Omron HCG-801 User manual

Type
User manual

This manual is also suitable for

Portable ECG Monitor
Model HCG-801
INSTRUCTION MANUAL
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
ESPAÑOL
ENGLISH
2
Before Using the Unit
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General Information About the Heart and ECG Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Important Safety Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Know Your Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Operating Instructions
How to Insert and Replace Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Battery Life and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
How to Attach the Battery Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
How to Set the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
How to Use the SD Memory Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Inserting and Ejecting SD Memory Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Formatting an SD Memory Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Important Notes on Taking a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
How to Take a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Common Causes of Inaccurate Measurement Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
How to Display ECG Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
How to Adjust the Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Basic Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Date Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Overwrite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
About the Overwrite Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
TABLE OF CONTENTS
How to Erase, Save and Organize ECG Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Erasing ECG Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Copying Data to an SD Memory Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Copying Data from the SD Memory Card to a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Copying Data from a PC to the SD Memory Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Data Description Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Criteria of Data Description Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
Heart Rate Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Action Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Lists and Explanations of Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
Notice Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Care and Maintenance
Trouble Shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Maintenance and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
FCC Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Warranty Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Sample Symptom Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
TABLE OF CONTENTS
3
4
Thank you for purchasing the OMRON
®
Portable ECG Monitor.
INTRODUCTION
Fill in for future reference.
DATE PURCHASED:________________
SERIAL NUMBER: _________________
Staple your purchase receipt here
Your OMRON HCG-801 is an easy-to-use electrocardiograph unit that can record and store
electrocardiogram (ECG) measurements of your heart rhythm. Each ECG reading records an approximately
30-second measurement and these ECG readings can help your doctor monitor your condition. It also
includes guidance screens to help you obtain good data when taking measurements. This unit is intended for
use on the adult population.
The monitor is supplied with an SD memory card that can store up to 300 measurements, including the ECG
data along with the date and time of measurement. Storing ECG data on the SD memory card, or in unit
memory, allows your doctor to view them later.
IMPORTANT: Please note that while the measurements from an ECG monitor are a useful guide for your
doctor, they cannot detect all changes in heart conditions. Always consult your doctor if you
notice any change in your condition, regardless of the measurement results. Consult a
doctor even if “Regular heart rate, Regular heart rhythm” is displayed as the data message,
if you have been diagnosed with a heart condition.
The HCG-801 comes with the following components:
• Main Unit • SD Memory Card • Storage Pouch • 2 × LR03(AAA) Batteries
• Instruction Manual* • Registration Card • Quick Reference Card
* The illustrations used in this manual may differ slightly from the appearance of the actual unit.
SAVE THESE INSTRUCTIONS
Before Using the Unit
The heart is a muscular pump controlled by electrical impulses generated by the body. It is divided into two
sides by the septa and each side has two chambers – an atrium and a ventricle – linked by a one-way valve.
The left atrium and ventricle control oxygenated blood, and the right atrium and ventricle de-oxygenated
(“used”) blood.
The electrical impulse that causes the heart to beat spreads across the atria, causing the left and right atria to
contract and pump blood into the left and right ventricles respectively. The two ventricles then contract and
pump blood out of the heart. The heart muscle then relaxes, or re-polarizes, allowing blood to fill up the
heart again.
An ECG monitor is able to measure the electrical impulse as it passes across and through the heart, causing
the heart to beat. An ECG does not measure the movement of your heartbeat, but rather the electrical
activity that causes the heart to beat. The measurements recorded by an ECG monitor, when combined with
a medical examination, can help your doctor monitor your heart condition. The ECG measurements
recorded by the unit are NOT designed or intended for medical diagnosis. Conditions such as arrhythmia
can only be diagnosed by a doctor through a special examination.
GENERAL INFORMATION ABOUT THE HEART AND ECG MEASUREMENTS
5
Left atrium
Left ventricle
Septa
Right atrium
Right ventricle
6
IMPORTANT SAFETY NOTES
To assure the correct use of the product basic safety measures should always be followed including the
warnings and cautions listed in this instruction manual.
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
Indicates a potentially hazardous situation which, if not avoided, could result
in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment
or other property.
WARNING
CAUTION
THIS UNIT IS DESIGNED FOR RECORDING THE ECG WAVEFORM IN THE ADULT
POPULATION. IT IS NOT DESIGNED OR INTENDED FOR MEDICAL DIAGNOSIS.
NEVER USE THE UNIT IN CRITICAL CARE SITUATIONS.
Messages after measurement report heart rate and heart rhythm only.
This device neither detects nor measures all heart rate, heart rhythm and heart waveform changes,
especially those related to ischemic heart conditions.
Always consult your doctor if you notice change in your condition, regardless of measurement results
and messages.
WARNING:
Do not use this device if you have a cardiac pacemaker or other implanted devices.
Do not use this device with a defibrillator.
Do not use in the presence of flammable anesthetics, drugs or pressurized oxygen (such as in a hyperbaric
chamber, ultraviolet sterilizer or oxygen tent).
Do not attempt self-diagnosis or self-treatment based on the measurement results and analysis.
Always consult your doctor. Self-diagnosis or self-treatment may lead to deterioration in your condition.
Do not use this device during an MRI scan.
WARNING:
Do not use this device on people with sensitive skin or allergies.
Do not attempt to use the device over or through clothing.
Do not attempt to use the device with your left hand.
When you experience symptoms potentially related to heart disease always consult your physician,
even if “Regular heart rate, Regular heart rhythm” is displayed.
If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult a doctor
immediately.
CAUTION:
Do not measure if dust is on the electrodes. Clean them first.
If the portion of the body where the electrode is applied has too much body fat, a successful reading
may not be possible.
Do not measure with the electrode upside down (Refer to the illustration on page 26).
Do not expose the unit to static electricity. Always disperse any static electricity from your body
before handling the unit.
Do not take measurements in a moving vehicle.
Do not use a cellular phone near the unit.
Do not disassemble, repair, or modify the unit.
Do not insert batteries with their polarities reversed.
The chest electrode will be impeded by excessive body hair. Regularly remove hair at the electrode
contact point. Omron does not recommend the use of contact gel.
Keep out of reach of infants, small children, and compromised people who cannot express their consent.
IMPORTANT SAFETY NOTES
7
CAUTION:
Do not use for any purpose other than obtaining an electrocardiograph.
Do not expose the unit to strong shocks or vibrations. Do not drop or step on the unit.
Do not use batteries of a type other than that specified in this manual.
Do not take measurements when there are drops of water on your skin, such as from sweat or after bathing.
Do not take measurements where the unit will be exposed to strong electromagnetic forces.
If the batteries are removed for more than one hour, the settings and any ECG data stored in the unit
may be erased. If the date and time setting is erased, be sure to reset it before using the unit again.
Any ECG data erased from the unit cannot be recovered.
Avoid extremes in temperature and humidity. Do not use this unit in locations subject to high or low
temperatures or humidity. Use at a temperature within +50 to +104°F (+10 to +40°C) and between 30
and 85%RH.
Do not sterilize this unit in an autoclave or gas sterilizer (EOG, formaldehyde, high density ozone, etc.)
Do not wash this unit with water.
Do not store the unit in the following ambient conditions:
• Locations exposed to direct sunlight.
• Locations subject to high temperatures and high humidity.
• Wet or damp locations where water may get on the unit.
• Dusty locations.
• Near fires or open flames.
• Locations exposed to strong vibration.
• Locations exposed to strong electromagnetic fields.
IMPORTANT SAFETY NOTES
8
Chest electrode
Place this against your
chest when taking a
measurement.
Display
Displays the wave pattern and
displays menus, messages and
display options.
MENU button
Displays or closes the menu screen.
button
Moves the desired menu or wave
pattern to the left or up and
decreases the number of the
selected field.
button
Moves the desired menu or wave
pattern to the right or down and
increases the number of the
selected field.
SD Memory card slot cover
Open this cover to insert the SD
memory card.
ON/OFF / STOP button
Turns the power on/off and
cancels measurement.
START button
Starts measurement.
Finger electrodes
Place your right index finger over
these when taking a measurement.
Battery cover
Open this cover to insert batteries.
SELECT button
Selects the highlighted menu
option or ECG data.
9
KNOW YOUR UNIT
MAIN UNIT:
MAIN UNIT
Display:
The screen is used to display basic instructions and results of ECG readings, to adjust settings, and to
review ECG data stored on the SD memory card or in unit memory.
Menus
KNOW YOUR UNIT
10
Menu Screen
The main menu screen is
displayed when you press
the MENU button.
SET menu
The SET menu is displayed
when you select the SET
menu and press the
SELECT button.
GRAPH menu
This allows you to view ECG readings stored on
the SD memory card or in unit memory. (page 30)
SET menu
This allows you to set the date and time, and
adjust other settings. (page 32)
GUIDE
This displays the guidance screens that show
you how to take a measurement.
Back
This option takes you back to the previous
screen, from the SET menu.
Close
This option closes the menu screen.
MAIN UNIT
Display:
ECG Graph
This screen is displayed when you use the GRAPH menu to select ECG data stored on the unit.
KNOW YOUR UNIT
11
NOTE: During measurement, the screen displays the ECG waveform. The progress bar at the bottom of the
screen displays measurement progress, and the number on the right shows the heart rate.
Depending on the waveform, situations can occur in which the unit cannot assess and display the
heart rate during the measurement.
Calibration guide
Serves as a guide when your
doctor analyzes the stored ECG
data.
Approximately 30-second
waveform
Measurement time indicator
SD memory card indicator
Displayed when the SD memory card is
in the card slot.
Battery level indicator
ECG waveform
Heart rate
(*bpm=beats per minute)
KNOW YOUR UNIT
12
COMPONENTS:
• SD memory card
(801SDMEMORY)
• Storage pouch
(801STORAGEPOUCH)
• Instruction Manual
(HCG801INST)
• 2 LR03(AAA)
Batteries
• Quick Reference Card
• Registration Card
1. Slide the battery cover in the direction of the arrow, and lift the end of
the cover upwards.
NOTE: Do not pull hard on the cover.
2. Insert two LR03(AAA) batteries so the + (positive)
and the - (negative) polarities match the polarities on the
battery compartment as illustrated.
NOTE: Always use LR03(AAA) batteries.
3. Close the battery cover.
Lower the battery cover (1), then while gently pressing the cover
down, slide it in the direction of the arrow (2).
HOW TO INSERT AND REPLACE BATTERIES
13
1
2
Operating Instructions
BATTERY LIFE AND REPLACEMENT
When the “Please insert new batteries.” message appears, replace both batteries with new ones.
Always turn the unit off before replacing the batteries.
Dispose of the used batteries according to applicable local regulations.
Battery life
Two new LR03(AAA) batteries will last for approximately 400 measurements, if measurements are taken
once a day at room temperature (72ºF (22ºC)).
The batteries enclosed in the package are intended for the purpose of demonstration. It is possible these
batteries will not last for 400 measurements.
The battery life can be confirmed in the upper right of the display.
HOW TO INSERT AND REPLACE BATTERIES
14
WARNING: If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
CAUTION:
If the batteries are removed for more than one hour, the settings and any ECG data stored
in the unit may be erased. If the date and time setting is erased, be sure to reset it before
using the unit again. Any ECG data erased from the unit cannot be recovered.
Do not use batteries not specified for this unit.
Do not insert the batteries with the polarities in the wrong direction.
Do not dispose of batteries in fire.
If battery fluid should get on your skin or clothing, immediately rinse with plenty
of clean water.
Remove the batteries from this unit when you are not going to use it for a long period of
time (approximately three months or more).
Do not use batteries of different types together.
Do not use new and used batteries together.
Full power. Half remaining.
Remaining power is low.
Replace the batteries with new ones.
If the battery cover is accidentally detached,
attach the battery cover as described in the
procedure below.
1. Align the pin of the battery cover with the groove
at the top of battery compartment on the START
button side (front) of the unit (1).
2. Align the other pin of the battery cover with the
groove on the other side of the battery
compartment (2).
While gently pushing the pin already inserted in
the START button side (front), insert the pin
in the groove on the opposite side of the battery
compartment.
NOTE: Do not push the pin in the START
button side of the battery compartment
hard. This could result in damage to the
battery cover.
3. Close the battery cover.
While pressing the battery cover, slide it in the direction of the arrow.
HOW TO ATTACH THE BATTERY COVER
15
2
1
Align pin with
groove
Align pin with
groove
Battery cover
START button side
Always set the date and time before using the unit for the first time.
Check that date and time are correct before using the unit, reset them if necessary. The date and time are
important indicators of when a measurement was taken.
NOTE: The date and time format can be changed.
(Refer to “Date Format” on page 38)
1. Press the ON/OFF / STOP button to turn the unit on.
When the unit is used for the first time, or if the Date and Time setting
has been reset, the Date and Time setting screen is automatically
displayed, and the month digits are highlighted.
2. Press the button or button to set the month.
Press the button to increase the number.
Press the button to decrease the number.
HOW TO SET THE DATE AND TIME
16
2006
3. Press the SELECT button to confirm the month setting.
The month setting is set and the day setting is highlighted.
The display cycles through the settings,
month -> day -> year -> hour -> minute -> AM/PM, with each
press of the SELECT button.
NOTE: If the setting for the date format has been changed ( refer to
“Date Format” on page 38), the order in which items are
highlighted may differ from the example above.
4. In the same way, set the day, year, hour, minute, and AM/PM.
After you have confirmed the last item on the date and time setting, press
the SELECT button to complete the date and time setting.
NOTE: If you need to set the date and time again, refer to page 34.
5. Press the ON/OFF / STOP button for two seconds to turn off the
power.
The unit beeps, then the power is turned off.
NOTE: If you forget to turn off the power, the unit will turn itself off
automatically after five minutes.
HOW TO SET THE DATE AND TIME
17
2006
Press for two seconds.
The unit includes an SD memory card for storing ECG data, which can then be viewed by your doctor later.
This section describes how to insert and eject an SD memory card, and how to format a card so it can be
used to store ECG data.
NOTES:
If you need to purchase a new SD memory card, contact Omron Healthcare Customer
Service 1-800-829-6427.
Do not store any data on the card from other devices rather than this unit.
ECG data stored on an SD memory card can be damaged by static electricity and other causes. Copy data
to PC or other device for additional safety.
INSERTING AND EJECTING SD MEMORY CARD
The SD ( ) mark is displayed in the top right corner of the screen while the SD memory card is inserted,
and the unit is turned on.
Inserting the SD memory card
1. Check that the unit is turned off.
CAUTION: Never insert an SD memory card while the unit is turned on.
HOW TO USE THE SD MEMORY CARD
18
2. Open the SD memory card slot cover and insert the SD memory card
into the slot.
NOTES:
• Check that the card is correctly aligned and insert it into the slot as shown.
• Insert the card until it clicks into position.
Write Protection (Locking the SD Memory Card)
• Make sure that the write-protect switch is not locked, before inserting the SD
memory card. If the write-protect switch is locked (down), you will not be able to
store new ECG data on the SD memory card.
CAUTION:
CAUTION: When inserting the SD memory card, do not try to force the
card into the slot if it is upside down, back to front, etc. This could damage the
card or the unit.
HOW TO USE THE SD MEMORY CARD
19
Back to front
Upside down Back to front and
upside down
CORRECT ALIGNMENT
3. Close the SD memory card slot cover.
HOW TO USE THE SD MEMORY CARD
20
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128

Omron HCG-801 User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document