  • Hello! I am an AI chatbot trained to assist you with the Endres+Hauser Cleanfit CPA450 Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
  • What is the intended application of the Cleanfit CPA450 assembly?
    How can I identify the assembly variant?
    What safety precautions should be taken while using the device?
    What safety features does the assembly offer?
    What are the installation requirements for using the assembly in pressurised systems?
BA 183C/07/en/11.99
Nr.50090675
probfit
CPA 450
Retractable Assembly for
pH/Redox Measurement
Hauser+Endress
The Power of Know How
Operating Instructions
Table of contents
1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.1 Symbols used ............................................................. 2
1.2 Storage and transport ....................................................... 2
1.3 Unpacking ................................................................ 2
1.4 Packaging and disposal ..................................................... 2
1.5 Product structure ........................................................... 3
2 Safety ............................................................... 4
2.1 Intended use .............................................................. 4
2.2 General safety notes ........................................................ 4
2.3 Installation, start-up, operation ................................................ 4
2.4 Safety features ............................................................. 5
2.5 Notes for installation in pressurised systems...................................... 5
3 Installation........................................................... 6
3.1 Measuring system .......................................................... 6
3.2 Dimensions and process connections........................................... 6
3.3 Sectional view ............................................................. 7
3.4 Assembly installation ........................................................ 8
4 Maintenance ........................................................ 11
4.1 Cleaning the electrode...................................................... 11
4.2 Calibration ............................................................... 12
4.3 Maintenance and replacement of wearing parts .................................. 13
5 Accessories and spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6 Technicaldata....................................................... 15
7 Index............................................................... 16
PA450E00.CHP
probfit CPA 450 Table of contents
Endress+Hauser 1
1 General information
1.1 Symbols used
Warning:
This symbol alerts to hazards.
Failure to observe these warnings
may result in injury or damage to
equipment.
Note:
This symbol indicates important
items of information. Ignoring this
information may result in malfunction.
1.2 Storage and transport
The packaging material used to store or
transport the assembly must provide shock
protection. Optimal protection is provided by
the original packaging materials.
The ambient conditions also have to meet the
requirements (see Technical data).
1.3 Unpacking
Verify that the packaging and contents are
undamaged! Inform the post office or
freight carrier of any damage. Damaged
merchandise must be retained until the matter
has been settled.
Check that the delivery is complete and
agrees with the shipping documents and your
order (refer to nameplate for type and variant).
The delivery comprises:
Assembly CPA 450
PMC mounting set
Inner pipe for electrode (inner pipe length
depends on electrode type and immersion
depth of assembly)
Operating instructions BA 183C/07/en.
Keep the original packaging materials for
future storage or shipping of the assembly.
If you have any questions, consult your
supplier or the Endress+Hauser sales
agency in your area (see back cover of
these operating instructions for addresses).
1.4 Packaging and disposal
Package the assembly properly for reuse at a
later point in time. Optimal protection is
provided by the original packaging materials.
Observe local regulations for disposal.
1 General information probfit CPA 450
2 Endress+Hauser
1.5 Product structure
You can identify the assembly variant by the
order code on the nameplate.
Immersion depth
1 250 mm (10 inch)
2 700 mm (28 inch)
Process connection
A G 1½ F (without adapter)
B Adapter SS 316S with G 1¼ M
C Adapter SS 316S with 1¼" NPT M
D Adapter with DN32, PN16 flange
E Adapter with ANSI 1¼" flange, 150 lbs
F Ball valve SS 316C with G1¼ F
G Ball valve SS 316S with G1¼ F
H Ball valve SS 316C with 1¼" NPT F
I Ball valve SS 316C with DN32, PN16 flange
K Ball valve SS 316C with ANSI 1¼" flange
Sealing material
1 Sealing material: EPDM
2 Sealing material: Viton
Equipment
10 Basic version
20 Assembly desiliconised, Pg 13.5
30 Certificate 3.1.B acc. to EN 10204
(not for process connections F, H, I, K)
CPA 450- complete order code
Retractable assembly CPA 450
PA450E01.CHP
ENDRESS+HAUSER
CPA 450
Spec.:
PN=4bar T=-15...130°C
Order code: CPA450-1F110
Serial no.:
Fig. 1.1 Nameplate of CPA 450
probfit CPA 450 1 General information
Endress+Hauser 3
2 Safety
2.1 Intended use
The manually operated retractable assembly
CPA 450 is intended for installation of pH/
redox sensors in tanks and pipelines.
Thanks to its special design, this assembly
can be used in pressurised systems (see
Technical data).
It is the operator’s responsibility to assure that
the following safety regulations are observed:
Regulations for explosion protection
Regulations for installation
Operating conditions for the device and its
materials
Local standards and regulations
2.2 General safety notes
The CPA 450 assembly has been designed
for safe operation according to the state of
the art in engineering and according to
current regulations and European standards
(see Technical data).
However, if used improperly or for purposes
other than the intended purpose, it may be
dangerous, e.g. due to incorrect installation
or incorrect operating conditions.
Warning:
Operating this assembly in
any way other than as described
in these instructions may
compromise the safety and
function of the measuring system.
The notes and warnings in these
operating instructions are to be
strictly adhered to!
2.3 Installation, start-up, operation
Warning:
This device may only be installed,
connected electrically,
commissioned, operated and
serviced by properly trained
personnel authorised by the
system operator.
The personnel must be familiar
with these operating instructions
and must adhere to the
instructions contained therein.
When this assembly is used in an
explosive atmosphere, adherence
to the applicable regulations is
mandatory.
Check that all connections have
been properly made before
powering up the system!
Damaged assemblies that may be
dangerous must not be operated
and should be clearly identified as
being defective.
Any troubleshooting of the
measuring system is to be
performed exclusively by
authorized, trained personnel.
If faults cannot be remedied, the
assembly must be removed from
service and secured to prevent
accidental start-up.
Repair work which is not
described in these operating
instructions must be carried out
directly by the manufacturer or by
the Endress+Hauser Service
Organization.
2Safety probfit CPA 450
4 Endress+Hauser
2.4 Safety features
Safety devices
The assembly is protected against external
influences and damage by means of the
following design measures:
media-resistant material
• stopcock
2.5 Notes for installation in pressurised systems
Warning:
The maximum operating pressure
of the assembly must not be
exceeded.
The pressure in the system is to
be released before assembly
installation or removal.
Couplings, cocks and lines are to
be checked for leakage and
damage at regular intervals.
PA450E02.CHP
probfit CPA 450 2 Safety
Endress+Hauser 5
3 Installation
3.1 Measuring system
The complete measuring system consists of:
Retractable assembly Probfit CPA 450
1 gel-filled pH combination electrode,
length 120 mm, with or without integrated
temperatur sensor, e.g. Orbisint CPS 11
pH/redox transmitter, e.g. Mycom CPM
152, Mypro CPM 431,
Liquisys S CPM 223 / 253
Measuring cable CPK 1, CPK 7 or CPK 9
(terminated)
Optional:
Junction box VBA for measuring cable
extension
Measuring cable CYK 71 (not terminated)
for extension
Cleaning device with injector CYR 10
3.2 Dimensions and process connections
PARAM
MEAS CAL
DIAG
Meas. instrument, e.g. Mycom CPM 152
Measuring cable CPK 9
Assembly
CPA 450
MESSEINR.CDR
Fig. 3.1
Complete measuring
system
Power supply
(e.g. 230 V AC / 50 Hz)
Output
(e.g. 0 or 4 ... 20 mA)
Setting collar for fixing the
immersion depth
Allen screw M5
Pg 11
Drill hole for
hook wrench
Rinse connection G ¼
Protection cage
CPA450-6.CDR
Fig. 3.2
Mechanical construction
of assembly CPA 450
3Installation probfit CPA 450
6 Endress+Hauser
3.3 Sectional view
CPA450-3.CDR
Process connection
A*
Process connection
B/C**
Process connection
F/G/H**
Process connection
I/K**
Process connection
D/E**
* Version A requires adapter for installation with ball valve
** Versions B - K: adapter with G 1/4 rinse connection
435 / 885
693 / 1143
Ø25
Ø27
G1½
DN 32, Ø140, 4xM16
ANSI 1¼" , Ø117.5, 4xM12
250 / 700
693 / 1143
DN 32
ANSI 1¼"
380 / 830
693 / 1143
693 / 1143
G 1¼
NPT 1¼"
285 / 735
693 / 1143
89
AF 55
inside
outside
inside
Fig. 3.3
Dimensions of the various
length and connection
variants (moved-in state,
i.e. sensor in process)
PA450-09.CDR
Flat gasket
59 x 50 x 2
O-ring seal
10.69 x 3.53
O-ring seal
24.99 x 3.53
O-ring seal
32.92 x 3.53
O-ring seal
18.72 x 2.62
Contact spring for
potential matching
Inner pipe
Detail A
Detail A enlarged
Fig. 3.4
Sectional view of assembly
CPA 450 showing seals
Variant: adapter with
flange ball valve and
4-pole electrode
(with Pt 100)
G 1¼
NPT 1¼"
380 / 830
PA450E03.CHP
probfit CPA 450 3 Installation
Endress+Hauser 7
3.4 Assembly installation
The CPA 450 assembly should preferably
be installed upright. See the next page for
mounting instructions.
The immersion assembly is to be installed in
the tank or pipeline at such a depth that the
electrode is sufficiently immersed in the
medium at all times.
The figure below shows the assembly in
the moved-in state (operation) and in the
retracted state (for electrode change,
calibration, rinsing).
Warning:
Depressurise the system before
assembly installation or removal.
The medium pressure in the tank
must not exceed the maximum
permissible assembly and
electrode pressures.
The assembly must not be
installed horizontally. The
minimum installation angle from
the horizontal is 15°.
Cable gland Pg 11
Setting collar
Union nut
Clamping ring
G ¼ rinse connection
Inner pipe
PM contact spring
Cable connector
Distance sleeve
Protection guard
Ball valve
Handle
Sleeve
Cable guide
Immersion tube
Pan head screw
M5x16 DIN 912
Flanged sleeve
Lock ring
Adapter
Flange
Hexagon head screw / washer
Electrode
PA450-04.CDR
Fig. 3.5
Overview of assembly
parts
ANSICHT.CDR
Fig. 3.6
Left
Assembly CPA 450 in
moved-in state (ball valve
open)
Assembly CPA 450 in
moved-out state (ball valve
closed)
Right
Permissible installation angle
of CPA 450 assembly
15°15°
PA450-01.CDR
Permissible
range Permissible
range
3Installation probfit CPA 450
8 Endress+Hauser
Instructions for assembly, electrode and measuring cable installation
Note:
Required mounting clearance:
min. 700 / 1150 mm.
Use proper inner pipe for
electrode. The inner pipe
facilitates cable threading
through the cable guide.
Assembly
immersion
depth
Inner pipe length
2-pole electrode 4-pole electrode
250 mm 524 mm 462 mm
700 mm 974 mm 912 mm
Warning:
Make sure that the O-ring seals
are properly seated.
Be sure to remove the pressure
ring before screwing in the
electrode.
1/2
8
1
9
2
10
3 4
5 6
7
Loosen union nut ½ revolution. Extend immersion tube completely.
Close ball valve. Unscrew lock ring with AF 55 or hook wrench.
Remove assembly. Loosen cable gland.
Unscrew protection cage. Thread measuring cable through inner
pipe and assembly from process side.
Remove pressure ring from new electrodes
(s. below). Push greased O-rings + distance sleeve
onto electrode.
O-ring
Distance sleeve
MONTAGE1.CDR
Inner pipe
Connect PM contact spring to contact cable.
Attach with 2 cable clips.
Contact spring
Cable clip
Pressure ringO-Ring
ELEKTROD.CDR Fig. 3.7
Pressure / O-ring
electrode
Contact cable
PA450E03.CHP
probfit CPA 450 3 Installation
Endress+Hauser 9
Mounting instructions (continued)
Warning:
Before you open the ball valve
make sure that the union nut is
only loosened slightly
(1/2 revolution).
For electrodes changing perform
steps 1 to 7 and 10 to 20. Be sure
to always loosen the cable gland
before unscrewing the sensor
protection cage.
1/2
17
11 12
13 14
15 16
18
20
19
Plug electrode head onto cable connector and
hand-tighten. Push electrode head into inner pipe
all the way to stop.
Plug protection cage onto electrode and
hand-tighten.
Tighten cable gland. Tighten completely with
AF 20. Insert tube assembly in ball valve.
Screw on lock ring.
Tighten with AF 55 or hook wrench. Make sure that union nut is only loosened
1/2 revolution.
Adjust setting collar to desired
immersion depth and lock in position.
Push in immersion tube all the way against
setting collar.
MONTAGE2.CDR
Tighten union nut.
Open ball valve.
3Installation probfit CPA 450
10 Endress+Hauser
4 Maintenance
Electrode soiling may impair measurement to
such an extent that the electrode ceases to
function at all, e.g. due to:
coatings on the pH-sensitive part of the
glass electrode these cause poor
response and low sensitivity or slope;
soiling or blocking of diaphragm this
causes poor response and unstable
measurement.
In order to guarantee reliable measurement,
the electrodes must be cleaned regularly. The
frequency and intensity of cleaning will
depend on the type of medium to be
measured.
4.1 Cleaning the electrode
Electrode cleaning is required:
before every calibration;
as necessary at regular intervals during
operation.
Cleaning may be performed manually by
removing the electrode or with the aid of an
injector using the integrated rinse chamber.
Note:
Do not use any abrasive cleaning
materials for the electrodes. These
may cause irreparable damage to
the measuring surfaces.
Rinse the rinse chamber
thoroughly with water (possibly
distilled or deionised) after
cleaning. Cleaning agent residue
may impair measurement.
Calibrate the measuring system
always after cleaning.
Note:
Clean Redox electrodes only
mechanically! Chemical cleaning
cause measured errors because a
potential is implicated. This does not
abate before a few hours.
Manual cleaning
All parts in contact with the medium, such as
the electrode, electrode holder and
immersion tube, require regular cleaning.
Remove slight soiling using a suitable
cleaning agent.
Use a soft brush and suitable cleaning
agent to remove more adhering dirt.
Stubborn dirt is to be dissolved by soaking
in cleaning fluid.
Selection of cleaning agents
The selection of cleaning agents will depend
on the type of soiling. The types of soiling and
the appropriate cleaning agents are listed in
the following table:
Type of soiling Cleaning agent
Fat, oil
Detergents or water-
soluble organic
solvents (e.g. alcohol)
Limestone deposits,
metal hydroxide
coatings,
heavy biol. deposits
manual:
3% HCl
via rinse chamber:
HCl 10% diluted to 3%
in the injector
Sulphide deposits
Mixture of 3% HCl and
thio-urea (usual
commercial)
Protein deposits
Mixture of 0.1 molar
HCl and pepsin (usual
commercial)
Fibres, suspended
materials
Pressurized water,
possibly with wetting
agents
Light biological
deposits Pressurized water
PA450E04.CHP
probfit CPA 450 4 Maintenance
Endress+Hauser 11
Cleaning via the rinse chamber
The integrated rinse chamber permits
cleaning via a hose connection set and
injector CYR 10 (see Accessories).
Instructions for electrode cleaning
4.2 Calibration
Careful and regular calibration is essential
for reliable and accurate measurement. The
calibration cycles depend on the area of
application and the desired accuracy.
The calibration cycles must be individually
determined for each application. More
frequent calibration is recommended at the
beginning, e.g. once a week, to study the
operating behaviour.
1/2
1 2
3 4
5 6
7
Loosen union nut 1/2 revolution. Extend immersion tube completely.
Close ball valve. Rinse electrode via rinse chamber.
Open ball valve. Push in immersion tube all the way
against setting collar.
Tighten union nut.
Rinse connection
MONTAGE3.CDR
4 Maintenance probfit CPA 450
12 Endress+Hauser
Instructions for electrode calibration
Refer to the operating instructions of the measuring transmitter
in question for the calibration procedure to be used.
Mount electrode and assembly in reversed dircetion as at unmounting.
4.3 Maintenance and replacement of wearing parts
The retractable assembly Probfit CPA 450
requires only little maintenance. However,
the following maintenance work is required to
guarantee safe operation:
Replace damaged assembly parts.
Keep O-rings and sealing surfaces free of
dirt.
Grease dry O-rings (particularly the
O-rings on the protection cage).
Inspect O-rings for damage regularly and
replace at appropriate intervals.
Replace adhering coatings from time to
time.
1/2
12
34
Loosen union nut ½ revolution. Extend immersion tube completely.
Remove assembly.
Close ball valve.
Screw off lock ring. Tighten with
AF 55 or hook wrench.
MONTAGE4.CDR
PARAM
MEAS CAL
DIAG
5 6 7 8
910 11
12
PARAM
MEAS CAL
DIAG
Push up lock ring
and union nut. Clean and dry the
electrode. Immerse the electrode
in buffer solution 1. Select required settings
on measuring
instrument.
Rinse the electrode with
distilled water.
Select required settings
on measuring
instrument.
Immerse the electrode
in buffer solution 2.
Rinse the electrode with
distilled water.
MONTAGE5.CDR
PA450E04.CHP
probfit CPA 450 4 Maintenance
Endress+Hauser 13
5 Accessories and spare parts
The following accessories may be ordered
separately:
Spray cleaning injector CYR 10
(see Techn. Information TI 046C/07/en)
Order no. 50014221)
Hose connection set (2 pcs.)
12 mm hose connector for rinse
connection on CPA 450
Order no. 50090491
Welding neck 320S G 1¼ straight
Order no. 51502284
Gel-filled pH combination electrode with or
without integrated temperature sensor,
e.g. Orbisint CPS 11:
Order no. CPS11-2AA2 TSA
(with temperature sensor)
Order no. CPS 11-1AA2 GSA
(without temperature sensor)
Gel-fillled Redox combination electrode
with gold or platinum electrode,
e.g. Orbisint CPS 12:
Order no. CPS12-0NA2 GSA
(with gold electrode)
Order no. CPS12-0PA2 GSA
(with platinum electrode)
Measuring cables for electrodes
e.g. CPK 1 (for electrodes without
temperature sensor)
Order no. CPK1-05/0/A (5 m)
e.g. CPK 7 (for electrodes with
temperature sensor)
Order no. CPK7-05/A (5 m)
e.g. CPK 9 (for electrodes with
temperature sensor and waterproof
TOP 68 connector)
Order no. CPK9-NAA1A (5 m)
Junction box VBA
for extension of measuring cable
between assembly and measuring
instrument
Order no. 50005276
Cable CYK 71 for measuring cable
extension
Order no. 50085333
Hook wrench DIN 1810 design B
D 58 - 62 mm
Order no. 50090687
Calibration solution CPY 2
for pH electrode calibration
Order no. according to pH value
Calibration solution CPY 3
for redox electrode calibration
e.g. 220 mV, pH 7:
Order no. CPY3-0
The following spare parts can be ordered:
PMC mounting set
Order no. 51501551
Protection cage
with EPDM O-rings
Order no. 51501533
Protection cage
with Viton O-rings
Order no. 51501534
O-ring sets:
Sealing material Order no.
EPDM 50090489
Viton 50090490
5 Accessories and spare parts probfit CPA 450
14 Endress+Hauser
6 Technical data
Note:
The operating limits of the entire
system are determined by the
operating limits of the components
used (assembly, sensor, cable,
accessories, etc.).
General specifications Manufacturer Endress+Hauser
Designation Probfit CPA 450
Installation Immersion depth 250 mm, 700 mm
No. of electrode positions in sensor holder 1 mounting position for combination gel electrode
Electrode length 120 mm
Electrode plug-in head Pg 11
Process connection A G 1½ internal thread
B G 1¼ external thread
C 1¼" NPT external thread
F, G G 1¼ internal thread
H 1¼" NPT internal thread
D, I DN 32, PN 16 flange
E, K 1¼" ANSI flange
Flush connection G ¼
Weight Without ball valve 2 kg
With threaded ball valve 5 kg
With flange ball valve 10 kg
Materials in contact with the medium Immersion tube stainless steel 316S
Rinse adapter plugs PVDF
O-rings EPDM / Viton
Ball valve stainless steel 316S or 316C
Ball valve sealings PTFE
Materials not in contact with Inner tube, lock ring stainless steel 304S
the medium Screws stainless steel 316S
Union nut PA66GF
Clamping ring PEEK
Distance sleeve PVDF
Handle PVC
Cable gland Brass
Operating data Operating pressure max. 4 bar
Max. pressure limits 20 bar at 20 °C, 5 bar at 130 °C
Temperature range -15 to +130 °C
Subject to modifications
200
4
8
12
16
-15
20
40 60 80 100 120 140
max. pressure limit
Permissible pressure in bar
Operating pressure
Temperature in °C
PT-DIAGR.CDR
Fig. 6.1
Pressure -temperature
diagram
PA450E06.CHP
probfit CPA 450 6 Technical data
Endress+Hauser 15
7 Index
A
Accessories . . . . . . . . . . . . . . . . 14
Assembly installation . . . . . . . . . . 8-9
C
Cable CYK 71 . . . . . . . . . . . . . . . 14
Cable installation . . . . . . . . . . . . . . 9
Calibration . . . . . . . . . . . . . . . . . 12
Calibration cycles . . . . . . . . . . . . . 12
Calibration solution CPY 2 . . . . . . . . . 14
Cleaning . . . . . . . . . . . . . . . . . . 11
Cleaning agent . . . . . . . . . . . . . . 11
Cleaning cycles . . . . . . . . . . . . . . 11
Cleaning, manual . . . . . . . . . . . . . 11
D
Damage . . . . . . . . . . . . . . . . . . . 2
Delivery, scope of . . . . . . . . . . . . . . 2
Disposal . . . . . . . . . . . . . . . . . . . 2
E
Electrode . . . . . . . . . . . . . . . 6, 9, 14
Electrode installation . . . . . . . . . . . . . 9
Electrode length . . . . . . . . . . . . . . 15
Equipment . . . . . . . . . . . . . . . . . . 3
G
General information . . . . . . . . . . . 2-3
H
Hazards . . . . . . . . . . . . . . . . . . . 2
Hook wrench . . . . . . . . . . . . . . . . 14
Hose connection set . . . . . . . . . . . . 14
I
Immersion depth . . . . . . . . . . . . . 3, 15
Inner pipe . . . . . . . . . . . . . . . . 2, 9
Installation ................6-10
Installation angle . . . . . . . . . . . . . . . 8
Intendeduse.................4
J
Junction box VBA . . . . . . . . . . . . 6, 14
M
Maintenance . . . . . . . . . . . . . . 11-13
Materials . . . . . . . . . . . . . . . . . . 15
Measuring cable CPK 1, CPK 7 . . . . . 6, 14
Measuring cable CYK 71 . . . . . . . . . . 6
Measuring cable extension . . . . . . . 6, 14
Measuring system . . . . . . . . . . . . . . 6
Medium pressure . . . . . . . . . . . . . . 8
Mounting instructions . . . . . . . . . . . 9
N
Nameplate . . . . . . . . . . . . . . . . . 3
O
O-rings . . . . . . . . . . . . . . 7, 9, 13-14
Operating limits . . . . . . . . . . . . . . . 15
Order code . . . . . . . . . . . . . . . . . 3
P
Packaging ................. 2
PM contact spring . . . . . . . . . . . . . 7
PM mounting set . . . . . . . . . . . . . . 14
Pressure . . . . . . . . . . . . . . . . . . 8
Pressure ring . . . . . . . . . . . . . . . . 9
Process connection . . . . . . . . . . . 3, 6
Protection guard . . . . . . . . . . . . . . 14
Q
Questions.................. 2
R
Regulations . . . . . . . . . . . . . . . . . 4
Regulations for installation . . . . . . . . . 4
Rinse chamber . . . . . . . . . . . . . . . 12
Rinse connection . . . . . . . . . . . . . . 8
S
Safety................... 4-5
Sealing material . . . . . . . . . . . . 3, 14
Seals .................... 7
Sectionalview ............... 7
Soiling ...................11
Spare parts . . . . . . . . . . . . . . . . . 14
Spray cleaning injector CYR 10 . . 6, 12, 14
Storage . . . . . . . . . . . . . . . . . . . 2
Symbols . . . . . . . . . . . . . . . . . . 2
T
Technicaldata ...............15
Temperature sensor . . . . . . . . . . . . 6
Transmitter . . . . . . . . . . . . . . . . . 6
Transport . . . . . . . . . . . . . . . . . . 2
U
Unpacking . . . . . . . . . . . . . . . . . 2
W
Wearing parts . . . . . . . . . . . . . . . 13
Weight ...................15
Index probfit CPA 450
16 E+H Conducta
Europe
Austria
Endress+Hauser Ges.m.b.H.
Wien
Tel.++43 (1) 88056-0, Fax (1) 88056-35
Belarus
Belorgsintez
Minsk
Tel.++375 (172) 263166, Fax (172) 263111
Belgium / Luxembourg
Endress+Hauser S.A./N.V.
Brussels
Tel.++32 (2) 2480600, Fax (2) 2480553
Bulgaria
INTERTECH-AUTOMATION
Sofia
Tel.++359 (2) 664869, Fax (2) 9631389
Croatia
Endress+Hauser GmbH+Co.
Zagreb
Tel.++385 (1) 6637785, Fax (1) 6637823
Cyprus
I+G Electrical Services Co.Ltd.
Nicosia
Tel.++357 (2) 484788, Fax (2) 484690
Czech Republic
Endress+Hauser GmbH+Co.
Praha
Tel.++420 (26) 6784200, Fax (26) 6784179
Denmark
Endress+Hauser A/S
borg
Tel.++45 (70) 131132, Fax (70) 132133
Estonia
Elvi-Aqua
Tartu
Tel.++372 (7) 422726, Fax (7) 422727
Finland
Endress+Hauser Oy
Espoo
Tel.++358 (9) 8596155, Fax (9) 8596055
France
Endress+Hauser
Huningue
Tel.++33 (3) 89696768, Fax (3) 89694802
Germany
Endress+Hauser Meßtechnik GmbH+Co.
Weil am Rhein
Tel.++49 (7621) 97501, Fax (7621) 975555
Great Britain
Endress+Hauser Ltd.
Manchester
Tel.++44 (161) 2865000,
Fax(161)9981841
Greece
I & G BuildingServices Automation S.A.
Athens
Tel.++30 (1) 9241500, Fax (1) 9221714
Hungary
Mile Ipari-Elektro
Budapest
Tel.++36 (1) 2615535, Fax (1) 2615535
Iceland
Vatnshreinsun HF
Reykjavik
Tel.++354 (5) 619616, Fax (5) 619617
Ireland
Flomeaco Company Ltd.
Kildare
Tel.++353 (45) 868615, Fax (45) 868182
Italy
Endress+Hauser Italia S.p.A.
Cernusco s/N Milano
Tel.++39 (02) 92106421,
Fax (02) 92107153
Latvia
Raita Ltd.
Riga
Tel.++371 (7) 312897, Fax (7) 312894
Lithuania
Agava Ltd.
Kaunas
Tel.++370 (7) 202410, Fax (7) 207414
Netherlands
Endress+Hauser B.V.
Naarden
Tel.++31 (35) 6958611, Fax (35) 6958825
Norway
Endress+Hauser A/S
Tranby
Tel.++47 (32) 859850, Fax (32) 859851
Poland
Endress+Hauser Polska Sp.z o.o.
Warszawy
Tel.++48 (22) 7201090, Fax (22) 7201085
Portugal
Tecnisis - Tecnica de Sistemas Industriais
Linda-a-Velha
Tel.++351 (1) 4172637, Fax (1) 4185278
Romania
Romconseng SRL
Bucharest
Tel.++40 (1) 4101634, Fax (1) 4101634
Russia
Endress+Hauser Moscow Office
Moscow
Tel.++709 (5) 1587571, Fax (5) 1589864
Slovak Republic
Transcom Technik s.r.o.
Bratislava
Tel.++421 (74) 4888684, Fax (74) 4887112
Slovenia
Endress+Hauser D.O.O.
Ljubljana
Tel.++386 (61) 1592217, Fax (61) 1592298
Spain
Endress+Hauser S.A.
Barcelona
Tel.++34 (93) 4803366, Fax (93) 4733839
Sweden
Endress+Hauser AB
Sollentuna
Tel.++46 (8) 55511600, Fax (8) 55511600
Switzerland
Endress+Hauser AG
Reinach/BL 1
Tel.++41 (61) 7157575, Fax (61) 7111650
Turkey
Intek Endüstriyel Ölcü ve Kontrol Sistemleri
Istanbul
Tel.++90 (212) 2751355,
Fax(212)2662775
Ukraine
Industria Ukraïna
Kiev
Tel.++380 (44) 26881, Fax (44) 26908
Yugoslavia
Meris d.o.o.
Beograd
Tel.++381 (11) 4446164, Fax (11) 4441966
Africa
Egypt
Anasia
Heliopolis/Cairo
Tel.++20 (2) 417900, Fax (2) 417900
Morocco
Oussama S.A.
Casablanca
Tel.++212 (2) 241338, Fax (2) 402657
Nigeria
J F Technical Invest.Nig. Ltd.
Lagos
Tel.++234 (1) 62234546, Fax (1) 62234548
South Africa
Endress+Hauser Pty. Ltd.
Sandton
Tel.++27 (11) 4441386, Fax (11) 4441977
Tunisia
Controle, Maintenance et Regulation
Tunis
Tel.++216 (1) 793077, Fax (1) 788595
America
Argentina
Endress+Hauser Argentina S.A.
Buenos Aires
Tel.++54 (1) 145227970,
Fax(1)145227909
Bolivia
Tritec S.R.L.
Cochabamba
Tel.++591 (42) 56993, Fax (42) 50981
Brazil
Samson Endress+Hauser Ltda.
Sao Paulo
Tel.++55 (11) 50313455,
Fax (11) 50313067
Canada
Endress+Hauser Ltd.
Burlington, Ontario
Tel.++1 (905) 6819292,
Fax(905)6819444
Chile
DIN Instrumentos Ltda.
Santiago
Tel.++56 (2) 2050100, Fax (2) 2258139
Colombia
Colsein Ltd.
Bogota D.C.
Tel.++57 (1) 2367659, Fax (1) 6107868
Costa Rica
EURO-TEC S.A.
San Jose
Tel.++506 (2) 961542, Fax (2) 961542
Ecuador
Insetec Cia.Ltda.
Quito
Tel.++593 (2) 269148, Fax (02) 461833
Guatemala
ACISA Automatizacion Y Control Industrial S.A.
Ciudad de Guatemala, C.A.
Tel.++502 (3) 345985, Fax (2) 327431
Mexico
Endress+Hauser I.I.
Mexico City
Tel.++52 (5) 568965, Fax (5) 568418
Paraguay
Incoel S.R.L.
Asuncion
Tel.++595 (21) 213989, Fax (21) 226583
Uruguay
Circular S.A.
Montevideo
Tel.++598 (2) 925785, Fax (2) 929151
USA
Endress+Hauser Inc.
Greenwood, Indiana
Tel.++1 (317) 5357138,
Fax(317)5358489
Venezuela
H.Z.Instrumentos C.A.
Caracas
Tel.++58 (2) 9440966, Fax (2) 9444554
Asia
China
Endress+Hauser Shanghai
Instrumentation Co. Ltd.
Shanghai
Tel.++86 (21) 54902300,
Fax (21) 54902303
Endress+Hauser Beijing Office
Beijing
Tel.++86 (10) 68344058,
Fax (10) 68344068
Endress+Hauser (H.K.) Ltd.
Hong Kong
Tel.++852 (2) 5283120, Fax (2) 8654171
India
Endress+Hauser India Branch Office
Mumbai
Tel.++91 (22) 8521458, Fax (22) 8521927
Indonesia
PT Grama Bazita
Jakarta
Tel.++62 (21) 7975083, Fax (21) 7975089
Japan
Sakura Endress Co., Ltd.
Tokyo
Tel.++81 (422) 540611, Fax (422) 550275
Malaysia
Endress+Hauser (M) Sdn.Bhd.
Petaling Jaya, Selangor Darul Ehsan
Tel.++60 (3) 7334848, Fax (3) 7338800
Pakistan
Speedy Automation
Karachi
Tel.++92 (21) 7722953, Fax (21) 7736884
Papua New Guinea
SBS Electrical Pty Limited
Port Moresby
Tel.++675 (3) 251188, Fax (3) 259556
Philippines
Brenton Industries Inc.
Makati Metro Manila
Tel.++63 (2) 6388041, Fax (2) 6388042
Singapore
Endress+Hauser (S.E.A.) Pte., Ltd.
Singapore
Tel.++65 (5) 668222, Fax (2) 666848
South Korea
Endress+Hauser (Korea) Co., Ltd.
Seoul
Tel.++82 (2) 6587200, Fax (2) 6592838
Taiwan
Kingjarl Corporation
Taipei R.O.C.
Tel.++886 (2) 27183938, Fax (2) 27134190
Thailand
Endress+Hauser Ltd.
Bangkok
Tel.++66 (2) 996781120, Fax (2) 9967810
Vietnam
Tan Viet Bao Co.Ltd.
Ho Chi Minh City
Tel.++84 (8) 8335225, Fax (8) 8335227
Iran
Telephone Technical Services Co.Ltd.
Tehran
Tel.++98 (21) 8746750, Fax (21) 8737295
Israel
Instrumetrics Industrial ControlLtd.
Tel-Aviv
Tel.++972 (3) 6480205, Fax (3) 6471992
Jordan
A.P.Parpas Engineering S.A.
Amman
Tel.++962 (6) 4643246, Fax (6) 4645707
Kingdom of Saudi Arabia
Anasia
Jeddah
Tel.++966 (2) 6710014, Fax (2) 6725929
Kuwait
Kuwait Maritime & Mercantile Co.K.S.C.
Safat
Tel.++965 (2) 441481, Fax (2) 441486
Lebanon
Nabil Ibrahim
Jbeil
Tel.++961 (3) 254052, Fax (9) 548038
Sultanate of Oman
Mustafa & Jawad Science & Industry Co.
L.L.C.
Ruwi
Tel.++968 (60) 2009, Fax (60) 7066
United Arab Emirates
Descon Trading EST.
Dubai
Tel.++971 (4) 653651, Fax (4) 653264
Yemen
Yemen Company for Ghee and Soap Industry
Taiz
Tel.++976 (4) 230664, Fax (4) 212338
Australia + New Zealand
Australia
ALSTOM Australia Ltd.
Sydney
Tel.++61 (2) 97224777, Fax (2) 97224888
New Zealand
EMC Industrial Instrumentation
Auckland
Tel.++64 (9) 4155110, Fax (9) 4155115
All other countries
Endress+Hauser GmbH+Co.
Instruments International
D-Weil am Rhein
Germany
Tel.++49 (7621) 97502, Fax (7621) 975345
BA 183C/07/en/11.99
Printed in Germany / CV5 / DT