COMPONENTS:
1 x 5013 Instruction
1 x 9198G Sample Tube, Graduated (25 mL) w/ cap & green dot, plastic
1 x R-0638G-C Phenolphthalein Indicator, 2 oz w/ green cap, DB
1 x R-0645-C Total Alkalinity Indicator, 2 oz, DB
1 x R-0687G-C Sulfuric Acid .12N, 2 oz w/ green cap, DB
1 x R-0736G-C Sulfuric Acid .6N, 2 oz w/ green cap, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
P/T Alkalinity Test
Water Supply & Rinses
1.Rinseandll25mLsampletube(#9198G)to25mLmarkwithwatertobetested.
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
2.Add3dropsR-0638GPhenolphthaleinIndicator.Swirltomix.Samplewillturnpink
(Fig.1)ifPalkalinityispresent—proceedtoStep3.Ifsampleiscolorless,proceed
toStep4.
3.AddR-0687GSulfuricAcid.12Ndropwise,swirlingandcountingaftereachdrop,
untilcolorjustchangesfrompinktocolorless.RecorddropsasPreading.
DROP TEST
P/T ALKALINITY (Laundry Washroom, 1 drop = 10 or 50 ppm)
4.Add5dropsR-0645TotalAlkalinityIndicator.Swirltomix.Samplewillturngreen
(Fig.2).
5.AddR-0687GSulfuricAcid.12Ndropwise,swirlingandcountingaftereachdrop,
untilcolorchangesfromgreentored(Fig.3).Recordtotaldrops(Steps3&5)
as T reading.
6.Multiply P reading by 10. Record as parts per million (ppm) P alkalinity as
calcium carbonate. Multiply T reading by 10. Record as ppm T alkalinity as
calciumcarbonate.
NOTE: For 14.6 mL sample, record P reading as grains per gallon (gpg) P
alkalinityascalciumcarbonate.Record TreadingasgpgTalkalinity as
calciumcarbonate.
Breaks & Suds
1.Rinseandll25mLsampletube(#9198G)to25mLmarkwithwatertobetested.
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
2.Add3dropsR-0638GPhenolphthaleinIndicator.Swirltomix.Samplewillturn
pink(Fig.4)ifPalkalinityispresent—proceedtoStep3.Ifsampleiscolorless,
proceedtoStep4.
3.AddR-0736GSulfuricAcid.6Ndropwise,swirlingandcountingaftereachdrop,
untilcolorjustchangesfrompinktocolorless.RecorddropsasPreading.
Fig. 1
Fig. 2
Fig. 3
Instr. #5013
(OVER)
DROP TEST
P/T ALKALINITY (Laundry Washroom, 1 drop = 10 or 50 ppm)
Fig. 4
Fig. 5
Fig. 6
Instr. #5013
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 8/16
4.Add5dropsR-0645TotalAlkalinityIndicator.Swirltomix.Samplewillturngreen
(Fig.5).
5.AddR-0736GSulfuricAcid.6Ndropwise,swirlingandcountingaftereachdrop,
untilcolorchangesfromgreentored(Fig.6).Recordtotaldrops(Steps3&5)as
T reading.
6.Multiply P reading by 50. Record as parts per million (ppm) P alkalinity as
calcium carbonate. Multiply T reading by 50. Record as ppm T alkalinity as
calciumcarbonate.
NOTE: For14.6mLsample,multiplyPreadingby5.Recordasgrainspergallon
(gpg)Palkalinityascalciumcarbonate.MultiplyTreadingby5.Recordas
gpgTalkalinityascalciumcarbonate.