ADC MGA3000 series Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the ADC MGA3000 series Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
MGA3000
MULTI-GAS ANALYSER
OPERATION MANUAL
MGA3000 Multi-Gas Analyser Operation Manual Iss. 1.00
Copyright 2007, ADC Gas Analysis Ltd. 1
CONTENTS
Contents ............................................................................................................ 1
1. Introduction ................................................................................................... 1
1.1 About the Instrument .............................................................................................1
1.2 Instrument Identification.........................................................................................1
1.3 The Display & Operating Controls.............................................................................2
1.3.1 The Display ...................................................................................................2
1.3.2 The Operating Controls ...................................................................................2
1.4 Electrical Outputs ...................................................................................................2
1.4.1 Analogue Outputs ..........................................................................................2
1.5 RS232 port............................................................................................................2
1.6 Operating Principles................................................................................................3
1.6.1 Infrared Absorption ........................................................................................3
1.6.2 The Optical Bench In Use ................................................................................3
1.6.3 The Oxygen Sensor in Use ..............................................................................4
2. Installation ..................................................................................................... 5
2.1 Installation Arrangement.........................................................................................5
2.2 Location & Ambient Conditions.................................................................................5
2.3 Power Supply Setting & Connections ........................................................................6
2.3.1 Power Supply Setting......................................................................................6
2.3.2 Power Supply Connections ..............................................................................6
2.4 Serial I/O (RS232C) ...............................................................................................7
2.5 Analogue Outputs ..................................................................................................7
2.5.1 Setting the analogue outputs...........................................................................8
2.6 Alarm (Trip) Outputs ..............................................................................................9
2.7 Gas Arrangements ............................................................................................... 10
2.7.1 The Gas Circuit ............................................................................................ 10
2.7.2 The Sample Gas........................................................................................... 10
2.7.3 The Span Gas .............................................................................................. 11
2.7.4 The Zero Gas............................................................................................... 11
2.7.5 The Exhaust Gas .......................................................................................... 12
2.7.6 Gas Pump(s) ............................................................................................... 12
2.8 Use of Chemicals (certain models only)................................................................... 12
2.8.1 Water Vapour .............................................................................................. 12
2.8.2 Carbon Dioxide ............................................................................................ 13
2.8.3 Precautions ................................................................................................. 13
2.9 Pre-operation Checks............................................................................................ 13
2.10 General Advice................................................................................................... 13
2.10.1 Use of Gas Cylinders................................................................................... 13
2.10.2 Gas Leaks ................................................................................................. 14
2.10.3 Exhausting & Ventilation ............................................................................. 14
3. Setting Up .....................................................................................................15
3.1 Preview............................................................................................................... 15
3.2 On Power Up ....................................................................................................... 15
3.3 Zero & Span Operations........................................................................................ 15
MGA3000 Multi-Gas Analyser Operation Manual Iss. 1.00
Copyright 2007, ADC Gas Analysis Ltd. 2
3.4 Display & Keypad ................................................................................................. 16
3.4.1 Display ....................................................................................................... 16
3.4.2 Function Keys .............................................................................................. 17
3.4.3 Menus......................................................................................................... 17
3.5 Analogue (Chart) Outputs ..................................................................................... 18
3.6 Alarm (Trip) Outputs ............................................................................................ 18
3.6.1 Setting Alarm Outputs .................................................................................. 18
3.7 RS232C Serial Port............................................................................................... 19
3.7.1 Protocol ...................................................................................................... 19
3.7.2 Serial Port Settings....................................................................................... 19
3.8 Measurement ...................................................................................................... 20
4. Calibration.....................................................................................................21
4.1 Initial Checks....................................................................................................... 21
4.2 Warm Up ............................................................................................................ 21
4.3 Zero Adjustment.................................................................................................. 21
4.3.1 Manual Zero ................................................................................................ 22
4.3.2 Auto Zero.................................................................................................... 22
4.4 Span Adjustment ................................................................................................. 23
4.4.1 Setting Span ............................................................................................... 23
5. Maintenance ..................................................................................................24
5.1 Routine Checks .................................................................................................... 24
5.1.1 Water Vapour .............................................................................................. 24
5.1.2 Dust ........................................................................................................... 24
5.2 Instrument Checks............................................................................................... 25
5.2.1 Source Block Purge (when fitted) ................................................................... 25
5.2.2 Filter Column (when fitted)............................................................................ 25
5.2.3 Sample Flow................................................................................................ 25
5.2.4 Throttles ..................................................................................................... 26
6. Optical Bench Descriptions ..............................................................................27
6.1 Introduction ........................................................................................................ 27
6.2 GC Bench (when fitted)......................................................................................... 27
6.3 SB Bench ............................................................................................................ 28
7. Electrical Specifications ...................................................................................30
7.1 Power Supply ...................................................................................................... 30
7.2 Analogue Outputs ................................................................................................ 30
7.2.1 When set for voltage output .......................................................................... 30
7.2.2 When set for current output .......................................................................... 30
7.3 Alarm (Trip) Contacts ........................................................................................... 30
7.4 Serial (RS 232) Port ............................................................................................. 31
8. Spare Parts ...................................................................................................32
9. Moisture Sensor Option...................................................................................33
9.1 General Description .............................................................................................. 33
9.1.1 MGA3000 Zero Calibration............................................................................. 33
9.1.2 MGA3000 Span Calibration ............................................................................ 33
9.1.3 Sensor Autocal............................................................................................. 33
9.1.4 Pre-conditioning the transmitter..................................................................... 34
9.1.5 Adjust the Auto Cal ...................................................................................... 34
MGA3000 Multi-Gas Analyser Operation Manual Iss. 1.00
Copyright 2007, ADC Gas Analysis Ltd. 3
9.1.6 Completing the Auto Cal ............................................................................... 34
9.1.7 Faults/Errors................................................................................................ 35
1. INTRODUCTION
IMPORTANT
Before operating the Instrument for the first time, ensure all the relevant installation
checks have been carried out [refer to Section 2.9].
1.1 About the Instrument
The MGA3000 Series Instrument is designed for bench or rack mounting and provides a
measurement of gas concentration according to specific user requirements.
Single, dual or triple gas versions are available utilising one or more IRGA optical benches
and/or chemical cells.
Most instruments incorporate IRGA benches that generate a signal proportional to the infrared
absorption of the measured gas. Providing the gas concentration lies within the designed range
of the instrument, the signal is periodically compared with reference gas (usually at zero and a
convenient mid-span condition). Calibration is achieved manually via simple front panel
controls.
Some instruments may include a chemical cell which reacts with the specific gas and which
provides an electrical output in proportion to the reaction.
Gas concentrations are indicated on a front display panel together with both analogue and
digital signal outputs.
Gas connections to and from the instrument are via suitable fittings on the rear panel. The gas
flow is maintained via an internal sample pump.
The MGA3000's electrical supply can be set at the rear panel to operate from a 110V, 220V or
240VAC mains supply.
1.2 Instrument Identification
Each Instrument is uniquely identified with a Serial Number. This is located on a label on the
rear panel alongside the Model Number for the particular Instrument.
Always quote the serial number and model number in any communications requiring
after sales support.
1.3 The Display & Operating Controls
1.3.1 The Display
The display is a LCD Graphic type that is
illuminated from behind to achieve good
contrast and visibility. Its primary function is
to provide large character readouts of the gas
concentrations together with sample flow
status and key functions.
1.3.2 The Operating Controls
The operating controls are menu driven and
accessed by the 5 keys on the keypad
adjacent to the LCD display. The individual
functions of the keys are clearly labelled on
the display.
Selecting a menu function (by pressing the appropriate key) will cause the function to be
selected, or the next menu to be displayed, as required. In general, the bottom menu key is
used to return to the previous menu, acknowledge a message, or cancel an operation.
1.4 Electrical Outputs
1.4.1 Analogue Outputs
MGA series analysers have up to four analogue outputs, numbered 1 to 4 for connection
purposes. One output is provided for each of the gas channels fitted. When fewer than four
gas channels are fitted, one or more outputs will be inactive or not connected.
Alarm (trip) outputs
Up to five alarm contacts are provided to indicate gas readings above/below/outside user set
concentrations
1.5 RS232 port
An industry-standard serial port is fitted to the rear panel to allow gas readings to be read and
zero to be set by a PC or other computer equipment.
Figure 1 - An example main screen (single gas
model)
1.6 Operating Principles
1.6.1 Infrared Absorption
With the exception of Oxygen, the instrument measures gas concentrations using the effect of
infrared absorption.
All heteroatomic gases absorb (or transmit) light energy in the infrared region at specific
wavelengths that are dependent upon the chemical composition of each type of gas.
Furthermore, the level of absorption is proportional to the mass of the gas present. This
characteristic is utilised in infrared gas analysis (IRGA) benches.
The benches incorporate a source of infrared energy (usually in the form of a heated filament),
which is then optically filtered to reduce the radiation spectrum to that of the absorption band
of the gas being measured.
A pyro-electric detector is used to measure the difference in energy between a reference zero
(no absorption) condition and that after some absorption in the gas has occurred. By
calibrating the bench at the zero condition and a known span concentration, a reliable measure
of concentration over the span range is obtained.
To improve the stability and to assist in the detection process, the infrared source is modulated
mechanically at 3-18hz depending on the type of bench being used.
The relationship between absorption and concentration is not linear. For this reason the non-
linear output from the detector is modified electronically to give a linear output to the chart
recorder and display outputs.
The performance of an IRGA bench is compromised by the presence of 'interference' gases in
the sample. Such gases, of which the most common is water vapour, have an overlapping
infrared response to that of the measured gas and their effects are minimised by careful
filtering and the selection of the absorption band.
The level of absorption measured is relative and therefore an IRGA provides a comparative
measurement, not an absolute one, against 'zero' and 'span' reference gases.
A 'zero' gas has no detectable concentration of the gas being measured. Depending on the
gas, fresh air, chemically stripped air or nitrogen may be used. A 'span' gas has a known
concentration of the measured gas and is used to set the gain (or span) of the analysis in the
mid-range region. Under normal circumstances, the concentration is equivalent to about 80%
of the displayed full-scale reading.
1.6.2 The Optical Bench In Use
In order to obtain and maintain the optimum performance from the instrument, a number of
guidelines are offered associated with the use of the instrument, and, in particular, those
elements under the control of the User.
The IRGA bench is essentially an optical system and can suffer some deterioration in the
performance if it is internally contaminated by the condition of the incoming gases.
The gas sample should be reasonably dry. If the presence of water vapour is or is not expected
it is recommended that the instrument is continuously switched 'ON' to prevent condensation.
The flow rate of the gas sample affects the response time (to change) of the measurement -
the measurement being on average the mix of the sample flowing through the cell. In practice
a sample flow rate of nominally 500 ml/min
1
provides satisfactory performance for any
instrument. The instrument is fitted with a flow sensor and the flow rate is indicated on the
display. There is a potentiometer and throttle fitted internally that allows this flow rate to be
reset if necessary.
Generally the sample gas will be pumped to achieve the required flow rate. An internal pump is
fitted in the instrument and can be switched on or off via the key labelled ‘pump’ on the front
panel. It is recommended that the pump is switched 'OFF' if the sample lines are being
attended to or the condition of the sample gas is suspect (for example it contains water).
The sample rate is factory pre-set for sample gas at ambient pressure.
The installation of pressurised cylinders must include suitable pressure regulators and carefully
controlled to avoid damage to pump or flow sensor due to over-pressure.
The instrument must be allowed to 'temperature stabilise' after first being switched on. The
display will show a thermometer symbol until the associated bench has reached operating
temperature. The instrument's enclosure plays a role in temperature control as well as
providing protection and the enclosure (top and front) should be securely in place before use.
Unless the instrument is only used occasionally it is advisable to leave the instrument switched
on, particularly over-night when reduced temperatures may result in condensation. In standby
conditions the sample pump should be switched off'.
1.6.3 The Oxygen Sensor in Use
As oxygen does not absorb infrared (it is diatomic), it's measurement is via a chemical oxygen
sensor. This operates in a similar way to a battery in that a voltage is generated between two
plates immersed in electrolyte. The sample gas is passed over a membrane which allows
oxygen molecules to pass through it into the electrolyte, stimulating the reaction, and resulting
in a voltage proportional to the oxygen concentration.
Typically the sensor is calibrated by adjusting for zero reading with nitrogen applied, and for
span using a cylinder containing 20.9% O
2
, though fresh air can be used for a quick check.
Both span and zero adjustments are made via front panel controls.
The sensor fitted has a long operational life and is very stable. If the output falls (i.e. the gas
reading is low), the sensor should be replaced. Sensor lifetime depends primarily conditions of
use (e.g. average sample oxygen concentration, sample flow rate, temperature), and so this
part is excluded from the instrument warranty. For typical applications, ADC recommends
routine replacement of the sensor every two years as part of the maintenance programme.
ADC (Sales & Service) Ltd. or their authorised local agents can supply replacement sensors with
the correct fittings and electrical characteristics. Fitting alternative parts may cause problems
and will invalidate the instrument warranty.
2. INSTALLATION
2.1 Installation Arrangement
The basic arrangement of the Instrument is a ‘3U’ high, 483mm (19”) chassis for rack
mounting, or alternatively, it may be supplied with a case for bench mounting.
All power input, output and gas connections are via the rear panel.
Power connection is via a standard three-way IEC power plug. A suitable plug and cable is
supplied.
Gas connection details are indicated in Section 2.7.
RS232C, analogue and alarm contact outputs are detailed in sections 2.4, 2.5 and 2.6
respectively.
2.2 Location & Ambient Conditions
The location must provide firm support and access to the rear of the instrument when the
instrument is mounted in the normal attitude on its base. When installed in a rack, the use of
runners is recommended.
It is not necessary to force ventilate the enclosure to cool the unit, but free circulation of air is
necessary for accuracy and reliability.
Although it is not necessary to maintain a controlled environment during use, a reasonably
stable ambient temperature gives best results. Sudden changes in temperature will disturb the
IRGA bench(s), causing a shift in readings until the compensation recovers. If a drop in
temperature is expected, it is best to leave the instrument energised (switched 'ON'), to avoid
possible condensation of water in the optical bench. If not being used in this condition, it is also
better to switch off the sample pump as this not only prolongs the life of the pump, but helps to
reduce condensation.
In all cases the ambient conditions must comply with the specifications. High
ambient conditions will affect reliability.
IMPORTANT: The MGA series analysers are not certified for operation in hazardous
environments.
2.3 Power Supply Setting & Connections
2.3.1 Power Supply Setting
The instrument operates from single phase, 110, 220 or 240 volts nominal AC, at 50 or 60 Hz.
The supply frequency is not critical, but the voltage must be set on the instrument to suit the
local power supply.
The voltage setting and connections are on the rear panel.
On delivery, the instrument is usually set for 240 volts operation. In this case, the number '240’
can be seen in the window of the voltage selector, just below the inlet socket.
If 220VAC or 110VAC operation is required, the voltage setting must be changed via the
combined voltage selector and fuse holder.
The selector is removed by pushing down the dip and pulling the fuse out. Pull the fuse holder
out and rotate it so that the voltage required shows in the selector window. If necessary,
change the fuse according to the table below.
Supply Fuse Ratings
110V 220V 240V
2A T 1A T 1A T
Spare fuses are provided which are 20 X 5 mm ceramic HRC ‘T’ (Time Delay) types.
Maximum Power consumption is 120VA.
2.3.2 Power Supply Connections
The power cable supplied must be connected to a suitable plug or distribution board
as follows:
BROWN - Line or High Voltage
BLUE
- Neutral or Low Voltage
GREEN/YELLOW - Earth or Ground
For safety reasons the Earth connection MUST provide a low resistance to Ground.
Further protection is provided by making a direct Earth-Bonded connection to the chassis of the
instrument, via a 4mm screw. This bonding is just below the voltage selector.
The instrument is provided with an ON/OFF switch under the power socket.
WARNING
If the front, top or bottom sections of the enclosure are removed, BEWARE of
HIGH VOLTAGES near the power connector, switch and internal supply.
When REMOVING or REPLACING the enclosure sections ALWAYS disconnect the
mains supply FIRST.
Avoid touching the electronic boards. If these are removed for any reason -
SWITCH OFF the instrument FIRST.
DO NOT HANDLE without ANTI-STATIC precautions being taken.
2.4 Serial I/O (RS232C)
An industry-standard serial port is fitted which allows the analyser to be connected to a wide
range of computer equipment for logging and remote control. The connector is similar to an
‘AT’ style PC serial port allowing the use of standard cables.
The serial port is configured from the front panel – see section 3.7.2. The serial port protocol is
detailed in section 3.7.1.
Serial (RS232C) Port Connections
Pin Name Function
1 - no connection
2 Rx Received data (input)
3 Tx Transmitted data (output)
4 DTR Handshake for Rx (output)
5 ground 0V (nominally chassis)
6 - no connection
7 RTS ‘true’ whenever unit is powered
8 CTS Handshake for Tx (input)
9 - no connection
Notes:
1. DTR is normally ‘true’
2. CTS is monitored only when ‘hardware handshake’ is selected
3. Screened cables must be used for RS232 connections
In common with all RS232C ports, long cable runs can cause problems with data corruption and
noise. Keep cables as short as practicable as – certainly no longer as 30 meters at 9600 baud.
Route signal cables well away from power cables, switchgear and other noise sources. For long
cable runs, and in electrically noisy environments, optical RS232C cable extenders are
recommended. Contact ADC (Spares & Service) Ltd. or their local authorised agents for advice.
2.5 Analogue Outputs
Analogue outputs are provided for connection to chart recorders, remote meters, analogue
logging equipment etc. Up to four outputs are available; one for each gas reading.
The outputs are arranged so that the reading shown at the top of the screen is presented on
analogue output 1, the reading on the next line down (if any) is presented on output 2 and so
on. Where there are fewer than four gas readings, unused analogue outputs will be inactive or
not fitted.
The analogue outputs are available via a circular ‘DIN’ connector on the rear panel, labelled
‘OUTPUTS’. A suitable connector is provided in the kit with each instrument, and further
supplies are available from ADC (Spares & Service) Ltd. or their authorised local agent.
mA
+ -
Channel 1 3 8
Channel 2 7 5
Channel 3 2 4
Channel 4 6 1
NOTES:
1. V is used to indicate pin functions when output is set to voltage mode
2. I is used to indicate pin functions when output is set to 4..20mA mode
3. Screened cable should be used for analogue output connections (screen to connector shell).
Each output can be set to one of:
0 – 1V ]
0 – 5V ]- voltage modes
0 – 10V ]
4 – 20mA - current mode
Independently, using jumper links inside the case as shown in the section below.
2.5.1 Setting the analogue outputs
The procedure is as follows:
1. Disconnect the mains supply from the analyser, together with all the signal cables.
2. Isolate gas supplies, then remove the gas connections if necessary, taking all necessary
precautions.
3. Remove the top cover from the instrument by undoing the six screws on the top face, and
two screws on the back edge of the cover plate. Lift the cover clear.
4. Locate the RS232C connector on the rear panel, then look inside to locate the connector on
the circuit board (PC 392). The jumper links are all near to the RS232 connector, as shown
on the drawing (Fig. 3).
Voltage
+ -
Channel 1 3 8
Channel 2 7 5
Channel 3 2 4
Channel 4 6 1
Figure 2 – Analogue Output
pin numbers
5. In the drawing, the jumper settings
shown illustrate the different settings as
follows:
Output 1 – 10V
Output 2 – 5V
Output 3 – 4.. 20mA
Output 4 – 1V
6. Set the links as required, by pulling
them off the pins with tweezers or a
small pair of pliers, and then very
carefully re-seat them in the new
positions. The links should slide on
easily if they are aligned correctly.
7. Re-fit the top panel completely before
making connections and applying
power.
WARNING
Applying voltages to analogue outputs set to voltage mode will cause damage!
Check link settings before making connections
2.6 Alarm (Trip) Outputs
The alarm outputs are isolated relay contacts which can be set to operate at certain gas levels,
or on sample flow error. Up to five contacts may be available depending on the specification of
the instrument. The number actually available is shown on the trip set-up screen (see section
3.6.1).
Alarm contact outputs are available via a 14 way Amphenol connector on the rear panel. A
suitable connector is provided in the kit with each instrument, and further supplies are available
from ADC (Spares & Service) Ltd. or their authorised local agent.
The connector pin functions are printed on the rear panel adjacent
to the connector, and the pin numbering scheme is shown in Fig 4.
See section 7.3 for specifications.
A nominal 24 Volt (±20%) DC supply is available at the alarm
connector which can be used to power larger relays, sirens,
indicator lamps etc. This power source is internally fused at 1A
and the maximum continuous load is 500mA. The +24V DC
supply is referenced to chassis ground.
Figure 3 - Position of the Voltage/Current selection
links on the PCB
(rear panel view)
1
7
8
14
Figure 4 - Alarm
connector pin numbers
bi
2.7 Gas Arrangements
2.7.1 The Gas Circuit
The gas connections within the instrument are shown on the gas circuit enclosed with this
manual. The gas circuit is quite specific to a particular instrument and shows how the gas is
routed from the inlets to individual analysis bench(s). Gas connections are via 'entries' that are
located on the rear panel.
All external connections are made with 6.35mm (1/4'') pipe fittings. The type of piping used is
recommended to be either polythene for general use or polytetraflurothalene (PTFE) or
stainless steel for corrosive gases.
All entries are fitted with in-line filters that are intended to stop any foreign matter entering the
instrument as a last resort. Unless the gas is perfectly clean and dry, particulate filters
and / or driers must be fitted externally. Contact ADC (Spares & Service) Ltd. or their
authorised local agents for advice.
The requirements for particular gases vary with the application and therefore these and the gas
circuit are specific to the delivered Instrument. Some guidelines follow, but it is the
responsibility of the installer to ensure that all safety and other requirements are met.
The Instrument Specification accompanying the Instrument defines the specific requirements
and this must be referred to during installation.
WARNING
All Gases are potentially toxic and hazardous to health.
All gas lines and connections must be leak-tight. Joints can be checked by
applying water containing some liquid soap and looking for bubbles.
Exhaust lines must be properly vented and arranged to prevent blockage.
Ventilate the enclosure to prevent a build-up of gas in the event of a leak.
2.7.2 The Sample Gas
The gas to be measured is the 'sample gas' labelled 'GAS IN'. After passing through the gas
circuit the measurement gases are exhausted at the entry labelled 'GAS OUT'. The analyser is
designed to receive sample gas at nominally atmospheric pressure, using the internal sample
pump to control the flow.
The Sample Gas supplied to the Instrument must be relatively clean, cool and of low moisture
content. Samples containing dust or particulates must be filtered externally and the filters
serviced at regular intervals.
Hot Samples with high moisture content should be passed through a water trap or dissector
after cooling.
To prevent the possibility of condensation in the Instrument if the Instrument is not in
continuous use, it is recommended that it is left Switched on with the Sample Pump Switched
off.
Samples may contain other gases which ‘interfere’ with the gas being measured. The extent to
which known interferents can affect the measurement are defined in the Instrument
Specification.
Sample gas pipes shall be inert to and shall not contaminate the gas.
The response time for a change in concentration will be affected by the sample flow rate.
Instrument response times are normally quoted for the recommended 500ml min
-1
flow rate,
ignoring the effect of sample pipe volume. Lower flow rates will increase the response time, as
will the volume of the sample piping. Higher flow rates will reduce the response time, but
excessive flow will pressurise the instrument which may cause damage or measurement errors.
If it is necessary to draw higher flows than recommended (e.g. due to long sample gas lines), it
will be necessary to provide some means of bypassing the excess flow to exhaust.
2.7.3 The Span Gas
The Span Gas is used for calibration and is a known concentration of the measured gas -
usually between 50% and 80% of the full-scale range of the Instrument. The gas is usually
obtained specially mixed in a pressurised cylinder and this must be fitted with a 2-stage
pressure regulator, with an output pressure indication of 0 - 0.3 Bar (0-5psig).
The Regulator should be set to give a nominal output pressure of 0.2 Bar (3psig). This will
usually give a suitable gas flow during a span operation (i.e. flow indicator approximately mid-
position). If necessary, a small adjustment to the pressure will provide correct flow.
WARNING – Application of gas pressures above 0.3 Bar / 5 psig will damage the
instrument.
2.7.4 The Zero Gas
The zero gas is used to check/set the reading from the optical benches in the zero
concentration condition. Dry nitrogen is the preferred zero gas as it is cheap and readily
available.
The zero gas must not contain any significant traces of the gas measured or interferents. For all
types of instruments and gases measured, nitrogen (from a cylinder) is usually specified. For
some gases however, fresh air may be used (for example on high concentration CO and CO
2
Instruments). When fresh air is used, it shall be drawn from 'outside’ air and away from any
possible contamination from such as Exhaust Ducts, Chimneys, etc.
In most respects fresh air should be treated in the same way as the sample gas with regard to
filtering, moisture content etc. See span gas, section 2.7.3.
2.7.5 The Exhaust Gas
Exhaust gas is the sample (or zero/span) gas that has passed through the analyser. For
hazardous gases (high concentrations, flammable or poisonous), exhaust gas must be carefully
routed to a safe venting point.
It is important to minimise ‘back pressure’ at the GAS OUT port, as this will affect readings and
reduce the efficiency of the sample pump. Where long runs of pipe are necessary, larger bore
pipe should be used via a suitable adapter. Where the exhaust pipe is connected to a forced air
extraction duct, make sure that any suction is minimal.
2.7.6 Gas Pump(s)
Gas pumps fitted in the instrument are of the vibrating diaphragm type and are capable of up
to 1.0 I/min
1
under pressures normally encountered.
The sample pump operates at fixed speed and at a flow rate of nominally 500 ml/m1n
1
. If it
should be necessary to adjust the flow, it can be set using the potentiometer VR2 on the power
supply board and / or the throttle (when fitted).
The internal sample pump is operated via the front panel 'PUMP' menu button.
Certain instrument configurations may be fitted with a second internal pump which is used to
circulate purge gas. This pump would be set to run continuously at a factory set rate.
2.8 Use of Chemicals (certain models only)
Certain instruments use chemicals to remove (or strip) certain gases from the air for zero
adjustment, or to purge the optics to prevent ambient air from affecting readings. These
chemicals must be replaced regularly to ensure efficient operation and hence good accuracy.
The chemicals are contained in glass ‘columns’ which are accessible from the front of the
instrument to aid inspection and refilling.
The chemicals that may be used are listed below under the gas application.
2.8.1 Water Vapour
An indicating form of calcium sulphate (CaSO
4
) known as 'DRIERITE' is normally supplied with
new instruments and replacement supplies are available from ADC (Spares & Service) Ltd or
their local authorised agents. In use, drierite changes from blue to pink as it becomes
exhausted, making it obvious when the container should be refilled with fresh chemical.
Drierite can be recovered by drying.
Magnesium Perchlorate (Mg (ClO
4
)
2
0 (also known as Anhydrone) is also suitable, but it is a
strong oxidising agent, which may restrict its use. It is an effective drier, but does not provide
any indication when it is exhausted and is difficult to recover. As more water vapour is
absorbed it takes on a solid appearance, at which time it should be replaced.
If water vapour is to be removed from gases external to the instrument, calcium chloride
(CaCl
2
) is an inexpensive and efficient chemical to use. When absorbing moisture however, it
forms a corrosive liquid and therefore the container should be made of glass, with a moisture
trap.
‘Silica gel’ is a good indicating drier, but can ‘hang on’ to some gases, especially C0
2
, giving
false readings, and is not generally recommended.
2.8.2 Carbon Dioxide
To remove C0
2
, usually in a C0
2
Instrument, indicating ‘soda lime’ is used. Soda lime consists
of about 80% calcium hydroxide ((Ca (0H)
2
)), 4% sodium hydroxide (NaOH) and 0.2%
indicator (green to brown).
In use, soda lime changes from green to brown and must be changed when green is no longer
visible. It is not recoverable. Replacement supplies are available from ADC (Spares & Service)
Ltd or their local authorised agents.
Other chemicals may produce corrosive by-products which can damage the instrument.
2.8.3 Precautions
Any chemical intended for use in the instrument is provided with it in a separate container,
labelled with international identification labels. Under normal use, the quantity supplied will last
for about 6 months to a year. Replacements can be obtained from ADC (Spares & Service)
Limited who will also advise on the use of chemicals obtainable by the User.
When handling chemicals, refer to the ‘Safety Letter’ enclosed in the manual for precautions
against possible health risk. For all chemicals, avoid skin and eye contact and do not taste or
swallow them.
As Magnesium Perchlorate is an oxidising agent, do not bring it into contact with
combustible material and dispose of it in a sealed glass container.
2.9 Pre-operation Checks
Before connection power to the instrument, check:
1. The voltage indicated on the voltage selector matches the local supply
2. That gas connections are leak free
3. That the exhaust gas venting arrangements are safe for the gases in the sample
4. That the instrument is properly grounded (earthed).
2.10 General Advice
2.10.1 Use of Gas Cylinders
Before use, Gas Cylinders must be checked to ensure that -
the cylinders are securely located.
the cylinders are fitted with a two-stage regulator.
the cylinders contain the correct gas mixture for use.
the cylinders are connected to the correct ‘entry’ on the Instrument.
the cylinder's Output Pressure regulator is 'OFF'.
During use, check -
the Instrument is 'ON'.
the Sample Pump is 'ON'.
the Output Pressure regulator is set for 0.2 Bar.
After use, check -
the Output Pressure regulator is turned 'OFF'.
the cylinder tap is turned 'OFF'.
if the Instrument is in ‘Standby’, turn 'OFF' the Sample Pump.
Following a cylinder change, it is advisable to initiate a Span or Zero cycle a few minutes after
the supply is re-connected.
2.10.2 Gas Leaks
All external-piping connections must be checked for gas leaks, to ensure that no gases are
leaking to the local area (which may be a health hazard), and that no uncontrolled air or gas is
drawn into the Instrument (and so affect the measurement).
Applying water containing a mild soap solution can check connections under pressure. Do not
increase the pressure to check for leaks (or for any other reason), since this may damage the
Instrument.
Connections of pipes under nominally ambient pressure, such as the sample piping, are difficult
to check for leaks, but can be checked for tightness.
2.10.3 Exhausting & Ventilation
Because any gas is potentially toxic, and, some may be inflammable, the gas must be
exhausted to where it will be rapidly dispersed. For the same reasons, the Instrument and its
environment including the location of gas cylinders must be properly ventilated to minimise the
dangers of gas leaks.
If Span cylinders contain hazardous or inflammable gases, they must carry suitable warning
labels, and if inflammable, must be located in a suitable ‘Inflammable Store’ when in use or in
storage.
In all cases the local safety regulations must be consulted and followed.
3. SETTING UP
3.1 Preview
This section describes the settings available from the front panel. Factory settings have been
chosen to suit the general application of the instrument, or in accordance with the particular
instructions specified previously by the user.
3.2 On Power Up
IMPORTANT: Before operating the Instrument, ensure all the relevant installation
checks have been carried out [refer to Section 2.9], and that if a filter column is
fitted, it is properly charged with the chemical supplied.
Immediately after power is applied, the display presents a screen for 10 seconds. This page
contains the software version number.
If the logo does not appear on the screen within 20 seconds of applying power, try
adjusting the ‘LCD’ control, using a 2mm flat bladed screwdriver through the labelled
hole in the rear panel.
The next screen is the main display, which will show the measured value(s), units and the Gas
type, sample flow status and key functions.
The measured value is initially replaced with a symbol of a thermometer until the instrument
has warmed up.
Pressing the key labelled ‘Set Up’ displays a second screen showing the gas(s) measured
together with the range and units.
3.3 Zero & Span Operations.
Before precise gas readings can be obtained, the analyser must be adjusted so that it reads
zero when there is no trace of the measured gas present, then, when a known concentration of
the gas is applied, the reading is adjusted to agree with this concentration. These adjustments
are known as ‘zero’ and ‘span’ respectively.
The gases used for zero and span adjustment are supplied through the inlets at the back of the
instrument (see section 2.7), and electrically operated valves are used to route the appropriate
gas in place of the sample automatically when zero or span adjustment is selected using the
front panel controls.
'How often' to Zero or Span varies with each instrument. The interval required depends on how
much re-adjustment is being carried out: this can be established by noting the discrepancy of
the displayed value prior to resetting.
If the discrepancy is less than 0.5% of full scale, the intervals can be increased, or if this is not
accurate enough, decreased. Generally Zeroing is required more often than Spanning, so that
typically a daily Zero, with weekly Span is adequate.
The instrument can be set to automatically adjust zero on a periodic basis. The period can be
set from 1 hour to 250 hours.
See section 4 for details.
3.4 Display & Keypad
3.4.1 Display
The normal operating layout is shown in Fig 5,
but note that this is an example; the gases
shown will vary with the specification ordered.
On the left of the screen are the gas names,
followed by the current reading(s). A
thermometer symbol indicates that the
analysis bench is warming up – readings will
be shown when warm-up is completed.
On the right, adjacent to the five ‘function’
keys, are the key labels. These describe the functions of the keys at any time, and the labels
change as different options are selected.
Between the gas readings and the function key labels, is the sample flow indicator, which is
designed to mimic a float flowmeter. Normally, the flow, as indicated by the ‘float’, should be
around the centre of the indicator. If the float is in a shaded region, the flow is too high or low
and a warning triangle is shown.
Flow fail and alarm trips are indicate at the bottom of the display.
The LCD display is affected by ambient
temperature and viewing angle. A contrast
control is provided to adjust the display to
give the best contrast under all viewing
conditions.
To adjust the contrast, Select the following
menu functions:
Set Up , Display
The contrast screen is displayed. Use the
lighter and darker controls to get the best
display. The bar shows the amount of
Figure 6 - screen contrast adjustment
Figure 5 - Example Main Screen (3 gas + ratio
model)
Comment [BC1]:
/