.02 pNaCl (@ 20 °C / 68 °F)
Calibration Manual, one point
Environment 0 to 50 °C (32 to 122 °F)
RH max. 95%
Battery type 1.5V Alkaline
Battery life Approximately 500 hours of continuous use
Dimensions 175 x 41 x 23 mm (6.9 x 1.6 x 0.9”)
Weight 95 g (3.4 oz.)
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before
using the meter. For more information about Hanna
Instruments and our products, visit www.hannainst.com
or e-mail us at [email protected].
For technical support, contact your local Hanna
Instruments office or e-mail us at [email protected].
Calibration
1. Remove the protective cap and slide the ON/OFF
switch to turn the meter on.
2. Immerse the meter in HI7081 30 g/L NaCl calibration
solution, paying attention not to exceed maximum
immersion level. Allow the reading to stabilize.
3. When the reading is stable,
use the supplied screwdriver
to adjust the calibration
trimmer on the side until the
display shows “.30”.
Calibration is complete.
To avoid calibration drift,
refer to Care & Maintenance
section. A solution of
known concentration can
also be used to recalibrate
the meter.
MAX.
LEVEL
MIN.
LEVEL
CALIB.
TRIMMER
ON/OFF
SWITCH
Measurement
1. Remove the protective cap.
2. Slide the ON/OFF switch to turn the meter on.
3. Immerse the meter into the solution to be tested, paying
attention not to exceed maximum immersion level.
4. Stir gently and wait for the reading to stabilize.
5. Use the chart to convert read value to g/L of NaCl.
6. After use, switch the meter off, rinse the electrode
with water, and replace the protective cap.
Note: For accurate readings keep the electrode
hydrated and calibrate frequently.
How to Use the Chart
The logarithmic chart below converts readings into g/L
of sodium chloride.
1. Locate the reading on the horizontal axis.
2. Move vertically upwards to intersect with the 45° line.
3. Move horizontally and read the g/L of NaCl value.
E.g. Read value = .42
converted g/L NaCl value = 22
1 Division = 0.02
g/L NaCl
SALINTEST Reading
/