Rinstrum C510 Industrial Weight Controller User manual

  • Hello! I have reviewed the work instruction for rinstrum's C510, C520, and C527 sealing methods. The document details how to seal devices using calibration stickers or wire and lead seals. I can answer your questions about the specific equipment and steps needed for these procedures.
  • What equipment is needed for sealing with calibration seal stickers?
    How do I seal the setup switch using stickers?
    What is used to seal the load cell connection?
    What equipment is needed for sealing with wire and Lead seals?
    How do I secure devices with wire and Lead seals?
WORK INSTRUCTION M500-605-100-M01 Rev 1.00
Page 1 M500-605-100-M01
Sealing Methods C510, C520 and C527
Sealing with calibration seal stickers
Equipment required.
• Destructible stickers
• Setup switch cover
• Load cell connection cover (M5010)
Method
1. Place the setup switch cover in position so that it covers the setup switch and 2
closure screws.
2. Place a sticker over the centre of the setup switch cover and onto the side of the
enclosure as shown in Figure 1.
WORK INSTRUCTION M500-605-100-M01 Rev 1.00
Page 2 M500-605-100-M01
Figure 1: Setup Switch and pluggable screw terminal load cell connection sealing
with a sticker.
3. Either
Place the load cell connector adaptor cover over the load cell connector
adaptor and place a sticker across the load cell connector adaptor cover as
shown in Figure 1.
Or
Place a sticker across the load cell connector as shown in Figure 2
WORK INSTRUCTION M500-605-100-M01 Rev 1.00
Page 3 M500-605-100-M01
Figure 2: Sealing DB9 loadcell connection.
Sealing with wire and Lead seals
The Setup switch cover and pluggable screw terminal load cell connection can also
be secured with wire and Lead seals if required. This will also seal access to two of
the enclosure closure screws.
Equipment required.
• Wire and Lead seals
• Setup switch cover
• Load cell connection cover
Method
Place the setup switch cover in position so that it covers the setup switch and 2
closure screws.
Insert wire seals through loops provided and close the seals as shown in Figure 3.
Place the load cell connection cover over the load cell connector.
Insert a wire seal through the connections cover provided and close the seals as
shown in figure 3.
WORK INSTRUCTION M500-605-100-M01 Rev 1.00
Page 4 M500-605-100-M01
Figure 3: Wire seal loop locations
/