Rosemount X-STREAM X1 Flameproof Addendum for Hazardous Area Owner's manual

Type
Owner's manual
Flameproof Gas Analyzer
Instruction Manual Addendum for
Hazardous Area Applications
www.EmersonProcess.com
Instruction Manual
HASADE-IM-Ex
5
th
edition
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Emerson Process Management (Rosemount Analytical) designs, manufactures and
tests its products to meet many national and international standards. Because these
instruments are sophisticated technical products, you MUST properly install, use, and
maintain them to ensure they continue to operate within their normal specications.
The following instructions MUST be adhered to and integrated into your safety program
when installing, using and maintaining Emerson Process Management (Rosemount
Analytical) products. Failure to follow the proper instructions may cause any one of the
following situations to occur: Loss of life; personal injury; property damage; damage to
this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson Process
Management (Rosemount Analytical) representative for clarication.
• Follow all warnings, cautions, and instructions marked on and supplied with the
product.
• Inform and educate your personnel in the proper installation, operation, and
maintenance of the product.
• Install your equipment as specied in the Installation Instructions of the
appropriate Instruction Manual and per applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualied personnel to install, operate, update,
program, and maintain the product.
• When replacement parts are required, ensure that qualied people use replacement
parts specied by Emerson Process Management (Rosemount Analytical).
Unauthorized parts and procedures can affect the product’s performance, place the
safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike
substitutions may result in re, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place,
except when maintenance is being performed by qualied persons, to prevent
electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Emerson Process Management GmbH & Co. OHG
Industriestrasse 1
D-63594 Hasselroth
Germany
T +49 (0) 6055 884-0
F +49 (0) 6055 884-209
Internet: www.EmersonProcess.com
5
th
edition 2008-04
Instruction Manual
HASDE-IM-Ex
04/2007
Emerson Process Management GmbH & Co. OHG S-1
Safety Instructions
X-STREAM FD
This instruction manual provides information about installing, operating and maintaining/
servicing X-STREAM series gas analyzers in hazardous (classied) areas and shall be read
in conjunction with the standard analyzer instruction manual only!
This instruction manual covers several X-STREAM series analyzer variations and therefore
may describe congurations and/or options not part of your specic analyzer.
PREAMBLE
The following denitions apply to WARNINGS, CAUTIONS and NOTES found throughout
this publication.
DEFINITIONS
Highlights an operation or maintenance procedure, practice, condition,
statement, etc.
If not strictly observed, could result in injury, death, or long-term health
hazards of personnel.
NOTE
Highlights an essential operating procedure, condition or statement.
Highlights an operation or maintenance procedure, practice, condition,
statement, etc.
If not strictly observed, could result in damage to or destruction of equipment,
or loss of effectiveness.
Instruction Manual
HASDE-IM-Ex
04/2007
Emerson Process Management GmbH & Co. OHGS-2
X-STREAM FD
Symbols
1. Where equipment or covers are marked with the symbol to the right, hazar-
dous voltages are likely to be present beneath. These covers should only be
removed when power is removed from the equipment — and then by trained
service personnel only.
2. Where equipment or covers are marked with the symbol to the right, there is a
danger from hot surfaces beneath. These covers should only be removed by
trained service personnel when power is removed from the equipment. Certain
surfaces may remain hot to the touch.
3. Where equipment or covers are marked with the symbol to the right, refer to
the Instruction Manual for instructions.
4. Further graphical symbols possibly used in this product:
SYMBOLS USED IN THIS MANUAL
Explosion Hazard!
Harmful (to Health)!
Elektrostatic discharge (ESD)
Toxic!
Heavy Instrument!
Disconnect from Power!
Instruction Manual
HASDE-IM-Ex
04/2007
Emerson Process Management GmbH & Co. OHG S-3
Safety Instructions
X-STREAM FD
TERMS USED IN THIS MANUAL
ATEX
Directive 94/9/EC, commonly called the
ATEX („Atmosphères Explosibles“) directive,
dealing with equipment intended to be used
in potentially explosive atmospheres.
This directive is valid for equipment to be
sold into and/or installed and operated in the
European Union (EU).
Area Classication
Zone 1
Where ignitable concentrations of ammable
gases can exist some of the time under nor-
mal operating conditions.
(A guideline value [not part of a standard ] is
10 to 1.000 hours per year.)
Zone 2
Where ignitable concentrations of ammable
gases are not likely to exist under normal
operating conditions.
(A guideline value [not part of a standard ] is
less than 10 hours per year.)
Lower Explosion Limit (LEL)
Volume ratio of ammable gas in air below
which an explosive gas atmosphere will not
be formed: the mixture of gas and air lacks
sufcient fuel (gas) to burn.
Upper Explosion Limit (UEL)
Volume ratio of ammable gas in air above
which an explosive gas atmosphere will not
be formed: the mixture of gas and air is too
rich in fuel (decient in oxygen) to burn.
Flammable Gas(es)
Gases and gas mixtures are assigned to be
ammable if they might become ignitable
when in a mixture with air.
Terms
Instruction Manual
HASDE-IM-Ex
04/2007
Emerson Process Management GmbH & Co. OHGS-4
X-STREAM FD
Safety Instructions
AUTHORIZED PERSONNEL
To avoid explosions, loss of life, personal injury and damage to this equipment and on-site
property, do not operate or service this instrument before reading and understanding this
instruction manual and receiving appropriate training.
SAFETY SUMMARY
If this equipment is used in a manner not specied in these instructions, protective sy-
stems may be impaired.
INTENDED USE STATEMENT
X-Stream series gas analyzers are intended to be used as analyzers for industrial pur-
poses. They must not be used in medical, diagnostic or life support applications nor as
safety devices, and no independent agency certications or approvals are to be implied
as covering such applications!
ADDITIONAL LITERATURE
This manual covers aspects specic for using ameproof X-STREAM FD analyzers in
hazardous (classied) areas, only.
To install, start-up, operate and maintain/service the instrument in a safe manner it is
MANDATORY to read all additional instruction manuals shipped together with the instru-
ment!
The following instruction manuals are available and/or referenced within this manual at
hand:
HASAxE-IM-HS X-STREAM series instruction manual
HASFAE-IM-H Flame arrestors series instruction manual
The original manufacturer´s cable gland or conduit instruction manual, depending on
what is used.
Contact your local service center or sales ofce when missing documents.
Save all instructions for future use!
SAFETY INSTRUCTIONS
Instruction Manual
HASDE-IM-Ex
04/2007
Emerson Process Management GmbH & Co. OHG S-5
Safety Instructions
X-STREAM FD
EXPLOSION HAZARD BY MODIFICATION
Any addition, substitution, or replacement of components installed on or
in this device, must be certied to meet the hazardous area classication
that the device was certied to prior to any such component addition,
substitution, or replacement. In addition, the installation of such device or
devices must meet the requirements specied and dened by the hazardous
area classication of the unmodied device.
Any modications to the device not meeting these requirements, will void
the product certication(s).
Contact Emerson Process Management‘s customer service center for return
authorization.
POSSIBLE EXPLOSION HAZARD
The X-STREAM analyzer may utilize not only sample gas but one or more
pressurized carrier gases and/or calibration gases.
If an external owmeter is required for ow control, legislative requirements
and instructions for installation in hazardous (classied) areas must be
considered.
Safety Instructions
POSSIBLE EXPLOSION HAZARD
Do not open instrument when energized.
Ensure that external circuitry is disconnected or de-energized before
opening the instrument.
Ensure that all gas connections are made as labeled and are leak free.
Improper gas connections could result in explosion and death.
Instruction Manual
HASDE-IM-Ex
04/2007
Emerson Process Management GmbH & Co. OHGS-6
X-STREAM FD
Safety Instructions
HOW TO STAY IN COMPLIANCE WITH THE
EUROPEAN DIRECTIVE 94/9/EC ("ATEX") WHEN PERFORMING GAS
ANALYSIS WITHIN A FLAMEPROOF ENCLOSURE.
Special conditions apply to using a ameproof enclosure analyzer under the scope of the
"European Directive for Equipment used in Explosive Atmosphere" (Directive 94/9/EC;
ATEX). To stay compliant to the directive please consider the following clarication sheet
released by the European ATEX Notied Body Group (see next page):
HEAVY INSTRUMENT
The analyzer model X-STREAM FD, to which this manual relates, intended
to be wall mounted and/or outdoor installed, weighs up to approx. 63 kg
(139 lbs), depending on included options!
Use two people and/or suitable tools for transportation and lifting these
instruments!
Take care to use anchors and bolts specied to be used for the weight of
the units!
Take care the wall or stand the unit is intended to be installed at is solid and
stable to hold the units!
Instruction Manual
HASDE-IM-Ex
04/2007
Emerson Process Management GmbH & Co. OHG S-7
Safety Instructions
X-STREAM FD
Safety Instructions
Instruction Manual
HASDE-IM-Ex
04/2007
Emerson Process Management GmbH & Co. OHGS-8
X-STREAM FD
TOC-1Emerson Process Management GmbH & Co. OHG
Table of Contents
Instruction Manual
HASDE-IM-Ex
07/2007
X-STREAM FD
Table of ConTenTs
Preamble S-1
Denitions S-1
Symbols used in this manual .............................................................................................S-2
Terms used in this manual
.................................................................................................S-3
Safety instructions
..............................................................................................................S-4
Chapter 1 Technical Description 1-1
1-1 Overview .................................................................................................................1-1
1-2 Design Features .....................................................................................................1-1
1-3 Protective Measures in Detail ...............................................................................1-1
1-4 Explosion Protection Compliances......................................................................1-3
1-5 Technical Data ........................................................................................................1-4
Chapter 2 Installation 2-1
2-1 Installing the Analyzer ...........................................................................................2-2
2-2 Connecting Gas Lines ...........................................................................................2-3
2-3 Electrical Installation .............................................................................................2-5
Chapter 3 Startup 3-1
Chapter 4 Service and Maintenance 4-1
4-1 Verications and Tests after Service or Maintenance ........................................4-3
4-1-1 Routine Tests ......................................................................................................4-3
4-1-1-1 Visual Inspection ............................................................................................4-3
4-1-1-2 Tests on Flame Arrestors ..............................................................................4-3
4-2 Replacement of Parts ............................................................................................4-4
Appendix A-1
A-1 EC Declaration of Conformity ...............................................................................A-2
A-2 ATEX EC Type Examination Certicate ...............................................................A-3
A-3 CSA Certicate of Compliance .............................................................................A-6
A-4 Cable gland ADL 1F installation instruction ......................................................A-11
A-5 Block diagram X-STREAM FD .............................................................................A-12
A-6 Assignment of Terminals and Socket ...............................................................A-13
TOC-2 Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASDE-IM-Ex
07/2007
X-STREAM FD
Table of Contents
Table of figures
Fig. 1-1: X-STREAM FD .....................................................................................................................................1-1
Fig. 1-2: X-STREAM FD, bottom view ..............................................................................................................1-2
Fig. 1-3: X-STREAM FD - dimensions .............................................................................................................1-4
Fig. 1-4: X-STREAM FD - power terminals / fuse holders ..............................................................................1-7
Fig. 1-5: X-STREAM FD - signals terminals ....................................................................................................1-8
Fig. 2-1: X-STREAM FD dimensions ................................................................................................................2-2
Fig. 2-2: Flame arrestor installed into instrument enclosure ........................................................................2-3
Fig. 2-3: Flame arrestor elements, considering as example FA 01 ...............................................................2-3
Fig. 2-4: X-STREAM FD - allocation of terminals ..........................................................................................2-10
Fig. 2-5: X-STREAM FD - Analog output terminals .......................................................................................2-12
Fig. 2-6: X-STREAM FD - Modbus interface terminals .................................................................................2-13
Fig. 2-7: X-STREAM F - Modbus over ethernet connector ...........................................................................2-14
Fig. 2-8: Relay status signals, block diagram ..............................................................................................2-15
Fig. 2-9: X-STREAM FD - Relay Status Terminals ........................................................................................2-15
Fig. 2-10: Digital Inputs & Outputs Terminals ...............................................................................................2-16
Fig. 2-11: Power terminals ..............................................................................................................................2-17
Fig. 2-12: Equipotential bonding conductor terminal ..................................................................................2-18
Fig. A-1: Terminals strip 1 ..............................................................................................................................A-13
Fig. A-2: Terminals strip 2 ..............................................................................................................................A-13
Fig. A-3: Power terminals ..............................................................................................................................A-13
Fig. A-4: Ethernet connector for Modbus .....................................................................................................A-13
X-STREAM FD
Emerson Process Management GmbH & Co. OHG 1-1
Instruction Manual
HASADE-IM-EX
07/2007
1 Technical Description
Chapter 1
Technical Description
1-1 Overview
The new X-STREAM FD gas analyzer is de-
signed to be used in hazardous areas. Its a-
meproof enclosure can be installed at Zone 1
locations without the need of any additional
protective measures, e.g. purge gas supply.
1-2 Design Features
Packaged into a cast aluminum enclosure, the
X-STREAM FD gas analyzer provides all the
measurement options available for general
purpose instruments, but for installation at lo-
cations where explosive atmosphere might be
present frequently and classied as Zone 1.
The basic concept used is to protect the sur-
rounding atmosphere from being ignited if an
internal failure results in high temperatures,
ames or even an explosion.
To provide adequate explosion protection
the X-STREAM FD analyzer features several
protection methods:
Fig. 1-1: X-STREAM FD
1-3 Protective Measures in Detail
The cast aluminum enclosure consists of
two parts: base and cover, connected by
hinges.
The area where the two parts are in contact
is designed to work as a ange, quenching
ames entering the small path between them.
When operated, the analyzer enclosure has to
be closed and secured by 20 screws evenly
arranged all over the ange.
The cast aluminum enclosure is designed to
withstand an internal explosion,
extinguish the ames resulting from an
internal explosion (thus preventing a
sur-rounding explosive atmoshere from
being ignited).
Flame arrestors avoide ame transmission
from the gas paths into the surrounding
atmosphere.
Approved cable glands (option: conduits)
protect the cable inlets and outlets.
1: Enclosure base
2: Screws
3: Enclosure cover
4: Flange
5: Eyebolts for lifting
6: Eyebolts for installation
7: Hinges
1
2
3
5
4
6
7
X-STREAM FD
Emerson Process Management GmbH & Co. OHG1-2
Instruction Manual
HASADE-IM-EX
07/2007
1-3 Protective Measures in Detail
The only openings penetrating the enclosure
are threads, to be used for gas and cable in-
and outlets:
Depending on the measurement application
the instrument provides up to 8 gas in- and
outlets, each protected by an approved ame
arrestor. These arrestors are installed into
threads at the bottom side of the enclosure
base. Two tting sizes are available for ex-
ternal connection of gas pipes with 3.18 mm
(1/8“) or 6.35 mm (1/4“) outer diameter (OD).
Optionally a clamping ring for 6 mm OD may
be used, replacing the 6.35 mm version.
Cables are fed into the enclosure utilizing up
to 4 cable glands, located at the enclosure`s
bottom right side. The approved glands accept
3 different internal elastomeric sealing rings
with different internal diameters, supporting
a wide range of cables.
For installation in North-America cable glands
are replaced by a combination of conduits and
metric-to-NPT thread adapters.
All threads provide a ame path of a length
ensuring that possibly entering ames are
extin-guished before reaching the external
atmos-phere.
Unused threads must be closed with plugs
when the instrument is operated to ensure
explosion protection.
Note!
See the X-STREAM series instruction manual
for more information about common X-STREAM
series gas analyzers features and special
features of the X-STREAM FD.
Fig. 1-2: X-STREAM FD, bottom view
1: Plugged when not used
2: Gas tting (part of ame arrestor)
3: Plug
4: Cable gland (or conduits)
1 2 3 4
X-STREAM FD
Emerson Process Management GmbH & Co. OHG 1-3
Instruction Manual
HASADE-IM-EX
07/2007
1 Technical Description
1-4 Compliances
1-4 Explosion Protection Compliances
This product is available in two different variations, separately certied by agencies for the use
in hazardous (classied) areas:
The one version, to be equipped with metric-to-NPT adapters and conduits (these components
are not part of the instrument certication), is certied by the Canadian Standards Association,
an „OSHA Nationally Recognized Testing Labor tory“ (NRTL), for Canada and USA and conforms
to the provisions of CAN/CSA-E60079-0:02 (R2006), CAN/CSA-E60079-1:02 (R2006), ANSI/ISA-
12.00.01-2002 (IEC 60079-0 Mod), ANSI/ISA-12.22.01-2002 (IEC 60079-1 Mod).
The second variation, to be equipped with cable glands, is certied by Fyzikálně technický
zkušební ústav, s.p (FTZÚ), an European Notied Body under the Directive 94/9/EC („ATEX“)
and conforms to the provisions of EN 60079-0 and EN 60079-1. See appendix for a copy of the
CE type examination certicate.
Depending on the variation, the following certication markings apply to the product:
USA
Class I, Zone 1, AEx d IIB + H
2
T3
Canada Class I, Zone 1, Ex d IIB + H
2
T3
European Union (EU) ATEX, category 2, Zone 1, Ex d IIB + H
2
T4
EC ATEX Type Examination Certicate: FTZU 06 ATEX 0186.
Conforms with the provisions of the „Equipment intended for use in Potentially Explosive Atmosphe-
res (ATEX)“ Directive 94/9/EC, EMC Directive 89/336/EEC and CE Directive 93/68/EEC.
X-STREAM FD
Emerson Process Management GmbH & Co. OHG1-4
Instruction Manual
HASADE-IM-EX
07/2007
1-5 Technical Data
1-5 Technical Data
Fig. 1-3: X-STREAM FD - dimensions
approx. in mm [inches]
Cable inlets
(enclosure threads; M20 x 1.5)
Flame arrestors with gas ttings
(enclosure threads: M18 x 1.5)
Eyebolt detail
Housing
Permissible ambient temperature range -22 F to 122 F ( -30 °C to +50 °C)
Weight: approx. up to 139 lbs (63 kg)
(depending on analyzer conguration)
Protection class: IP 66 (EN 60529) / NEMA 4X
for outdoor installation
Analyzer must not be exposed to direct
sun light
Gas ttings: quantity: max. 8
specication: ame arrestors with ttings
6/4 mm or 1/4“, stainless steel
X-STREAM FD
Emerson Process Management GmbH & Co. OHG 1-5
Instruction Manual
HASADE-IM-EX
07/2007
1 Technical Description
1-5 Technical Data
Site of installation
Humidity (non condensing) < 90 % r. h. at 68 F (+20 °C)
< 70 % r. h. at 104 F (+40 °C)
Pollution degree 2
Installation category I I
Altitude 0 to 6560 ft (2000 m) above sea level
Sourrounding atmosphere Analyzers must not be operated in
corrosive atmosphere.
Power supply
Rated input voltageated input voltage
100 - 240 V 50/60 Hz, wide range input
Power supply voltage uctuations are
not to exceed +/- 10 % of the nominal
supply voltage!
Input voltage range 85 - 264 V , 47 - 63 Hz
Rated input current
standard 0,75 - 0,35 A max.
with thermostated physics 2 - 1 A max.
General Purpose Compliances
Electrical safety CAN / USA CSA-C/US, based on
CAN/CSA-C22.2 No. 61010-1-04 /
UL 61010-1, 2
nd
Edition
Europe CE, based on EN 61010-1
Electromagnetic compatibility
Europe CE, based on EN 61326
other NAMUR
X-STREAM FD
Emerson Process Management GmbH & Co. OHG1-6
Instruction Manual
HASADE-IM-EX
07/2007
1-5 Technical Data
Gas parameters
X-STREAM Instruction Manual, chapter 3 Measuring Principles
Interfaces, signal inputs / outputs
2 analog outputs channel 4 (0) - 20 mA (R
B
< 500 )
(optically isolated; congurable by keypad
start and end concentration user congurable)
Modbus interface RS 485 (2- or 4-wire)
optional:
RS 232 with or without optical isolation
Ethernet (RJ45 socket)
3 relay outputs (option „status signals“) “Failure”
according NAMUR NE 107 “Maintenance required / Off specication“
“Function check”
dry contacts, max. 30 V; 1 A; 30 W resi-
stive
Digital Inputs and Outputs (option)
7 digital inputs zero calibration ch1 & ch2,
(common ground) span calibration ch1,
span calibration ch2,
open valve V4
open valve V1
open valve V2
open sample gas valve V3 /
switch off sample pump
max. 30 V, internally limited to 2.3 mA
H level: min. 4 V; L level: max. 3 V
8 digital outputs 2 thresholds per channel,
(optically isolated, common ground) sample gas valve,
zero gas valve V4,
span gas valve V1,
span gas valve V2
“Open Collector”, max. 30 V / 30 mA
X-STREAM FD
Emerson Process Management GmbH & Co. OHG 1-7
Instruction Manual
HASADE-IM-EX
07/2007
1 Technical Description
Fig. 1-4: X-STREAM FD - power terminals / fuse holders
1-5 Technical Data
Power Connection
Connection via internal screw terminals near
cable entries, (g. 2-8).
Cross section: max. 10 AWG (4 mm
2
), using conductor
sleeves is not required
Cable entry via 1 cable gland, classied IP 68 or suitable
conduit with metric-to-NPT adaptor
Permissible outer cable diameter for power
cord when provided with cable glands: 0.11 to 0.5 inch (3-13 mm),
depending on used cable gland sealing ring
Power fuses
The power terminals integrate fuse holders.
Fuse ratings: AC 230 V, 3.15 A, 5x20 mm
1 Power terminals with integrated
fuse holders
2 Protective earth terminal (PE)
3 Power cable entry
4 EMI power lter
324 1
X-STREAM FD
Emerson Process Management GmbH & Co. OHG1-8
Instruction Manual
HASADE-IM-EX
07/2007
Fig. 1-5: X-STREAM FD - signals terminals
1-5 Technical Data
Signal inputs / outputs
All signal lines need to be connected to internal
screw terminals (g. 1-5), except the optional RJ45 ethernet connector.
Cross section: max. 12 AWG (2.5 mm
2
), using conductor
sleeves is not required
Cable entry via
3 cable glands, classied IP 68 or suitable
conduits with metric-to-NPT adaptors
Permissible outer cable diameter for signal
cables when provided with cable glands: 0.11 to 0.5 inch (3-13 mm)
depending on used cable gland sealing rings
Available signals: standard: Analog signal outputs
Relay status signals
Modbus interface (RS232; RS 485)
optional: Digital inputs/outputs
Modbus RJ45 ethernet connector
Detailed pin assignment 2 Installation
The optional ethernet connector (RJ45),
located on the electronics
main board BKS
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64

Rosemount X-STREAM X1 Flameproof Addendum for Hazardous Area Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI