Computer Assembly Kit for FTB690 Series Meter
M-4410B /0307
InstallatIon InstructIons
1. Your meter should be installed with the arrow pointed in the
direction of flow. You can rotate the computer electronics
180 degrees before you install the computer.
2. Check that the batteries are secure in the back of the
computer.
3. Mount the computer on the turbine with the four screws at
the corners of the faceplate. Make sure the O-ring is fully
seated before tightening the screws.
4. Check meter accuracy before using. Make sure there is no air
in the system by starting the flow until it runs steadily. Then
stop the flow using a valve or nozzle.
5. Hold down DISPLAY button for three seconds to zero the
meter’s Batch Total. When zeroes appear, release the button.
6. Meter an exact known volume into an accurate container.
For best results, meter with one continuous full stream.
7. Check the readout. If the amount metered is accurate, field
calibration is not necessary. If not, refer the Calibration
Section in your Owner’s Manual.
This Kit includes:
• Replacement Computer Assembly with Two Lithium
Batteries
• Four Screws
• One O-Ring
Use with Kits:
FTB691-CK
FTB692-CK
FTB693-CK
FTB694-CK
FTB695-CK