ReliOn Micro User Instruction

Category
Measuring, testing & control
Type
User Instruction
User
Instruction
Manual
Blood Glucose Monitoring System
MICRO
CAUTION
Before using any product to test your blood sugar (blood glucose),
read all instructions and practice the test.
Do all quality control checks as directed and consult with a diabetes
healthcare professional.
2
Introduction ........................................................................................... 5
Intended Use .................................................................................... 6
Important Safety Instructions ........................................................... 6
Warnings, Precautions and Limitations ............................................ 8
Understanding the ReliOn™ Micro Blood Glucose Meter ................... 11
Changing the Battery...................................................................... 14
Meter Set-up ........................................................................................ 17
Basic Rules ..................................................................................... 19
Audible Beep .................................................................................. 20
Year ................................................................................................. 22
Month/Day ...................................................................................... 23
Hour/Minute ................................................................................... 24
Test Average Display ...................................................................... 25
Ending Set-up ................................................................................. 26
Blood Glucose Testing ......................................................................... 27
Getting a Drop of Blood.................................................................. 27
Performing a Blood Glucose Test ................................................... 29
Flagging Test Results ...................................................................... 33
TABLE OF CONTENTS
3
4
Cleaning and Disinfecting Meter and Lancing Device ......................... 35
Alternate Site Testing (AST) ................................................................. 41
Understanding Test Results ................................................................. 50
Performing a Control Solution Test ...................................................... 55
Troubleshooting Control Solution Test ................................................. 62
Recalling Results - Entering Memory Mode ........................................ 65
Maintenance ........................................................................................ 69
Storage and Handling ..................................................................... 69
Display Screen and Error Messages ............................................... 71
Product Information ............................................................................. 76
 ProductSpecications .................................................................... 76
Warranty ......................................................................................... 77
Thank you for choosing the ReliOn™ Micro Blood Glucose Monitoring
System. This system measures blood glucose in fresh capillary whole
blood. Results are shown as plasma values. This makes it easy to
compare the ReliOn™ Micro blood glucose meter and lab results.
Please read this manual before using the system. If you have any
questions, call Customer Service at 855.776.0662, 7 a.m. to 7 p.m.
central time, 7 days a week. Contact your healthcare professional with
questions if you are unable to reach Customer Service.
INTRODUCTION
5
6
Intended Use
The ReliOn™ Micro Blood Glucose Monitoring System is intended for
the quantitative measurement of glucose in fresh capillary whole blood
samplesdrawnfromthengertiporpalm.Testingisdoneoutsidethe
body (In Vitro diagnostic use). It is indicated for use at home by persons
with diabetes as an aid to monitor the effectiveness of diabetes control.
It is not intended for the diagnosis of or screening for diabetes mellitus,
and is not intended for use on neonates. It is intended for single patient
use and should not be shared with other individuals.
Important Safety Instructions
WARNING
• The meter and lancing device are for single-patient use
• DO NOT share them with anyone including other family members
DO NOT use on multiple patients
• All parts of the kit are considered biohazardous and can potentially
transmit infectious diseases, even after you have performed cleaning
and disinfection .¹,²
7
Cleaning and Disinfecting Meter and Lancing Device
All parts of the blood glucose meter and lancing device are
considered biohazardous and can potentially transmit
infectious diseases.
• CLEANING removes dirt and other visible material.
DISINFECTING destroys organisms that can cause infection or
disease such as hepatitis B, hepatitis C, and HIV.
For complete cleaning and disinfecting instructions see page 35
of this manual.
1.FDAPublicHealthNotication.UseofFingerstickDevicesonMorethanOnePersonPosesRiskforTransmitting
Bloodborne Pathogens: Initial Communication. FDA. Published: Nov. 2010. Accessed: Nov. 2016.
www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm
2. CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting
Bloodborne Pathogens. CDC. Published: Feb. 2011. Accessed: Nov. 2016.
www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html
Warnings, Precautions and Limitations
Before testing, it is important to wash hands with soap and warm water.
Dryhandsthoroughly.Ifyouusealcoholwipestocleansengers,make
suretheareaisdrybeforelancingnger.Unwashedhandsorresidual
alcohol may lead to inaccurate readings.
Before testing, always make sure the meter is at operating temperature
between 50–104°F (10–40°C).
If the meter has been exposed to extreme temperatures, move the meter
to an area within operating range, and allow the meter to warm up or
cool down slowly (20 minutes) until the temperature warning symbol
goes away.
Nevermakesignicantchangestoyourdiabetescontrolprogramorignore
physical symptoms without consulting with your healthcare professional.
• Patients undergoing oxygen therapy may yield false results.
Severe dehydration (excessive water loss) may cause false low results.
If you believe you are suffering from dehydration, consult your
healthcare professional right away.
Extremes in hematocrit may affect test results. Hematocrit levels less
than 30% may cause falsely high readings. Hematocrit levels greater
than 55% may cause falsely low readings.
8
•Triglyceridesupto3,000mg/dLdonotsignicantlyaffecttestresults.
However, glucose values in specimens beyond this triglycerides level
should be interpreted with caution.
Normal endogenous (within body) natural levels of uric acid, ascorbate
(vitamin C), bilirubin, triglycerides, and hemoglobin do not interfere with
your blood glucose results obtained.
Externally taken drugs L-dopa, dopamine, methyl-dopa, acetaminophen,
and ibuprofen will not interfere with ReliOn™ Micro blood glucose results
when taken at therapeutic concentrations.
The ReliOn™ Micro should not be used for the diagnosis of diabetes.
• Use only fresh capillary whole blood.
Do not use the ReliOn™ Micro Blood Glucose Monitoring System to test
neonates. It has not been validated for neonatal use.
•Alwaysinserttheteststripintothemeterrst,thenprickthenger.Not
doing so may result in an“E-6” appearing on the display or a test result
that is higher than the actual value.
• Do not use at altitudes higher than 10,000 ft (3,048 m). 9
Inaccurate results may occur in severely hypotensive individuals or patients
in shock. Inaccurate low results may occur for individuals experiencing
a hyperglycemic-hyperosmolar state, with or without ketosis. This meter
should not be used to test critically ill patients.
The System Includes
• Meter
• Carrying case
• User Instruction Manual
• Quick Reference Guide
• Self-testing Logbook
• Lancing device
• Lancets
Test strips and control solution are required but not
provided and must be purchased separately.
10
Carrying case
UNDERSTANDING THE RELION™ MICRO
BLOOD GLUCOSE METER
Test Strip Port
Top
Display
Battery
Compartment
Cover
Strap Hole
Strap Hole
Button
Button
Bottom
11
Distributed by
Wal-Mart Stores, Inc.
Bentonville, AR 72716
P/N 7835-02 Rev 11/14
S/N 1941KE000000
MADE IN PHILIPPINES
12
Explanation of the Full Display Screen
NOTE
The correct unit of measurement in the U.S.A. is mg/dL. If your
meter is not set to mg/dL at the time of purchase contact ReliOn
Customer Service at 855.776.0662, 7 a.m. to 7 p.m. central time,
7 days a week.
13
ReliOn™ Conrm/Micro Plus Test Strips
These test strips need only 0.3 µL (microliter) of blood. Apply blood to
the sample application tip. The test strip draws in the blood by capillary
action. The blood glucose reacts with the enzyme on the test strip. This
produces a current proportional to the blood glucose level. The meter
detects the current and converts it into a blood glucose reading.
Test Strip Bottle
Front Back
Test Strip
Sample
Application Tip
( Apply blood to
end of test strip.
Capillary action
will pull blood to
reaction site).
Insert
this way
into
meter
Blood Glucose Test Strips
PLUS
Contact Bars
14
CAUTION
Replacing the battery will not delete meter settings or test results
stored in the meter unless you:
• Take over 1 minute to replace battery
• Load the battery incorrectly
• Load a used or weak battery
Changing the Battery
The ReliOn™ Micro blood glucose meter comes with one lithium battery
CR2032 (or DL2032).
WARNING
Keep all batteries out of childrens reach. If a battery is swallowed,
consult a doctor immediately.
NOTE
To save battery power, the meter turns itself off 3–5 minutes after
last use.
Replace batteries when low battery symbol appears. Otherwise
inaccurate results could be obtained.
• Discard used batteries as per local guidelines.
15
To replace battery
Step 1: Turn meter off.
Step 2: Remove the battery cover on the back of the meter.
Step 3: Slide the battery to the right (A) and remove it (B).
Step 4: Load a new battery into the meter and slide the
battery to the left (C). Place the ‘+’ side upward.
CAUTION
Make sure the battery is loaded in the proper position.
Otherwise, the battery cover will not close.
A
B
C
16
Step 5: Replace battery cover.
If the clock setting has been deleted, ‘SET’ will appear on the screen.
If ‘SET’ appears, set the time and date. If you do not set the time and
date, all subsequent results will be stored with a test date and time of
January 1, 2008, 12:00 a.m.
Meter set-up allows you to adjust the audible beep volume and to set
time, date, and test averaging.
BeforeusingtheReliOn™Microbloodglucosemeterforthersttime,
set the time and date. Do not perform a blood glucose test until you
have set up the meter.
METER SETUP
Default Settings
17
Audible Beep Volume High (3)
Average Off
18
Setup mode Audible Beep Date and time Test average display End
See page 20 See page 22 See page 25
19
Note
If you need to go back to a setting, you must scroll through each
setup option. You cannot go backwards in setup mode.
In setup mode, the meter will turn off automatically after 3 minutes
of nonuse.
Basic Rules for Meter Setup
Step 1: Turn off meter.
Step 2: Press the and buttons at the same time
for 5 seconds. A beep sounds and the meter turns on.
Next, a second beep sounds and “SET” shows on
the screen.
Step 3: Use the button to change the setup screen. When you press
the button, the setup screen changes in the order shown in the
diagram to the left. Press the button to change a setting within a
setup screen menu.
Setup mode Audible Beep Date and time Test average display End
See page 20 See page 22 See page 25
20
Setting the Audible Beep
If the audible beep is turned ‘on, the beep will sound when:
• You insert a test strip into the meter
• You apply blood or control solution to a test strip
• The test has ended and the result is displayed
The meter is turned on, but you do not use it for several minutes
Meter will beep and automatically shut off
• You go into Memory Mode
Make sure the meter is off. Press the and
buttons at the same time for 5 seconds.
A beep sounds and the meter turns on. Next,
a second beep sounds and ‘SET’ shows on
the screen.
Press the button. The screen will display
‘bEP’ and the current beep setting.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78