Roche cobas infinity central lab User manual

Category
Software
Type
User manual
cobas
®
infinity central lab
User Guide - Microbiology module
Version 1.3
Software version 2.0
cobas
®
infinity microbiology module
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
2
Publication information
Edition notice
This publication is intended for operators of the cobas
®
infinity central lab.
Every effort has been made to ensure that all the
information contained in this publication is correct at the
time of publishing. However, the manufacturer of this
product may need to update the publication information
as output of product surveillance activities, leading to a
new version of this publication.
Where to find information The User Assistance contains all information about the
product, including the following:
• Safety
• Routine operation
• Configuration information
The User Guide focuses on routine operation. The
chapters are organized according to the normal operation
workflow.
!
General attention
To avoid incorrect results, ensure that you are familiar
with the instructions and safety information.
r Pay particular attention to all safety notices.
r Always follow the instructions in this publication.
r Do not use the software in a way that is not described
in this publication.
r Store all publications in a safe and easily retrievable
place.
Training Do not carry out operation tasks or maintenance actions
unless you have received training from Roche
Diagnostics. Leave maintenance, installation, or service
that is not described in the manuals to trained Roche
Service personnel.
Screenshots The screenshots in this publication have been added
exclusively for illustration purposes. Configurable and
variable data, such as tests, results, or path names visible
therein must not be used for laboratory purposes.
Publication
version
Software version Revision date Change description
1.0 2.0 June 2016 First edition
1.1 2.0 June 2016 New features of software version 2.0
1.2 2.0 June 2016 Minor updates
1.3 2.0 October 2016 New features of software version 2.0 and new chapters added
y Revision history of User Guide
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
3
Warranty Any customer modification to the system renders the
warranty or service agreement null and void.
For conditions of warranty, contact your local sales
representative or refer to your warranty contract partner.
Always leave software updates to a Roche Service
representative, or perform such updates with their
assistance.
Copyright © 2013-2016, F. Hoffmann-La Roche Ltd. All rights
reserved.
License information
cobas
®
infinity central lab software is protected by
contract law, copyright law, and international treaties.
cobas
®
infinity central lab contains a user license
between F. Hoffmann-La Roche Ltd. and a license holder,
and only authorized users may access the software and
use it. Unauthorized use and distribution may result in
civil and criminal penalties.
Open Source and Commercial Software
cobas
®
infinity central lab may include components or
modules of commercial or open-source software. For
further information on the intellectual property and other
warnings, as well as licenses pertaining to the software
programs included in cobas
®
infinity central lab, refer to
the electronic distribution included with this product.
This open source and commercial software and cobas
®
infinity central lab as a whole can constitute a device
regulated in accordance with applicable law. For more
detailed information, refer to the user manual and
labeling.
Please note that the respective authorization is no longer
valid according to the corresponding legislation should
any unauthorized changes be made to cobas
®
infinity
central lab.
Trademarks The following trademarks are acknowledged:
COBAS and LIFE NEEDS ANSWERS are trademarks of
Roche.
All other trademarks are the property of their respective
owners.
Feedback Every effort has been made to ensure that this publication
fulfills the intended use. All feedback on any aspect of
this publication is welcome and is considered during
updates. Contact your Roche representative, should you
have any such feedback.
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
4
Contact addresses
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Made in Spain
Distributed in USA by:
Roche Diagnostics
9115 Hague Road
Indianapolis
Indiana, USA
Table of contents 5
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
Table of contents
Publication information 2
Contact addresses 4
Table of contents 5
Intended use 7
Use environment and intended users 8
Symbols and abbreviations 8
Safety classifications 10
System safety information 11
21 CFR Part 11 compliance 16
Interface overview
1 Interface overview
Accessing the software 23
About the software framework 26
Performing basic tasks in the software 28
Filtering, sorting and exporting table
information 36
About the common interface items 40
Routine tasks
2 General tasks
Viewing and assigning comments 47
Printing barcode labels manually 51
Patient management 57
Checking connections 61
Searching for order and patient information 63
Releasing tests or orders 66
3 Pre-analysis
Order and pre-order management 69
Test management 80
Handling samples 82
4 Sample analysis
Printing a worklist 87
Monitoring Microbiology orders and
samples 89
5 Result entry and validation
Entering and editing results manually 95
Entering results manually in the
Microbiology module 97
Declaring results 110
Verifying and validating results in the
Microbiology module 114
List of result entry and validation items in
the Microbiology module 117
6 Post-analysis
Reports 121
Appendix
Index 129
6 Table of contents
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
7
Intended use
The cobas® infinity IT solutions application is intended to
be used for:
• the configuration and connectivity management of
instruments and software systems
• the management of data regarding
- Samples
- Technical validation including automatic release
- Quality Control (both qualitative and quantitative)
- Test results and their entry (offline workplaces)
• the management and storing of information, such as
- Samples Archiving Storage information
- Rule engine for technical validation
- Notifications from any part of the system
- Reagent and Calibrator management
- Turn Around Time management
- Production statistics
In addition to the above intended use, the cobas infinity IT
solutions application is intended for:
• the management of data regarding
- Order Data
- Patient Data
- Medical Validation support
- Result Consolidation and Reporting
- Billing support
• the management and storing of information, such as
- General statistics (Data Warehouse)
- Microbiology workflows and data for:
(Microbiology module)
- human samples
• Microbiology workflows and data for: (Microbiology
module)
- human samples
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
8
Use environment and intended users
The cobas
®
infinity central lab software is intended for
Clinical Laboratories.
You can find different user profiles using the cobas
®
infinity central lab software.
Field Service Engineer (global and local) Field Service Engineers configure the system and master
data (test, test groups, senders, instruments, reports,
interfaces, etc.) according to specific requirements of the
customers with regards to connected hosts, instruments
and sample workflow.
Access level depends on user rights.
Lab Technician Lab technicians utilize the solution to perform technical
validation on patient and QC results, manually edit orders
(add or remove tests, edit patient-demographic data),
enter patient and QC results that have been obtained
from off-line workplaces, print-out patient and QC
reports, archive samples, retrieve samples from the
archive, etc.
Access level depends on specific user rights.
Lab IT Administrator Lab IT Administrators maintain users and authorization,
and maintain tests and senders.
Access level depends on specific user rights.
GP/Hospital doctor/Hospital
nurse/Community nurse
These users are not employers of the laboratory, thus
their access is restricted to the Lab Link module only.
Users enter orders manually within the Lab Link module,
release print-outs of barcode labels for the tubes for
positive patient/sample identification, release print-outs
of reports for the patient for blood drawing, view and print
the patient result report of their patients.
Phlebotomist These users are employers of the laboratory, but their
access is restricted to the Lab Link module only.
Phlebotomists confirm in the system that the samples
have been taken according to the order.
Symbols and abbreviations
Product names Except where the context clearly indicated otherwise, the
following product names and descriptors are used.
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
9
Symbols used in the publication
Abbreviations The following abbreviations are used.
Product name Descriptor
cobas
®
infinity central lab
software
cobas
®
infinity general lab
module
cobas
®
infinity emergency
lab
module
cobas
®
infinity lab flow
module
cobas
®
infinity lab link
module
cobas
®
infinity microbiology
module
cobas
®
infinity total quality
management
module
y Product names
Symbol Explanation
o List item
u Related topics containing further information
q
Tip. Extra information on correct use or useful
hints.
r Start of a task
I Extra information within a task
f Result of an action within a task.
c Frequency of a task.
n Duration of a task.
d Materials that are required for a task.
j Prerequisites of a task.
u Topic. Used in cross-references to topics.
p Task. Used in cross-references to tasks.
w
Figure. Used in figure titles and cross-
references to figures.
y
Table. Used in table titles and cross-references
to tables.
z
Equation. Used in cross-references to
equations.
k
Code example. Used in code titles and cross-
references to codes.
[ ] Square bracket. Used in the elements name as
defined by your system administrator
y Symbols used in the publication
Abbreviation Definition
ANSI American National Standards
Institute
EN European standard
n/a not applicable
UPS Uninterruptible Power Supply
y Abbreviations
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
10
Safety classifications
The safety precautions and important user notes are
classified according to the ANSI Z535.6 standard.
Familiarize yourself with the following meanings and
icons.
These symbols and signal words are used for specific
hazards:
WARNING
!
Warning...
r ...indicates a hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
!
Caution...
r ...indicates a hazardous situation which, if not avoided,
could result in minor or moderate injury.
NOTICE
Notice...
r ...indicates a hazardous situation that, if not avoided,
may result in damage to the system.
Important information that is not safety relevant is
indicated with the following icon:
q Tip...
...indicates additional information on correct use or useful
tips.
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
11
System safety information
Roche Diagnostics has established a series of
recommendations with the aim of allowing the user to
work with the software under safe conditions and
guaranteeing the correct operation and proper
performance of the communication network in which this
product has been installed.
q Read the following recommendations carefully for
the correct operation of the software.
Security
WARNING
!
Secure infrastructure
Risk of privacy violations.
r A secure infrastructure must be established for the
software network and security policies must be
defined to address potential problems or system
failures.
r Access is controlled by a logon. Every access to the
software is registered and logged, including the
unsuccessful ones.
r Pay special attention when configuring the number of
logon attempts, the password expiry period, and
session management.
WARNING
!
Enabled security parameters
r All security parameters (profiles, access restrictions,
etc.) are enabled by default. Remember that disabling
security parameters may lead to potentially major
risks, e.g. unqualified staff may change instruments or
system configurations, or unauthorized users may
access confidential information.
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
12
UPS
WARNING
!
UPS usage and software shutdown
Risk of data becoming lost or corrupted in the event of a
power failure
r It is strongly recommended to use a UPS and to
perform regular UPS maintenance.
r It is strongly recommended to keep the journaling
feature activated.
r There should be a standard procedure to shut down
the software. Users must exit the software before
shutting down the computer on a daily basis. Having a
UPS that can perform a safe shutdown installed
mitigates the risk of data loss due to power failure.
CAUTION
!
UPS usage
Risk of data loss in the event of a power failure
r It is strongly recommended to use a UPS.
r It is strongly recommended to keep the journaling
feature activated.
User settings
WARNING
!
Restricting access to critical functions
Risk of confidential information being accessed
r It is recommended to grant users the minimum rights
necessary to perform their tasks to protect
confidential data and configuration items in the
database.
WARNING
!
Access to the software and user accounts
r Access to the software should only be granted to user
who have been trained and assigned a user name and
password. The password must be confidential and
must comply with the usual security principles. It must
be changed periodically to prevent unauthorized
parties from gaining access.
r There must be one account per user. Do not create
generic user accounts to be used by more than one
person. Create a different account for each user, even
if they have the same access rights.
r It is recommended that you develop SOP to ensure
that only qualified staff access the software.
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
13
WARNING
!
Passwords, access, and system shutdown
r For security reasons and for patient confidentiality, all
users must exit the software or lock their computer
before leaving their workplace.
r It is recommended to correctly configure the
automatic log-off time.
CAUTION
!
System performance
Risk of delayed results
r To ensure the operation of the system at full capacity,
avoid connecting more users or instruments than your
infrastructure supports.
q System configurations
The preconfigured settings in the system must only be
used as a reference. It is recommended that they are not
used as default settings since they should be modified to
suit the system and the health center requirements.
General
CAUTION
!
Servers and software performance
Risk of delayed results
r Use servers that are correctly dimensioned to the
needs of your installation to guarantee the best
performance of the software.
r It is recommended to configure the servers using a
high availability and data redundancy system. The
existence of a second server prevents data loss in the
event of severe or critical failures in the system.
CAUTION
!
Alarms on validation screens and reports
Risk of incomplete results that could lead to a wrong
diagnosis
r To prevent users from validating results without taking
into account important information, always configure
alarms for validation screens and reports.
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
14
WARNING
!
SOP
Risk of results becoming confused
r It is recommended to establish a suitable standard
operating procedure (SOP) to prevent tubes from
being mishandled.
CAUTION
!
Barcodes
Risk of delayed results
r It is recommended to use barcodes for tubes and trays
to easily identify and locate samples.
q All the changes made in the software are recorded
using an audit trace.
Backups
CAUTION
!
Backup protocol
Risk of data loss
r It is strongly recommended to activate regular cache
backup and recovery features (preferably daily).
r It is strongly recommended to store (daily) cache
backup files in a safe place.
r It is strongly recommended to use a hardware
mirroring system
WARNING
!
Backups and updates
r Before upgrading the software, make backups of all
databases and system configurations.
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
15
Instruments and reading devices
WARNING
!
The software is not prepared to receive different
units for the same test and host, or from the same
test and different hosts. When receiving the
information, the software applies the test units that
have been configured.
Risk of incorrect results
r Be careful when configuring the test primary and
secondary units. If the units in the software do not
match the ones received from the external host or
instrument, ensure that the correct modification is
applied for every result received.
r If the units change from one message to another for
the same test, you must apply the necessary result
modification in the system to match the test units.
WARNING
!
Repeated values
Risk of incorrect results
r In the event that instruments do not show the time
when results were sent, there is the possibility that
results could be repeated due to data being
transmitted twice and not because a test was
repeated.
q Reading devices
Risk of results being misinterpreted
o It is recommended that you use reading devices
(optical readers, barcode readers, card readers, etc.)
to take results. Check the reading devices are
operating correctly before using them in real
environments.
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
16
21 CFR Part 11 compliance
The Food and Drug Administration (FDA) of the United
States under Title 21 of the US Code of Federal
Regulation Part 11 (21 CFR Part 11) establishes the
requirements to ensure that electronic records and
electronic signatures are trustworthy, reliable and
equivalent to paper records with traditional handwritten
signatures.
cobas
®
infinity central lab provides the capability to
meet the requirements of part 11 compliance if the
system is configured and used properly within the
customer's processes. This section describes the steps
and settings that are necessary to ensure that the
software is configured to meet the requirements of this
regulation.
The final responsibility of 21 CFR Part 11 compliance
relies on the adequate implementation from the customer
ensuring both procedural controls (i.e. Standard
Operating Procedures, training) and administrative
controls.
In case customers need to be compliant with 21CFR Part
11 regulatory requirements the software should be
configured according the customers policies and
procedures. Otherwise, the following parameters are
recommended:
General parameters Administration > General > General parameters
ID Parameter Description Recommended value
1 Accessing to the application via
Direct Access without having to
log on
When accessing the software using a link or
externally, this parameter allows you to access only
the screens marked as accessible without having
to enter the password and user name on the logon
screen.
No (default value)
2 Backup path (Folder) This parameter allows you to configure the
directory where the backups of the application
databases are stored.
Folder path configured
3 Session timeout (min) This parameter allows you to specify how long the
session is kept open (in minutes) since activity was
last detected. Once this time is exceeded, the
software logon screen appears, and the user name
and password must be entered again.
15
4 Direct access session timeout
(min)
This parameter only applies to sessions accessed
from an external environment. This parameter
allows you to specify how long the session is kept
open (in minutes) since activity was last detected.
Once this time is exceeded, the session closes.
15
5 Days configuration trace is kept Indicates the days the configuration trace will be
kept before deleting it.
9999
6 History retention period (d) Days to delete the order traces. 9999
y Parameter configuration for 21 CFR Part 11 compliance
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
17
Reports Administration > Reports > Definition
The following settings must be properly configured to
comply with this regulation:
7 Delete orders and patients after
the retention period (h) has
passed.
When deleting orders and patients, the application
sends them to the recycle bin. These orders /
patients remain in the recycle bin until they exceed
the number of hours set in this parameter. The
system then proceeds to permanently delete these
items from the database via the end-of-day
process.
24 (default value)
8 Number of logon attempts
before the user is blocked.
Indicates the amount of consecutive mistakes that
the user can do when logging in.
5
9 Password renewal period (d) Period of time when the system will request to
change the password for a user.
90 (default value)
10 Password expiry warning (d) Period of time when the system will start reminding
the user to change the password before the
password renewal period.
10
11 Minimum number of uppercase
letters in password.
Minimum number of uppercase letters that the
password must contain.
1 (default value)
12 Minimum number of digits or
symbols in password.
Minimum number of digits or symbols letters that
the password must contain.
1 (default value)
13 Minimum password length. Minimum length required for the password 6 (default value)
14 Request password when
accessing the Validation screen
(positive user identification) and
the Rule Engine Administration
screen.
Indicates if the password for the user will be
requested when entering into the validation screen
or Rule Engine validation screen.
Yes
15 Enable automatic data removal
in database.
Indicates if the system will remove order’s data
automatically on the database.
No (default value)
16 Enable administration trace. Indicates if configuration actions will be tracked or
not.
Yes (default value)
17 Add comment when editing
validated test (for all required
applications: General Lab,
Emergency Lab and
Microbiology)
When enabling this parameter, the application
forces the user who edits validated results and
wants to save them, to enter a comment. The call
out that appears for entering the comment cannot
be left empty. It is possible to enter a coded
comment.
These comments do not appear by default in the
report. If the user want to view them there, the
check box of the Imp./Host column on the
Comments screen should be selected.
Yes
18 Request confirmation when
changing results (for all
required applications: General
Lab and Emergency Lab)
If this parameter is enabled, the system requests
the user to confirm whether the test result change
should actually be saved in the database. This
option helps the user to avoid mistakes.
Yes
ID Parameter Description Recommended value
y Parameter configuration for 21 CFR Part 11 compliance
ID Configuration Procedure
19 Report configuration includes the user that validates the
result.
1. Choose the corresponding report.
2. Choose the Design button.
3. Include the Validation user name list field in the
report.
y Report configuration for 21 CFR Part 11 compliance
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
18
To ensure that these settings are properly configured,
contact your Roche Service Representative.
It is recommended that customers keep evidences of this
configuration according the required regulations.
ID Configuration
20 Check Backup List (Database configuration)
21 Check Backup task (Database configuration)
22 The communication channel uses SSL or HTTPS
between client and server
y Backup and communication configuration for 21 CFR Part
11 compliance
Roche Diagnostics
cobas infinity central lab · Software version 2.0 · User Guide - Microbiology module · Version 1.3
Interface overview
1 Interface overview .................................................................................................. 21
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130

Roche cobas infinity central lab User manual

Category
Software
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI