Leuze ODS 96 M/S Laser Technical And Software Description

Category
Measuring & layout tools
Type
Technical And Software Description

This manual is also suitable for

Leuzeelectronic
Optical Distance Sensors ODS 96
Technical Description
Software Description
© All rights reserved, in particular the rights of reproduction and translation. Duplication or reproduction in any form
(print, photocopies, microfilm or data) may only be carried out with the expressed written consent of Leuze elec-
tronic GmbH & Co.
We reserve the right to make changes for the technical improvement of the product.
Leuze electronic Contents
Leuze electronic ODS 96 3
1 General Information .......................................................................... 5
1.1 Explanation of Symbols ......................................................................5
1.2 Important Terms .................................................................................. 5
1.3 Declaration of Conformity ................................................................... 6
2 Safety Notices ................................................................................... 6
2.1 Safety Standard .................................................................................. 6
2.2 Intended Use .......................................................................................6
2.3 Working Safely .................................................................................... 7
2.4 Organizing Measures .......................................................................... 7
3 Description ........................................................................................ 8
3.1 General description ............................................................................. 8
3.2 Typical Areas of Application for the ODS 96 ....................................... 8
3.2.1 Continuous Distance Measuring .................................................... 8
3.2.2 Positioning Tasks ...........................................................................9
3.2.3 Filling Level Control for Bulk Goods ............................................... 9
3.3 ODS 96 Models ................................................................................. 10
3.3.1 ODS 96 M/V with Analog Output ................................................. 10
3.3.2 ODS 96 M/D with Serial Output ................................................... 11
3.3.3 ODS 96 M/S with Two Switching Outputs ....................................13
4 Technical Data ................................................................................. 14
4.1 Optical Data ...................................................................................... 14
4.2 LED Indicators .................................................................................. 14
4.3 Electrical Data, Installation Data ....................................................... 15
4.4 Dimensioned and Connection Drawings ........................................... 16
4.5 Accessories ....................................................................................... 17
5 Installation ....................................................................................... 18
5.1 Storage, Transportation ....................................................................18
5.2 Mounting ........................................................................................... 18
5.3 Teach-In ............................................................................................ 19
6 Software ........................................................................................... 20
6.1 Connection of the PC ........................................................................ 20
6.2 Installation of the Programming Software ......................................... 20
6.3 Starting the Program ......................................................................... 21
6.3.1 Description of the Menu Commands ............................................22
6.3.2 Programming and Adjustment Possibilities .................................. 22
7 Appendix .......................................................................................... 27
7.1 How to update the parameter files of the ODS 96 software .............. 27
Contents
4 ODS 96 Leuze electronic
Figures and Tables Leuze electronic
Figure 3.1: Application example "Positioning Tasks" ............................................ 9
Figure 3.2: Application example "Filling Level Control" ....................................... 10
Figure 3.3: Behavior of the analog output, ODS 96 M/V (LED) .......................... 10
Figure 3.4: Behavior of the analog output, ODS 96 M/V (laser) .......................... 11
Figure 3.5: Serial output, ODS 96 M/D ............................................................... 11
Figure 3.6: Behavior of the switching outputs, ODS 96 M/S ............................... 13
Figure 4.1: Dimensioned drawing ODS 96 devices ............................................ 16
Figure 4.2: Electrical connection, ODS 96 M/V ................................................... 16
Figure 4.3: Electrical connection, ODS 96 M/D ................................................... 17
Figure 4.4: Electrical connection, ODS 96 M/S ................................................... 17
Figure 5.1: Preferred movement of the objects ................................................... 18
Figure 5.2: Preferred mounting, objects with structured surface ......................... 19
Figure 5.3: View through a chase ....................................................................... 19
Figure 6.1: Installation directory .......................................................................... 21
Figure 6.2: Device selection ................................................................................ 21
Figure 6.3: Start menu before measurement ...................................................... 22
Figure 6.4: Parameterization mode, ODS 96 with analog output ........................ 23
Figure 6.5: Programming window ODS 96 with two switching outputs ............... 24
Table 1: Optical data, Error limits, Timing ....................................................... 14
Table 2: LED indicators .................................................................................. 14
Table 3: Electrical, Mechanical, and Environmental data ............................... 15
Table 4: Accessories ...................................................................................... 17
Table 5: Regulation values of the sensors ...................................................... 24
Figures and Tables
Leuze electronic General Information
Leuze electronic ODS 96 5
1 General Information
1.1 Explanation of Symbols
The symbols used in this operating manual are explained below.
Attention
This symbol appears in front of text which must be carefully observed. Failure
to heed this information can lead to injuries to personnel or damage to the
equipment.
Attention Laser Radiation
This symbol warns of possible danger through hazardous laser radiation.
Notice
This symbol indicates text which contains important information.
1.2 Important Terms
Triangulation
Distance measuring procedure, which determines the distance of an object
by the incidence angle of the light reflected from the object.
Absolute measurement
accuracy
Shows the possible divergence of the measurement value from the anticipat-
ed value through changes in the environmental conditions during the mea-
suring process. Higher accuracy is given at constant environmental condi-
tions.
Repeatability
Measuring distance change with repeated measurement at the same output
signal (observe the same peripheral conditions as with resolution).
Resolution
The smallest possible distance change of the measured object, which causes
a definite change in the output signal. Resolution is higher in the proximate
area than in the distant area. Small objects can be recognized better in the
proximate area.
Remission
Return and/or degree of reflection of the radiated light.
Integration time
The integration time of the ODS 96 is comparable to the exposure time in a
camera. It is automatically adjusted to the intensity of the reflected light and
thus depends on the reflectance factor of the measured object. It is inversely
proportional to the measurement frequency.
Measurement
frequency
The measurement frequency represents the number of measurements per
second. When adjusting the integration time to the reflectivity independency,
the measurement frequency changes corresponding to the reflectivity value.
Response time
The time period required by the ODS to obtain stable measurements after
change of the reflectivity behavior.
Delay before
start-up
The delay before start-up indicates the point in time when the first valid mea-
surement can be obtained after switching on.
Safety Notices Leuze electronic
6 ODS 96 Leuze electronic
Light switching/
Dark switching
Indicates the behavior of the switching output when an object is inside the
tought/parametered switching distance. At light switching, the switching out-
put is active (high), at dark switching inactive.
Insensitivity towards
extraneous light
Indicates the insensitivity of the measurement result towards extraneous
light. The ODS 96 is reliably measuring even with extraneous light intensity
of 5 kLux. Typical light intensity in a work place is only 1 kLux.
1.3 Declaration of Conformity
The optical distance sensors of the series ODS 96 have been manufactured
observing current European standards and guidelines.
Notice
The corresponding declaration of conformity can be requested from the
manufacturer.
The manufacturer of the product, Leuze electronic GmbH & Co. in D-73277
Owen/Teck, possesses a certified quality assurance system in accordance
with ISO 9001.
2 Safety Notices
2.1 Safety Standard
The optical distance sensors of the series ODS 96 have been developped,
manufactured and tested, observing current safety standards. They corre-
spond to the state of the art.
2.2 Intended Use
Attention
The protection of personnel and the device cannot be guaranteed if the de-
vice is operated in a manner not corresponding to its intended use.
Optical distance sensors of the series ODS 96 are intelligent, adjustable sen-
sors with CCD element for distance measuring.
In particular, unauthorized uses include:
rooms with explosive atmospheres (zone 0, 1). Operation in zone 2 is
possible with a manufacturer's declaration.
operation for medical purposes
Leuze electronic Safety Notices
Leuze electronic ODS 96 7
Fields of application
The optical distance sensors of the series ODS 96 have been designed for
the following areas of application:
Distance measurement
Contour determination
Stack positioning
Filling level measurement
Packet conveying machines and many more
2.3 Working Safely
Attention Laser Radiation
The ODS 96 M/D, V, S ... are laser devices of the Laser Protection Class 2.
Do not look directly into the laser beam.
Observe the applicable legal and local regulations for the operation of laser
units.
Attention
Access to or making changes to the device, except where positively de-
scribed in this operating manual, is not authorized.
2.4 Organizing Measures
Documentation
All entries in this operating manual must be heeded, in particular those in sec-
tion 2 "Safety Notices".
Carefully store this operating manual where it is accessible at all times.
Safety regulations
Observe the locally applicable legal regulations and the rules of the employ-
er's liability insurance association.
Qualified personnel
Mounting, commissioning and maintenance of the device may only be carried
out by qualified personnel.
Electrical work must be carried out by a certified electrician.
Repair
Repairs must only be carried out by the manufacturer or an authorized rep-
resentative.
Description Leuze electronic
8 ODS 96 Leuze electronic
3 Description
3.1 General description
The ODS 96 is a distance measuring device with a large area of application.
The equipment is available as LED or laser version with analog or digital out-
puts. The distance measuring device works on the triangulation principle and
uses a CCD line for evaluating the measurement signal.
An integrated microcontroller enables parameter programming via an option-
al programming software. Except for the RS 485 interface, the switching
points of the switching outputs of all other types can be set without software
via a teach-in input.
Through adjustment of the integration time (exposure time) to the intensity of
the objects’ reflected light, a high degree of independence from the reflectivity
properties of the measured object is achieved. In case of low reflectivity (dark
objects) a lower measurement frequency results.
Accessories
To expand the functionality of the ODS 96, a programming software is avail-
able.
With regard to their dimensions, the ODS 96 distance sensors are identical
with the sensors of the series 96 of Leuze electronic. Particularly, the mount-
ing accessories of the series 96 can be used for the ODS 96. Details can be
found in section 4 "Technical Data".
3.2 Typical Areas of Application for the ODS 96
3.2.1 Continuous Distance Measuring
All ODS 96 types with analog or digital output can be used for continuous dis-
tance measuring. In order to use all features of the ODS 96, use of the pro-
gramming software is recommended.
Depending on position or settings of the ODS, various applications are pos-
sible:
Measuring the thickness of planks with two opposing sensors and a dif-
ferential of the two measured values.
Stack/Object height measuring
Even with difficult surfaces, the stack/ object height of moving objects
can be measured. Averaging is recommended here.
Contour determination through controlled passing movement of an
object through the beam of the ODS 96.
Volume measuring by taking measurements on two levels during the
concurrent movement of the object. Leuze electronic offers a system
solution: the VMS volume measurement system.
In case of ODS 96 types with analog outputs, it is recommended to limit the
working range of the analog output to the required distance range. The ana-
log output will then be activated within the distance range from 1 ... 10 V or
4 - 20 mA. Distances outside this range will automatically have an output
voltage of < 1 V, 4 mA or > 10 V, 20 mA.
Leuze electronic Description
Leuze electronic ODS 96 9
3.2.2 Positioning Tasks
For simple positioning tasks, all ODS 96 M/S types with two teachable
switching outputs are suitable.
The ODS 96 is mounted in a way to enable positioning in the direction of the
measuring beam. Both switching outputs are tought to the required position.
In case the distance is smaller than the tought position, switching output 1 is
active and switches the positioning motor to enlarge the distance. In case the
distance is bigger than the tought position, switching output 2 is active and
switches the positioning motor to reduce the distance.
Using this method, positioning tasks can be performed without difficulties.
Figure 3.1: Application example "Positioning Tasks"
3.2.3 Filling Level Control for Bulk Goods
For the filling level control in connection with control of material flow, all ODS
96 types with one switching output are suitable. However, the programming
software is required to adjust the hysteresis.
The ODS 96 is mounted ensuring that its measuring beam meets the surface
of the bulk good vertically. Using the software, function of Q1 is set to "light
switching" and the minimal filling level entered as lower limit. The upper limit
is set to the maximum range of the ODS (600 mm or 2000 mm, depending on
the ODS type). The distance between minimum and maximum filling level is
entered as hysteresis.
Old Stack
New Stack
Positioning
Description Leuze electronic
10 ODS 96 Leuze electronic
Material flow is switched on/off by the ODS as soon as the filling level falls
below the lower limit or reaches the maximum.
Figure 3.2: Application example "Filling Level Control"
Notice
For mounting instructions please refer to section 5.2 "Mounting".
3.3 ODS 96 Models
Types
Two different types of the ODS 96 are available:
Infrared distance sensor
Measurement range 100 - 600 mm
Laser distance sensor with visible red light
Measurement range 200 - 2000 mm
3.3.1 ODS 96 M/V with Analog Output
Analog output of the
LED device
Figure 3.3: Behavior of the analog output, ODS 96 M/V (LED)
I (mA) U (V)
app.21 app.10,5
20 10
4 1
1 0
0 90 100 600 app.625 app.800 - 1500
Meas. distance
(mm)
undefined
Linearity
undefined
Measurement range
Linearity
undefined
Object
present
Range
no object
detected
Leuze electronic Description
Leuze electronic ODS 96 11
Analog output of the
laser device
Figure 3.4: Behavior of the analog output, ODS 96 M/V (laser)
Behavior of the output,
ODS 96 M/V
The ODS 96 M/V has an analog output with linear behavior inside the respec-
tive measurement range. The user can choose between current output
(4 - 20 mA) and voltage output (1 - 10 V). Above and below the linear range,
linearity is lost however, the output values signify an upper deviation (>
20 mA respectively > 10 V) or a lower deviation (< 4 mA respectively < 1 V)
of the measurement range.
Additionally, a switching output is available with the ODS 96 M/V. The posi-
tion within the measuring range, at which the switching output is active can
be set as needed via a teach-in lead. The breadth of the active range is
± 2 mm (LED) respectively ± 10 mm (laser).
Using the optional programming software, the declivity of the output charac-
teristic curve can be changed (steep progression with concurrent reduction
of the measurement range). Furthermore, the switching behavior of the addi-
tional switching output can be individually set.
3.3.2 ODS 96 M/D with Serial Output
Serial output of the LED
and laser device
Figure 3.5: Serial output, ODS 96 M/D
Serial output of the
ODS 96 M/D
The serial output of the ODS 96 M/D delivers a continuous data stream. The
measurement value is transmitted with two byte. The LSB signifies the High-
and the Low-Byte, to ensure recognition of definite coherence. For transmis-
sion, no telegram frame, but only the normal serial protocol is used. The
transmission protocol consists of 8 databit, 1 startbit, 1 stopbit, no parity. The
I (mA) U (V)
app.21 app.10,5
20 10
4 1
1 0
0 180 200 2000 2100 app.3000
Meas. distance
(mm)
undefined
Linearity
undefined
Measurement range
Linearity
undefined
Object
present
Range
no object
detected
Measurement value = 14 Bit
1. Low-Byte (Bit 0 = 0) 2. High-Byte (Bit 0 = 1)
LSB MSB
Description Leuze electronic
12 ODS 96 Leuze electronic
ODS 96 M/D has a digital output which is realized either through the RS 232
or the RS 485 interface. The transmission rate is 9600 baud at the RS 232
interface and 9600 baud without termination at the RS 485 interface.
A "C" program example demonstrates how bytes are read and processed for
further usage:
Program example in "C"
//Start of the program for measurement value generation-------------------------------
Rxbyte = inportb(RXB(COM2.port_adr)); //Read transmitted byte
if (flag==0) //First byte has to be a
//Low-Byte
{
if ((Rxbyte & 0x01) == 0) //Test on Low-Byte
{
value = (Rxbyte & 0xFE) >> 1; //Insert measurement value
//point-aligned
flag = 1; //Next received
//byte has to be a
//High-Byte
}
}
else
{
if ((Rxbyte & 0x01) == 1) //Test on High-Byte
{
value|=((Rxbyte&0xFE) << 6); //Insert point-aligned in
//measurement value
flag = 0; //14 Bit measurement value
//is generated
measurement value = value; //Save measurement value
}
}
// End of program of measurement value generation--------------------------------------
The behavior of the switching output of the ODS M/D can be set as described
with the ODS 96 M/V. However, the ODS 96 M/D type with RS 485 interface
does not offer a teach input.
Leuze electronic Description
Leuze electronic ODS 96 13
3.3.3 ODS 96 M/S with Two Switching Outputs
Switching outputs of
the LED and laser
device
Figure 3.6: Behavior of the switching outputs, ODS 96 M/S
ODS 96 M/S with two
switching outputs
The two switching outputs of the ODS 96 M/S are working independently
from each other. Through the teach input, the second signal edge can be
tought at output 1, the first signal edge is situated at the beginning of the mea-
surement range. The first signal edge can be tought at output 2, the second
signal edge is situated at the end of the measurement range.
A teach line is available for both switching outputs, meaning the switching
outputs are tought alternatingly. The presently tought output is displayed
through the simultaneous or alternating flashing of the LEDs.
The outputs can be adjusted independently of each other by using the option-
al programming software.
+UB
2. switching output
1. switching output
LED: 100 600
Laser: (200) (2000)
Measuring distance
(mm)
Technical Data Leuze electronic
14 ODS 96 Leuze electronic
4 Technical Data
4.1 Optical Data
4.2 LED Indicators
ODS 96
M/V, M/D, M/S ODS 96
M/V, M/D, M/S Laser
Optical data
Measurement range 1
1.Reflectance factor 6%...90%, over the whole temperature range, measured object
50x50 mm
2
100-600 mm 200-2000 mm
Resolution 0,5 mm 5 mm
Light source LED (modulated light) Laser (modulated light)
Wavelength 880 nm (infrared) 660 nm (visible redlight)
Light-spot diameter approx. 10 mm (over whole
measuring range) divergent min. 3x12 mm in
2000 mm distance
Error limits
Absolute measure-
ment accuracy 1± 2% ± 2%
Repeatability 2
2.Same object, measured object
50x50 mm
2
± 0,5% ± 0,5%
b/w detection thresh-
olds (6%/90%) 1% 1%
Timing
Switching frequency 20-100 Hz 10-100 Hz
Response time 100 ms
Delay before start-up 300 ms
Table 1: Optical data, Error limits, Timing
LED ODS 96 M/V, M/V Laser;
ODS 96 M/D, M/D Laser ODS M/S, M/S Laser
teach-in on GND teach-in on +UBteach-in on GND teach-in on +UB
permanent
green light Device Ready Device Ready
flashing green Error Teach-in proce-
dure1
1.The teach-in process is described in detail in section 5.3 "Teach-In"
Error Teach-in
procedure1
green off no voltage no voltage
yellow perma-
nent light
Object inside
teach-in measure-
ment distance
Object inside
measurement
range
yellow flash-
ing Teach-in
procedure1
Object outside
measurement
range
Teach-in
procedure1
yellow off Object outside
teach-in measure-
ment distance
no object
detected
Table 2: LED indicators
Leuze electronic Technical Data
Leuze electronic ODS 96 15
4.3 Electrical Data, Installation Data
M/V; M/V Laser M/D; M/D Laser M/S; M/S Laser
Electrical data
Operating voltage UB 18...30 V DC (incl.
residual ripple) 10...30 V DC (incl. residual ripple)
Residual ripple 15% of UB
Bias current 150 mA
Switching outputs 1
1.
Inversion possible through programming software
PNP transistor, high-active 2 PNP transistor out-
puts, high active
Signal voltage high/low (UB - 2V) / 2V
Analog output
RL 2 kOhm
(Voltage)
1…10 V
RL 500 Ohm
(Current)
4…20 mA
Output current max. 100 mA per
transistor output
Digital output RS 232 9600 Baud
RS 485 9600 Baud, no ter-
mination
Transmission protocol 2 Byte transm.,
cont. data stream
Mechanical data
Housing Diecast zinc
Optical cover Glass
Weight 380g
Connection type Terminals or M 12 dia. connectors
Environmental data
Ambient temp. (opera-
tion./storage) -20°C..+50°C / -30°C..+70°C
Extraneous light limit 5 kLux
Protective circuit 2
2.
1=transient protection, 2= polarity reversal protection, 3= short circuit protection for all outputs
1,2,3
VDE safety class 3
3.
Rating voltage 250 V AC
II, all-insulated
Protection class IP 67
Impact resistance Semi-sinusoidal, 30 gn, 11 ms (VDE 0660T208)
Vibration resistance 10-55 Hz, max. 7,5 gn (VDE 0660T208)
Electromagnetic compatibil-
ity Severity level 3 (IEC 60947-5-2)
Table 3: Electrical, Mechanical, and Environmental data
Technical Data Leuze electronic
16 ODS 96 Leuze electronic
4.4 Dimensioned and Connection Drawings
All ODS 96 devices
Figure 4.1: Dimensioned drawing ODS 96 devices
ODS96 M/V
Figure 4.2: Electrical connection, ODS 96 M/V
green yellow
Indicator LEDs
optical axis
optical axis
Device Plug
Cable Screwing PG 11 for dia. 5-10 mm
Parameter Plug
Connection Terminals
Cable lead-in
Countersinking for
SK nut M5, 4.2 deep
Leuze electronic Technical Data
Leuze electronic ODS 96 17
ODS 96 M/D
Figure 4.3: Electrical connection, ODS 96 M/D
ODS 96 M/S
Figure 4.4: Electrical connection, ODS 96 M/S
4.5 Accessories
The following accessories are available for the ODS 96:
Designation Order No. Short Description
KB-ODS96-1500 50082007 Parameterizing cable 1,5 m
KB-ODS96-6000 50061428 Parameterizing cable 6 m
ODS96-PS 50082006 Programming software
KB-095-5000-5 50020500 Connection lead (M12, angled, 5 m)
KB-095-5000-5A 50020499 Connection lead (M12, axial, 5 m)
BT 96 50025570 Mounting part
UMS 96 50026204 Universal mounting system
Table 4: Accessories
Installation Leuze electronic
18 ODS 96 Leuze electronic
5 Installation
5.1 Storage, Transportation
Unpacking
©
Check the packaging for any damage. If damage is found, notify the post
office or shipping agent as well as the supplier.
©
Check the delivery contents using your order and the delivery papers:
delivered quantity
device type and model as indicated on the name-plate
accessories
operating manual
©
Save the original packaging for later storage or shipping.
If you have any questions concerning your shipment, please contact your
supplier or your local Leuze electronic sales office.
©
Observe the applicable local regulations when disposing of the packaging
materials.
5.2 Mounting
Mounting systems are available which have to be ordered separately at
Leuze electronic. For order numbers please refer to section 4.5 "Accesso-
ries". Apart from this, the through-borings can be used for individual mounting
of the ODS 96, depending on the area of application.
Mounting of the ODS 96
To avoid errors while the object enters the measurement beam, correct
movement direction of the objects has to be observed. The following graphics
show instructions on the installation of the ODS 96:
Preferred movement of
the objects
Figure 5.1: Preferred movement of the objects
Leuze electronic Installation
Leuze electronic ODS 96 19
Preferred mounting in
connection to objects
with structured surface
Figure 5.2: Preferred mounting, objects with structured surface
View through a chase
Figure 5.3: View through a chase
If the ODS 96 has to be installed behind a cover, the chase has to have at
least the size of the optical glass cover. Otherwise, a correct measurement is
not possible or can not be guaranteed.
5.3 Teach-In
Notice
The teach-in procedure can be extended and facilitated through the optional
programming software.
Switching points can also be set through teach-in without using the software.
Connected to the teach-in procedure, differences occur between the ODS 96
M/S types with two switching outputs and the analog and digital device:
Teach-In procedure
ODS 96 M/V, ODS 96
M/D
©
Position the measured object at the desired distance. Connect the teach
input for
2
sec. to +U
B
. After that, connect the teach input to GND. The
switching output is tought.
The tought switching point represents the middle of the switching range of the
output.
These default values are preset:
Function characteristics of the switching output: "light switching"
Switching range: ± 2 mm with infrared devices and ± 10 mm with laser
devices
Hysteresis: ± 1 mm with infrared devices and ± 5 mm with laser devices
These values can only be changed by using the programming software.
Software Leuze electronic
20 ODS 96 Leuze electronic
Teach-In procedure
ODS 96 M/S
©
Position the measured object at the first desired distance. Connect the
teach input for
2
sec. to +U
B
. The LEDs are flashing simultaneously. Re-
connect the teach input to GND. The first switching output is tought. Now,
position the measured object at the second desired distance. Connect the
teach input for
2
sec to +U
B
. The LEDs are flashing alternatingly. Recon-
nect the teach input to GND. The second switching output is tought. In
non-operational mode, the teach input is connected to GND.
6 Software
General description
The programming software can be used either for online programming of the
connected ODS 96 or for the offline generation of configurations. For this pur-
pose, the command Type!, explained in section 6.3.1 "Description of the
Menu Commands" can be used. After starting the program a dialog is dis-
played in which you have to choose a device (see section 6.3 "Starting the
Program"). After the offline generation of a parameter configuration, this con-
figuration can be transmitted to the ODS 96 after connection to the PC has
been established.
The order number for the software can be found in section 4.5 "Accessories".
6.1 Connection of the PC
A special parameter plug is situated behind the plastic cover on the backside
of the ODS 96. Connection between the ODS 96 and the PC is established
by means of the cable included in the shipment.
©
Connect the cable to the parameter plug on the ODS 96 and to a serial
interface on the PC. More information on the definition of the interface can
be found in section 6.3.1 "Description of the Menu Commands".
6.2 Installation of the Programming Software
Requirements for the installation of the programming software:
Windows 3.1 or Windows 95/98/NT,
486 processor or faster,
4 MByte RAM,
2 MByte free disk space
and a CD-Rom drive.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29

Leuze ODS 96 M/S Laser Technical And Software Description

Category
Measuring & layout tools
Type
Technical And Software Description
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI