Basics on I/Q data acquisition and processing
R&S®ESW
18User Manual 1177.6317.02 ─ 09
The R&S ESW is equipped with an overload protection mechanism. This mechanism
becomes active as soon as the signal level at the input mixer exceeds the specified
limit. It ensures that the connection between RF input and input mixer is cut off.
In this case, you must decrease the level at the RF input connector and then close the
message box. Then measurements are possible again.
●Using the RF attenuator..........................................................................................18
●Using the preamplifier............................................................................................. 19
●Using the preselector.............................................................................................. 19
4.1.1 Using the RF attenuator
The first tool provided by the R&S ESW to control measurement sensitivity is the RF
attenuator.
The RF attenuator is available in all hardware configurations of the R&S ESW.
Attenuation has the following effects on the measurement:
●High attenuation protects the input mixer: the main purpose of the attenuator is to
protect the input mixer.
●High attenuation makes sure that the measurement results are reliable (signals
that are stronger than allowed can distort the results)
●High attenuation helps you to avoid intermodulation
●High attenuation increases inherent noise (i.e. the noise floor) and thus decreases
measurement sensitivity: if you increase attenuation by 10 dB, the sensitivity is
reduced by 10 dB (in other words: the displayed noise increases by 10 dB)
Depending on the required test setup, you must find a compromise between a high
sensitivity, low intermodulation and input mixer protection. We recommend to let the
R&S ESW determine the ideal attenuation automatically.
You can determine the attenuation automatically with the auto ranging feature in the
receiver application and the auto attenuation feature in the other applications. Deter-
mining the attenuation automatically might not necessarily utilize the maximum
dynamic range, but still yields valid and reliable results.
When you select the attenuation manually and are measuring unknown signals, espe-
cially DUTs with a high RFI voltage, always select the highest possible attenuation
level before you apply the signal.
If you need a better sensitivity or signal-to-noise ratio, make sure that the applied sig-
nal does not exceed the specified limits, before you lower the attenuation.
For further protection of the input mixer, the R&S ESW does not allow you to select
attenuation levels of less than 10 dB unless you explicitly turn on this feature ("10 dB
Minimum Attenuation").
Protecting the input mixer
1. NOTICE! EMC measurements often measure unknown signals that contain pulses
with possibly strong signal levels. Strong signal levels can damage the input mixer.
Increasing measurement sensitivity (or avoiding an input mixer overload)