Lafayette Instrument 80225M/80226M Series LED Stimulus Light Owner's manual

  • Hello! I've analyzed the user's guide for the Lafayette Instrument LED Stimulus Lights, specifically the 80225M and 80226M series. This document details the specifications and wiring for both single and variable intensity models used in animal conditioning with computer-controlled systems. I'm here to answer any questions you may have about these lights.
  • What is the power requirement for these LED stimulus lights?
    How many intensity levels does the 80226M series have?
    What is the output illuminance for the 80225M series?
    What is the output illuminance range for the 80226M series?
3700 Sagamore Pkwy N
Lafayette, IN 47904
Tel: (765) 423-1505
www.lafayetteinstrument.com
Copyright © 2023 Lafayette Instrument Company, Inc. All Rights Reserved: Rel. 5.16.23
80225M/80226M Series
LED Stimulus Light
User’s Guide
MAN80225/6
[email protected] www.lafayetteinstrument.com
80225M/80226M Series LED Stimulus Light Users Guide
Description
The 80225M and 80226M series LED Stimulus lights are used for the conditioning of animals with
computer controlled operant systems using ABET, ABET II or other computer-controlled interface
and software. The 80225M and 80225M14 LED Stimulus lights are congured as a single intensity
light. The 80226M and 80226M14 LED Stimulus lights are congured as variable intensity light with 7
intensities.
Specications
Power: +24volts – 28volts DC
Output Illuminance
80225M/M14: 8 lux
80226M/M14: 8 - 60 lux
Wiring Instructions
80225M series LED Stimulus light
Red: Power (+24volts - 28 volts)
Black: Ground
White: Operate (Active low, shunt to Ground)
80226M series LED Stimulus light
Use the supplied 6 pin modular cable with stripped and tinned wire and wire as follows.
Pin Wire Color Description
1 White Output 3
2 Black Ground
3 Red Output 2
4 Green Ground
5 Yellow Output 1
6 Blue +24 volts – 28 volts
Table 1: Wiring RJ12 – 6 PIN Modular
Pin 1
Pin 2
Pin 3
Pin 5
Pin 4
Pin 6
Intensity
Level
Binary Values
(Output 1, Output 2, Output 3) Output Illuminance
0 0 0 0 0 lux, output o
1 0 0 1 8 lux
2 0 1 0 15 lux
3 0 1 1 20 lux
4 1 0 0 30 lux
5 1 0 1 40 lux
6 1 1 0 50 lux
7 1 1 1 60 lux
Table 2: Intensity
Note: All outputs are
Active low, shunt to
Ground
Note: Value 1 indicates
output # is turned “On”
/