MRC DIST-984 User manual

Type
User manual

MRC DIST-984 is an automatic distillation system designed according to the classical Kjeldahl method. This instrument provides lab staff with great convenience in measuring the content of nitrogen and protein, owning features such as high reliability, uncompromised safety, hassle-free operation, and time and energy-saving performance.

With a measurable amount of solid sample of <6g per sample and a measurable amount of liquid sample of <16 mL per sample, this device has a measurement range of 0.1~240 mg nitrogen and a recovery rate of ≥99.5%.

MRC DIST-984 is an automatic distillation system designed according to the classical Kjeldahl method. This instrument provides lab staff with great convenience in measuring the content of nitrogen and protein, owning features such as high reliability, uncompromised safety, hassle-free operation, and time and energy-saving performance.

With a measurable amount of solid sample of <6g per sample and a measurable amount of liquid sample of <16 mL per sample, this device has a measurement range of 0.1~240 mg nitrogen and a recovery rate of ≥99.5%.

Contents
I. INTRODUCTION ....................................................................................................- 1 -
II. SPECIFICATIONS.................................................................................................- 2 -
III. NAMES OF INSTRUMENT PARTS..................................................................- 3 -
IV. INSTALLATION INSTRUCTION........................................................................- 4 -
V. OPERATION INSTRUCTION..............................................................................- 7 -
VI. REGULAR MAINTENANCE............................................................................- 29 -
VII. APPENDIX.........................................................................................................- 29 -
VIII. COMMON PROBLEMS AND TROUBLESHOOTING...............................- 30 -
IX. ATTENTION........................................................................................................- 31 -
Warning: The instrument can not provide the designed protection for
operators who do not follow the right procedures and requirements given by
the manufacturer.
Warning: All solutions must be handled with care according to the
lab’s safety regulation. Please make a reference to the related material safety
data sheet. Wear the lab-gown, goggle and rubber gloves all the time. Be
care of hot reagents.
Warning: Be aware of the risk of electric shock. Only the trained
professionals are permitted to open the face panel or back cover.
I. Introduction
1. Application
The Kjeldahl method is the classical method of nitrogen and protein
determination. Currently, ',67984 Auto Distiller plays important role in
determining the nitrogen and protein contents of soil, food, agricultural
products and feeding stuff etc. ',67984 Auto Distiller can be widely used in
areas of food processing, feeding stuff production, tobacco, farming, soil
fertilizer, environment monitoring, agriculture, scientific research, education,
and quality control and so on, performing either the nitrogen and protein
analysis to macro and semi-micro samples, or tests of ammonium salt and
volatile fatty acid/alkali etc. There are three processes including digestion,
distillation and titration when Kjeldahl method is adopted to test samples. The
distillation is the main determining process used in ',67984 Auto Distiller,
which is an automatic distillation system designed according to the classical
Kjeldahl method. This instrument provides the lab staff with a great
convenience in measuring the content of nitrogen and protein, owning
features such as high reliability, uncompromised safety, hassle-free operation
and time and energy-saving performance etc. The user-friendly interface
written in English is easy to operate and can display the sufficient information,
making users easily understand how to handle the instrument.
2. Theory
According to the Kjeldahl theory, there are three processes including
digestion, distillation and titration in the test. It is able to execute the
distillation process automatically.
- 1 -
HotSteam
The chemical reactions listed below occur after a sample has been
completely digested.
[1] (NH4)2 SO4+2NaOH Na2SO4+2H2O+2NH3

[2] 2NH3+4H3BO3= (NH4)2B4O7+5H2O
Via the condensing pipe, the ammonia gas and steam released during the
reaction are condensed and collected into the receiving flask where the boric
acid (with mixed indicator) is added.
- 2 -
II. Specifications
1. Technical specifications
a. Measurable amount of the sample solid sample < 6g per sample;
liquid sample<16 mL per sample
b. Measurement range0.1~240 mg nitrogen
c. Time for measurement3~6 minutes per sample
d. Recovery rate99.5%;
e. Repetition raterelative error of average value<0.5%
2. Working Condition
a. Rated power1.6kW
b. Input voltageAC 220V±10% 50Hz
Reliable ground wire is needed.
c. Cooling water pressure:>0.15Mpa
d. Cooling water temperature: 20
e. Temperature of environment: +10~28
Working Humidity: 35 and Max. relative humidity 67%; 31 and
Max. relative humidity 80%;
III. Names of Instrument Parts
The instrument is a special system, which can automatically distill the
completely-digested samples and show the workflow in real time mode. The
system mainly consists of microcontroller, steam generator, distillation
system, and automated reagent dispensing system. The exterior structure is
shown in the following diagram:
Figure 1 Front view
1. LCD 2. Connector for liquid-collecting tube
3. Waste receiving tank 4. Protection door 5.Digestion tube
- 3 -
- 4 -
IV. Installation Instruction
1. Pre-installation check
After the instrument is unpacked, check the whole instrument and all marked
spare parts according to the attached packing list and check whether they are
damaged or not. If something is damaged, please contact the manufacturer
timely. (Please keep the damaged parts well).
2. Installation condition:
The instrument should be installed in neither excessively hot, cold or humid
places nor places with excessive sunlight exposure. Commonly, the room
temperature should be maintained between 1028. The instrument
should be installed on the clear, level and stable platform, which should be
close to the water source and draining sink. The distance from the
instrument to the water valve, draining sink and power supply should be
shorter than one meter to ensure the convenient operation. The distance
between the instrument and the wall or other devices should be at least 40
cm. For your safety, do not store anything on the top of the instrument and
put any container, chemical agent or devices etc. behind the instrument.
The pressure and temperature of the water supply should meet the certain
requirement, and the water should be clear and free of impurities (see
Chapter 2 Item 2 for details). The draining sink should be lower than the
discharge outlet of instrument to ensure the good water-discharging
performance.
The power supply should meet the specified requirement, which must have
ground wire, independent power switch and fuse protection to guarantee the
safety of operators.
The instrument should be far away from any big-scale electric equipment,
and its working environment should be free of vibration and electromagnetic
interference, in which there is not corrosive substance.
3. Installation procedures:
According to the view of left side (figure 2):
Figure 2 View of left side
6. Cooling water intake 7. Water discharge outlet (connect to drainage
sink) The water intake6is commonly connected to the faucet of tap
water. The water discharge outlet (7) is connected to drainage sink and
make sure that the discharging performance is good.
According to the view of right side (figure 3):
Figure 3 View of right side
- 5 -
8. Ventilation hole for alkali bottle 9. Ventilation hole for boric acid bottle
10. Ventilation hole for distilled water bottle 11. Intake of alkali solution 12.
Intake of boric acid 13. Intake of distilled water
The ventilation hole for alkali bottle (8) is connected to the air tube of alkali
bottle.
The ventilation hole for boric acid bottle (9) is connected to the air tube of
boric acid bottle.
The ventilation hole for of distilled water bottle (10) is connected to the air
tube of distilled water bottle.
The intake of alkali solution (11) is connected to delivery tube of alkali
solution bottle.
The intake of boric acid solution (12) is connected to delivery tube of boric
acid solution bottle.
The intake of distilled water (13) is connected to delivery tube of distilled
water bottle.
According to the back view (figure 4):
Figure 4 Back view
14. Power switch 15. Power cord socket 16. Fuse device
The power cord can be plugged into the power cord socket (15). The blown
fuse can be replaced in the fuse device (16) (There is a spare fuse in the fuse
socket). Slide the power switch (14) to turn on the instrument.
- 6 -
V. Operation Instruction
1. Preparation of chemical reagent
(1) Add distilled water into the distilled water bottle (with a blue label marked
with “H
2
O”) and then tighten the bottle cap.
(2) Prepare the sodium hydroxide solution (NaOH ) (concentration: 30%~
40%), and add it into the alkali solution bottle( with a yellow label marked
with “NaOH”) and then screw the bottle cap firmly. (The concentration of
35% is recommended because the solution with this concentration is not
subject to the crystallization and does not cause the pipe clogging when
the room temperature fluctuates).
(3) According to the need of experiment, make up Bromocresol Green/Methyl
Red Indicator
(4) Prepare the boric acid (H 3BO 3with the concentration of 2%, and then
add Bromocresol Green/Methyl Red Indicator at the proportion of 100:1.
After they are well mixed, pour the solution into the boric acid bottle(with a
red label marked with “H 3BO ”), and screw the bottle cap tightly at last.
3
2. Operation panel:
Figure 5
- 7 -
- 8 -
1. Easy -touch keypad and keys with the protective film.
2. Environment-friendly LCD with the blue background light which can
display in Chinese or English
3. Heating indicator, which can be on or off according to the distillation
process.
4. Warning indicator, which is on when there is any malfunction in the
system or any necessary factor is missed in the distillation stage.
Attention: in the warning condition, warning words are showed on the
screen when the warning sound goes off at the same time.
5. Power light, which can indicate the state of power supply and is on
when the instrument is turned on.
3Key functions:
(1)Restart】:
Press this key to reset the instrument to its original state when there is
any malfunction or error in the system. After the reset, the interface
showing theAutoand Manual options will be recovered on the
display screen.
(2)Stop】:
For the maximal protection, press this key to stop the instrument
completely when any maloperation or other emergency occurs. After
problems have been taken care, press Backkey to return to the
interface on which the Autoand Manualoptions are displayed,
bringing the instrument back to its normal standby state.
(3)Amend】:
After entering interface in which the auto and manual parameters can
be set , press Amendto modify the current parameters if necessary.
(4)Back】:
Press this key to go back to the previous menu. When the Emergency
Stop has been executed, press Backto return to the mode-selecting
interface.
(5)Enter】:
Press this key to confirm any correction and related operation.
(6)0~9】:
Press these keys to input 0~9 and decimal point.
(7)Auto】:
Under the main mode-selecting interface, press this key to enter the
auto mode.
(8)Manual】:
Under the main mode-selecting interface, press this key to enter the
auto mode.
(9)Debug】:
If a non-quantitative adding is required or any part in the system is needed be
tested, press this key to enter the testing interface. With this key, any
operational part in the system can be turned on or off. On the right side of the
interface, the real-time information about the system auto test will be
displayed.
Attention: Before use, check the solution amount in the boric acid bottle,
alkali solution bottle and distilled water bottle. Refill bottles if necessary to
avoid the miscalculation or system malfunction.
4Start-up:
The instrument is turned on. After two seconds, the mode-selecting interface
will be displayed automatically.
Figure 6
- 9 -
5. Debug
In the mode-selecting interface, pressDebugkey located on the keypad and
the information about “whether the flask is in the position” will be displayed,
and then press “ confirm” to enter the next menu.
Figure 7
In the above interface, press【∧】or 【∨】key to highlight each option one
by one (→【Boric Acid】→【Alkali Solution】→【Distillation】→【Leach
Liquid DrainingDilutionAir PumpCooling WaterDistilled
Water】→)
Function of Each Option
1) Manually start or stop the boric acid filling (Non-quantitatively)
2) Manually start or stop the alkali solution filling (Non-quantitatively)
3)Manually start or stop the distilling operation(Non-quantitatively)
(Attention: the distilling operation is prohibited by the system when any
item in the system condition test that is displayed at the right side of screen
is in unsatisfied condition.)
4) Manually start or stop the leaching operation. (Non-quantitatively)
5) Manually start or stop the liquid draining operation (Non-quantitatively)
6) Manually start or stop dilution operation (Non-quantitatively)
7) Manually start or stop the air pump
8) Manually start or stop the cooling water supply
9) Manually start or stop the distilled water supply
- 10 -
Real-time information about the system condition test
10) Current state of protection door
11) Current state of nitrogen tube
12) Current state of distilled water supply
13) Current state of cooling water supply
press【∧】 or 【∨】 key to highlight each option one by one, and when
the option is highlighted, press Enterkey to change its current state (On
Off or OffOn) . PressBackkey to return to the mode-selecting
interface.
Note:
With the rinse function, the condensing pipes can be cleaned to avoid any
nitrogen loss.
With the discharge function, the distilled water in the heating cup can be
discharged (when the heating cup needs cleaning or the instrument is
moved).
With the dilution function, the digested sample in the nitrogen tube can be
diluted with the distilled water.
6. Operation
In the mode-selecting interface (shown in figure 8), press【∧】 or 【∨】 key
to select the operation mode.
Figure 8
Once the operation mode is selected, pressEnterkey to enter. User also
- 11 -
can press Auto or Manual to enter the corresponding mode.
1Auto ModeThe system first displays the interface of current parameters
showed in figure 9:
Figure 9
Press Amendkey to enter the new parameters to the selected item that
can be chosen by pressing【∧】 or 【∨】 key, where the blinking cursor will
stay at. After the item is selected, the user can press <and>to move
the cursor to the position where the new digit will be entered, and then users
can either press0~9key to type
in the new number .
After the desired parameters are set, the user can press the Enterkey to
save all inputted parameters and system will go to next step. It is shown in
figure 10:
Figure 10
The user can press the Enter again. The current item users are operating
- 12 -
and related data are displayed by the system in a real-time. Once the
operation for one item is done, the system starts the operation for the next
item automatically, following the item sequence displayed on the screen
(from top to bottom).
First the dilution operation is started if the parameter is not 0.
Figure 11
But if its parameter is 0, the operation will be skipped.
Figure 12
After the dilution operation is performed, the system starts the boric acid
filling operation. If the parameter of boric acid filling is set to 0, the system
skips this operation. (Showed in Figure13)
- 13 -
Figure 13
If the parameter is not 0, the system initializes this filling operation. (Showed
in Figure 14)
Figure 14
After the boric acid filling operation is performed, the system starts the alkali
solution filling operation. If the parameter of alkali solution filling is set to 0,
the system skips this operation. (Showed in Figure 15)
- 14 -
Figure 15
If the parameter is not 0, the system initializes this alkali solution filling
operation. (Showed in Figure 16)
Figure 16
The current item users are operating and related data are displayed by the
system in a real-time mode. Once the operation for one item is done, the
system starts the operation for the next item automatically, following the item
sequence displayed on the screen (from top to bottom).
After the alkali solution filling operation is performed, the system starts the
distillation operation. If the parameter of distillation is set to 0, the system
skips this operation. (showed in Figure 17)
- 15 -
Figure 17
If the parameter is not 0, the system initializes this distillation operation.
(showed in Figure 18)
Figure 18
The current item users are operating and related data are displayed by the
system in a real-time mode while conditions of protection door, nitrogen tube,
distilled water and cooling water are being checked by the system. If any
problem exists in any above-mentioned part, the distillation operation will stop
and can not continue until this problem is solved.
After the distillation operation is performed, the system starts the rinse
operation. If the parameter of rinse is set to 0, the system skips this operation.
(showed in Figure 19)
- 16 -
Figure19
If the parameter is not 0, the system initializes this rinse operationshowed in
Figure 20
Figure 20
After operations for all items have been done, “operation ends” will appear on
the screen (showed in Figure 21). After two seconds, the system
automatically goes back to the mode-selecting interface.
- 17 -
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34

MRC DIST-984 User manual

Type
User manual

MRC DIST-984 is an automatic distillation system designed according to the classical Kjeldahl method. This instrument provides lab staff with great convenience in measuring the content of nitrogen and protein, owning features such as high reliability, uncompromised safety, hassle-free operation, and time and energy-saving performance.

With a measurable amount of solid sample of <6g per sample and a measurable amount of liquid sample of <16 mL per sample, this device has a measurement range of 0.1~240 mg nitrogen and a recovery rate of ≥99.5%.