MSA TRIGARD® Gas Monitoring System Owner's manual

Category
Carbon monoxide (CO) detectors
Type
Owner's manual
TRIGARD
®
Gas Monitor
Instruction Manual
THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL
HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE PRODUCT. Like any piece
of complex equipment, this instrument will perform as designed only if it is used and
serviced in accordance with the manufacturers instructions. OTHERWISE, IT COULD
FAIL TO PERFORM AS DESIGNED AND PERSONS WHO RELY ON THIS PRODUCT FOR
THEIR SAFETY COULD SUSTAIN SEVERE PERSONAL INJURY OR LOSS OF LIFE.
The warranties made by Mine Safety Appliances Company with respect to the product are
voided if the product is not used and serviced in accordance with the instructions in this
manual. Please protect yourself and others by following them. We encourage our cus-
tomers to write or call regarding this equipment prior to use or for any additional infor-
mation relative to use or repairs.
In North America, to contact your nearest stocking location, dial toll-free 1-800-MSA-INST
To contact MSA International, dial (724) 776-8626
© MINE SAFETY APPLIANCES COMPANY 2013 - All Rights Reserved
This manual is available on the internet at www.msasafety.com
Manufactured by
MSA NORTH AMERICA
1000 Cranberry Woods Drive, Cranberry Township, Pennsylvania 16066
(L) Rev 3 10072857
"! WARNING
MSA Instrument Warranty
1. Warranty- Seller warrants that this product will be free from
mechanical defect or faulty workmanship for a period of eighteen
(18) months from date of shipment or one (1) year from installation,
whichever occurs first, provided it is maintained and used in
accordance with Seller’s instructions and/or recommendations.
This warranty does not apply to expendable or consumable parts
whose normal life expectancy is less than one (1) year such as,
but not limited to, non-rechargeable batteries, sensor elements,
filter, lamps, fuses etc. The Seller shall be released from all
obligations under this warranty in the event repairs or modifications
are made by persons other than its own or authorized service
personnel or if the warranty claim results from physical abuse or
misuse of the product. No agent, employee or representative of the
Seller has any authority to bind the Seller to any affirmation,
representation or warranty concerning the goods sold under this
contract. Seller makes no warranty concerning components or
accessories not manufactured by the Seller, but will pass on to the
Purchaser all warranties of manufacturers of such components.
THIS WARRANTY IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESSED, IMPLIED OR STATUTORY, AND IS
STRICTLY LIMITED TO THE TERMS HEREOF. SELLER
SPECIFICALLY DISCLAIMS ANY WARRANTY OF
MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR
PURPOSE.
2. Exclusive Remedy- It is expressly agreed that Purchaser’s sole
and exclusive remedy for breach of the above warranty, for any
tortious conduct of Seller, or for any other cause of action, shall be
the repair and/or replacement at Seller’s option, of any equipment
or parts thereof, which after examination by Seller is proven to be
defective. Replacement equipment and/or parts will be provided at
no cost to Purchaser, F.O.B. Seller’s Plant. Failure of Seller to
successfully repair any nonconforming product shall not cause the
remedy established hereby to fail of its essential purpose.
3. Exclusion of Consequential Damage- Purchaser specifically
understands and agrees that under no circumstances will seller be
liable to purchaser for economic, special, incidental or
consequential damages or losses of any kind whatsoever, including
but not limited to, loss of anticipated profits and any other loss
caused by reason of nonoperation of the goods. This exclusion is
applicable to claims for breach of warranty, tortious conduct or any
other cause of action against seller.
General Warnings and Cautions
1. The Trigard Gas Monitors described in this manual must be
installed, operated and maintained in strict accordance with their
labels, cautions, warnings, instructions, and within the limitations
stated.
2. The Trigard Gas Monitor is designed to detect gases or vapors in
air. It cannot measure the concentration of gases or vapors in
steam or inert or oxygen-deficient atmospheres. The oxygen
sensor can measure oxygen-deficient atmospheres.
3. Electrochemical sensors are sealed units which contain a corrosive
electrolyte. Should a sensor develop leakage, it must be
immediately removed from service; then, remove it from the
sensing head and discard it properly. Caution must be exercised so
that the electrolyte does not contact skin, eyes, clothing or circuitry;
otherwise, serious personal injury (burns) and/or equipment
damage may result.
4. Use only genuine MSA replacement parts when performing any
maintenance procedures provided in this manual. Failure to do so
may seriously impair instrument performance. Repair or alteration
of the Trigard Gas Monitor, beyond the scope of these
maintenance instructions or by anyone other than an authorized
MSA service personnel, could cause the product to fail to perform
as designed and persons who rely on this product for their safety
could sustain serious personal injury or death.
5. Do not locate the general-purpose enclosure models in an area
which may contain a flammable mixture of gas and air; otherwise,
an explosion may occur. The general-purpose Trigard Gas
Monitors can be a source of ignition and must not be mounted in
an area where a flammable mixture of combustible gas and air
may become present; otherwise, an explosion may occur.
Failure to follow the above can result in serious personal
injury or death.
"
WARNING
i
1. As with all gas monitors of these types, high levels of, or long
exposure to, certain compounds in the tested atmosphere could
contaminate the sensors. In atmospheres where Trigard Gas
Monitor may be exposed to such materials, calibration must be
performed frequently to ensure that operation is dependable and
display indications are accurate.
2. The Trigard Gas Monitor must not be painted. If painting is done in
an area where a Monitor is located, care must be exercised to
ensure that paint is not deposited on the sintered, metal flashback
arrestor in the inlet fitting of the Trigard Gas Monitor, if so
equipped. Such paint deposits would interfere with the diffusion
process, whereby a sample of the atmosphere being monitored
diffuses into the Monitor.
3. The only absolute method to ensure proper overall operation of
Trigard Monitor is to check it with a known concentration of the gas
for which it has been calibrated. Consequently, calibration checks
must be included as part of the routine inspection of the system.
4. Protect the Trigard Gas Monitor from extreme vibration. Do not
mount the sensing head in direct sunlight as this may cause
overheating of the sensor.
Failure to follow the above can result in injury, product
damage and/or an unsafe condition.
"
CAUTION
ii
Table of Contents
Chapter 1,
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Identifying Your Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Installing Your Gas Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Installing the Trigard Gas Monitor . . . . . . . . . . . . . . . .1-3
Electrical Connections for Trigard Gas Monitors . . . . . . . . . . .1-3
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
For Milliamp Output . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Typical Trigard Gas Monitor Wiring . . . . . . . . . . . . . . . . . . . . .1-5
Installing the Trigard Remote Sensor Module . . . . . . . . . . . . .1-7
Electrical Connections for the Trigard
Remote Sensor Module . . . . . . . . . . . . . . . . . . . .1-9
At the Trigard Remote Sensor Location: . . . . . . . . . .1-10
Chapter 2,
Start-up and Calibration for 4-20 mA Output . .2-1
Initial Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Calibration Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Chemicals that Reduce Catalytic Sensor Sensitivity . .2-4
Ultima Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Ultima Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Note on Resetting Latched Alarms
with Controller or Calibrator . . . . . . . . . . . . . . . . . . . . .2-6
Trigard Gas Monitor Calibration Output Signal . . . . . .2-6
Calibration Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Trigard Gas Monitor Calibration Procedure . . . . . . . . . . . . . . .2-7
INITIAL Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Regular Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Zeroing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Spanning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
Calibration Documentation . . . . . . . . . . . . . . . . . . . .2-12
iii
Chapter 3,
Start-up and Calibration for MODBUS output . .3-1
Typical ModBUS Network Topography (FIGURE 3-1). . . . . . .3-1
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
RS485 Communications . . . . . . . . . . . . . . . . . . . . . . .3-1
Transmitter Sensor Distance . . . . . . . . . . . . . . . . . . .3-1
Typical Communication Cable Wiring . . . . . . . . . . . . . . . . . .3-5
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Alarm Status Screen . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Ultima Controller or Ultima Calibrator: . . . . . . . . . . . . .3-8
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
ModBUS Communications . . . . . . . . . . . . . . . . . . . . .3-10
ModBUS Addressing . . . . . . . . . . . . . . . . . . . . . . . . .3-10
ModBUS Communications . . . . . . . . . . . . . . . . . . . . .3-11
ModBUS Base Address (Read/Write) . . . . . . . . . . . .3-12
Chapter 4,
Additional Features . . . . . . . . . . . . . . . . . . . . . . .4-1
Internal Relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Installing the Trigard Gas Monitor
with Internal Relays . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Trigard Gas Monitor Internal Relays . . . . . . . . . . . . . .4-2
Relay Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Alarm Relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Fault Relay or Trouble . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Calibration/RESET Push-button . . . . . . . . . . . . . . . . . . . . . . .4-4
Push-button Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Relay Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Horn Relay Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
To Activate the Horn Relay . . . . . . . . . . . . . . . . . . . . .4-7
To Reset the Horn Relay . . . . . . . . . . . . . . . . . . . . . . .4-7
Chapter 5,
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
iv
Chapter 6,
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Replacing Trigard Sensor . . . . . . . . . . . . . . . . . . . . . .6-1
Obtaining Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . .6-6
Appendix A,
Optional Features . . . . . . . . . . . . . . . . . . . . . . . .A-1
Strobes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Internal Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Backup Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
List of Figures
Figure 1-1. Trigard Gas Monitor . . . . . . . . . . . . . . . . . . . . . . .1-1
Figure 1-2. Circuit Board (3-Wire Version) . . . . . . . . . . . . . . .1-6
Figure 1-3. Three-Wire 4 to 20 mA Operation . . . . . . . . . . . . .1-7
Figure 1-4. Remote Sensor Cable Assemblies . . . . . . . . . . . .1-7
Figure 1-5. General-Purpose Remote Sensor Assemblies . . .1-8
Figure 1-6. Explosion-Proof Remote Sensor Assemblies . . . .1-8
Figure 2-1. LCD Gas Concentration Display . . . . . . . . . . . . . .2-1
Figure 2-3. Ultima Controller . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Figure 2-2. Ultima Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Figure 2-4. Apply Zero Gas Flag . . . . . . . . . . . . . . . . . . . . . . .2-9
Figure 2-5. Apply SPAN Gas Flag . . . . . . . . . . . . . . . . . . . . .2-10
Figure 2-6. Calibration End Display . . . . . . . . . . . . . . . . . . . .2-11
Figure 3-1. Typical ModBUS Network Topography . . . . . . . . .3-2
Figure 3-2. Typical Communications Wiring Scheme . . . . . . .3-4
Figure 3-3. Sensor Display Screens . . . . . . . . . . . . . . . . . . . .3-6
Figure 3-4. Alarm Status Screen . . . . . . . . . . . . . . . . . . . . . . .3-8
Figure 3-5. ModBus PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
FIGURE 4-1. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Figure 6-1. "Change Sensor" Scrolls Across the Display . . . .6-1
Figure 6-2. Sensor Assembly and Sensor Guard for General
Purpose Model . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Figure A-1. Trigard Gas Monitor with Additional Options . . . .A-1
Figure A-2. Trigard Gas Monitor Power Supply and Battery
Backup Options . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
v
List of Tables
Table 1-1. Power Cable Distances for the Trigard Gas
Monitor with Internal Relays (4-20 mA Model) . . .1-5
Table 1-2. Cable Length and Wire Size
(Power Supply 24 VDC) (Toxic Gas or Oxygen)
Sensor, 4-20 mA Signal Output
(Three Wire Sensor) . . . . . . . . . . . . . . . . . . . . . .1-5
Table 1-3. Remote Module Wiring and Placement . . . . . . . . .1-9
Table 1-4. Remote Sensor Wiring Cable . . . . . . . . . . . . . . . .1-10
Table 1-5. Low Temperature Wiring Cable . . . . . . . . . . . . . .1-10
Table 2-1. Instrument Operation . . . . . . . . . . . . . . . . . . . . . . .2-2
Table 3-1. Maximum Power Cable Length . . . . . . . . . . . . . . .3-3
Table 3-2. Transmitter Power Consumption
(7-30 VDC Supply) . . . . . . . . . . . . . . . . . . . . . . . .3-3
Table 3-3. Remote Sensor Power Consumption
(AC or DC Operation) . . . . . . . . . . . . . . . . . . . . .3-3
Table 3-4. Status Indication Codes . . . . . . . . . . . . . . . . . . . . .3-8
Table 3-5. Supported ModBUS Function Codes . . . . . . . . . .3-11
Table 3-6. ModBUS Memory Map Overview . . . . . . . . . . . . .3-11
Table 3-7. ModBUS Bas Address (Read/Write) . . . . . . . . . .3-12
Table 3-8. ModBUS Factory Configuration Data
(Read Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
Table 3-9. ModBUS User Configuration Data (Read/Write) .3-14
Table 3-10. Alarm Function Codes - Word 1(Read/Write at
Address Base+137) . . . . . . . . . . . . . . . . . . . . . .3-15
Table 3-11. Alarm Function Codes - Word 2(Read/Write at
Address Base+138) . . . . . . . . . . . . . . . . . . . . . .3-16
Table 3-12. ModBUS Device Status (Read only) . . . . . . . . .3-17
Table 3-13. ModBUS General Status Bits
(Read Only at address Base+201) . . . . . . . . . .3-19
Table 3-14. ModBUS Fault Status Bits
(Read Only at address Base+202) . . . . . . . . . .3-19
Table 3-15. Control Words (Read/Write) . . . . . . . . . . . . . . . .3-20
Table 3-16. ModBUS Command Word 1 (Read at address
Base+301/Write Coils 1 through 16) . . . . . . . . .3-20
Table 13-17 ModBUS Command Word 2 (Read at address
Base+302/Write Coils 17 through 32) . . . . . . . .3-21
Table 3-18. Sensor Type . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-22
Table 3-19. Sensor Engineering Units . . . . . . . . . . . . . . . . . .3-23
vi
Table 3-20. Information Flags Word #1 –
(Read at address Base+254) . . . . . . . . . . . . . . .3-23
Table 3-21. Information Flags Word #2
(Read at address Base+255) . . . . . . . . . . . . . . .3-24
Table 3-22. Information Flags Word #3
(Read at address Base+26) . . . . . . . . . . . . . . . .3-24
Table 3-23. Information Flags Word #4
(Read at address Base+257) . . . . . . . . . . . . . . .3-25
Table 3-24. Alternate Gas Readings (Read/Write
at address Base+258 to Base+260) . . . . . . . . .3-25
Table 4-1. Relay Specifications . . . . . . . . . . . . . . . . . . . . . . . .4-2
Table 4-2. Push-button Calibration . . . . . . . . . . . . . . . . . . . . .4-5
Table 5-1. Performance Specifications . . . . . . . . . . . . . . . . . .5-1
Table 5-2. Sensor Response to Interferants . . . . . . . . . . . . . .5-4
Table 6-1. Operational Display Messages . . . . . . . . . . . . . . . .6-3
Table 6-2. Configuration Display Messages . . . . . . . . . . . . . .6-4
Table 6-3. Troubleshooting Guidelines . . . . . . . . . . . . . . . . . .6-5
Table 6-4. Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . .6-6
vii
Chapter 1,
Installation
General Description
The Trigard Gas Monitor is designed to sample the environment where
mounted and alert you to potentially dangerous levels of your target
gas, depending on your particular model. The unit is factory-calibrated
and shipped ready for installation.
Identifying Your Unit
The Trigard Gas Monitor is housed in a rugged, plastic general-
purpose enclosure (FIGURE 1-1).
The Trigard Gas Monitor comes standard with internal relays, a
Cal/Reset pushbutton, and a 95 dB horn. See Chapter 4 for details.
The gas sensor type and range are located on the sensor module
label. Unscrew the sensor assembly to view the label.
Any options ordered, such as Strobes, Battery Backup or Internal AC
Power Supply, will be included.
Figure 1-1. TRIGARD Gas Monitor
1-1
Installing Your Gas Monitor
NOTE: Reference installation outline drawings listed in Chapter 3,
"Specifications".
Generally, the Trigard Gas Monitors or remote sensing module should
be mounted close to the area where a leak is likely to occur or where
the gas is expected. Install the Trigard Gas Monitors or the remote
sensor at a high level (ceiling) or low level (floor), depending on the
density of the gas most likely to be found. Install the unit so that the
front display is not blocked or hidden from view.
Mount the Trigard Gas Monitor or remote sensor with the
sensor inlet fitting (FIGURE 1-1) pointed downward; other-
wise, the inlet may become clogged with particulate matter
or liquids.
Do not paint the Trigard Gas Monitors. If painting is done in
an area where a sensor is located, exercise CAUTION to
ensure paint is not deposited on the sensor inlet fitting.
Such paint deposits would interfere with the diffusion
process, whereby a sample of the monitored atmosphere
diffuses into the sensor. In addition, solvents in the paint
may cause an alarm condition to occur.
Protect the Trigard Gas Monitors from extreme vibration. Do
not mount sensing head in direct sunlight as this may cause
overheating of the sensor.
Do not locate the general-purpose enclosure models in an
area which may contain a flammable mixture of gas and air;
otherwise, an explosion may occur. The general-purpose
Trigard Gas Monitors can be a source of ignition and must
not be mounted in an area where a flammable mixture of
combustible gas and air may become present.
"
WARNING
"
WARNING
1-2
Installing the Trigard Gas Monitor
Remove lid and drill enclosure for power, signal and optional relay cable
entry. Use one of the following methods to mount the general-purpose
Trigard Gas Monitor.
Using customer-installed wiring holes, install the Trigard Gas
Monitor to the end of rigid conduit.
Use mounting holes in the corners of the Trigard enclosure to
mount directly to a wall.
Electrical Connections for Trigard Gas Monitors
Before wiring the Trigard Gas Monitors, disconnect power
source supplying the monitor; otherwise, electrical shock
could occur.
NOTE: For information on internal relays, see Chapter 4.
For Trigard units with ModBUS, see Chapter 3.
This three-wire assembly is marked to identify power, ground and signal
connections for all combustible, toxic, and oxygen models.
Wiring
Install wiring in accordance with the electrical code of the country in use
and UL 61010-A1 or CSA C22.2 No. 1010.1, as applicable. In these
installations, twisted-pair, instrument quality cable is recommended.
Shielded cable is recommended for cable runs where interferences from
radio frequency interference (RFI), electromagnetic interference (EMI)
or other noise sources exist (such as motors, welding equipment,
heaters, etc.).
NOTE: See Installation Outline Drawings for wiring details.
Conduit may also be needed in areas where large amounts of electrical
noise is expected.
Use caution when selecting a cable size. TABLES 1-1 and 1-2 express
the maximum cable length when only using the Trigard Gas Monitors.
Trigard options may take additional power which requires a heavier
cable or a short cable run.
"
WARNING
1-3
When selecting cable size, consider future needs (i.e., addition of
sensors and/or options available with the Trigard Monitors). See
Chapter 3, "Specifications" for proper input voltage.
Ensure that water and dirt are not able to enter the unit via the wire or
conduit. If the unit is installed in a location known to be wet or damp, it
is good practice to loop or bend the entry into the unit that prevents
water incursion.
For Milliamp Output
The Trigard Gas Monitor (three-wire version) may be connected to any
device capable of accepting 4 to 20 mA analog signals such as:
Model 9010/9020 Controller
GasGard XL Controller
Programmable controllers
DCS’s, etc.
An external power supply is needed if the optional internal power supply
is not ordered. (For power requirements, see Chapter 3,
"Specifications".) All connections should be made by following
appropriate wire code procedures.
See TABLES 1-1 through 1-2 for typical cable length and wire size
for installation.
When using any of the the Trigard accessories (such as
relays) with the 4 to 20 mA output Trigard Gas Monitor, a
three-wire connection must be used. Failure to use a three-
wire connection could damage the electronics within the
Trigard Gas Monitor which can result in serious personal
injury or death.
Be sure to install your Trigard Gas Monitor according to National
Electrical and local procedural codes. Failure to do so can result in an
unsafe condition.
NOTE: See Chapter 3 for ModBUS output.
"
WARNING
1-4
Table 1-1.
Power Cable Distances for the Trigard Gas Monitor with Internal
Relays (4-20 mA Model)
SENSOR TYPE POWER WIRE SIZE MAXIMUM MAXIMUM LOAD
SUPPLY CABLE RESISTANCE
VOLTAGE LENGTH (mA OUTPUT
(IN FEET) ONLY)
CATALYTIC
COMBUSTIBLE 12 VDC 16 AWG 900 300 OHMS
CATALYTIC
COMBUSTIBLE 24 VDC 16 AWG 3000 600 OHMS
TOXIC OR OXYGEN 12 VDC 16 AWG 2500 300 OHMS
TOXIC OR OXYGEN 24 VDC 16 AWG 8000 600 OHMS
See TABLE 1-6 for remote sensor distances.
In all installations, twisted instrument-quality cable is
recommended.
Shielded cable is recommended in situations where radio
frequency interference (RFI), electro-magnetic interference (EMI)
or other electrical noise sources exist or are anticipated.
Typical Trigard Gas Monitor Wiring
Table 1-2.
Cable Length and Wire Size (Power Supply 24 VDC)
(Toxic Gas or Oxygen) Sensor,
4-20 mA Signal Output (Three Wire Sensor)
WIRE SIZE MAXIMUM CABLE LENGTH IN FEET MAXIMUM LOAD RESISTANCE
22 AWG 12,000 600 Ohms
Three-wire Trigard Monitors operate in the current source mode
(see FIGURE 1-3).
1. Identify the main pc board:
It is the round pc board, located inside the lid assembly.
1-5
2. Connect 8 to 30 VDC power lead to J8-1 (see FIGURE 1-2).
3. Connect J8-2 to 4 to 20 mA input on remote system.
4. Connect the signal ground to J8-3.
5. Connect the sensor module to labeled connector J-1 on the main
pc board.
6. Wire for relays (see Chapter 4).
7. Assemble lid on enclosure.
Figure 1-2. Circuit Board (3-Wire Version)
1-6
Installing the Trigard Remote Sensor Module
The Remote Sensor Modules are available in a variety of options.
Pre-wired general-purpose sensors are available in lengths of 25, 50 or
100 feet. These remote sensors are not housed in an enclosure, which
makes it easier to mount in small places.
Figure 1-4. Remote Sensor Cable Assemblies
Figure 1-3. Three-Wire 4 to 20 mA Operation
1-7
Remote sensors can be ordered with general purpose (GP) remote
sensor assemblies.
Remote sensors can also be ordered with explosion-proof (XP) remote
sensor assemblies.
Figure 1-6. Explosion-Proof Remote Sensor Assemblies
Figure 1-5. General-Purpose Remote Sensor Assemblies
1-8
The GP or XP Remote Sensor Modules can be mounted in a manner
similar to the gas monitor installation in the preceding procedure and at
a maximum distance outlined in TABLE 1-3.
Table 1-3. Remote Module Wiring and Placement
GAS TYPE MINIMUM WIRE SIZE MAXIMUM DISTANCE
Toxic and Oxygen 20 AWG 100 FEET
Catalytic Combustible 18 AWG 50 FEET
16 AWG 100 FEET
IR Combustible 16 AWG 50 FEET
12 AWG 100 FEET
Permanently connect 1/4" ID tubing to the post on the windguard. Route
this tubing to the Trigard Gas Monitor, ensuring that there are no kinks,
leaks or other obstructions. Secure this tubing near the monitor; it is
used to deliver check gas to the sensor module during calibration.
Tubing used for reactive gases (Cl
2
, HCl, ClO
2
, HF, NH
3
,
ETO, F
2
, B
2
H
6
, Br) should be as short as possible (18 inch-
es maximum) to ensure that the gas reaches the sensor.
Excessive tubing can result in no gas sensor reading.
Electrical Connections for the Trigard Remote
Sensor Module
Before wiring the Trigard Remote Sensor Module, discon-
nect the power source feeding the Remote Sensor Module
and the Trigard Gas Monitor/Less Sensor; otherwise, electri-
cal shock could occur.
When installing Trigard Remote Sensor Module with its mat-
ing Trigard Gas Monitor/Less Sensor, follow National
Electrical and local procedural Codes; failure to do so can
result in an unsafe condition.
Five conductors are required for the Trigard Remote Sensor Module.
The Trigard Monitor has a five-wire terminal to accommodate up to #16
AWG conductors.
"
CAUTION
"
WARNING
"
CAUTION
1-9
Some installations require metal pipe or metallic conduit. In these
cases, separate conductors or unshielded cable may be used.
For open wiring, shielded wire or cable should be used to minimize the
possibility of noise interference and contact with other voltages.
Selection of this shielded cable must comply with local requirements.
TABLE 1-4 shows suggested cables for Trigard installations; other
cables are available which are also adequate.
Table 1-4. Remote Sensor Wiring Cable
SUPPLIER CATALOG NUMBER DESCRIPTION
ALPHA WIRE CORP. 5525 5 cond., shielded, 18 AWG
5535 5 cond., shielded, 16 AWG
5514 4 cond., shielded, 20 AWG
Table 1-5. Low Temperature Wiring Cable
SUPPLIER CATALOG NUMBER DESCRIPTION
ALPHA WIRE CORP. 45525 5 cond., shielded, 18 AWG
45366 6 cond., shielded, 16 AWG
45545 5 cond., shielded, 14 AWG
At the Trigard Remote Sensor Location:
1. Open the Trigard Remote Sensor cover by removing lid.
2. Route the cable from the Gas Monitor through a customer-supplied
opening in the enclosure and wire it to the terminal block
(FIGURE 1-2).
3. Verify the identity of each conductor of the cable and connect the
wire to the terminal block.
4. Re-install the cover of the Trigard Remote Sensor
NOTES:
Incoming power and signal cable shield should be earth grounded
at the power source. Connect power and remote sensor cable
shields to shield terminals on main pc board. Provide shield
terminations inside the sensor housing as indicated on Installation
Outline Drawings for Remote Sensor; see Table 5-1 for Installation
Outline Drawing document numbers.
Cables larger than #16 AWG will require a splice of smaller cable
to fit the connector.
1-10
Chapter 2,
Start-up and Calibration
for 4-20 mA Output
Initial Start-up
The Trigard Gas Monitors are factory-calibrated. A bump check
prior to use is recommended.
Once power is applied to the unit, the LCD shows a test of all
display words. The software version number displays; then, a 30-
second (self-check) countdown for sensor stability begins.
During the 30-second countdown, the output signal is the same as
the calibration signal when enabled during a normal calibration.
This is described later in this chapter under "Trigard Gas Monitor
Calibration Output Signal".
The Alert red LED will be solid ON during the 30-second
countdown.
After the 30-second countdown, observe that the gas type
and gas concentration (ppm, % Gas, or % LEL) alternately flash
(FIGURE 2-1).
1).
The Normal green LED will be solid ON after the 30-second
countdown.
A complete listing of instrument operation features can be found in
TABLE 2-1.
Figure 2-1. LCD Gas Concentration Display
2-1
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79

MSA TRIGARD® Gas Monitoring System Owner's manual

Category
Carbon monoxide (CO) detectors
Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI