TAP1500
1.5 GHz, 10X Active Probe
ZZZ
Instruction Manual
*P071180901*
071-1809-01
TAP1500
1.5 GHz, 10X Active Probe
ZZZ
Instruction Manual
xx
www.tek.com
071-1809-01
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by
national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously
published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tek.com to nd contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If
any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts
and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for
warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make
suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service
center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a
location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties,
taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix
shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix
representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to
repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modied or integrated with
other products when the effect of such modication or integration increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR
IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY
PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY
INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 15AUG04]
Table of Contents
Table of Contents
General safety summary ................................................................................................................... iii
Service safety summary..................................................................................................................... v
Environmental considerations.............................................................................................................. vi
Preface..................................................................................................................................... vii
Documentation....................................................................................................................... vii
Conventions used in this manual.................................................................................................... vii
Returning the probe for servicing.................................................................................................... viii
Key features ................................................................................................................................ 1
Installation .................................................................................................................................. 2
Connecting to the host instrument.................................................................................................... 2
Probe controls and indicators......................................................................................................... 3
Functional check. ........................................................................................................................... 6
Required equipment .................................................................................................................. 6
Basic oper ation ............................................................................................................................. 9
Probe head assembly................................................................................................................. 9
Probe i nput . . ......................................................................................................................... 10
Probe offset .......................................................................................................................... 11
Accessories and options .................................................................................................................. 13
Using sta ndard accessories ......................................................................................................... 13
TAP1500 1.5 GHz, 10X Active Probe Instruction Manual i
Table of Contents
Optional accessories. . ............................................................................................................... 22
Options............................................................................................................................... 24
Probing principles.......................................................................................................................... 25
Ground lead length . . . ............................................................................................................... 25
Ground lead inductance ............................................................................................................. 26
Low-inductance grounding........................................................................................................... 27
SureFoot grounding ............................................................................................................... 28
Probe tip test points.................................................................................................................. 29
Probe tip stabilization................................................................................................................ 30
Specications.............................................................................................................................. 31
Warranted characteristics............................................................................................................ 32
Typical characteristics ............................................................................................................... 33
Nominal characteristics.............................................................................................................. 40
Performance verication................................................................................................................... 41
Equipment required .................................................................................................................. 41
Equipment setup..................................................................................................................... 43
DC gain accuracy.................................................................................................................... 44
Test record ........................................................................................................................... 47
Maintenance............................................................................................................................... 48
Error condition ....................................................................................................................... 48
Replacement parts................................................................................................................... 48
Index
ii TAP1500 1.5 GHz, 10X Active Probe Instruction Manual
General safety summary
General safety summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specied.
Only qualied personnel should perform service procedures.
To avoid re or personal injury
Connect and disconnect properly.
Connect the probe output to the measurement instrument before
connecting the probe to the
circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe input. Disconnect t
he probe
input and the probe reference lead from the circuit under test before disconnecting the probe from the measureme nt ins
trument.
Observe all terminal ratings.
To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product
manual for further ratings information before making connections to the product.
Do not app ly a potential to any terminal, including the commo n terminal, that exceeds the maximum rating of that terminal.
Do not operate without covers.
Do not operate this produ ct with covers or panels removed.
Do not operate with suspected failures.
If you suspect that there is damage to this product, have it inspected by qualied
service personnel.
Avoid exposed circuitry.
Do not touch exposed connections and components when power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
TAP1500 1.5 GHz, 10X Active Probe Instruction Manual iii
General safety summary
Terms in this manual
These terms may appear in this manual:
WARNING.
Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION.
Caution statements identify c onditions or practices that could result i n damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
iv TAP1500 1.5 GHz, 10X Active Probe Instruction Manual
Service safety summary
Service safety summary
Only qualied personnel should perform service procedures. Read this
Service s afety summary
and the
General safety summary
before performing any service procedures.
Do not service alone.
Do not perform internal service or adjustments of this product unless another person capable of rendering
rst aid and resuscitation is present.
Disconnect power.
To avoid electric shock, switch off the instrument power, then disconnect the power cord from the mains
power.
Use care when servicing with power on.
Dangerous voltages or currents may exist in this product. Disconnect power, remove
battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
TAP1500 1.5 GHz, 10X Active Probe Instruction Manual v
Environmental considerations
Environmental considerations
This section p rovides information about the environmental impact of the product.
Product End-of-Life handling
Observe the following guidelines when recycling an instrument or component:
Equipment recycling.
Production of this equipment required the extraction and use of natural resources. The equipment may
contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. To
avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this
product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to
Directives 2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For
information about recycling options, check the Tektronix Web site (www.tek.com/productrecycling).
vi TAP1500 1.5 GHz, 10X Active Probe Instruction Manual
Preface
Preface
This manual describes the installation and operation of the TAP1500 active probe. Basic probe operations and concepts are
presented in this manual. You can also access the Tektronix Web site for this document and other related information.
Documentation
To read about Use these documents
1
TAP1500 Probe: First Time Operation, Functional Check,
Operating Basics, Specications, Performance Verication
Read this Instruction Manual.
In-depth oscilloscope operation, user interface help, GPIB
commands
Access the online help from the Help menu on the host
instrument.
1
To access the documentation that is installed on your instrument, click
Start
in the taskbar and select
Programs > TekApplications
.
Conventions used in this manual
The following icon is used throughout this manual to indicate a step sequence.
TAP1500 1.5 GHz, 10X Active Probe Instruction Manual vii
Preface
Returning the probe for servicing
If your probe requires servicing, you must return the probe to Tektronix. If the original packaging is unt for use or not available, use
the following packaging guidelines:
Preparation for shipment
1.
Use a corrugated cardboard shipping
carton having inside dimensions at
least one inch greater than the probe
dimensions. The box should have a carton
test strength of at least 200 pounds.
2.
Put the probe into an antistatic bag or wrap
it to protect it from dampness.
3.
Place the probe into the box and stabilize it
with light packing material.
4.
Seal the carton with shipping tape.
5.
Refer to
Contacting Tektronix
at the
beginning of this manual for the shipping
address.
viii TAP1500 1.5 GHz, 10X Active Probe Instruction Manual
Key features
Key features
The TAP1500 act ive probe enables you to make accurate measurements with minimal circuit loading from DC to 1.5 GHz, using
oscilloscopes featuring the new Tektronix TekVPI oscilloscope interface. Key features include:
DC to >1.5 GHz Bandwidth
Risetime <267 ps
±8 Volts Dynamic Range with ±10 volt
offset capability
1M Input Resistance
<1 pF Input Capacitance
10X Attenuation
TekVPI Interface
Small, low-mass probe head for probing
dense circuitry
TAP1500 1.5 GHz, 10X Active Probe Instruction Manual 1
Installation
Installation
Connecting to the host instrument
1.
Slide t he probe into the TekVPI receptacle.
The probe snaps when fully engaged.
When the probe is connected, the host
instrument reads information from the
probe and identies the device.
2.
To disconnect, press the latch release
button and pull away from the instrument.
2 TA P1500 1.5 GHz, 10X Active Probe Instru ction Manual
Installation
Probe controls and indicators
Status LED
When the probe is powered on, the multicolor
Status LED:
Glows green after successfully
completing the power-on self test
routine. The probe is in normal
operating m ode.
Glows red if an error condition exists.
(See page 48,
Error condition
.)
Quick Tip
If the LED does not light, and other probes are
connected to the ho st instrument, the available
probe power may be limited. Try disconnecting
another probe from the instrument to reduce
the load.
TAP1500 1.5 GHz, 10X Active Probe Instruction Manual 3
Installation
Menu button
1.
Press the probe
Menu
button to display the
Probe Setup screen on the oscilloscope.
2.
Use the buttons on the instrument t o set
the probe parameters.
3.
Press the probe
Menu
button again to
close the Probe Setup screen.
4 TA P1500 1.5 GHz, 10X Active Probe Instru ction Manual
Installation
AutoZero
We r ecommend that you run the p robe
AutoZero routine:
After the 20 minute warm-up period
When the operating temperature of the
probe changes by ±5 °C
1.
Press the probe
Menu
button to display the
Probe Setup screen on the oscilloscope.
2.
Short the probe tip to ground.
3.
Press the
AutoZero
button on the
instrument to execute the AutoZero routine.
4.
Press the
Menu
button again to close the
Probe Setup screen.
TAP1500 1.5 GHz, 10X Active Probe Instruction Manual 5
Functional check
Functional check
Use the following procedure to check that your probe is functioning properly. If you want to verify that your probe meets the
warranted specications, refer to the
Performance Verication
procedures. (See page 41,
Performance verication
.)
Required equipment
Description and quantity Performance requirement Recommended example
1
Oscilloscope TekVPI Interface Tektronix DPO4000 Series
Y-Lead adapter
0.025 inch square pins for probe tip
connections
196-3463-XX
2
SMT KlipChip adapters (2)
0.025 inch square pins-to-mini clips 206-0364-XX
2
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers.
2
Standard probe accessory.
6 TA P1500 1.5 GHz, 10X Active Probe Instru ction Manual
/